K Number
K214056

Validate with FDA (Live)

Date Cleared
2022-09-20

(267 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue/purple colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: S, M, L, XL. The subject device is non-sterile.

AI/ML Overview

The provided document describes the acceptance criteria and the results of a study for "Disposable Nitrile Inspection Gloves (Tested for Use with Chemotherapy Drugs)".

Here's the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodPurposeAcceptance CriteriaReported Device Performance (Blue Gloves)Reported Device Performance (Purple Gloves)
ASTM D6319Physical Dimensions TestLength(mm): S: ≥220; M/L/XL: ≥230.(S: ≥220; M/L/XL: ≥230) - Pass(S: ≥220; M/L/XL: ≥230) - Pass
Width(mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10.S: 84-86/Pass; M: 96-98/Pass; L: 104-106/Pass; XL: 114-116/PassS: 88-90/Pass; M: 99-100/Pass; L: 102-104/Pass; XL: 115-117/Pass
Finger Thickness: ≥0.05 mm0.10-0.13 mm / Pass0.12-0.15 mm / Pass
Palm Thickness: ≥0.05 mm0.06-0.09 mm / Pass0.07-0.09 mm / Pass
ASTM D5151Watertightness Test for Detection of HolesAQL 2.50/125 / Pass0/125 / Pass
ASTM D6124Powder Content< 2.0 mg0.3 mg / Pass0.1 mg / Pass
ASTM D412Physical properties (Before Aging)Tensile Strength: ≥14 MPa15.3-19.0 MPa30.0-37.8 MPa
Ultimate Elongation: ≥500%521-569%505-539%
ASTM D412Physical properties (After Aging)Tensile Strength: ≥14 MPa14.3-18.0 MPa29.2-38.6 MPa
Ultimate Elongation: ≥400%417-524%453-504%
ISO 10993-5CytotoxicityNon- In Vitro CytotoxicityDevice extract is cytotoxic. (Further evaluated by ISO 10993-11)Device extract is cytotoxic. (Further evaluated by ISO 10993-11)
ISO 10993-11Acute Systemic ToxicityNon-acute systemic toxicityDid not show acute systemic toxicity in vivo / PassDid not show acute systemic toxicity in vivo / Pass
ISO 10993-10IrritationNon-irritatingNot an irritant / PassNot an irritant / Pass
ISO 10993-10SensitizationNon-sensitizingNot a sensitizer / PassNot a sensitizer / Pass
ASTM D6978-05 (Chemotherapy Permeation)See detailed tables below for each drugBreakthrough Detection Time in MinutesSee tables in section 6.1See tables in section 6.2

Chemotherapy Drug Permeation (Blue Gloves):

Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes (Acceptance implicitly >0, ideally >240 for most drugs)Reported Device Performance (>240 is generally considered passing for clinical relevance)
Carboplatin10.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Carmustine3.3 mg/ml> 240 Minutes (implicit desired)18.4 Minutes / Caution (Low permeation time)
Cisplatin1.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Cyclophosphamide20.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Doxorubicin HCI2.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Etoposide20.0 mg/ml> 240 Minutes (implicit desired)202.7 Minutes / Caution (Low permeation time)
Fluorouracil50.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Paclitaxel6.0 mg/ml> 240 Minutes> 240 Minutes / Pass
ThioTepa10.0 mg/ml> 240 Minutes (implicit desired)26.6 Minutes / Caution (Low permeation time)

Chemotherapy Drug Permeation (Purple Gloves):

Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes (Acceptance implicitly >0, ideally >240 for most drugs)Reported Device Performance (>240 is generally considered passing for clinical relevance)
Carboplatin10.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Carmustine3.3 mg/ml> 240 Minutes (implicit desired)20.9 Minutes / Caution (Low permeation time)
Cisplatin1.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Cyclophosphamide20.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Doxorubicin HCI2.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Etoposide20.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Fluorouracil50.0 mg/ml> 240 Minutes> 240 Minutes / Pass
Paclitaxel6.0 mg/ml> 240 Minutes> 240 Minutes / Pass
ThioTepa10.0 mg/ml> 240 Minutes (implicit desired)33.5 Minutes / Caution (Low permeation time)

2. Sample size used for the test set and the data provenance

The document provides details of the tests conducted but does not explicitly state the exact sample sizes for each test (e.g., number of gloves tested for physical dimensions, biocompatibility, or chemotherapy permeation). It mentions quantities like "0/125/Pass" for the watertightness test, implying 125 units were tested.

