K Number
K213168

Validate with FDA (Live)

Date Cleared
2022-01-11

(105 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A medical glove is a disposable device intended for medical purposes that is worn on hand of the examiner's hand to prevent contamination between the patient and the examiner.

Device Description

The proposed device is a powder free medical glove. The device is blue in color. The device meets the requirements of ASTM D6319-19: Standard specification for Nitrile Examination Gloves for Medical Application. The proposed gloves are available in five sizes, which are XS, S, M, L, XL, it could be selected by the user depended on size of hand. The difference between each size is just in the dimension. The proposed device is provided non-sterile.

AI/ML Overview

The provided text describes the acceptance criteria and the study conducted for "Medical Examination Gloves" (K213168). However, it does not describe a study involving an AI/Machine Learning device or a multi-reader multi-case (MRMC) comparative effectiveness study, but rather the performance testing for a physical medical device (gloves). Therefore, many of the requested points related to AI/ML device testing will not be applicable and will be marked as "N/A".

Here's the breakdown of the acceptance criteria and study as described in the document:

1. A table of acceptance criteria and the reported device performance

Test MethodologyPurposeAcceptance CriteriaReported Device Performance (Results)
ASTM D6319-19, ASTM D3767-03 (2020)Physical DimensionsWidth:XS: 70±10mmS: 80±10mmM: 95±10mmL: 110±10mmXL: 120±10mmLength:XS: 220mm minS: 220mm minM: 230mm minL: 230mm minXL: 230mm minThickness:Palm: 0.05mm minFinger: 0.05mm minPass
ASTM D6319-19, ASTM D412-16Physical PropertiesBefore Aging:Tensile Strength: 14MPa minUltimate Elongation: 500% minAfter Aging:Tensile Strength: 14MPa minUltimate Elongation: 400% minPass
ASTM D5151-19Freedom from HolesBe free from holes when tested in accordance with ASTM D5151Pass
ASTM D6124-06Powder free residueLess than 2mg per glovePass
ISO 10993-10:2010IrritationNon-IrritatingPass
ISO 10993-10:2010SensitizationNon-SensitizingPass
ISO 10993-11:2009System toxicityNon-toxicityPass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the exact sample sizes for each physical test. It merely states that "Non clinical tests were conducted to verify that the proposed device met all design specifications and acceptance criteria in the test methodology and the standards." The provenance of the data (country of origin, retrospective/prospective) is not explicitly stated, but given that the manufacturer is in China and the testing is for FDA clearance, it's reasonable to infer the tests were conducted or overseen by the manufacturer. The tests are prospective in nature, as they are performed on the device to prove its compliance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as the device is a physical medical glove undergoing performance testing against standardized physical and biocompatibility criteria, not an AI/ML device requiring expert ground truth for image/data interpretation. Ground truth is established by the test methodologies themselves (e.g., measuring dimensions, tensile strength, etc.).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are typically for subjective assessments or disagreements in interpretative tasks, which do not apply to the objective physical and chemical testing of medical gloves.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This study is for a physical medical glove, not an AI/ML device. Therefore, an MRMC comparative effectiveness study was not conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm, but a physical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is derived from standardized test methodologies and specifications (e.g., ASTM D6319-19 for dimensions and physical properties, ASTM D5151-19 for holes, ASTM D6124-06 for powder residue, and ISO 10993-10/11 for biocompatibility). These standards define the objective measurements and criteria that the device must meet.

8. The sample size for the training set

Not applicable. This is a physical device, not an AI/ML model that requires a training set.

9. How the ground truth for the training set was established

Not applicable. There is no training set for a physical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

January 11, 2022

Shandong Huiwosheng Health Technology Co.,Ltd Haitao Wu QA Director 50m East Of Chaoyang Road, Zhanqian Street, Linyi Chemical Industrial Park, Linzi Town Dezhou. Shandong 250000 China

Re: K213168

Trade/Device Name: Medical Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: August 18, 2021 Received: September 28, 2021

Dear Haitao Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K213168

Device Name Medical Examination Gloves

Indications for Use (Describe)

A medical glove is a disposable device intended for medical purposes that is worn on hand of the examiner's hand to prevent contamination between the patient and the examiner.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

PSC Publishing Services (301) 443-6740

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K213168

    1. Date of Preparation: 01/11/2022
    1. Sponsor Identification

Shandong Huiwosheng Health Technology Co.,Ltd

50m east of Chaoyang Road, Zhanqian street, Linyi Chemical Park, Linzi Town, Linyi County, Dezhou City, Shandong Province, China

Establishment Registration Number: Not yet registered

Contact Person: Haitao Wu QA Director Tel: +86-18865233273 Email: haitao@kbt-cn.com

    1. Identification of Proposed Device
      Trade Name: Medical Examination Gloves Common Name: POWDER FREE EXAMINATION GLOVES

Regulatory Information

Classification Name: polymer patient examination glove Classification: I; Product Code: LZA; Regulation Number: 21CFR 880.6250 Review Panel: General Hospital;

Indication for Use:

A medical glove is a disposable device intended for medical purposes that is worn on hand of the examiner's hand to prevent contamination between the patient and the examiner.

