Intended Use

The Cardio Flow Peripheral Guide Wire is intended for temporary placement in peripheral vasculature the placement and exchange of diagnostic and therapeutic devices during percutaneous intravascular procedures. This guidewire device is intended for peripheral vascular use only.

Device Description

The Cardio Flow Peripheral Guide Wire (Guidewire), Model GW1001 is a 304V stainless steel mandrel guidewire with a fixed distal spring coil. The distal coil is radiopaque to fluoroscopy. The coil is medium-stiffness in load-force and the distal end is coated in hydrophobic silicone to facilitate tracking. The Guidewire has a maximum outer diameter of 0.36 mm (0.014-inch). The GW1001 Guidewire has overall nominal length of 325 cm.

AI/ML Overview

The Cardio Flow Peripheral Guide Wire (Model GW1001) underwent several non-clinical performance tests to demonstrate substantial equivalence to its predicate device, the Asahi Astato XS 20 Guide Wire (K103057).

Here's the breakdown of the acceptance criteria and the study information:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria (based on implied compliance with standards/guidance)Reported Device Performance
Corrosion ResistanceCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Dimensional VerificationCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Kink ResistanceCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
LubricityCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Particulate EvaluationCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
RadiopacityCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Silicone Coating IntegrityCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Simulated Use (Compatibility)Compliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Tensile StrengthCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Tip FlexibilityCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Tip PullCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Torque StrengthCompliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Torqueability (Torque Response)Compliance with FDA Guidance: Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling (October 10, 2019)Pass
Flex ResistanceResistant to damage by flexing in accordance with ISO 11070:2014 Annex G criterion, between two cylindrical mandrels with diameters equal to 20 times the guidewire diameter.Pass
Fracture ResistanceShall not fracture or fail when tested in accordance with ISO 11070:2014 Annex F criterion, test article wrapped around a mandrel with a diameter 10 times larger than the guidewire outside diameter.Pass
Tip LoadGuidewire tip load buckling force inducing buckling deformation with a 20 mm gauge length from the distal tip is documented. (Implied acceptance is demonstration of acceptable force for intended use)Pass
SterilitySterilization validated according to ANSI/AAMI/ISO 11135:2014 with a minimum sterility assurance level (SAL) of 10-6.Pass
Sterile barrier integrityPost terminal sterilization simulated distribution, seal peel (per ASTM F88), and bubble emission testing (per ASTM F2096) results demonstrate integrity.Pass
BiocompatibilityCompliance with ISO 10993-4, -5, -10, -11 (for Cytotoxicity, Complement Activation, Intracutaneous Reactivity, Sensitization, Material Mediated-Pyrogenicity, Acute Systemic Toxicity, and Standard in Vivo Thrombogenicity in Canine), ASTM F756 (for Hemolysis), and ASTM F2888 (for Heparinized Platelet and Leukocyte Count).Pass
Shelf life (2 years)Confirmation of device functional performance and sterile barrier pouch integrity (seal peel per ASTM F88 and bubble leak per ASTM F2096) with accelerated aging and simulated distribution per ASTM 4169-16.Pass

2. Sample Size Used for the Test Set and Data Provenance

The document describes bench evaluations and non-clinical performance tests. This implies that physical samples of the Cardio Flow Peripheral Guide Wire were tested in a laboratory setting.

  • Sample Size: Not explicitly stated for each individual test. However, it is standard practice for such tests to use a statistically significant number of samples to ensure robust results.
  • Data Provenance: The tests were conducted internally by or on behalf of Cardio Flow, Inc., likely at their facilities or a contracted testing lab, following established international and FDA-recognized standards. There is no mention of country of origin for test data, but the company is based in Mahtomedi, MN, USA. The data is prospective for the device under review.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. This was a physical device performance study, not a study requiring expert clinical assessment for ground truth. The "ground truth" for these tests is defined by the objective performance criteria specified in the referenced standards (e.g., ISO, ASTM, FDA guidance).

4. Adjudication Method for the Test Set

Not applicable, as this was a physical device performance study, not a clinical study involving human judgment that would require adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This device is a medical guidewire, not a diagnostic AI system that evaluates medical images or data. Therefore, an MRMC study comparing human readers with and without AI assistance is irrelevant.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This device is a physical medical guidewire, not an algorithm.

