K Number
K213363

Validate with FDA (Live)

Date Cleared
2022-04-13

(183 days)

Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CARESCAPE B450 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility.
The CARESCAPE B450 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE B450 is indicated for monitoring of:
· hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, regional oxygen saturation, total hemoglobin concentration, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation),
· respiratory (impedance respiration, airway gases (CO2, O2, N2O, and anesthetic agents), spirometry, gas exchange), and
· neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission).
The CARESCAPE B450 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.
The CARESCAPE B450 is able to detect and generate alarms for ECG arrhythmias: atrial fibrillation, accelerated ventricular rhythm, asystole, bigeminy, bradycardia, ventricular couplet, missing beat, multifocal premature ventricular contractions (PVCs), pause, R on T, supra ventricular tachycardia, trigeminy, ventricular bradycardia, ventricular fibrillation/ventricular tachycardia, ventricular tachycardia, and VT>2. The CARESCAPE B450 also shows alarms from other ECG sources.
The CARESCAPE B450 also provides other alarms, trends, snapshots and calculations, and can be connected to displays, printers and recording devices.
The CARESCAPE B450 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility

Device Description

CARESCAPE B450 is a new version of a portable multiparameter patient monitoring system. The CARESCAPE B450 includes the monitor itself with built-in CPU, power unit, a 12 inch touch display, the CARESCAPE Software and one or two batteries. CARESCAPE B450 is equipped with an ePort interface that supports use of PDM or CARESCAPE ONE patient data acquisition modules for patient monitoring. CARESCAPE B450 is also equipped with one module slot where patient data acquisition modules (E-Modules), can be connected to perform patient monitoring. The CARESCAPE B450 includes features and subsystems that are optional or configurable.

AI/ML Overview

Based on the provided text, here's an analysis of the acceptance criteria and the study that proves the device meets them:

The document describes the CARESCAPE B450, a multiparameter patient monitor. This submission is for a new version of the device, primarily focusing on updated software and minor hardware modifications.

The document does not contain details about specific acceptance criteria for performance metrics (e.g., sensitivity, specificity, accuracy for arrhythmia detection) or a study proving the device meets those criteria with statistical significance. Instead, it primarily focuses on demonstrating substantial equivalence to a predicate device (K191249 CARESCAPE B450) and compliance with general safety and performance standards through non-clinical testing.

Here's a breakdown of the requested information based on the available text:

  1. A table of acceptance criteria and the reported device performance

    This information is not explicitly provided in the document. The submission aims to show that the new CARESCAPE B450, with its updated software and minor hardware, is "substantially equivalent" to its predicate device. This implies that its performance is expected to meet the same standards as the predicate, but specific performance metrics and acceptance thresholds for those metrics are not detailed.

  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided. The document states that "Bench testing related to software, hardware and performance including applicable consensus standards was conducted on the CARESCAPE B450, demonstrating the design meets the specifications." It also notes that "The subject of this premarket submission, CARESCAPE B450 did not require clinical studies to support substantial equivalence." This indicates that the primary validation was through non-clinical bench testing, not through studies on patient data.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided. As clinical studies were not required and the validation was primarily non-clinical bench testing, the concept of "ground truth" derived from expert consensus on patient data (as would be typical for AI/ML performance studies) is not applicable or described in this document.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided. Since no clinical studies or evaluations of diagnostic performance against a "ground truth" established by experts on a test set are detailed, adjudication methods are not mentioned.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done or reported. The device is a patient monitor with arrhythmia detection, not an AI-assisted diagnostic tool for human readers in the context of an MRMC study.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document states that "Bench testing related to software, hardware and performance... was conducted," implying that the device's inherent functional performance was tested. The phrase "algorithm only" isn't explicitly used, but the testing would effectively assess the device's standalone operation. However, no specific performance metrics (like those one would expect for an AI algorithm, e.g., sensitivity/specificity for specific arrhythmias) are reported. The device features "EK-Pro arrhythmia detection algorithm EK-Pro V14", and its performance is assumed to be equivalent to the predicate using the same algorithm version.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical bench testing, the "ground truth" would likely be based on established engineering specifications, simulated physiological signals, and validated test protocols inherent to medical device performance testing, rather than expert consensus, pathology, or outcomes data from human subjects. This type of detail is not further elaborated in the document.

  8. The sample size for the training set

    This information is not provided. As the submission is for a new version of an existing device primarily involving software updates and minor hardware changes, and the algorithm (EK-Pro V14) itself is listed as "Identical" to the predicate, details about a training set for a new or significantly retrained algorithm are not discussed.