Device Description

The proposed device is a powder free medical glove. The device is blue in color. The device meets the requirements of ASTM D6319-19: Standard specification for Nitrile Examination Gloves for Medical Application. The proposed gloves are available in five sizes, which are XS, S, M, L, XL, it could be

{4}------------------------------------------------

selected by the user depended on size of hand. The difference between each size is just in the dimension. The proposed device is provided non-sterile.

    1. Identification of Predicate Device
      510(k) Number: K031384

Product Name: MULTIPLE PRIVATE LABELED, NON-STERILE, POWDER-FREE, POLYURETHANE WHITE COLOR, EXAMINATION GLOVES

  • ર. Technological Characteristics Comparison
ITEMProposed DevicePredicate DeviceRemark
Product CodeLZALZASame
Regulation Number21 CFR 880.625021 CFR 880.6250Same
ClassIISame
Indication for useA medical glove is a disposable device intended for medical purposes that is worn on hand of the examiner's hand to prevent contamination between the patient and the examiner.This is a medical glove to be worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patient.Same
MaterialPolyurethanePolyurethaneSame
ColorBlueWhiteAnalysis1
SterilityNon-sterileNon-sterileSame
Single-useYesYesSame
SizeXS, S, M, L, XLXS, S, M, L, XLSame
Dimensions(ASTM D6319-19)Width
XS70±10mmXS
S80±10mmS
M95±10mmM
L110±10mmL
XL120±10mmXL
Length
XS220mm minXS
S220mm minS
M230mm minM
  • Table 1 Comparison of Technology Characteristics

{5}------------------------------------------------

L230mm minLall sizes
XL230mm minXL
Thickness
Palm0.05mm minPalmMin 0.13mm
Finger0.05mm minFingerMin 0.11mm
Before Aging
Tensile Strength16.2 - 19.7MPaTensile Strength16 – 23MPaAnalysis3
Physical Properties(ASTM D6319-19And ASTM D412-16)UltimateElongation642% - 754%UltimateElongation600% — 800%
After Aging
Tensile Strength15.2 - 19.1MPaTensile Strength20 — 23MPa
UltimateElongation552% - 662%UltimateElongation600% – 720%
Powder free residue
(ASTM D6319-19And ASTM D6124-06)Less than 2mg per gloveBelow 2mg/gloveSame
Freedom from Holes(ASTM D5151-19)Meet AQL 2.5Pass GIAQL=2.5Same
Biocompatibility
Skin IrritationNo IrritationNo Irritation
SensitizationNo SensitizationNo SensitizationAnalysis4
System ToxicityNo Toxicity/

Analysis 1- Color

The color of the proposed device is different from the predicate device.

Analysis 2- Dimensions

The dimension of the proposed device is not exactly same as the predicate device. The user can select appropriate model depended on size of user's hand.

Analysis 3- Physical Properties (ASTM D6319-19 and ASTM D412-16) The physical properties of the proposed device are not exactly same as the predicate device.

Analysis 4- Biocompatibility

The biocompatibility test item of the proposed device is different from the predicate device.

  • Non-Clinical Test Conclusion 6.
    Non clinical tests were conducted to verify that the proposed device met all design specifications and

{6}------------------------------------------------

acceptance criteria in the test methodology and the standards. The test results demonstrated that the proposed device complies with the following standards and test methodology:

  • ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for MedicalApplication

  • ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves

  • ASTM D3767-03 (2020) Standard Practice for Rubber-Measurement of Dimensions

  • ASTM D412-16 Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers-Tension

  • A ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
  • ISO 10993-10:2010 Biological Evaluation of Medical Devices-Part 10: Tests for Irritationand Skin Sensitization;

  • A ISO 10993-11:2017 Biological evaluation of medical devices-Part 11: Tests for systemic toxicity;
Test MethodologyPurposeAcceptance CriteriaResults(Pass/Fail)
ASTM D6319-19,ASTM D3767-03(2020)Physical DimensionsWidthXS$70\pm10$ mmPass
S$80\pm10$ mm
M$95\pm10$ mm
L$110\pm10$ mm
XL$120\pm10$ mm
LengthXS220mm min
S220mm min
M230mm min
L230mm min
XL230mm min
ThicknessPalm0.05mm min
Finger0.05mm min
ASTM D6319-19,ASTM D412-16Physical PropertiesBefore AgingTensileStrength14MPa minPass
UltimateElongation500% min
After AgingTensileStrength14MPa min
UltimateElongation400% min
ASTM D5151-19Freedom from HolesBe free from holes when tested in accordance withASTM D5151Pass
ASTM D6124-06Powder free residueLess than 2mg per glovePass

Table 2 Summary of non-clinical performance testing

{7}------------------------------------------------

ISO 10993-10:2010IrritationNon-IrritatingPass
ISO 10993-10:2010SensitizationNon-SensitizingPass
ISO 10993-11:2009System toxicityNon-toxicityPass

7. Clinical Test Conclusion

No clinical study is included in this submission.

    1. Conclusion:
      The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe,

as effective, and performs as well as or better than the legally marketed predicate device K031384.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.