7. Type of Ground Truth Used

The "ground truth" for this type of device performance testing is established by objective engineering and performance specifications defined in recognized national and international standards (e.g., ISO, ASTM) and FDA guidance documents. These standards dictate acceptable ranges, thresholds, or qualitative outcomes (e.g., "Pass/Fail") for various physical and material properties.

8. Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

March 18, 2022

Cardio Flow, Inc. Michael Kallok Chief Executive Officer Quality & Regulatory Associates, LLC 888 East Evenuse, Mahtomedi, MN 55115 USA

Re: K213834

Trade/Device Name: Cardio Flow Peripheral Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: February 15, 2022 Received: February 17, 2022

Dear Michael Kallok:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213834

Device Name Cardio Flow Peripheral Guidewire

Indications for Use (Describe)

The Cardio Flow Peripheral Guide Wire is intended for temporary placement in peripheral vasculature the placement and exchange of diagnostic and therapeutic devices during percutaneous intravascular procedures. This guidewire device is intended for peripheral vascular use only.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) summary was prepared to provide an understanding of the basis for the determination of substantial equivalence in accordance with the requirements 21 CFR 807.92.

Submitters Name:Cardio Flow, Inc.888 East AvenueMahtomedi, MN 55115
Contact Person:Michael J. Kallok, Ph.D., FACC, FAHA; Chief Executive Officer,Director, Cardio Flow, Inc.
Contact Phone:(800) 294-5517
Date Summary Prepared:December 8, 2021
Device Trade Name:Cardio Flow Peripheral Guide Wire
Common Name:Peripheral Guide Wire
Classification Name:21 CFR 870.1330, Catheter Guide Wire, Class IIProduct Code: DQX
Predicate Device:K103057, Asahi Astato XS 20 Guide WireAsahi Intecc, Co. Ltd

Device Description

The Cardio Flow Peripheral Guide Wire (Guidewire), Model GW1001 is a 304V stainless steel mandrel guidewire with a fixed distal spring coil. The distal coil is radiopaque to fluoroscopy. The coil is medium-stiffness in load-force and the distal end is coated in hydrophobic silicone to facilitate tracking. The Guidewire has a maximum outer diameter of 0.36 mm (0.014-inch). The GW1001 Guidewire has overall nominal length of 325 cm.

Intended Use of the Device

The Cardio Flow Peripheral Guide Wire is intended for temporary placement in peripheral vasculature to facilitate the placement and exchange of diagnostic and therapeutic devices during percutaneous intravascular procedures. This guidewire device is intended for peripheral vascular use only.

Summary of Technological Characteristics

The following table provides a side-by-side comparison of the Guidewire to the predicate device applied to support this pre-market notification.

Substantial Equivalence Technical Characteristics
FeatureCardio Flow Peripheral GuideWire(Under Review)Asahi Astato XS 20 GuideWire(Predicate: K10357)EquivalenceComments
Product Code,ClassificationDQX,21 CFR 870.1330, Catheter GuideWire, Class IIDQX,21 CFR 870.1330, Catheter GuideWire, Class IIIdentical to predicate.
Substantial Equivalence Technical Characteristics
FeatureCardio Flow Peripheral Guide Wire(Under Review)Asahi Astato XS 40 GuideWire(Predicate: K153443K)EquivalenceComments
Intended UseGuidewire supporting guidingdiagnostic and interventional devicesthrough peripheral vasculature.Guidewire supporting guidingdiagnostic and interventional devicesthrough peripheral vasculature.Same
Indications forUseThe Cardio Flow peripheral guidewireis intended for temporary placement inperipheral vasculature to facilitate theplacement and exchange of diagnosticand therapeutic devices duringpercutaneous intravascular procedures.This guidewire device is intended forperipheral vascular use only.The ASAHI Astato XS 20 PeripheralGuide Wire is intended to facilitatethe placement and exchange ofdiagnostic and therapeutic devicesduring intravascular procedures.This device is intended for peripheralvascular use only.Same
Prescription UseOnlyYesYesSame
Single use,disposableYesYesSame
Provided sterileYesYesSame
SterilizationmethodEthylene OxideEthylene OxideSame
SterilityAssurance Level$\leq$ 10-6$\leq$ 10-6Same
Sterile barrierpackageTyvek Flexible PouchTyvek Flexible BarrierSimilar
BloodcontactingmaterialsPt-Ni, 304V Stainless SteelAu-Sn (Solder)SiliconePt-Ni, 304 Stainless SteelAu-Sn and Ag-Sn (Solder)PTFEHydrophilic coating on coil tipSimilar
BiocompatibilityevaluationsCytotoxicity ISO L929 MEM Elution,Cytotoxicity Neutral Red Uptake,ASTM Hemolysis Assay,Complement Activation Assay SC5b-9,Intracutaneous Reactivity,Sensitization Test ISO Guinea PigMaximization,Material Mediated-Pyrogenicity,Acute Systemic Toxicity Test: AcuteSystemic Injection Assay,Hemocompatibility: HeparinizedPlatelet and Leukocyte Count Assay,Standard In Vivo Thrombogenicity inCanineCytotoxicity Study,In Vitro Hemolysis Study,C3a Complement Activation Study,SC5b-9 Complement ActivationStudy,Intracutaneous Study,Sensitization Study,Pyrogen Study,Systemic Toxicity Study,Plasma Recalcification TimeCoagulation Study,In Vivo Thromboresistance StudySimilar
Outer Diameter0.36 mm (0.014 inches)Maximum Outer Diameter0.36 mm (0.014 inches)Maximum Outer DiameterSimilar
Overall Length325 cm180 to 300 cmSimilar
Tip Length35 mm17 cmSimilar
Tip Flexibility(Tip Load)10 to 25 gf(13 gf average)20 gfSimilar
Shelf life2 Years3 YearsSimilar