The data provenance is not specified regarding the country of origin. The study appears to be bench testing (non-clinical), meaning it was performed in a lab setting rather than with human subjects, making it akin to a controlled experimental study rather than retrospective or prospective clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable as the document describes bench testing for physical properties, chemical permeation, and biocompatibility of gloves. These tests rely on established scientific methods and instrumental measurements, not on expert human interpretation or consensus to establish ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable as the study did not involve human interpretation or subjective assessments that would require adjudication. The performance is determined by objective measurements against predefined criteria in the standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is an examination glove, not an AI-assisted diagnostic or therapeutic tool. Therefore, MRMC studies and AI-related effectiveness metrics are irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not applicable. The device is a physical product (gloves) and does not involve any algorithms or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The 'ground truth' for these tests is defined by industry standards and their specified acceptance criteria. For example:

  • Physical Properties: ASTM D6319, ASTM D412 specify acceptable ranges for length, width, thickness, tensile strength, and elongation.
  • Freedom from Holes: ASTM D5151 sets the Acceptable Quality Level (AQL) for watertightness.
  • Powder Content: ASTM D6124 defines the maximum allowable powder.
  • Biocompatibility: ISO 10993 standards define methods and acceptable responses for cytotoxicity, irritation, sensitization, and systemic toxicity.
  • Chemotherapy Permeation: ASTM D6978-05 provides the standard practice for assessing breakthrough times. The "ground truth" here is the measured time until breakthrough, compared against a desired clinical threshold (e.g., >240 minutes for safe handling).

8. The sample size for the training set

This is not applicable. There is no "training set" as the device is a physical product and the study involves direct measurements and testing against specified standards, not machine learning model training.

9. How the ground truth for the training set was established

This is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 20, 2022

Mezorrison Medical Technology (Dongying) Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161 Lujiazui East Rd.,Pudong Shanghai. 200120 China

Re: K214056

Trade/Device Name: Disposable Nitrile Inspection Gloves (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: August 22, 2022 Received: August 23, 2022

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bifeng Oian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K214056

Device Name

Disposable Nitrile Inspection Gloves (Tested for Use with Chemotherapy Drugs)

Indications for Use (Describe)

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

  1. The following is the resistance to permeation by chemotherapy drugs for the blue gloves.
Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes
Carboplatin10.0 mg/ml> 240 Minutes
Carmustine3.3 mg/ml18.4 Minutes
Cisplatin1.0 mg/ml> 240 Minutes
Cyclophosphamide20.0 mg/ml> 240 Minutes
Doxorubicin HCI2.0 mg/ml> 240 Minutes
Etoposide20.0 mg/ml202.7 Minutes
Fluorouracil50.0 mg/ml> 240 Minutes
Paclitaxel6.0 mg/ml> 240 Minutes
ThioTepa10.0 mg/ml26.6 Minutes

Please note that the following drugs have low permeation times: Carmustine 3.3 mg/ml 18.4 Minutes ThioTepa 10.0 mg/ml 26.6 Minutes Etoposide 20.0 mg/ml 202.7 Minutes Warning: Please do not use with Carmustine, ThioTepa and Etoposide.

  1. The following is the resistance to permeation by chemotherapy drugs for the purple gloves.
Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes
Carboplatin10.0 mg/ml> 240 Minutes
Carmustine3.3 mg/ml20.9 Minutes
Cisplatin1.0 mg/ml> 240 Minutes
Cyclophosphamide20.0 mg/ml> 240 Minutes
Doxorubicin HCI2.0 mg/ml> 240 Minutes
Etoposide20.0 mg/ml> 240 Minutes
Fluorouracil50.0 mg/ml> 240 Minutes
Paclitaxel6.0 mg/ml> 240 Minutes
ThioTepa10.0 mg/ml33.5 Minutes

Please note that the following drugs have low permeation times: Carmustine 3.3 mg/ml 20.9 Minutes ThioTepa 10.0 mg/ml 33.5 Minutes Warning: Please do not use with Carmustine and ThioTepa.

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary K214056

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: Mezorrison Medical Technology (Dongying) Co., Ltd. Address: Room 202, No. 166 of East Campus, Nanyi Road, the Development Zone of Dongying, Shandong, China Contact: Ping Chen Date of Preparation: Aug.22, 2022

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 Lujiazui East Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Disposable Nitrile Inspection Gloves (Tested for Use with Chemotherapy Drugs) Patient Examination Gloves Common name: Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL

3.0 Classification

Production code: LZA,LZC Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Ever Growth (Vietnam) Co., Ltd. Device: Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy

{5}------------------------------------------------

Drugs, Orange Color 510(k) number: K190860

5.0 Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue/purple colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: S, M, L, XL. The subject device is non-sterile.

