K Number
K213426

Validate with FDA (Live)

Date Cleared
2022-08-16

(299 days)

Product Code
Regulation Number
864.7750
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HemosIL ReadiPlasTin is an in vitro diagnostic thromboplastin reagent, based on recombinant human tissue factor, for the quantitative determination, in human citrated plasma, of Prothrombin Time (PT) and Fibrinogen, on the ACL TOP Family and ACL TOP Family 50 Series of analyzers. The product is intended to be used for the extrinsic coagulation pathway and the monitoring of Oral Vitamin K Antagonist Therapy.

Device Description

The thromboplastin reagent included in the ReadiPlasTin kit, after mixing with the ReadiPlasTin Diluent, is a liposomal preparation that contains recombinant human tissue factor (RTF), re-lipidated in a synthetic phospholipid blend. In the PT test, the addition of the tissue thromboplastin (ReadiPlasTin reagent) to the patient plasma in the presence of calcium ions initiates the activation of the extrinsic pathway. This results ultimately in the conversion of fibrin, with formation of a solid gel. The fibrinogen is quantitated (PT-based method) by relating the absorbance or light-scatter during clotting to a calibrator.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device called "HemosIL ReadiPlasTin." This document doesn't describe an AI/ML-based device, but rather an in vitro diagnostic (IVD) reagent used for Prothrombin Time (PT) and Fibrinogen determination. Therefore, many of the requested categories related to AI/ML device testing (e.g., number of experts for ground truth, adjudication method, MRMC study, training set information) are not applicable to this type of submission.

However, I can extract the relevant information regarding acceptance criteria and performance studies for this IVD device.

Here's a breakdown of the requested information, adapted for an IVD reagent:

Device: HemosIL ReadiPlasTin (In vitro diagnostic thromboplastin reagent)
Purpose: Quantitative determination of Prothrombin Time (PT) and Fibrinogen in human citrated plasma.
Reason for Submission (K213426): Reformulation of the existing HemosIL ReadiPlasTin by adding EDTA as a stabilizer and removing unnecessary fillers.


1. Table of Acceptance Criteria and Reported Device Performance

For this IVD, "acceptance criteria" are typically defined by the method validation standards (e.g., CLSI guidelines) and the equivalence to the predicate device. The performance data presented are the results of meeting these criteria.