{4}------------------------------------------------

{5}------------------------------------------------

Non-clinical Performance Tests to Demonstrate Substantial Equivalency

To establish the technical equivalency of the Guidewire, bench evaluations were conducted to confirm compliance with performance requirements.

TestTest Method SummaryResult
Corrosion ResistanceFDA Guidance: Coronary, Peripheral, and NeurovascularPass
Dimensional VerificationGuidewires - Performance Tests and Recommended Labeling,(October 10, 2019)Pass
Kink ResistancePass
LubricityPass
Particulate EvaluationPass
RadiopacityPass
Silicone Coating IntegrityPass
Simulated Use (Compatibility)Pass
Tensile StrengthPass
Tip FlexibilityPass
Tip PullPass
Torque StrengthPass
Torqueability (Torque Response)Pass
Flex ResistanceThe guidewire will be resistant to damage by flexing in accordanceto ISO 11070:2014 Annex G criterion, between two cylindricalmandrels with diameters equal to 20 times the guide wire diameter.Pass
Fracture ResistanceThe guide wire shall not fracture or fail when tested in accordanceto ISO 11070:2014 Annex F criterion, test article wrapped around amandrel with a diameter 10 times larger than the guidewire outsidediameter.Pass
Tip LoadGuidewire tip load buckling force inducing buckling deformationwith a 20 mm gauge length from the distal tip is documented.Pass
SterilitySterilization of the subject device was validated according toANSI/AAMI/ISO 11135:2014, Sterilization of healthcare products- Ethylene oxide - Requirements for development, validation androutine control of a sterilization process for medical devices, with aminimum sterility assumed level (SAL) of 10-6.Pass
Sterile barrier integrityThe sterile barrier package is a flexible pouch. The methods appliedto evaluate the sterile barrier package integrity included postterminal sterilization simulated distribution, seal peel and bubbleemission testing.Pass
BiocompatibilityISO 10993-4, -5, -10, -11 appliable standard applied for:Cytotoxicity, Complement Activation, Intracutaneous Reactivity,Sensitization, Material Mediated-Pyrogenicity, Acute SystemicToxicity, and Standard in Vivo Thrombogenicity in Canine. ASTMF756 applied for Hemolysis. ASTM F2888 applied for HeparinizedPlatelet and Leukocyte Count.Pass
Shelf life two (2) yearsConfirmation of device functional performance and sterile barrierpouch integrity (seal peel per ASTM F88 and bubble leak perASTM F2096) with accelerated aging and simulated distributionper ASTM 4169-16 and bubble leak (ASTM F2096).Pass

Animal and Clinical data are not needed to support substantial equivalence.

Conclusion

The Guidewire meets performance requirements equivalent to the predicate device. The intended use and technology of the Guidewire is the same as the predicate device.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.