K Number
K213107

Validate with FDA (Live)

Date Cleared
2022-02-17

(146 days)

Product Code
Regulation Number
880.6250
Age Range
0 - 21
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner

Device Description

The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical properties. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151. This is a disposable device, intended for medical purposes, that is worn on the examiner's hand to prevent contamination between patient and examiner.

AI/ML Overview

The provided document is a 510(k) summary for "Powder Free Blue Nitrile Examination Gloves" (K213107) and does not describe an AI medical device. Therefore, many of the requested fields are not applicable. This summary focuses on the safety and performance of physical medical gloves, primarily through non-clinical testing against established standards.

Here's an analysis based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

Test Method / PropertyAcceptance CriteriaReported Device Performance
ASTM D 5151 (Pinhole)Free from holes, AQL 1.5Pass
ASTM D6319 (Physical)Before Aging: Tensile Strength: Min 14 MPa, Elongation: Min 500%Pass
After Aging: Tensile Strength: Min 14 MPa, Elongation: Min 400%Pass
Watertight (1000mL)Single Sampling in accordance with ISO 2859 G1: AQL 2.5Pass G1: AQL 2.5
Length (mm)Min 220 (S), Min 230 (M, L, XL)240mm minimum for all sizes
Palm width (mm)80 ± 10 (S), 95 ± 10 (M), 110 ± 10 (L), 120 ± 10 (XL)84-86 (S), 95-96 (M), 108-109 (L), 114-115 (XL)
Thickness (mm)Min 0.05 (Finger, Palm, Cuff)Min 0.06 (Finger), Min 0.06 (Palm), Min 0.05 (Cuff)
ASTM D412 (Physical)Before Aging: Tensile Strength: Min 14 MPa, Elongation: Min 500%17-23 (Tensile Strength), 540-607 (Elongation)
After Aging: Tensile Strength: Min 14 MPa, Elongation: Min 400%21-27 (Tensile Strength), 480-565 (Elongation)
ASTM D6124 (Powder)Max 2.0 mg/gloveBelow 2.0 mg/glove
Guinea Pig SensitizationNo deviations noted, no positive allergic reactionNo reaction observed, no positive allergic reaction. Meets conformance requirements.
Primary Dermal Irritation (Rabbits)No deviations noted, identify primary irritantsPrimary Irritation Index was 0. Meets conformance requirements.
Acute Systemic Cytotoxicity (Rats)No adverse reaction notedAll animals survived, gained weight, no abnormalities in necropsies. Meets conformance requirements.

2. Sample size used for the test set and the data provenance

The document does not explicitly state the exact sample sizes for each specific test (e.g., number of gloves tested for pinholes, physical properties, or biocompatibility). However, it refers to standards such as "Single Sampling in accordance with ISO 2859 G1" for the watertight test, which implies a statistically determined sample size based on the Acceptable Quality Level (AQL). The tests appear to be performed on the final and finished devices.

  • Data Provenance: Not explicitly stated, but given the applicant's location (Novid PPE Sdn Bhd, Malaysia), the testing was likely conducted in Malaysia or a lab contracted by them. The tests are non-clinical, meaning they involve physical product testing and animal studies, not human patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a non-AI medical device (examination gloves). Ground truth for these tests is established by adherence to recognized international and national standards (e.g., ASTM, ISO) and laboratory protocols, not by expert consensus on clinical data.

4. Adjudication method for the test set

Not applicable. The tests are objective measurements against predefined criteria in established standards, not subjective assessments requiring expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI medical device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI medical device.

7. The type of ground truth used

For the performance and physical properties, the "ground truth" is defined by the specified minimum or maximum values and methodologies outlined in the referenced ASTM and ISO standards (e.g., ASTM D6319 for physical properties, ASTM D5151 for pinhole detection, ASTM D6124 for powder content). For biocompatibility, the ground truth is established by the absence of adverse biological reactions using standard animal testing protocols (ISO 10993-10, ISO 10993-11).

8. The sample size for the training set

Not applicable. This is not an AI medical device, so there is no "training set."