Performance Study TypeAcceptance Criteria (Implied/Standard)Reported Device PerformanceComments
Precision (Repeatability & Within-Laboratory)CV% within acceptable limits for PT and Fibrinogen based on CLSI EP05-A3 guidelines and clinical utility.PT (Seconds): Repeatability CV 0.5-1.0%; Within Lab CV 0.8-1.7% (ACL TOP Family); Repeatability CV 0.7-1.1%; Within Lab CV 0.9-1.7% (ACL TOP Family 50 Series) PT (INR): Repeatability CV 0.7-1.7%; Within Lab CV 0.9-2.0% (ACL TOP Family); Repeatability CV 0.8-1.5%; Within Lab CV 1.0-2.0% (ACL TOP Family 50 Series) Fibrinogen (mg/dL): Repeatability CV 0.6-1.8%; Within Lab CV 0.8-2.0% (ACL TOP Family); Repeatability CV 0.6-2.0%; Within Lab CV 0.9-2.3% (ACL TOP Family 50 Series)All reported values fall within the expected range for good precision for coagulation assays. The text explicitly states: "The testing below and on the following pages met all acceptance criteria as follows."
Fibrinogen LinearityResults must support the labeled linearity claim of 60 to 700 mg/dL."The results for all 3 lots on both systems met acceptance criteria, supporting the labeled fibrinogen linearity claim of 60 to 700 mg/dL."Tested across 3 lots on both instrument families per CLSI EP06, 2nd Ed.
InterferenceNo significant interference from specified substances at given concentrations for PT and Fibrinogen measurements.No interference claimed for: - PT: UFH (1.0 IU/mL), LMWH (1.4 IU/mL), Hemoglobin (500 mg/dL), Triglycerides (1000 mg/dL), Bilirubin (Conjugated & Unconjugated) (50 mg/dL), Daptomycin (100 µg/mL) - Fibrinogen: UFH (1.5 IU/mL), LMWH (1.7 IU/mL), Hemoglobin (500 mg/dL), Triglycerides (600 mg/dL), Bilirubin (Conjugated & Unconjugated) (50 mg/dL), Daptomycin (200 µg/mL)New claims for daptomycin and conjugated bilirubin interference were added. The study used 2 clinical sample levels for both PT and Fibrinogen.
Method ComparisonStrong correlation and agreement between the subject device and the predicate device (HemosIL RecombiPlasTin 2G). Slope should be near 1, intercept near 0, and correlation coefficient (r) close to 1.ACL TOP Family: - PT (INR): Slope 1.031 (95% CI 1.009, 1.053), Intercept -0.043 (-0.068, -0.018), r 0.997 - Fibrinogen (mg/dL): Slope 0.975 (95% CI 0.963, 0.986), Intercept 7.171 (3.842, 10.50), r 0.995 ACL TOP Family 50 Series: - PT (INR): Slope 1.021 (95% CI 0.999, 1.043), Intercept -0.034 (-0.060, -0.009), r 0.996 - Fibrinogen (mg/dL): Slope 1.015 (95% CI 1.003, 1.027), Intercept -0.811 (-4.148, 2.527), r 0.994All method comparison results demonstrated excellent correlation and agreement, supporting substantial equivalence.
Open Vial StabilityMaintain performance for 10 days at 2-8°C in closed original vial after preparation."The results support the following labeled open vial stability claim: Once prepared for use, 10 days at 2-8℃ in closed original vial"Tested across 3 lots per CLSI EP25-A.
On-board Instrument StabilityMaintain performance for 10 days at 15°C on the ACL TOP Family and ACL TOP Family 50 Series after preparation."The results support the following labeled on-board instrument stability claim: Once prepared for use, 10 days at 15°C on the ACL TOP Family and ACL TOP Family 50 Series"Tested across 3 lots per CLSI EP25-A.
Real-time Shelf-life StabilityDevice maintains stated performance throughout its claimed shelf-life."The study will continue to a point past final claim." (Ongoing assessment to support shelf-life).Tested across 3 lots per CLSI EP25-A.

2. Sample Size Used for the Test Set and Data Provenance

  • Precision:

    • Sample Size: 80 measurements per instrument per lot (n=80/instrument/lot) for PT (controls + 6 native patient samples) and Fibrinogen (controls + 6 fibrinogen sample pools at 3 levels). Tested across 3 lots and representative members of both ACL TOP Family and ACL TOP Family 50 Series.
    • Data Provenance: Not explicitly stated, but "native (unadulterated) patient samples" and "fibrinogen sample pools" suggest human biological samples. Typically, these studies are conducted in a controlled laboratory setting (Likely within the manufacturer's R&D facilities or a contract research organization). The document is submitted to the US FDA, implying relevance to the US market. The retrospective/prospective nature is generally prospective for these types of validation studies.
  • Interference:

    • Sample Size: Two clinical sample levels each for PT (normal pooled plasma and a high INR clinical sample) and Fibrinogen (normal pooled plasma and a low fibrinogen sample). Specific "n" per concentration/sample type is not given, but refers to CLSI EP07, 3rd Ed and CLSI EP37, 1st Ed which define the study design.
    • Data Provenance: Not explicitly stated, but "clinical sample levels" implies human biological samples.
  • Method Comparison:

    • Sample Size:
      • PT (INR): 160 samples (normal and abnormal) for both ACL TOP Family and ACL TOP Family 50 Series.
      • Fibrinogen (mg/dL): 135 samples for ACL TOP Family, 134 samples for ACL TOP Family 50 Series.
    • Data Provenance: Not explicitly stated, but "normal and abnormal samples" implies human biological samples. The study was "in-house."
  • Open Vial and On-board Instrument Stability:

    • Sample Size: For PT, controls and four native (unadulterated) patient samples were tested in eight replicates at each time interval. For Fibrinogen, controls and six fibrinogen sample pools at three levels were tested in eight replicates at each time interval.
    • Data Provenance: Not explicitly stated, but "native (unadulterated) patient samples" and "fibrinogen sample pools" imply human biological samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This is not applicable for this type of IVD device. The ground truth for chemical assays like PT and Fibrinogen is established through precise measurement methods and reference materials, not through expert consensus of visual or diagnostic interpretations. The "truth" is the quantitative value derived from the reference method or calibrator.


4. Adjudication Method for the Test Set

Not applicable, as this is an IVD reagent and not an AI/ML-based diagnostic system requiring human interpretation or adjudication.


5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable, as this is an IVD reagent and not an AI/ML-based diagnostic system involving human interpretation.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable, as this is an IVD reagent. Its performance is inherent to the chemical reaction and the analytical instrument it is used with.


7. The Type of Ground Truth Used

The "ground truth" for this IVD is established by:

  • Reference Methods/Materials: For PT and Fibrinogen, this would rely on internationally recognized standards and calibrators, and the values obtained from a validated reference method (or the predicate device in method comparison studies).
  • Known Concentrations: For linearity and interference studies, samples are often spiked with known concentrations of analytes or interfering substances.
  • Validated Predicate Device: In the method comparison study, the predicate device (HemosIL RecombiPlasTin 2G) serves as the comparator for verifying the new formulation's performance.

8. The Sample Size for the Training Set

Not applicable. This is an IVD reagent, not an AI/ML model that requires a "training set." The development of the reagent involves chemical formulation and optimization, not data-driven machine learning.


9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no "training set" for an IVD reagent.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 16, 2022

Instrumentation Laboratory Co. Carol Marble Sr. Regulatory Affairs Director 180 Hartwell Road Bedford, Massachusetts 01730

Re: K213426

Trade/Device Name: HemosIL ReadiPlasTin Regulation Number: 21 CFR 864.7750 Regulation Name: Prothrombin Time Test Regulatory Class: Class II Product Code: GJS Dated: October 20, 2021 Received: October 21, 2021

Dear Carol Marble:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Min Wu Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213426

Device Name HemosIL ReadiPlasTin

Indications for Use (Describe)

HemosIL ReadiPlasTin is an in vitro diagnostic thromboplastin reagent, based on recombinant human tissue factor, for the quantitative determination, in human citrated plasma, of Prothrombin Time (PT) and Fibrinogen, on the ACL TOP Family and ACL TOP Family 50 Series of analyzers. The product is intended to be used for the extrinsic coagulation pathway and the monitoring of Oral Vitamin K Antagonist Therapy.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.

Submitter's InformationInstrumentation Laboratory Company180 Hartwell RoadBedford, MA 01730-2443 (USA)
------------------------------------------------------------------------------------------------------------------
Contact PersonCarol MarbleSenior Director of Quality Assurance and Regulatory AffairsPhone: 781-861-4467Fax: 781-861-4207Email: cmarble@werfen.com
----------------------------------------------------------------------------------------------------------------------------------------------------------------------
Preparation DateAugust 16, 2022
-----------------------------------
Predicate DeviceHemosIL ReadiPlasTinK122584
-------------------------------------------------
Regulatory Information
AnalyteRegulationSectionRegulatory DescriptionClassProductCodePanel
Prothrombin Time864.7750Prothrombin time testIIGJS81
Fibrinogen864.7340Fibrinogen determination systemIIGIS

{4}------------------------------------------------

Device Description

The thromboplastin reagent included in the ReadiPlasTin kit, after mixing with the ReadiPlasTin Diluent, is a liposomal preparation that contains recombinant human tissue factor (RTF), re-lipidated in a synthetic phospholipid blend. In the PT test, the addition of the tissue thromboplastin (ReadiPlasTin reagent) to the patient plasma in the presence of calcium ions initiates the activation of the extrinsic pathway. This results ultimately in the conversion of fibrin, with formation of a solid gel. The fibrinogen is quantitated (PT-based method) by relating the absorbance or light-scatter during clotting to a calibrator.