  9. How the ground truth for the training set was established

    This information is not provided, for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 13, 2022

GE Healthcare Finland Oy Joel Kent Senior Regulatory Affairs Manager Kuortaneenkatu 2 Helsinki, Helsinki FI-00510 Finland

Re: K213363

Trade/Device Name: CARESCAPE B450, E-musb Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, BZK, BZL, BZQ, CAP, CBQ, CBR, CBS, CCK, CCL, DPS, OEM, DOA, DOK, DRT, DSI, DSJ, DSK, DXG, DXN, FLL, GWJ, GWQ, KOI, KRB, MLD, MUD, DPZ, NHO, NHP, NHQ, OLT, OLW, ORT, OMC Dated: March 11, 2022 Received: March 14, 2022

Dear Joel Kent:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213363

Device Name CARESCAPE B450

Indications for Use (Describe)

The CARESCAPE B450 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility.

The CARESCAPE B450 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE B450 is indicated for monitoring of:

· hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, regional oxygen saturation, total hemoglobin concentration, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation),

· respiratory (impedance respiration, airway gases (CO2, O2, N2O, and anesthetic agents), spirometry, gas exchange), and

· neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission).

The CARESCAPE B450 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.

The CARESCAPE B450 is able to detect and generate alarms for ECG arrhythmias: atrial fibrillation, accelerated ventricular rhythm, asystole, bigeminy, bradycardia, ventricular couplet, missing beat, multifocal premature ventricular contractions (PVCs), pause, R on T, supra ventricular tachycardia, trigeminy, ventricular bradycardia, ventricular fibrillation/ventricular tachycardia, ventricular tachycardia, and VT>2. The CARESCAPE B450 also shows alarms from other ECG sources.

The CARESCAPE B450 also provides other alarms, trends, snapshots and calculations, and can be connected to displays, printers and recording devices.

The CARESCAPE B450 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility

Contraindications for using the monitor

The CARESCAPE B450 is not intended for use in a controlled MR environment.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

GE Healthcare Finland Oy Kuortaneenkatu 2
00510 Helsinki Finland T: +358 10 39411

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Owner/Contact/Date (807.92(a)(1)):

Date:October 12, 2021
Owner/Submitter:GE Healthcare Finland Oy.Kuortaneenkatu 200510 HelsinkiFINLANDPhone: +358 10 39411
Primary Contact Person:Joel KentSenior Regulatory Affairs ManagerGE HealthcarePhone: 617-851-0943E-mail: joel.kent@ge.com
Secondary Contact Person:Karin MårtensonRegulatory Affairs LeaderGE Healthcare Finland OyKuortaneenkatu 200510 HelsinkiFinlandPhone: + 358 50 384 6646E-mail: karin.martenson@ge.com
Device names (807.92(a)(2)):

Trade Name: CARESCAPE B450 Common/Usual Name: Multiparameter patient monitor (monitor, physiological, patient (with arrhythmia detection or alarms))

{5}------------------------------------------------

Classification Names: 21 CFR 870.1025 Arrhythmia detector and alarm (including STsegment measurement and alarm) 21 CFR 868.1850 spirometer, monitoring (w/wo alarm) 21 CFR 868.1730 computer, oxygen-uptake 21 CFR 868.2375 monitor, breathing frequency 21 CFR 868.2600 monitor, airway pressure (includes gauge and/or alarm) 21 CFR 868.1500 analyzer, gas, enflurane, gaseous-phase (anesthetic conc.) 21 CFR 868.1500 analyzer, gas, desflurane, gaseous-phase (anesthetic conc.) 21 CFR 868.1500 analyzer, gas, sevoflurane, gaseous-phase (anesthetic conc.) 21 CFR 868.1500 analyzer, gas, isoflurane, gaseous-phase (anesthetic conc.) 21 CFR 868.1700 analyzer, gas, nitrous-oxide, gaseous phase (anesthetic conc.) 21 CFR 868.1620 analyzer, gas, halothane, gaseous-phase (anesthetic conc.) 21 CFR 868.1400 analyzer, gas, carbon-dioxide, gaseous-phase 21 CFR 868.1720 analyzer, gas, oxygen, gaseous-phase 21 CFR 870.2340 electrocardiograph 21 CFR 870.2710 oximeter, ear 21 CFR 870.2700 oximeter 21 CFR 870.1425 computer, diagnostic, programmable 21 CFR 870.2300 monitor, cardiac (incl. cardiotachometer & rate alarm) 21 CFR 870.1025 detector and alarm, arrhythmia 21 CFR 870.1100 alarm, blood-pressure 21 CFR 870.1110 computer, blood-pressure 21 CFR 870.1435 computer, diagnostic, pre-programmed, single-function 21 CFR 870.1130 system, measurement, blood-pressure, noninvasive 21 CFR 870.2910 thermometer, electronic, clinical 21 CFR 882.1900 stimulator, auditory, evoked response 21 CFR 882.1400 full-montage standard electroencephalograph 21 CFR 870.2700 Oximeter, Tissue Saturation 21 CFR 868.2775 stimulator, nerve, peripheral, electric 21 CFR 870.1915 probe, thermodilution 21 CFR 870.1025 monitor, st segment with alarm 21 CFR 882.1400 non-normalizing quantitative electroencephalograph software 21 CFR 882.1400 index-generating electroencephalograph software 21 CFR 882.1400 reduced- montage standard electroencephalograph 21 CFR 882.1400 burst suppression detection software for electroencephalograph 21 CFR 870.2700 Cerebral Oximeter