9. How the ground truth for the training set was established

Not applicable. This is not an AI medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 17, 2022

Novid PPE Sdn. Bhd. Er Chuan QA Manager Lot 6071, Jalan Haji Abdul Manan, Batu 5 1/2, Jalan Meru Klang, Selangor 41050 Malaysia

Re: K213107

Trade/Device Name: Powder Free Blue Nitrile Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 7, 2021 Received: September 24, 2021

Dear Er Chuan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K213107

Device Name Powder Free Blue Nitrile Examination Gloves

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K213107

510K Summary

As required by 21 CFR 807.92

1) Submission Information:

Date: 18th September 2021
Type of 510(k) Submission: Traditional
Basis for 510(k) Submission: New Device
Applicant: Novid PPE Sdn BhdLot 6071, Jalan Haji Abdul Manan,Batu 5 ½, Jalan Meru,41050 Klang, Selangor, Malaysia
Contact Person: Er Chai Chuan (QA Manager)Lot 6123, Jalan Haji Salleh,Batu 5 ½, Jalan Meru, 41050 Klang,Selangor, Malaysia.E-mail: chuan@notionventure.comTel: +6016 3689637

2) Device:

Proprietary Name: Powder Free Blue Nitrile Examination Gloves Classification Name: Examination Gloves Regulation Number: 880.6250 Product Code: LZA Device Class: I Review Panel: General Hospital

3) Device Description

The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical properties. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151. This is a disposable device, intended for medical purposes, that is worn on the examiner's hand to prevent contamination between patient and examiner.

{4}------------------------------------------------

4) Identification of the Legally Marketed Devices

Class 1 Nitrile Patient Examination Gloves LZA, powder free that meets all the requirements of ASTM standard D6319 and FDA water leak test.

5) The Intended Use of Gloves

A Nitrile Examination Glove is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Comparison of Technological Characteristics with the Predicate Device: Table 1: General Comparison

Technological characteristics Comparison to Predicate Device
Subject DevicePredicate DeviceComparison
Novid PPE Sdn Bhd'sPowder Free BlueNitrile ExaminationGlovesMercator Medical(Thailand) LTD'smCare Powder FreeNitrile BlueExamination Gloves -
510K NumberK213107K172930
Indications for UseThe device is adisposable deviceintended for medicalpurposes that is wornon the examiner's handorfinger to preventcontamination betweenpatient and examinerThe device is adisposable deviceintended for medicalpurposes that are wornon the examiner's handto preventcontamination betweenpatient and examiner.Same
Design SpecificationMeet ASTM D6319-10Meet ASTM D6319-10Same
PerformancePhysical PropertiesMeet ASTM D6319-10Meet ASTM D6319-10Same
Material ofCompositionSynthetic NitrileRubberSynthetic NitrileRubberSame
Biocompatibility:Animal IrritationTest RabbitUnder the condition ofstudy, not an irritantUnder the condition ofstudy, not an irritantSame
Dermal SensitizationTest- Guinea PigUnder the condition ofstudy, not a sensitizerUnder the condition ofstudy, not a sensitizerSame
Acute SystemicCytotoxicityUnder the condition ofthe study, no adversebiological reactionUnder the condition ofthe study, no adversebiological reactionSame
ColorSynthetic gloves withembedded colorant-BlueSynthetic gloves withembedded colorant-BlueSame
SterilityNon-SterileNon-SterileSame
Powder FreeMeets applicable definition for Powder free; ≤ 2mg per gloveMeets applicable definition for Powder free; ≤ 2mg per gloveSame
LabellingInformationSingle Use indication, Powder free, device name, gloves size, quantity, Patient examination gloves, Non-sterileSingle Use indication, Powder free, device name, gloves size, quantity, Patient examination gloves, Non-sterileSame
PhysicalPropertiesas perASTMD6319-10Before AgingTensile StrengthMin 14 MPaUltimate ElongationMin 500%After AgingTensile StrengthMin 14 MPaUltimate ElongationMin 400%Before AgingTensile StrengthMin 14 MPaUltimate ElongationMin 500%After AgingTensile StrengthMin 14 MPaUltimate ElongationMin 400%Same
Dimensionas perASTMD6319-10Finger Thickness:0.06 - 0.10 mmLength: min 230 mmFinger Thickness:Min 0.05 mm Length:min 230 mmSimilar
Freedomfrom holesAQL per CFR21.800.20Test as per ASTMD5151AQL per CFR21.800.20Test as per ASTMD5151Same
ResidualPowderTested to ASTMD6124 and meetsrequirementTested to ASTMD6124 and meetsrequirementSame

