K Number
K213347

Validate with FDA (Live)

Date Cleared
2022-01-25

(109 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ADI Medical Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The ADI Medical Surgical Face Masks are single use, three-layer, flat -folded masks with ear loops and nose clamp. The ADI Medical Surgical Face Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose clamp in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of malleable polyethylene wire. ADI Medical Surgical Face Masks will be provided in blue. ADI Medical Surgical Face Masks are sold non-sterile and are intended to be single use, disposable devices.

AI/ML Overview

The provided text describes the regulatory filing for a surgical face mask, not an AI-powered diagnostic device. Therefore, the questions related to AI device performance, expert ground truth, MRMC studies, and training/test set sample sizes are not directly applicable.

However, I can extract the information relevant to the device's acceptance criteria and the study proving it meets those criteria based on the provided document.


Device: ADI Medical Surgical Face Mask
Regulatory Filing: K213347

Acceptance Criteria and Reported Device Performance

The acceptance criteria for the ADI Medical Surgical Face Mask are defined by various ASTM and EN standards for surgical masks. The performance tests were conducted to demonstrate that the device meets these standards.

Table of Acceptance Criteria and Reported Device Performance:

Performance Test (Standard)Acceptance CriteriaReported Device Performance (ADI Medical Surgical Face Mask)Result
Fluid Resistance (ASTM F1862)29 out of 32 samples pass at 120 mmHg94 out of 96 samples pass at 120 mmHg (from 3 non-consecutive lots)PASS
Particulate Filtration Efficiency (ASTM F2299)≥ 98%Average: 99.47% (96 out of 96 samples pass)PASS
Bacterial Filtration Efficiency (ASTM F2101)≥ 98%Average: 99.48% (96 out of 96 samples pass)PASS
Differential Pressure (Delta P) (EN 14683 Annex C)< 6.0 mmH2O/cm²Average: 3.52 mmH2O/cm² (96 out of 96 samples pass)PASS
Flammability (16 CFR 1610)Class 1Class 1PASS
Biocompatibility: CytotoxicityNon-CytotoxicNon-CytotoxicPASS
Biocompatibility: IrritationNon-IrritatingNon-IrritatingPASS
Biocompatibility: SensitizationNon-SensitizingNon-SensitizingPASS

Details of the Study Proving Device Meets Acceptance Criteria

Given that this is a physical medical device (surgical face mask) and not an AI/software as a medical device (SaMD), many of the typical questions for an AI study are not applicable. Below are the relevant answers based on the provided document.

  1. A table of acceptance criteria and the reported device performance: (See table above)

  2. Sample size used for the test set and the data provenance:

    • Sample Size: The non-clinical tests were conducted on 3 non-consecutive lots, with each lot containing 32 samples. This means a total of 96 samples (3 lots * 32 samples/lot) were tested for most performance criteria (Fluid Resistance, PFE, BFE, Differential Pressure, Flammability). Biocompatibility testing is typically done on specific extracts or samples, not necessarily a batch of 96.
    • Data Provenance: Not specified in terms of country of origin for the data. The tests are "non-clinical tests" meaning laboratory assessments. The document doesn't indicate if they were prospective or retrospective in the sense of a clinical trial; rather, they are standard product performance verification tests.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable. The "ground truth" for a surgical mask's performance is established by the specified ASTM/EN standards and test methods, not by human expert interpretation or consensus. The tests involve objective measurements (e.g., pressure, filtration efficiency, flammability rating).
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. As described above, the test results are objective measurements against defined standards, not subjective interpretations requiring adjudication.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a physical device, not an AI-driven medical imaging or diagnostic device. No human reader studies (with or without AI assistance) were part of this premarket notification. The document explicitly states: "No clinical study is included in this submission."
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI algorithm. Its performance is inherent to its physical properties and material construction.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" (or more accurately, the performance benchmark/standard) for this device is based on established national and international performance standards (ASTM F1862, ASTM F2299, ASTM F2101, EN 14683 Annex C, 16 CFR 1610) and recognized biocompatibility standards (e.g., ISO 10993 for Cytotoxicity, Irritation, Sensitization). These standards define the acceptable physical and biological properties of surgical masks.
  8. The sample size for the training set:

    • Not applicable. This is not an AI device that requires a training set.
  9. How the ground truth for the training set was established:

    • Not applicable. Since there is no training set for an AI model, there is no "ground truth" for it to be established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

January 25, 2022

Asia Dynamics Inc., d/b/a ADI Medical Timothy Kania Official Correspondant Mdi Consultants Inc. 55 Northen Blvd, Suite 200 Great Neck. New York 11021

Re: K213347

Trade/Device Name: ADI Medical Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: December 27, 2021 Received: December 29, 2021

Dear Timothy Kania:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213347

Device Name ADI Medical Surgical Face Masks

Indications for Use (Describe)

ADI Medical Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material.

These face masks are intended for use in infection control practices to reduce the posure of the wearer to blood and body fluids.