Intended Use / Indications for Use

HemosIL ReadiPlasTin is an in vitro diagnostic thromboplastin reagent, based on recombinant human tissue factor, for the quantitative determination, in human citrated plasma, of Prothrombin Time (PT) and Fibrinogen, on the ACL TOP Family and ACL TOP Family 50 Series of analyzers. The product is intended to be used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Vitamin K Antagonist Therapy.

Reason for Submission

This Traditional 510(k) is being submitted to reformulate HemosIL ReadiPlasTin, adding EDTA as a stabilizer to improve reagent stability and also removing unnecessary fillers used to make acceptable appearing cakes after lyophilization. These filler ingredients are a carryover from a previous generation of lyophilized reagents. Since HemosIL ReadiPlasTin kit components are liquid, these filler ingredients serve no intended purpose.

There were no changes to the following with this submission:

  • Intended use/indications for use .
  • Operating principle .
  • Labeled performance characteristics, except for the addition of daptomycin interference claims ●

{5}------------------------------------------------

Summary Comparison of Technological Characteristics (Predicate)
ItemPredicate (K122584)Subject Device
Similarities
Trade NameHemosIL ReadiPlasTinSame
ManufacturerInstrumentation Laboratory CompanySame
Intended Use/Indications for UseHemosIL ReadiPlasTin is an in vitro diagnostic thromboplastin reagent, based on recombinant human tissue factor, for the quantitative determination, in human citrated plasma, of Prothrombin Time (PT) and Fibrinogen, on the ACL TOP Family and ACL TOP Family 50 Series of analyzers. The product is intended to be used for the evaluation of the extrinsic coagulation pathway and the monitoring of Oral Vitamin K Antagonist Therapy.Same
Test PrincipleProthrombin Time (PT): In the PT test, the addition of the tissue thromboplastin to the patient plasma, in the presence of calcium, initiates the activation of the extrinsic pathway. This results in the conversion of fibrinogen to fibrin, with the formation of a solid gel.PT-derived Fibrinogen: Fibrinogen is quantitated (PT-based method) by relating the absorbance or light scatter during clotting to a calibrator.Same
Sample Type3.2% and 3.8% Citrated PlasmaSame
MeasurementQuantitativeSame
Reporting UnitsPT: Seconds, % Activity, INRFibrinogen: mg/dL, g/LPT: Seconds, INRFibrinogen: mg/dL, g/L
Summary Comparison of Technological Characteristics (Predicate) (Cont.)
ItemPredicate (K122584)Subject Device
Similarities (Cont.)
InstrumentationACL TOP Family K160276ACL TOP Family 50 Series K150877Same
Quality ControlAutomated QCSame
Vial ContentLiquidSame
On-Board Stability10 days at 15°C on the instrumentSame
Open Vial Stability10 days at 2-8°C in closed original vialSame
Differences
FormulationEach ReadiPlasTin kit consists of:ReadiPlasTin Reagent: A solution of recombinant human tissue factor, synthetic phospholipids with stabilizers, preservative and bufferReadiPlasTin Diluent: An aqueous solution of calcium chloride, polybrene and a preservative.Same except the following formulation changes:1. Addition of EDTA to ReadiPlasTin Reagent as a stabilizer for improved stability.2. Removal of bovine gamma globulin (BGG) and trehalose from ReadiPlasTin Reagent and trehalose from ReadiPlasTin Diluent as inactive ingredients (fillers) with no intended purpose in liquid reagents. These ingredients are a carryover from a previous generation of lyophilized reagents.

{6}------------------------------------------------

{7}------------------------------------------------

Performance and Stability Studies

Performance and stability studies were performed to establish the safety and effectiveness of the reformulated HemosIL ReadiPlasTin. These studies included repeatability, fibrinogen linearity, conjugated bilirubin and daptomycin interference, method comparison, open vial and on-board instrument stability and real-time shelf-life stability.