6.0 Indication for Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs

Chemotherapy DrugConcentrationBreakthrough DetectionTime in Minutes
Carboplatin10.0 mg/ml> 240
Carmustine3.3 mg/ml18.4
Cisplatin1.0 mg/ml> 240
Cyclophosphamide20.0 mg/ml> 240
Doxorubicin HCI2.0 mg/ml> 240
Etoposide20.0 mg/ml202.7
Fluorouracil50.0 mg/ml> 240
Paclitaxel6.0 mg/ml> 240
ThioTepa10.0 mg/ml26.6

6.1 The following is the resistance to permeation by chemotherapy drugs for the blue gloves.

Please note that the following drugs have low permeation times:

Carmustine (BCNU) 3.3 mg/ml 18.4 Minutes;

Thio Tepa 10.0 mg/ml 26.6 Minutes.

Etoposide 20.0 mg/ml 202.7 Minutes

Warning: Please do not use with Carmustine, ThioTepa and Etoposide.

6.2 The following is the resistance to permeation by chemotherapy drugs for the purple gloves.

{6}------------------------------------------------

Chemotherapy DrugConcentrationBreakthrough DetectionTime in Minutes
Carboplatin10.0 mg/ml> 240
Carmustine3.3 mg/ml20.9
Cisplatin1.0 mg/ml> 240
Cyclophosphamide20.0 mg/ml> 240
Doxorubicin HCI2.0 mg/ml> 240
Etoposide20.0 mg/ml> 240
Fluorouracil50.0 mg/ml> 240
Paclitaxel6.0 mg/ml> 240
ThioTepa10.0 mg/ml33.5

Please note that the following drugs have low permeation times:

Carmustine (BCNU) 3.3 mg/ml 20.9 Minutes;

Thio Tepa 10.0 mg/ml 33.5 Minutes.

Warning: Please do not use with Carmustine (BCNU) and ThioTepa.

7.0 Technological Characteristic Comparison Table

ItemSubject Device(K214056)Predicate Device(K190860)Remark
Product CodeLZA, LZCLZA, LZCSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseA patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves toA patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves toSame

Table1-General Comparison

{7}------------------------------------------------

Powdered or Powered freeChemotherapy Drugs.Chemotherapy Drugs.
Design FeatureAmbidextrousAmbidextrousSame
SterilityNon-SterileNon-SterileSame
Labeling InformationSingle-useindication,powder free, device color,device name, glove sizeand quantity,Non-Sterile,a statement of standardASTMD6978-05complianceand asummary of the testingresults.Single-useindication,powder free, device color,device name, glove sizeand quantity, Non-Sterile,a statement of standardASTMD6978-05complianceand asummary of the testingresults.Same
Dimensions(mm)Length:S:≥220;M/L/XL: ≥230;Width:S: 80±10;M: 95±10;L: 110±10;XL: 120±10.Length:XS/S/M/L/XL: ≥230;Width:XS:70±10;S: 80±10;M: 95±10;L: 110±10;XL: 120±10SimilarAnalysis1
Thickness(mm)Finger: ≥0.05;Palm: ≥0.05Finger: ≥0.05;Palm: ≥0.05Same
ColorantBlue, PurpleWhite, OrangeDifferentAnalysis2
PhysicalPropertiesBeforeAgingTensileStrength14MPa,minTensileStrength14MPa,minSame
UltimateElongation500% minUltimateElongation500% minSame
After AgingTensileStrength14MPa,minTensileStrength14MPa,minSame
UltimateElongation400%minUltimateElongation400%minSame
Freedom from HolesBe free from holes whentested in accordancewithASTMD5151AQL=2.5Be free from holes whentested in accordancewithASTMD5151AQL=2.5Same
Powder Content0.1~0.3 mg per glove,Meet the requirementsof ASTM D6124Meet the requirementsof ASTM D6124SimilarAnalysis3
BiocompatibilityISO 10993-10;Under the conditions ofthe study, not an irritantISO 10993-10;Under the conditions ofthe study, not an irritantSame

{8}------------------------------------------------

or a sensitizeror a sensitizer
ISO 10993-5Under conditions of thestudy, device extract iscytotoxicISO 10993-11;Under thecondition of acutesystemic toxicity test,the test article did notshow acute systemictoxicity in vivo.ISO 10993-5Under conditions of thestudy, device extract isnot cytotoxicDifferentAnalysis 4
Carboplatin10.0 mg/ml>240 Minutes/DifferentAnalysis 5
Carmustine3.3 mg/mlBlue: 18.4 MinutesPurple: 20.9 MinutesWhite:11.8 Minutes;Orange:31.6MinutesSimilar
Cisplatin1.0 mg/ml>240 Minutes>240 MinutesDifferentAnalysis 5
ChemotherapyDrugs Testedwith MinimumBreakthroughDetection Timeas Tested perASTM D 6978Cyclophosph-amide20.0 mg/ml>240 Minutes>240 MinutesSame
Dacarbazine10.0 mg/ml/>240 MinutesDifferentAnalysis 5
DoxorubicinHCI2.0 mg/ml>240 Minutes>240 MinutesSame
Etoposide20.0 mg/mlBlue: 202.7 MinutesPurple: >240 Minutes>240 MinutesDifferentAnalysis 5
Fluorouracil50.0 mg/ml>240 Minutes>240 MinutesSame
Paclitaxel6.0 mg/ml>240 Minutes>240 MinutesSame
ThioTepa10.0 mg/mlBlue: 26.6 MinutesPurple: 33.5 MinutesWhite:16.9 Minutes;Orange: 72.5 MinutesDifferentAnalysis 5

