K Number
K213125

Validate with FDA (Live)

Date Cleared
2021-10-27

(30 days)

Product Code
Regulation Number
878.3300
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PERI-STRIPS Dry Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers when staple line reinforcement is needed.

PSDV-SG can be used for reinforcement of staple lines during bariatric, gastric, small bowel, colon and colorectal procedures.

Device Description

PERI-STRIPS DRY Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology is prepared from dehydrated bovine pericardium procured from cattle under 30 months of age, originating from the United States.

There are two types of PSDV-SG configurations based on the tissue buttress thickness range. The Standard configuration has a tissue buttress thickness range of 0.20-0.60mm. The Thin configuration has a tissue buttress thickness range of 0.20-0.40mm.

The product consists of a loading unit which includes two (2) buttresses, one for the anvil and one for the cartridge side of the stapler. The buttresses are held in a foam/sheath configuration for loading of the buttress to the stapler jaws. Each buttress has acrylic adhesive on one side for attachment to the stapler surfaces. Each PSDV-SG loading unit is packaged sterile in a separate pouch.

PSDV-SG utilizes animal tissue; patient must be informed prior to any procedure.

PSDV-SG is provided sterile and intended for single use. Sterilization is accomplished via ethylene oxide (ETO). The bovine pericardium buttress and acrylic adhesive are considered permanent implants per ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.

AI/ML Overview

This document describes the PERI-STRIPS Dry Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology (PSDV-SG) device, which is a surgical mesh intended for staple line reinforcement. It is a Special 510(k) submission, meaning the changes are relatively minor compared to a predicate device.

Based on the provided text, here's a breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a quantitative, pass/fail manner with numerical targets for clinical performance metrics (e.g., specific reduction in anastomotic leaks, burst pressure thresholds). Instead, it focuses on demonstrating substantial equivalence to a predicate device by showing that the technological characteristics and performance (functional, manipulation, dimensional, usability, and shelf-life) are similar.

The "performance" described is largely comparative to the predicate device and verification/validation results. The core of the performance claim is that the device maintains the same intended use, indications for use, materials, and principles of operation as the predicate, despite changes in size and design for stapler compatibility.

Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit from text)Reported Device Performance
Intended UseSame as predicate device (K192615)SAME
Indications for UseSame as predicate device (K192615)SAME
MaterialsSame as predicate deviceSAME (bovine pericardium, acrylic adhesive)
Principle of OperationSame as predicate deviceSAME (non-cross-linked collagen tissue technology for neo-collagen formation, neovascularization, and remodeling)
Manipulation TestingPerformed to demonstrate device's handling characteristicsPerformed, specific results not detailed but support substantial equivalence.
Functional TestingPerformed to demonstrate device's operational capabilitiesPerformed, specific results not detailed but support substantial equivalence.
Dimensional TestingPerformed to ensure correct sizing and fit for stapler compatibilityPerformed, specific results not detailed but support substantial equivalence.
Usability TestingPerformed to assess ease of use and user interactionPerformed, specific results not detailed but support substantial equivalence.
Shelf-lifeStability over time1 year with testing to 3 years ongoing (predicate had 3 years). This is a difference, but presumably accepted by FDA.
Stapler CompatibilityCompatibility with specified stapler models (new models compared to predicate)Predicate: Ethicon Echelon ENDOPATH Staplers (K163455, K160521, K140560). Subject Device: Medtronic Endo GIA Staplers (K111825) and Intuitive Surgical SureForm Staplers (K173721). This is a key difference addressed by the Special 510(k).

Note: The document states "Verification and validation of the product have been completed using the same test methods as the predicate device, or accepted industry standard methods. No new issues of safety and effectiveness have been identified." This implies that the results from these tests met the internal acceptance criteria established by the manufacturer, which were aligned with the predicate or industry standards. However, the specific numerical acceptance criteria (e.g., minimum burst pressure, maximum staple pull-through force) are not provided in this summary.

2. Sample Size Used for the Test Set and Data Provenance

The document states "Testing performed to support the new product models subject of this submission included manipulation, functional, and dimensional testing, as well as usability and shelf-life." It does not provide details on the sample sizes used for these tests.