{6}------------------------------------------------

Product Code: MHX

Subsequent Product Codes:BZK, BZL, BZQ, CAP, CBQ, CBR, CBS, CCK, CCL, DPS, DPZ, DQA, DQK, DRT, DSI, DSJ, DSK, DXG, DXN, FLL, GWJ, GWQ, KOI, KRB, MLD, MUD, NHO, NHP, NHQ, OLT, OLW, OMC, ORT, QEM
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Predicate Device(s) The primary predicate for this submission is K191249 CARESCAPE B450 (807.92(a)(3)):

Additional predicates / reference devices: K182868, INVOS PM7100 Patient Monitor, INVOS Adult rSO2 Sensor K110028, MASIMO RADICAL Y PULSE CO-OXIMETER

Device Description (807.92(a)(4)):

CARESCAPE B450 is a new version of a portable multiparameter patient monitoring system. The CARESCAPE B450 includes the monitor itself with built-in CPU, power unit, a 12 inch touch display, the CARESCAPE Software and one or two batteries. CARESCAPE B450 is equipped with an ePort interface that supports use of PDM or CARESCAPE ONE patient data acquisition modules for patient monitoring. CARESCAPE B450 is also equipped with one module slot where patient data acquisition modules (E-Modules), can be connected to perform patient monitoring. The CARESCAPE B450 includes features and subsystems that are optional or configurable.

Intended Use: (807.92(a)(5)):

Indications for Use (from labeling)

The CARESCAPE B450 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility.

The CARESCAPE B450 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time.

The CARESCAPE B450 is indicated for monitoring of:

· hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, regional oxygen saturation, total hemoglobin concentration, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation),

{7}------------------------------------------------

· respiratory (impedance respiration, airway gases (CO2, O2, N2O, and anesthetic agents), spirometry, gas exchange), and

· neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission).

The CARESCAPE B450 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.

The CARESCAPE B450 is able to detect and generate alarms for ECG arrhythmias: atrial fibrillation, accelerated ventricular rhythm, asystole, bigeminy, bradycardia, ventricular couplet, irregular, missing beat, multifocal premature ventricular contractions (PVCs), pause, R on T, supra ventricular tachycardia, tachycardia, trigeminy, ventricular bradycardia, ventricular fibrillation/ventricular tachycardia, ventricular tachycardia, and VT>2. The CARESCAPE B450 also shows alarms from other ECG sources.

The CARESCAPE B450 also provides other alarms, trends, snapshots and events, and calculations, and can be connected to displays, printers and recording devices.

The CARESCAPE B450 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.

Contraindications for using the monitor

The CARESCAPE B450 is not intended for use in a controlled MR environment.

Technology (807.92(a)(6)): The CARESCAPE B450 incorporates updated software and minor modifications to the hardware.

The fundamental function and operation of the proposed CARESCAPE B450 monitor are unchanged compared to CARESCAPE B450 (K191249).

A summary of the main changes compared to the predicate are listed below in the comparison table.

