K Number
K213664

Validate with FDA (Live)

Date Cleared
2022-02-11

(81 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Introcan Safety® 2 IV Catheter is inserted into the patient's vascular system for short term use to sample blood, monitor blood pressure, administer fluids and blood intravascularly. The catheters maybe used intravascularly with power injectors at a maximum pressure of 300 psi with a luer lock connection only.

Device Description

The Introcan Safety® 2 IV Catheter consists of an over-the-needle, peripheral catheter made of radiopaque polyurethane, an integrated one directional septum, and a passive safety needle-shielding mechanism. Introcan Safety® 2 is designed to reduce blood exposure at insertion until first connection of an infusion line or luer device to protect clinicians and patients from blood exposure. During needle withdrawal through a septum that seals after the needle has been removed, blood is thus contained within the Introcan Safety® 2 device. The pressure exerted on the needle as it passes through the septum wipes blood from the needle further reducing potential blood exposure. The passive safety needle-shielding mechanism of the Introcan Safety® 2 is located inside the catheter hub. Upon withdrawal of the needle, the safety shield engages as the needle passes through the catheter hub and deploys automatically to shield the needle tip. The safety shield protects during disposal, aiding in the prevention of needlestick injuries. Once the safety shield engages and shields the needle tip, the user is unable to re-insert the needle which aids in the prevention of catheter shearing. This device may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy. The catheters may be used intravascularly with power injectors for which the maximum pressure setting is 300 psi with a luer lock connection only.

AI/ML Overview

The provided text is a 510(k) Summary for the B. Braun Introcan Safety® 2 IV Catheter. This document primarily focuses on establishing substantial equivalence to a predicate device and does not contain information about an AI/ML powered device. As such, information typically found in acceptance criteria and study descriptions for AI/ML devices (like sample sizes for test/training sets, expert qualifications, adjudication methods, or MRMC studies) is not present.

However, I can extract the acceptance criteria and the types of studies performed that validate the device's performance based on the provided FDA 510(k) summary.

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria / Test PerformedDevice Performance / Outcome
Material Composition (Addition of Polyoximethylene)Performance and biocompatibility evaluation demonstrates that the addition of Polyoximethylene does not raise additional questions of safety and effectiveness.
Flowrate through capillary (per ISO 10555-1, Annex E)Successful completion; Performs as intended. Specific values were provided in the comparison table for various sizes and lengths.
Burst Pressure (per ISO 10555-1:2013, Annex F)Successful completion; Performs as intended.
Septum opener retention force (to internal specifications)Successful completion; Performs as intended.
Biocompatibility (per ISO 10993-1)Successful completion; Performs as intended.
Shelf life1 year.
MRI labelingMRI Conditional.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document provides no information regarding sample sizes for test sets. The studies performed were bench tests and biocompatibility evaluations, not clinical studies involving patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable as the studies conducted were non-clinical bench tests and biocompatibility assessments, which do not typically involve human experts establishing ground truth in the way described for AI/ML performance evaluation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable. Adjudication methods are relevant for studies involving subjective human interpretation, which is not the case for the bench testing described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was performed. The device is a physical intravascular catheter, not an AI-powered diagnostic or assistive tool. The document explicitly states, "No clinical testing was performed as these devices does not require clinical studies to demonstrate substantial equivalence with the predicate device."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the bench tests, the "ground truth" consisted of established engineering and medical device standards (e.g., ISO 10555-1, ISO 10993-1) and internal specifications. For instance, flow rate was measured against an expected flow rate for that catheter size, and burst pressure was tested against a standard definition of failure.

8. The sample size for the training set

This is not applicable as there is no training set for a physical medical device.

9. How the ground truth for the training set was established

This is not applicable as there is no training set for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 11, 2022

B. Braun Medical Inc. Tracy Larish Sr. Regulatory Affairs Specialist 901 Marcon Blvd. Allentown, Pennsylvania 18109

Re: K213664

Trade/Device Name: B. Braun Introcan Safety 2 IV Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: January 12, 2022 Received: January 13, 2022

Dear Tracy Larish:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Gang Peng for Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213664

Device Name Introcan Safety® 2 IV Catheter

Indications for Use (Describe)

The Introcan Safety® 2 IV Catheter is inserted into the patient's vascular system for short term use to sample blood, monitor blood pressure, administer fluids and blood intravascularly. The catheters maybe used intravascularly with power injectors at a maximum pressure of 300 psi with a luer lock connection only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K213664. 510(k) SUMMARY

SUBMITTER INFORMATION:

Name:B. Braun Medical Inc.
Address:901 Marcon BoulevardAllentown, PA 18109-9341
Telephone Number:610-266-0500, ext. 2966
Contact Person:Tracy Larish, Sr. Regulatory Affairs Specialist
Telephone Number:(610) 596-2941
Fax Number:(610) 849-9286
Email:tracy.larish@bbraunusa.com
Date Prepared:February 12th, 2022

DEVICE NAME:

Device Trade Name:Introcan Safety® 2 IV Catheter
Common Name:Safety Intravascular Catheter
Classification Name:Catheter, intravascular, therapeutic, short-term less than
30 day. 21 CFR §880.5200: Class II. Product code FOZ

PREDICATE DEVICE:

  • K192676 Introcan Safety® 2 IV Catheter, B. Braun Medical, Inc. ●
    • Product code FOZ o
    • 0 21 CFR §880.5200 Catheter, intravascular, therapeutic, short-term less than 30 day