{5}------------------------------------------------

{6}------------------------------------------------

6) Non-clinical test was performed on the proposed device:

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

  • a) ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Tests for Irritation and Skin Sensitization
  • b) ISO 10993-11- Biological Evaluation of Medical devices- Part 11: Tests for Systemic toxicity
  • c) ASTM D6124-06 (Reapproved 2017), Standard test method for Residual powder on medical gloves
  • d) ASTM D5151-06 (Reapproved 2015), Standard Test Method for Detection of Holes in medical gloves
  • e) ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application

Performance data of gloves based on ASTM D6977 and FDA Watertight Test In addition the proposed device was tested and conformed to the following standards and requirements stated in guidance for industry passed and FDA staff - Premarket Notification [510(k)] Submission issued on March 5, 2004:

Table 2: Performance Testing Performance data of gloves and Biocompatibility Testing based on animal studies, biocompatibility studies i) Skin irritation in rabbits, Guinea pig sensitization (Buehler) and acute systemic cytotoxicity test were conducted on Novid PPE Sdn Bhd's final and finished Powder Free Blue Nitrile Examination Gloves.

Test MethodPurposeAcceptanceCriteriaPowder Free Blue NitrileExaminationGloves Results
ASTM D 5151Pinhole TestFree from holes,AQL 1.5Pass
ASTM D6319Physical PropertiesBefore AgingTensile Strength:Min 14 MPaElongation: Min500%After AgingTensile Strength:Min 14 MPaElongation: Min400%Pass

{7}------------------------------------------------

1. Watertight (1000mL)in accordance withASTM D5151Leak TestSingle Samplingin accordancewith ISO 2859G1 : AQL 2.5Pass G1 : AQL 2.5
2. Length (mm) SizePhysical Properties
SMin 220240mm minimum for allsizes
MMin 230
LMin 230
XLMin 230
3. Palm width (mm) SizePhysical Properties
S80 ± 1084 – 86
M95 ± 1095-96
L110 ± 10108 – 109
XL120 ± 10114 - 115
4. Thickness (mm)Single Layer FingerPhysical Properties
PalmMin 0.05Min 0.06
CuffMin 0.05Min 0.06
Min 0.05Min 0.05
5. Physical Properties inaccordance with ASTM D412Physical Properties
Before Aging
Tensile Strength (MPa)Min 1417 – 23
Ultimate Elongation (%)Min 500540 – 607
After Aging
Tensile Strength (MPa)Min 1421 – 27
Ultimate Elongation (%)Min 400480 - 565
6. Powder Content in accordancewith ASTM D6124Powder ResidueMax 2.0 mg/ gloveBelow 2.0 mg/ glove
I. Guinea PigSensitizationThis test wasdesigned todetermine if thetest article is apotential sensitizerto guinea pigswhen appliedatopically.No deviations werenoted, observednor requireclarification.No reaction was observedupon removal of the testmaterial and there was nopositive allergic reactionobserved on the test ofguinea pigs during thechallenge phase. None ofthe guinea pigs wassensitized.Conclusion: meetsconformancerequirements
II. Primary DermalThis test wasdesigned toNo deviations werenoted, observe norEach test wasindividually
Irritation in Rabbitsidentify substances which are primary irritants to rabbit skinrequire clarification.examined and scored at 24 $\pm$ 2,48 $\pm$ 2 and 72 $\pm$ 2 hours for erythema and edema using the Draize skin scoring scale. Results obtained as Primary Irritation Index was 0. Conclusion: meets conformance requirements.
III.Acute Systemic Cytotoxicity in RatsThis test was designed to identify any adverse biological reaction following administration of the extracts of the test item on the rats.No adverse reaction was noted, observed nor require clarification.For the 4 days observation done on the test subject by doing:1) Cage- side observation- all animals survived and appeared healthy and active through out of 4 days.2) Body weight- all animals gained body weight through out of the 4 days.

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

3) Pathology-
At sacrifice
times, gross
necropsies
showed no
abnormalities
for any of the
animals.
Conclusion:
meets
conformance
requirements.

7) Clinical Test Conclusion

No clinical study is included in this submission

8) Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the subject device is as safe, as effective, and perform as well as or better than the legally marketed predicated K172930.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.