The testing below and on the following pages met all acceptance criteria as follows.

Precision

Repeatability and within laboratory precision was assessed in accordance with CLSI EP05-A3 for 20 days, with 2 runs per day and 2 replicates per run for each sample level (n=80/instrument/lot), using 3 lots of HemosIL ReadiPlasTin on representative members of the ACL TOP Family and ACL TOP Family 50 Series.8 For PT, controls, as well as six native (unadulterated) patient samples, were tested; for fibrinogen, controls, as well as six fibrinogen sample pools at three levels, were tested.

The following tables include PT seconds, INR and fibrinogen precision data for representative systems each with one reagent lot.

Prothrombin Time (PT) Precision
SystemSampleMean PT(Seconds)RepeatabilityCVWithinLaboratory CV
ACL TOP FamilyNormal Control11.40.9%1.1%
Low Abnormal Control22.51.0%1.3%
High Abnormal Control37.50.6%0.8%
Sample 120.80.8%1.1%
Sample 231.40.9%1.7%
Sample 334.50.5%0.8%
Sample 439.60.7%0.8%
Sample 552.11.0%1.7%
Sample 648.30.9%1.4%
SampleMeanINRRepeatabilityCVWithinLaboratory CV
Sample 11.811.0%1.2%
Sample 22.761.7%2.0%
Sample 33.040.7%0.9%
Sample 43.520.7%0.8%
Sample 54.661.2%1.7%
Sample 64.311.2%1.5%

Prothrombin Time (PT) Precision

{8}------------------------------------------------

Prothrombin Time (PT) Precision
SystemSampleMean PT (Seconds)Repeatability CVWithin Laboratory CV
ACL TOP Family50 SeriesNormal Control11.80.7%1.0%
Low Abnormal Control23.30.8%0.9%
High Abnormal Control38.41.0%1.0%
Sample 121.40.7%1.0%
Sample 233.00.8%1.0%
Sample 336.20.7%1.0%
Sample 441.60.7%1.2%
Sample 553.21.1%1.7%
Sample 648.60.9%1.6%
SampleMean INRRepeatability CVWithin Laboratory CV
Sample 11.790.8%1.1%
Sample 22.791.0%1.0%
Sample 33.070.9%1.0%
Sample 43.541.2%1.4%
Sample 54.561.5%2.0%
Sample 64.161.2%1.7%
Fibrinogen (Fib) Precision
SystemSampleMean Fib(mg/dL)RepeatabilityCVWithinLaboratory CV
ACL TOP FamilyNormal Control3291.1%1.2%
Low Abnormal Control1711.2%1.5%
Low Fibrinogen Control1301.8%2.0%
Sample 11111.1%1.2%
Sample 21161.1%1.5%
Sample 33150.6%0.8%
Sample 43380.8%0.9%
Sample 56260.6%0.8%
Sample 66360.8%1.0%

{9}------------------------------------------------

Fibrinogen (Fib) Precision
SystemSampleMean Fib(mg/dL)RepeatabilityCVWithinLaboratory CV
ACL TOP Family50 SeriesNormal Control3191.3%1.4%
Low Abnormal Control1571.0%1.4%
Low Fibrinogen Control1222.0%2.3%
Sample 11061.2%1.4%
Sample 21101.1%1.3%
Sample 33070.9%1.2%
Sample 43300.6%0.9%
Sample 56240.7%1.1%
Sample 66350.8%1.2%

Fibrinogen Linearity

Fibrinogen linearity was assessed in accordance with CLSI EP06, 2nd Ed, using 3 lots of HemosIL ReadiPlasTin on representative members of the ACL TOP Family and ACL TOP Family 50 Series. The results for all 3 lots on both systems met acceptance criteria, supporting the labeled fibrinogen linearity claim of 60 to 700 mg/dL.