Regarding data provenance:

  • Country of Origin: Not specified for the testing data.
  • Retrospective or Prospective: Not specified. These appear to be benchtop and perhaps simulated use tests, not clinical studies involving patients.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

N/A. This submission focuses on engineering and bench testing performance for substantial equivalence, not clinical diagnostic accuracy or efficacy. There is no mention of "ground truth" derived from expert review in the clinical sense.

4. Adjudication Method for the Test Set

N/A. As there's no mention of expert review or clinical studies for ground truth establishment, no adjudication method is described.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No. There is no mention of a human reader or AI comparative effectiveness study. This device is a surgical mesh, not a diagnostic imaging or AI-assisted interpretation device.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

N/A. This device is a physical surgical mesh, not a software algorithm.

7. The Type of Ground Truth Used

For the engineering and functional tests conducted, the "ground truth" would be established by:

  • Measurement Standards: Adherence to predefined dimensional and physical property specifications.
  • Performance Specifications: Meeting functional requirements (e.g., tear strength, adhesion, burst pressure if these were tested) derived from predicate device performance or industry standards.
  • Visual Inspection: For manipulation and usability tests, often a qualitative or semi-quantitative assessment against predefined criteria.

No pathology or outcomes data is mentioned as "ground truth" in this context of bench testing for substantial equivalence for the device itself.

8. The Sample Size for the Training Set

N/A. This device is not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

N/A. Not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 27, 2021

Synovis Life Technologies. Inc. (A Subsidiary Of Baxter International Inc.) Megan Sajjad Senior Manager, Regulatory Affairs 2575 University Ave. W St. Paul, Minnesota 55114

Re: K213125

Trade/Device Name: PERI-STRIPS Dry Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTM, OXE Dated: September 24, 2021 Received: September 27, 2021

Dear Megan Sajjad:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Deborah Fellhauer, RN, BSN Assistant Director (acting) DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K213125

Device Name

PERI-STRIPS Dry Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology

Indications for Use (Describe)

PERI-STRIPS Dry Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staple line reinforcement is needed.

PSDV-SG can be used for reinforcement of staple lines during bariatric, gastric, small bowel, colon and colorectal procedures.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

PERI-STRIPS Dry Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology

I. SUBMITTER

Synovis Life Technologies, Inc. (A Subsidiary of Baxter International Inc.) 2575 University Avenue West St. Paul, MN 55114-1024

Contact Person: Megan Sajjad, Sr. Manager, Regulatory Affairs Phone: 651-796-7410

Date Prepared: September 24, 2021

II. DEVICE

Device Name:PERI-STRIPS Dry Staple Line Reinforcement with VERITASCollagen Matrix with SECURE GRIP Technology
Common Name:Staple Line Reinforcement
Classification Name:MESH, SURGICAL; MESH, SURGICAL, COLLAGEN, STAPLELINE REINFORCEMENT
Regulation Number21 CFR 878.3300
Product Codes:FTM, OXE

III. PREDICATE DEVICE

PERI-STRIPS Dry Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology, K192615.

This predicate has not been subject to any market recalls.

IV. DEVICE DESCRIPTION

PERI-STRIPS DRY Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology is prepared from dehydrated bovine pericardium procured from cattle under 30 months of age, originating from the United States.

There are two types of PSDV-SG configurations based on the tissue buttress thickness range. The Standard configuration has a tissue buttress thickness range of 0.20-0.60mm. The Thin configuration has a tissue buttress thickness range of 0.20-0.40mm.

The product consists of a loading unit which includes two (2) buttresses, one for the anvil and one for the cartridge side of the stapler. The buttresses are held in a foam/sheath configuration for loading of the buttress to the stapler jaws. Each buttress has acrylic adhesive

{4}------------------------------------------------

PERI-STRIPS Dry Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology SPECIAL 510K

on one side for attachment to the stapler surfaces. Each PSDV-SG loading unit is packaged sterile in a separate pouch.

PSDV-SG utilizes animal tissue; patient must be informed prior to any procedure.

PSDV-SG is provided sterile and intended for single use. Sterilization is accomplished via ethylene oxide (ETO). The bovine pericardium buttress and acrylic adhesive are considered permanent implants per ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.

