K Number
K213898

Validate with FDA (Live)

Date Cleared
2023-03-01

(442 days)

Product Code
Regulation Number
870.1330
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ProTrack™ Pigtail Wires are intended for use to introduce and position catheters and other interventional devices within the chambers of the heart, including those used within transcatheter aortic valve procedures (TAVR).

Device Description

The ProTrack™ Pigtail Wire is a single used for the patients requiring diagnostic and interventional procedures like transcatheter aortic valve replacement (TAVR). It is used to facilitate the introduction and placement of devices in catheterization procedures primarily by Electrophysiologists and Interventional Cardiologists. Procedures using the devices are performed in fully equipped catheter labs with imaging equipment, including fluoroscopy and echocardiography under sterile technique.

The device consists of a stainless-steel core wire with a flexible, spiral shaped stainless steel distal coil. A green PTFE coating is present on the entire length of the device and provides a lubricious surface for smooth tracking through accessory devices and vasculature. The Wire has a radiopaque coil (304V Stainless Steel) around the distal curve to allow visualization under appropriate imaging guidance during procedures. The device is offered in multiple models with two different lengths (230 cm and 275 cm), diameters (0.032″ and 0.035″), and distal curve shapes (regular and small). The device is provided sterilized).

The wire is classified as an external communicating device, contacting circulating blood, with limited exposure and contact up to a maximum of 24 hours. The patient contacting materials are PTFE coating, stainless steel and adhesive.

AI/ML Overview

The provided FDA 510(k) summary (K213898) for the ProTrack™ Pigtail Wire does not contain information about the acceptance criteria or results from a study that typically applies to AI/ML software as a medical device (SaMD). This document is for a physical medical device (guidewire), and the performance testing described focuses on mechanical, physical, biocompatibility, sterilization, pyrogen, and packaging verification, as well as bench-top validation for user needs.

Therefore, the requested information about acceptance criteria for AI/ML performance, sample sizes for test sets, data provenance, expert ground truth adjudication, MRMC studies, standalone algorithm performance, ground truth types, and training set details are not applicable to the content of this document.

The document describes the following types of performance testing and their general outcomes:

1. Table of Acceptance Criteria and Reported Device Performance (as interpreted for a physical device):

Acceptance Criteria CategorySpecific Tests Performed by Baylis Medical Company Inc.Reported Device Performance
Mechanical Performance- Corrosion Resistance (per 11070:2014/Amd:2018)All test requirements met
- Flex Test (per 11070:2014/Amd:2018)All test requirements met
- Fracture test (per 11070:2014/Amd:2018)All test requirements met
- Peak tensile force (per 11070:2014/Amd:2018)All test requirements met
- Tensile strength (per 11070:2014/Amd:2018)All test requirements met
- Tip Pull (per 11070:2014/Amd:2018)All test requirements met
- Dimensional verification (per FDA guidance)All test requirements met
- Visual inspection (per FDA guidance)All test requirements met
- Repeated Simulated Use (per FDA guidance)All test requirements met
- Torque Strength (per FDA guidance)All test requirements met
- Coating integrity (per FDA guidance)All test requirements met
- Particulate testing (per FDA guidance)All test requirements met
- Lubricity/Friction testing (per FDA guidance)All test requirements met
- Kink resistance (per FDA guidance)All test requirements met
- Tip flexibility (per FDA guidance)All test requirements met
- Radiopacity (per FDA guidance)All test requirements met
General Physical Performance- Curve Retention (per self-enforced requirements)All test requirements met
- Flexural rigidity (per self-enforced requirements)All test requirements met
Biocompatibility- ISO Cytotoxicity studyAll test requirements met
- ISO Maximization Sensitization StudyAll test requirements met
- ISO Intracutaneous Reactivity StudyAll test requirements met
- ISO/USP Material Mediated PyrogenicityAll test requirements met
- ISO Hemocompatibility StudyAll test requirements met
- ISO Acute Systemic Toxicity StudyAll test requirements met
Sterilization- EO Sterilization to SAL 10-6 (per ISO 11134-1:2014)All test requirements met
- Residual limits (per ISO 10993-7:2008)All test requirements met
Pyrogenicity- LAL testing (Kinetic Chromogenic method) (per ANSI/AAMI ST72:2019 and FDA guidance)Meets current FDA and USP pyrogen limit specifications
Packaging- Ship testingIntegrity verified
- Seal strengthCompliance with ISO 11607-1:2020 and ISO 11607-2:2017 verified
- Sterile barrier integrityCompliance with ISO 11607-1:2020 and ISO 11607-2:2017 verified
User Needs/Usability- Bench-top Validation with users familiar with catheter-based interventional proceduresMeets intended use, supporting substantial equivalence

The following information is not present in the provided document, as it pertains specifically to AI/ML software performance studies and not to a physical medical device like the ProTrack™ Pigtail Wire:

  1. Sample size used for the test set and the data provenance.
  2. Number of experts used to establish the ground truth for the test set and their qualifications.
  3. Adjudication method for the test set.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
  7. The sample size for the training set.
  8. How the ground truth for the training set was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

March 1, 2023

Baylis Medical Company Inc. May Tsai Director of Regulatory Affairs 5825 Explorer Dr. Mississauga, Ontario L4W 5P6 Canada

Re: K213898

Trade/Device Name: ProTrack Pigtail Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DOX Dated: February 1, 2023 Received: February 1, 2023

Dear May Tsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer Bastijanic -S

for Jaime Raben Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213898

Device Name ProTrack™ Pigtail Wire

Indications for Use (Describe)

The ProTrack™ Pigtail Wires are intended for use to introduce and position catheters and other interventional devices within the chambers of the heart, including those used within transcatheter aortic valve procedures (TAVR).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

Submitter Information

A. Company Name:Baylis Medical Company Inc.
B. Company Address:5825 Explorer DrMississauga, Ontario, L4W 5P6Canada
C. Company Phone:+1 (905) 602-4875
D. Company Facsimile:+1 (905) 602-5671
E. Contact Person:May TsaiDirector, Regulatory Affairs
F. Summary Prepared on:23-Feb-2023

Device Identification

A. Device Trade Name:ProTrack™ Pigtail Wire
B. Device Common Name:Guidewire
C. Classification Name:Wire, Guide, Catheter (21 CFR 870.1330)
D. Product Code:DQX
E. Review Panel:Cardiovascular
F. Device Class:Class II

Identification of Legally Marketed Device

A. Predicate Device:Medtronic Confida™ Brecker Curve™ Guidewire
B. Manufacturer:Medtronic
C. 510(k)K132623
D. Indications for UseThe Medtronic Confida™ Brecker Curve™ Guidewireis intended for use to introduce and positioncatheters during diagnostic and-interventionalprocedures within the chambers of the heart,including, transcatheter aortic valve implantation(TAVI).

{4}------------------------------------------------

Intended Use

The ProTrack™ Piqtail Wires are intended to facilitate the placement of devices during diagnostic and interventional procedures.

Indications for Use

The ProTrack™ Pigtail Wires are intended for use to introduce and position catheters and other interventional devices within the chambers of the heart, including those used within transcatheter aortic valve procedures (TAVR).

Device Description

The ProTrack™ Pigtail Wire is a single used for the patients requiring diagnostic and interventional procedures like transcatheter aortic valve replacement (TAVR). It is used to facilitate the introduction and placement of devices in catheterization procedures primarily by Electrophysiologists and Interventional Cardiologists. Procedures using the devices are performed in fully equipped catheter labs with imaging equipment, including fluoroscopy and echocardiography under sterile technique.

The device consists of a stainless-steel core wire with a flexible, spiral shaped stainless steel distal coil. A green PTFE coating is present on the entire length of the device and provides a lubricious surface for smooth tracking through accessory devices and vasculature. The Wire has a radiopaque coil (304V Stainless Steel) around the distal curve to allow visualization under appropriate imaging guidance during procedures. The device is offered in multiple models with two different lengths (230 cm and 275 cm), diameters (0.032″ and 0.035″), and distal curve shapes (regular and small). The device is provided sterilized).

The wire is classified as an external communicating device, contacting circulating blood, with limited exposure and contact up to a maximum of 24 hours. The patient contacting materials are PTFE coating, stainless steel and adhesive.

Comparison of Technological Characteristic with Predicate Device

The subject ProTrack™ Pigtail Wire and predicate device Confida™ Brecker Curve™ Guidewire have the same intended use, fundamental scientific technology, sterility and principle of operation. Differences in design and technological characteristics between the subject and predicate device do not raise any new types of questions of safety and effectiveness. The results of verification and validation testing provide reasonable assurance of the safe and

{5}------------------------------------------------

effective use of the ProTrack™ Pigtail Wire to establish substantial equivalence for its intended use to the predicate device. Table 1 below shows technological comparison of subject device with predicate device.