Analysis 1:

The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

Analysis 2:

The color of the subject device is different with that of the predicate. The subject device was evaluated according to ISO 10993-1 standards, and there were no risks identified.

{9}------------------------------------------------

Analysis 3:

Powder Content of subject device is similar with that of the predicate, because the predicate did not publish the exact results of the powder content. But they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

Analysis 4:

Under conditions of the ISO 10993-5 study, subject device extract is cytotoxic, but under the condition of the acute systemic toxicity test, the test article did not show systemic toxicity in vivo. Under conditions of the study, predicate device extract is not cytotoxic. Both of the subject device and predicate device have passed the toxicity study, so it does not raise any new safety or performance questions.

Analysis 5:

And Breakthrough detection times of Carboplatin, Carmustine, Dacarbazine, Etoposide, and Thio Tepa of subject device are different with those of the predicate. The Chemotherapy Labeling Claims has clearly defined on the labeling. So it does not raise any new safety or performance questions.

8.0 Summary of Non-Clinical Testing

Biocompatibility Testing

The biocompatibility evaluation for Nitrile Patient Examination Gloves (Tested for Use with Chemotherapy Drugs) was conducted in accordance with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests

for systemic toxicity.

Performance Testing (Bench)

Physical performance qualities of the proposed device were evaluated per ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2019), Standard Practice

{10}------------------------------------------------

for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs.

In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • ASTM D6124-06 (Reapproved 2017), Standard Test Method for -Residual Powder on Medical Gloves
  • ASTM D5151-19, Standard Test Method for Detection of Holes in -Medical Gloves.
  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
  • ASTM D6978-05 (Reapproved 2019), Standard Practice for -Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
TestMethodPurposeAcceptance CriteriaResults
ASTMD6319PhysicalDimensionsTestLength(mm):S: ≥220;M/L/XL: ≥230.Length(mm):S: ≥220;M/L/XL: ≥230.
Width(mm):S: 80±10;M: 95±10;L: 110±10;XL: 120±10.Width(mm):Blue:S: 84-86/PassM: 96-98/ PassL: 104-106/PassXL:114-116/ PassPurple:S: 88-90/PassM: 99-100/ PassL: 102-104/ PassXL:115-117/ Pass
Finger: ≥0.05;Palm: ≥0.05Thickness (mm):Finger:Blue:0.10-0.13/PassPurple:0.12-0.15/Pass

Table 2 - Summary of non-clinical performance testing

{11}------------------------------------------------

Palm:
Blue:
0.06-0.09/Pass
Purple:
0.07-0.09/Pass
ASTMD5151WatertightnessTest forDetection ofHolesMeet the requirements of ASTM D5151AQL 2.50/125/Pass
ASTMD6124PowderContentMeet the requirements of ASTM D6124 < 2.0mgBlue: 0.3mg/Pass;Purple: 0.1mg/Pass
ASTMD412PhysicalpropertiesBeforeAgingTensileStrength≥14MPaBlue:15.3-19.0MPaPurple:30.0-37.8MPa
UltimateElongation≥500%Blue:521-569%Purple:505-539%
AfterAgingTensileStrength≥14MPaBlue:14.3-18.0MPaPurple:29.2-38.6MPa
UltimateElongation≥400%Blue:417-524%Purple:453-504%
ISO10993-5CytotoxicityNon- In Vitro CytotoxicityUnder conditions ofthe study, deviceextract is cytotoxic.
ISO10993-11CytotoxicityNon- acute systemic toxicityUnder conditions ofthe study, did notshow acute systemictoxicity in vivo / Pass
ISO10993-10IrritationNon-irritatingUnder conditions ofthe study, not an

{12}------------------------------------------------

irritant. / Pass
ISO10993-10SensitizationNon-sensitizingUnder conditions ofthe study, not asensitizer. / Pass

9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Disposable Nitrile Inspection Gloves (Tested for Use with Chemotherapy Drugs) is as safe, as effective, and performs as well as or better than the legally marketed predicated device under K190860.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.