V. INTENDED USE/INDICATIONS FOR USE

The Intended Use and Indications for Use remain the same as the predicate device.

Intended Use:

PSDV-SG is intended to be used as a staple line buttress.

Indications for Use:

PERI-STRIPS Dry Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers when staple line reinforcement is needed.

PSDV-SG can be used for reinforcement of staple lines during bariatric, gastric, small bowel, colon, and colorectal procedures.

TECHNOLOGICAL CHARACTERISTICS AND COMPARISON WITH VI. THE PREDICATE DEVICE

PSDV-SG products are implantable biologic meshes comprised of non-cross-linked bovine pericardium. PSDV-SG is the VERITAS Collagen Matrix (K002233) tissue that has been dried and cut to specification for use with staplers to provide buttress reinforcement at the staple line. PSDV-SG allows for neo-collagen formation and neo-vascularization of the implanted device and permits replacement of the device with host tissue, or remodeling.

The technological characteristics and specifications of PSDV-SG have been evaluated against its predicate device to determine equivalence. A summary of the comparison between the predicate and subject devices is provided Table 5-1 below.

AttributePredicate DeviceSubject Device
Product NamePERI-STRIPS Dry Staple Line Reinforcement with VERITASCollagen Matrix with SECURE GRIP TechnologySAME
510(k)-holderSynovis Life Technologies, Inc.(A Subsidiary of Baxter International Inc.)SAME
510(k) NumberK192615K213125
Product CodeFTM, OXESAME
Regulation21 CFR 878.3300SAME
ClassificationClass IISAME
Intended UsePSDV-SG is intended to be used as a staple line buttress.SAME
Indications forUsePERI-STRIPS Dry Staple Line Reinforcement with VERITASCollagen Matrix with SECURE GRIP Technology is intended forSAME

Table 5-1. PSDV-SG Predicate and Subject devices - Comparison Table

{5}------------------------------------------------

PERI-STRIPS Dry Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology SPECIAL 510K

AttributePredicate DeviceSubject Device
use as a prosthesis for the surgical repair of soft tissuedeficiencies using surgical staplers when staple linereinforcement is needed.PSDV-SG can be used for reinforcement of staple lines duringbariatric, gastric, small bowel, colon and colorectal procedures.
TissueBovine pericardium procured from cattle under 30 months of age,originating from the United StatesSAME
ComponentsTissue strips, adhesive, liner pull tab, loading unit (foam, sheath),ship clip and packaging (inner pouch, outer pouch, labeling andcarton)SAME
Thickness perFiring (mm)Standard: 0.45 – 1.26Thin: 0.45 – 0.86SAME
Principle ofOperationNon-cross-linked collagen tissue technology allows for neo-collagen formation and neovascularization of the implanteddevice and permits infiltration of the device with host tissue, orremodeling. PSDV-SG is designed to provide strengththroughout the healing process, particularly where remodelingwith the patient's' own tissue is preferred.SAME
StaplerCompatibilityEthicon Echelon ENDOPATH Staplers (K163454, K160521,K140560)Medtronic EndoGIA Staplers(K111825)and IntuitiveSurgical SureFormStaplers (K173721)
How SuppliedSingle-useSterileSAME
PackagingSterile double pouchSAME
SterilizationETOSAME
Shelf-life3 years1 year with testingto 3 years ongoing

VII. PERFORMANCE DATA

Testing performed to support the new product models subject of this submission included manipulation, functional, and dimensional testing, as well as usability and shelf-life. Materials evaluation was not necessary as the new models are manufactured and packaged utilizing the same materials and components, from the same suppliers, used with the predicate device.

VII. CONCLUSION

The PSDV-SG product family maintains the same intended use, indications for use, materials, and principles of operation. The line extension introducing new product models does not impact these product characteristics. The differences in size and design are made for stapler compatibility. Verification and validation of the product have been completed using the same test methods as the predicate device, or accepted industry standard methods. No new issues of safety and effectiveness have been identified.

The new PSDV-SG models have been found substantially equivalent to the predicate device with respect to intended use, design, materials, principle of operation and overall technological characteristics.

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.