Interference

Interference was assessed for conjugated bilirubin and daptomycin in accordance with CLSI EP07, 3d Ed, and CLSI EP37, 1* Ed, using 1 lot of HemosIL ReadiPlasTin on a representative member of the ACL TOP Family. The studies used two clinical sample levels each for PT (normal pooled plasma and a high NR clinical sample at 2.0-3.0 INR) and fibrinogen (normal pooled plasma and a low fibrinogen sample at ~100 mg/dL).

The results from these additional studies, along with the original interference studies under K122584, support the following labeled claims of no interference for:

UFHLMWHHemoglobinTriglyceridesBilirubin(ConjugatedandUnconjugated)Daptomycin
Prothrombin Time (PT)
1.0 IU/mL1.4 IU/mL500 mg/dL1000 mg/dL50 mg/dL100 µg/mL
Fibrinogen
1.5 IU/mL1.7 IU/mL500 mg/dL600 mg/dL50 mg/dL200 µg/mL

{10}------------------------------------------------

Method Comparison

In-house method comparison was performed in accordance with CLSI EP09c on normal and abnormal samples, comparing HemosIL ReadiPlasTin to HemosIL RecombiPlasTin 2G on a representative member of the ACL TOP Family and a representative member of the ACL TOP Family 50 Series, with the following result(s):

SystemAssaynSlope(95% CI)Intercept(95% CI)r
ACL TOPFamilyPT(INR)1601.031(1.009, 1.053)-0.043(-0.068, -0.018)0.997
PT(INR)511.042(1.006, 1.078)0.011(-0.072, 0.094)0.998
Fibrinogen(mg/dL)1350.975(0.963, 0.986)7.171(3.842, 10.50)0.995
ACL TOPFamily 50SeriesPT(INR)1601.021(0.999, 1.043)-0.034(-0.060, -0.009)0.996
PT(INR)511.029(0.992, 1.067)0.023(-0.061, 0.107)0.997
Fibrinogen(mg/dL)1341.015(1.003, 1.027)-0.811(-4.148, 2.527)0.994

Open Vial Stability

Open vial stability was assessed in accordance with CLSI EP25-A, using 3 lots of HemosIL ReadiPlasTin on a representative member of the ACL TOP Family. For PT, controls and four native (unadulterated) patient samples, were tested in eight replicates at each time interval to a point past claim; for fibrinogen, controls and six fibrinogen sample pools at three levels, were tested in eight replicates at each time interval to a point past claim.

The results support the following labeled open vial stability claim:

  • Once prepared for use, 10 days at 2-8℃ in closed original vial -

On-board Instrument Stability

On-board instrument stability was assessed in accordance with CLSI EP25-A, using 3 lots of HemosIL ReadiPlasTin on a representative member of the ACL TOP Family. For PT, controls and four native (unadulterated) patient samples, were tested in eight replicates at each time interval to a point past claim; for fibrinogen, controls and six fibrinogen sample pools at three levels, were tested in eight replicates at each time interval to a point past claim.

The results support the following labeled on-board instrument stability claim:

  • Once prepared for use, 10 days at 15°C on the ACL TOP Family and ACL TOP Family 50 Series

{11}------------------------------------------------

Real-time Shelf-life Stability

Real-time shelf-life stability continues to be assessed in accordance with CLSI EP25-A, using 3 lots of HemosIL ReadiPlasTin on a representative member of the ACL TOP Family. For PT, controls and four native (unadulterated) patient samples were tested in eight replicates at Time 0 and this testing continues at each predefined time interval; for fibrinogen sample pools at three levels, were tested in eight replicates at Time 0 and this testing continues at each predefined time interval. The study will continue to a point past final claim.

Conclusion

Results of the performance and stability studies for HemosIL ReadiPlasTin, with the modified formulation, demonstrate that the subject device is substantially equivalent to the predicate device, HemosIL ReadiPlasTin, last FDA cleared under K122584.

§ 864.7750 Prothrombin time test.

(a)
Identification. A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).(b)
Classification. Class II (performance standards).