CharacteristicComparison ResultComment
Intended UseIdenticalThe subject and predicate device are bothintended to facilitate the placement ofdevices during diagnostic andinterventional procedures.
Indications for UseSimilarThe indications for use of the subjectdevice is similar to that of the predicatedevice, with minor wording differences toclarify the devices that can be positioned.The terms TAVI (used in predicate devicelabelling) and TAVR (used in subjectdevice labelling) are consideredequivalent.
Fundamental ScientificTechnologyIdenticalBoth subject and predicate device rely onuser controlled mechanical manipulationto percutaneously gain access tovasculature and chambers of heart.
Operating PrinciplesIdenticalBoth subject and predicate devicefacilitate the introduction and positioningof one catheter and/or interventionaldevice for another, while maintainingguidewire position in the chambers of theheart.
Mechanism of ActionIdenticalThe mechanism of action for both subjectand predicate device is mechanical force.
Environment of UseIdenticalBoth subject and predicate device are usedin facilities equipped and staffed to
perform diagnostic and interventionalprocedures.
MaterialsSimilarThe subject and predicate device are bothcomprised of the following materials:• Stainless steel• PTFE coatingThe following material is different:• Adhesive
TechnologicalcharacteristicsSimilarBoth subject and predicate device sharethe same characteristic:• Distal tip configurationThe following characteristics are different:• Device usable Length• Outer diameter• Distal tip spiral diameter• Compatibility with accessorydevice
PackagingSimilarThe subject and predicate devicepackaging system differ in:• Packaging material• Packaging configuration
SterilizationIdenticalBoth subject and predicate device areSingle Use, Ethylene Oxide sterilized

Table 1: Comparison of Technological Characteristics with Predicate Device

{6}------------------------------------------------

{7}------------------------------------------------

Performance Testing Summary

Performance Testing has been completed to demonstrate substantial equivalence of the subject device and predicate device. All test requirements were met as specified by applicable standard and the test protocols. The device was subjected to the following verification and validation activities.

Mechanical Testing

Mechanical verification testing was conducted for the ProTrack™ Pigtail Wire to ensure compliance with the requirements per FDA guidance and ISO 11070:2014/Amd:2018. The following mechanical tests were performed for the device:

Per 11070:2014/Amd:2018:

  • Corrosion Resistance .
  • Flex Test .
  • Fracture test ●
  • Peak tensile force
  • Tensile strength .
  • Tip Pull ●

Per FDA guidance Coronary, Peripheral, and Neurovascular Guidewires -Performance Tests and Recommended Labeling:

  • . Dimensional verification
  • Visual inspection .
  • Repeated Simulated Use .
  • Peak tensile force/Tensile strength ●
  • Tip pull .
  • Torque Strength .
  • Coating integrity ●
  • Particulate testing ●
  • Lubricity/Friction testing
  • Corrosion resistance
  • Kink resistance .
  • Tip flexibility .
  • Radiopacity ●

{8}------------------------------------------------

General Physical Testing

General physical verification testing was conducted for the ProTrack™ Pigtail Wire to ensure compliance with Baylis Medical Company Inc. self-enforced requirements. The following tests were performed:

  • . Curve Retention
  • . Flexural rigidity

Biocompatibility Verification

The biological safety of the ProTrack™ Pigtail Wire was verified in accordance with the requirements of ISO 10993-1:2020 and the FDA guidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". The following biocompatibility tests were performed:

  • . ISO Cytotoxicity study
  • ISO Maximization Sensitization Study .
  • ISO Intracutaneous Reactivity Study
  • ISO/USP Material Mediated Pyrogenicity .
  • ISO Hemocompatibility Study ●
  • . ISO Acute Systemic Toxicity Study

Sterilization Verification

Sterilization verification was completed for the ProTrack™ Pigtail Wire to the requirements of ISO 11134-1:2014. Sterilization is performed with Ethylene Oxide to a Sterility Assurance Level (SAL) of 10-6. Residual limits are in accordance with ISO 10993-7:2008.

Pyrogen Testing

The ProTrack™ Pigtail Wire is supplied non-pyrogenic. Limulus Amoebocyte Lysate (LAL) testing was evaluated using the Kinetic Chromogenic method, as per ANSI/AAMI ST72:2019 and the FDA guidance document, "Guidance for Industry – Pyrogens and Endotoxins Testing: Questions and Answers," to verify the subject device meets current FDA and USP pyrogen limit specifications.

{9}------------------------------------------------

Packaging Verification

Ship testing was evaluated to verify the integrity of the subject device packaging through the rigors of shipping and handling as well as storage over time. The seal strength and sterile barrier integrity was also evaluated to verify compliance with the current applicable requirements of ISO 11607-1:2020 and ISO 11607-2:2017 over the proposed intended shelf life of the subject device.

Bench-top Validation

User needs for ProTrack™ Piqtail Wire were validated through benchtop validation activities. Users drawn from populations familiar with catheter-based interventional procedures were presented with a sequence of actions to perform with a benchtop model representative of a clinical workflow involving the device. Observations regarding ease of use and errors made were recorded and evaluated to conclude if customer requirements were met. The validation testing demonstrated that the ProTrack™ Pigtail Wire meets the intended use to support substantial equivalence.

Conclusion

The intended use and fundamental scientific technology, including principles of operation and mechanism of action, of the subject ProTrack™ Pigtail Wire are identical to those of the predicate Medtronic Confida™ Brecker Curve™ Guidewire (K132623). Differences in design and technological characteristics do not raise any different questions of safety and effectiveness. The results of verification and validation activities support the substantial equivalence of the subject and the predicate device.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.