DEVICE DESCRIPTION

The Introcan Safety® 2 IV Catheter consists of an over-the-needle, peripheral catheter made of radiopaque polyurethane, an integrated one directional septum, and a passive safety needle-shielding mechanism. Introcan Safety® 2 is designed to reduce blood exposure at insertion until first connection of an infusion line or luer device to protect clinicians and patients from blood exposure. During needle withdrawal through a septum that seals after the needle has been removed, blood is thus contained within the Introcan Safety® 2 device. The pressure exerted on the needle as it passes through the septum wipes blood from the needle further reducing potential blood exposure. The passive safety needle-shielding mechanism of the Introcan Safety® 2 is located inside the catheter hub. Upon withdrawal of the needle, the safety shield engages as the needle passes through the catheter hub and deploys automatically to shield the needle tip. The safety shield protects during disposal, aiding in the prevention of needlestick injuries. Once the safety shield engages and shields the needle tip, the user is unable to re-insert the needle which aids in the prevention of catheter shearing. This device may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy. The catheters may be used intravascularly with power injectors for which the maximum pressure setting is 300 psi with a luer lock connection only.

INTENDED USE:

The Introcan Safety® 2 Intravascular Catheter is for short term use to sample blood, monitor blood pressure or administer fluids and blood intravascularly.

{4}------------------------------------------------

INDICATIONS FOR USE:

The Introcan Safety® 2 IV Catheter is inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids and blood intravascularly. The catheters maybe used intravascularly with power injectors at a maximum pressure of 300 psi with a luer lock connection only.

TECHNOLOGICAL CHARACTERISTICS:

The Introcan Safety 2 IV Catheters have the same indications for use, the same intended use, the same principle of operation, the identical safety clip and the same fundamental scientific technology as the predicate device.

The difference between the proposed Introcan Safety® 2 IV Catheter and predicate Introcan Safety® 2 IV Catheter is the septum holder material. This difference does not raise new issues of safety and effectiveness.

Proposed DeviceIntrocan Safety® 2 IV CatheterPredicate Device (K192676)Introcan Safety® 2 IV CatheterComparison
IndicationsThe Introcan Safety® 2 IV Catheter is forshort term use to sample blood, monitorblood pressure, or administer fluids andblood intravascularly and administer fluidssubcutaneously. The catheters may be usedintravascularly with power injectors at amaximum pressure of 300 psi with a luerlock connection only.The Introcan Safety® 2 IV Catheter is forshort term use to sample blood, monitorblood pressure, or administer fluids andblood intravascularly and administer fluidssubcutaneously. The catheters may be usedintravascularly with power injectors at amaximum pressure of 300 psi with a luerlock connection only.Same
ConfigurationSingle Lumen, Tapered Tip, septum withone-time blood controlSingle Lumen, Tapered Tip, septum withone-time blood controlSame
MaterialCompositionPolyurethane, Polypropylene, Stainlesssteel, MABS, Polyisoprene,PolyoximethylenePolyurethane, Polypropylene, Stainlesssteel, MABS, PolyisopreneDifferent:Performance andbiocompatibilityevaluationdemonstrates thatthe addition ofPolyoximethylenedoes not raiseadditionalquestions of safetyand effectiveness.
Catheter Sizes18ga-24ga from 9/16" (14mm) -2" (50mm)18ga-24ga from 9/16" (14mm) -2" (50mm)Same
Gravity FlowRate18ga x 32 mm105 mL/min18ga x 45 mm100 mL/min20ga x 25 mm65 mL/min20ga x 32 mm60 mL/min20ga X 50 mm55ml/min22ga x 25 mm35 mL/min24ga x 14 mm26 mL/min24ga x 19 mm22 mL/min(all flow rates are for both winged andwingless versions)18ga x 32 mm105 mL/min18ga x 45 mm100 mL/min20ga x 25 mm65 mL/min20ga x 32 mm60 mL/min20ga X 50 mm55ml/min22ga x 25 mm35 mL/min24ga x 14 mm26 mL/min24ga x 19 mm22 mL/min(all flow rates are for both winged andwingless versions)Same
SterilizationEthylene OxideEthylene OxideSame

{5}------------------------------------------------

Proposed DeviceIntrocan Safety® 2 IV CatheterPredicate Device (K192676)Introcan Safety® 2 IV CatheterComparison
Shelf life1 year1 yearSame
Bench TestingTesting according to ISO 10555-1:2013,Section 4.10, Annex E. Flowrate throughcapillary, ISO 10555-1:2013 Annex F BurstPressure and Test for septum openerretention forceTesting according to ISO 10555-1:2013,Section 4.10, Annex E. Flowrate throughcapillary and Test for septum openerretention forceDifferent:Bench testingdemonstrated thatthe differences donot raise additionalquestions of safetyand effectiveness
BiocompatibilityclassificationExternally communicating blood pathindirect prolonged contactExternally communicating blood pathindirect prolonged contactSame
MRI labelingMRI ConditionalMRI ConditionalSame

NONCLINICAL TESTING

Bench testing performed on Introcan Safety® 2 IV Catheters demonstrates that the device performs as intended. No clinical testing was performed as these devices does not require clinical studies to demonstrate substantial equivalence with the predicate device. The following testing has been successfully completed for the proposed devices:

  • Flowrate through capillary per ISO 10555-1 Intravascular catheters Sterile and ● single-use intravascular catheters - Part 1: General requirements, Annex E
  • Burst Pressure per ISO 10555-1:2013 Annex F ●
  • Septum opener retention force to internal specifications .
  • Biocompatibility per ISO 10993-1 ●

CONCLUSION:

Results of the testing conducted on the proposed devices demonstrate that the Introcan Safety® 2 IV Catheters are substantially equivalent to the predicate device.

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).