{8}------------------------------------------------

SpecificationCARESCAPE B450(K191249)ProposedCARESCAPE B450Differences
Patient typeAdult, pediatric &neonatalAdult, pediatric &neonatalIdentical
UseenvironmentsWithin a professionalhealthcare facility (Notintended for MRI)Within a professionalhealthcare facility (Notintended for MRI)Identical
IntrahospitaltransportIntrahospital transportwithin a professionalhealthcare facility.Intrahospital transportwithin a professionalhealthcare facility.Identical
Size (H x W xD) & Weight290 x 310 x 160 mm(11.4 x 12.2 x 6.3 in) (H xW x D) and weight 4.7 kg(10.4 lbs) with twobatteries inserted butwithout modules inserted.290 x 310 x 160 mm(11.4 x 12.2 x 6.3 in) (H xW x D) and weight 4.7 kg(10.4 lbs) with twobatteries inserted butwithout modules inserted.Identical
Module HousingOne E-module slot andoptional recorder. Oneslide mount foracquisition module.One E-module slot andoptional recorder. Oneslide mount foracquisition module.Identical
Display/screenIntegrated 12" TFTcolorLCD DisplayIntegrated 12" TFTcolorLCD DisplayIdentical
Waveforms andparameterwindowsStandard view: Up to 6individual waveforms andup to 16 parameterwindows, if horizontalparameter area turned on.Standard view: Up to 6individual waveforms andup to 16 parameterwindows, if horizontalparameter area turned on.Identical
ModulesE-BIS, E-COP, E-COPSv, E-PiCCO, E-EEGX, E-Entropy, E-Masimo, E-miniC, E-NMT, E-NSATX, E-PP,E-PT, E-sCAiO , E-sCAiOV, E-sCAiOVX,E-sCO, E-sCOV, E-sCOVX, PDM,CARESCAPE ONEE-BIS, E-COP, E-COPSv, E-PiCCO, E-EEGX, E-Entropy, E-Masimo, E-miniC, E-NMT, E-NSATX, E-PP,E-PT, E-sCAiO , E-sCAiOV, E-sCAiOVX,E-sCO, E-sCOV, E-sCOVX, PDM,CARESCAPE ONEIdentical
Optional systemcomponents-Remote Control-CARESCAPE D19KTVER01 optional display-CARESCAPE RAD,Remote Alarm Device-Keyboard-Mouse-Barcode scanner-Laser Printer-Remote Control-CARESCAPE D19KTVER01 optional display-CARESCAPE RAD,Remote Alarm Device-Keyboard-Mouse-Barcode scanner-Laser Printer-E-musb Interface moduleEquivalentAdded E-musb Interface modulethat provides a communication pathfor the OEM owned devicesCARESCAPE rSO2 – INVOS andCARESCAPE CO2 – Microstreamto the CARESCAPE Bx50 hostmonitors
AvailablemeasurementparametersECG, ST segment,arrhythmia detection,ECG diagnostic analysisand measurement,invasive pressure, non-invasive blood pressure,pulse oximetry, cardiacoutput (thermodilutionand pulse contour),temperature, mixedvenous oxygen saturation,and central venousoxygen saturation,impedance respiration,airway gases (CO2, O2,N2O, and anestheticagents), spirometry, gasexchange,electroencephalography,Entropy, Bispectral Index(BIS), neuromusculartransmission.ECG, ST segment,arrhythmia detection,ECG diagnostic analysisand measurement,invasive pressure, non-invasive blood pressure,pulse oximetry, regionaloxygen saturation, totalhemoglobinconcentration , cardiacoutput (thermodilutionand pulse contour),temperature, mixedvenous oxygen saturation,and central venousoxygen saturation,impedance respiration,airway gases (CO2, O2,N2O, and anestheticagents), spirometry, gasexchange,electroencephalography,Entropy, Bispectral Index(BIS), neuromusculartransmission.Equivalent
EK-ProarrhythmiadetectionalgorithmEK-Pro V14EK-Pro V14Identical
PrintingBuilt-in or central andnetworked laser printerPrintings for waveforms,alarms waveforms,numeric trends.Built-in or central andnetworked laser printerPrintings for waveforms,alarms waveforms,numeric trends.Identical
MountingoptionsMultiple GCX mountingsystems, roll stand, QuickMount, Bed Rail Hook,BRHMultiple GCX mountingsystems, roll stand, BedRail Hook, BRHEquivalentQuick Mount is obsolete.
AlarmsAlarm management corefunctionalities:Classification andnotification of alarmsAdjustment of alarmsettingsPossibility to set criticalalarm limitsAlarm On/Offfunctionality and audiosilencingAlarm management corefunctionalities:Classification andnotification of alarmsAdjustment of alarmsettingsPossibility to set criticalalarm limitsAlarm On/Offfunctionality and audiosilencingIdentical
Battery operationRechargeable Lithium-Ion batteriesRechargeable Lithium-Ion batteriesIdentical
NetworkingcapabilityCARESCAPE NetworkLAN/VLANOptional WLANCARESCAPE NetworkLAN/VLANOptional WLANIdentical
Networkinterface10baseT, 100baseT,802.11 abgn, IEEE802.11r fast roaming issupported.10baseT, 100baseT,802.11 abgn, IEEE802.11r fast roaming issupported.Identical

Subject Device and Predicate Device Comparison

{9}------------------------------------------------

{10}------------------------------------------------

Determination ofSubstantial Equivalence(807.92(b)(1)):Summary of Non-Clinical Tests:
Bench testing related to software, hardware and performanceincluding applicable consensus standards was conducted on theCARESCAPE B450, demonstrating the design meets thespecifications.
The hardware bench testing included electromagneticcompatibility, electrical safety, environmental, WLAN, andusability.
Software testing included software design, development,verification, validation and traceability.
Clinical (807.92(b)(2)):Summary of Clinical Tests:
The subject of this premarket submission, CARESCAPE B450did not require clinical studies to support substantial equivalence.
Conclusion (807.92(b)(3)):GE Healthcare considers the CARESCAPE B450 to be as safe, aseffective, and the performance to be substantially equivalent tothe predicate device.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.