This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

1. Submitter's Identification:

Applicant:Asia Dynamics Inc., d/b/a ADI Medical
Address:1565 South Shields Drive Waukegan, IL 60085
Contact Person:Jonathan Zhu
Tel:(847)-688-9968
Email:Jon@adimedical.com
Date Summary Prepared:December 27, 2021
Official Correspondent:Mr. Tim Kania
Mdi Consultants, Inc.
Address:
Tel:732-796-4565
Email:tim@mdiconsultants.com

Name of the Device: ನ

Proprietary Name:ADI Medical Surgical Face Mask
Classification Name:Mask,Surgical
Common Name:Surgical Mask
Regulatory Class:II
Product Code:FXX
Regulation Name:Surgical Apparel
Regulation Number:878.4040
Review Panel:General Hospital

Information for the 510(k) Cleared Device (Predicate Device): ನೆ

Disposable Medical Face Mask, K202513

Surgical Face Mask

Shenzhen Jinko Industrial Co., Ltd..

No reference devices were used in this submission.

4. Device Description:

The ADI Medical Surgical Face Masks are single use, three-layer, flat -folded masks with ear loops and nose clamp. The ADI Medical Surgical Face Masks are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose clamp in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of malleable polyethylene wire. ADI Medical Surgical Face Masks will be

{4}------------------------------------------------

provided in blue. ADI Medical Surgical Face Masks are sold non-sterile and are intended to be single use, disposable devices.

5. Indications for Use:

ADI Medical Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Technological Characteristics Comnarison: la

DeviceProposed DevicePredicate DeviceResult
ManufacturerAsia Dynamics Inc., d/b/a ADI MedicalShenzhen Jinko Industrial Co., Ltd.-
510K numberK213347K202513-
Model NameADI Medical Surgical Face MaskDisposable Medical Face MaskSimilar
Model No.76120FM-04-
ClassificationClass II Device, FXX (21 CFR878.4040)Class II Device, FXX (21 CFR878.4040)Same
Intend useADI Medical Surgical Masks are intended to be worn to protect both the patientand healthcare personnel from transfer of microorganisms, body fluids and particulate material.These face masks are intended for use in infection control practices to reduce the potential exposure of the wearer to blood and body fluids.This is a single use, disposable device(s), provided non-sterile.The Medical Masks are intended to be worn to protect both the patientand healthcare personnel from transfer of microorganisms, body fluids and particulate material.These face masks are intended for use in infection control practices to reduce the potential exposure of thewearer to blood and body fluids.This is a single use, disposable device(s), provided non-sterile.Same
ModelEar Loops, Flat Pleated, 3 layersEar Loops, Flat Pleated, 3 layersSame
MaterialOuter facing layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Middle layerMelt blown polypropylene filterMelt blown polypropylene filterSame
Inner Facing layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Nose ClampMalleable polypropylene with steel wireMalleable polypropylene with steel wireSame
Ear LoopsPolyester, SpandexPolyester, SpandexSame
ColorBlueCAS#147-14-8Variety (include blue)Similar
Dimension(Length)$175 \pm 2.5$ mm$175 \pm 8$ mmSimilar
Dimension(Width)$95 \pm 2.5$ mm95+4.5 mmSimilar
OTC useYesYesSame
SterilityNon-SterileNon-SterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
ASTM F2100 LevelLevel 2Level 2Same
Biocompatibility(ISO10993)Non-Cytotoxic, Non-Sensitizing, Non-IrritatingNon-Cytotoxic, Non-Sensitizing, Non-IrritatingSame

Table 1: Comparison to Predicate Device

{5}------------------------------------------------

{6}------------------------------------------------

7.Summary of Non-Clinical Tests:

The following testing was conducted to demonstrate that the subject device (3 nonconsecutive lots with each lot containing 32 samples) met the acceptance criteria and specification of the standard shown below.

StandardResults of tests onsubject devicePass CriteriaResults
Fluid ResistancePerformanceASTMF186294 out of 96 pass at 120 mmHg, 3 non-consecutive lots tested29 out of 32 pass at 120 mmHgPASS
Particulate FiltrationEfficiencyASTM F229996 out of 96 passAverage: 99.47%$\ge$ 98%$\ge$ 98%PASS
Bacterial FiltrationEfficiencyASTM F210196 out of 96 passAverage: 99.48%$\ge$ 98%$\ge$ 98%PASS
Differential Pressure (DeltaP)EN 14683 Annex C96 out of 96 passAverage: 3.52< 5.0mmH2O/cm2< 6.0mmH2O/cm2PASS
Flammability 16 CFR 1610Class 1Class 1PASS
Table 2: Performance testing
----------------------------------

Table 3: Biocompatibility Testing

ItemProposed deviceAcceptance CriteriaResult
CytotoxicityUnder theconditions of thestudy, thedevice is non-cytotoxic.Non-CytotoxicPASS
IrritationUnder theconditions of thestudy, the deviceis non-irritating.Non-IrritatingPASS
SensitizationUnder theconditions of thestudy, the deviceis non-sensitizingNon-SensitizingPASS

8.Discussion of Clinical Tests Performed:

No clinical study is included in this submission.

9.Conclusions:

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(k) submission K213347, the ADI Medical Surgical Masks, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K202513.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.