(84 days)
The Nitrile Examination Gloves (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed device, Nitrile Examination Gloves (Powder free, Blue) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size. The color of the proposed device is Blue.
The proposed device is provided non-sterile. The proposed device is made of Nitrile.
The provided text describes the acceptance criteria and the study results for "Nitrile Examination Gloves (Powder free, Blue)". This is a physical medical device, not a software-driven AI device. Therefore, the specific questions regarding AI-related performance metrics (like sample size for test set/training set, data provenance, number of experts for ground truth, adjudication method, MRMC studies, standalone algorithm performance) are not applicable to this document.
However, I can extract the acceptance criteria and reported device performance from the provided tables.
Acceptance Criteria and Reported Device Performance
| Test Method | Purpose | Acceptance Criteria | Reported Device Performance | Comments |
|---|---|---|---|---|
| Physical Performance (ASTM D6319, D412, D573) | ||||
| Freedom from Holes (ASTM D5151) | Testing for Freedom from holes | AQL 2.5 | No water leakage is inspected from 200 samples | Meets acceptance criteria. |
| Residual Powder (ASTM D6124) | Determine the powder residue for powder free gloves | < 2.0 mg per glove | Size XS: Average 0.28 mg; Size S: Average 0.31 mg; Size M: Average 0.34 mg; Size L: Average 0.31 mg; Size XL: Average 0.40mg | Meets acceptance criteria. |
| Tensile Strength (Before Aging) | Physical property characteristics | 14 MPa min. | ≥ 19MPa | Meets acceptance criteria. |
| Ultimate Elongation (Before Aging) | Physical property characteristics | 500% min. | ≥ 500%. | Meets acceptance criteria. |
| Tensile Strength (After Aging) | Physical property characteristics | 14 MPa min. | ≥ 18 MPa | Meets acceptance criteria. |
| Ultimate Elongation (After Aging) | Physical property characteristics | 400% min. | ≥ 473%. | Meets acceptance criteria. |
| Physical Dimensions (ASTM D412, D3767) | ||||
| Length (XS, S) | Physical dimensions specification | 220 mm min. | Size XS: ≥ 223 mm; Size S: ≥ 224 mm | Meets acceptance criteria. |
| Length (M) | Physical dimensions specification | 230 mm min. | Size M: ≥ 233 mm | Meets acceptance criteria. |
| Length (L, XL) | Physical dimensions specification | 235 mm min. | Size L: ≥ 236 mm; Size XL: ≥ 236 mm | Meets acceptance criteria. |
| Width (XS) | Physical dimensions specification | 70 ± 10 mm | 75 ± 2 (73-77) mm | Meets acceptance criteria. |
| Width (S) | Physical dimensions specification | 80 ± 10 mm | 85 ± 2 (85-87) mm | Meets acceptance criteria. |
| Width (M) | Physical dimensions specification | 95 ± 10 mm | 95 ± 3 (95-98) mm | Meets acceptance criteria. |
| Width (L) | Physical dimensions specification | 110 ± 10 mm | 115 ± 4 (115-119) mm | Meets acceptance criteria. |
| Width (XL) | Physical dimensions specification | 120 ± 10 mm | 125 ± 2 (115-117) mm | Meets acceptance criteria. |
| Cuff Thickness | Physical dimensions specification | ≥ 0.05 mm | Size XS, S, M, L, XL: ≥ 0.06 mm | Meets acceptance criteria. |
| Palm Thickness | Physical dimensions specification | ≥ 0.05 mm | Size XS, S, M, L, XL: ≥ 0.07 mm | Meets acceptance criteria. |
| Finger Thickness | Physical dimensions specification | ≥ 0.05 mm | Size XS, S, M, L, XL: ≥ 0.10 mm | Meets acceptance criteria. |
| Biocompatibility (ISO 10993-5, ISO 10993-10) | ||||
| Cytotoxicity (ISO 10993-5) | Evaluate the endpoint of Cytotoxicity for biocompatibility | No evidence of cytotoxic potential from the extract. | No evidence of systemic toxicity from the extract. | Meets acceptance criteria. |
| Irritation (ISO 10993-10) | Evaluate the endpoint of irritant for biocompatibility | No irritation on the skin. | No irritation on the skin. | Meets acceptance criteria. |
| Sensitization (ISO 10993-10) | Evaluate the endpoint of sensitization for biocompatibility | No evidence of causing delayed dermal contact sensitization. | No evidence of causing delayed dermal contact sensitization in the guinea pig. | Meets acceptance criteria. |
Study Information (where applicable based on the provided document):
-
Sample size used for the test set and the data provenance:
- Freedom from Holes (ASTM D5151): 200 samples were inspected.
- Biocompatibility (ISO 10993-10 for sensitization): Guinea pig was used for sensitization test. The document does not specify the exact number of guinea pigs.
- Data Provenance: The document does not explicitly state the country of origin where the tests were performed or if they were retrospective or prospective studies. The company is based in China.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. These are physical and chemical bench tests, not involving expert interpretation of data for ground truth.
-
Adjudication method for the test set: Not applicable.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable, as this is a physical device, not an AI or imaging device involving human readers.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable, as this is a physical device.
-
The type of ground truth used: For physical and chemical tests, the "ground truth" is defined by the technical specifications outlined in the referenced ASTM and ISO standards (e.g., specific tensile strength values, elongation percentages, powder residue limits for ASTM standards, and bio-response criteria for ISO biocompatibility standards).
-
The sample size for the training set: Not applicable, as this is a physical device.
-
How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 14, 2021
Weihai Hongyu Nonwoven Fabric Products Co., Ltd. Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China
Re: K213029
Trade/Device Name: Nitrile Examination Gloves (Powder free, Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 21, 2021 Received: September 21, 2021
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801: medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K213029
Device Name
Nitrile Examination Gloves (Powder free, Blue)
Indications for Use (Describe)
The Nitrile Examination Gloves (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
__ Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: K213029
-
- Date of Preparation: 11/23/2021
-
- Sponsor
Weihai Hongyu Nonwoven Fabric Products Co., Ltd.
No.567, Gushan Road, Area of Economy and Technique,264207, Weihai City, Shandong Province, P.R. China. Contact Person: Dong Bin Position: General Manager Tel: +86-15662383633 Fax: +86-631-3636942 Email: dbfeel@163.com
-
- Submission Correspondent
Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China,102401 Contact Person: Ray Wang Position: General Manager Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com
- Proposed Device Identification 4.
Trade Name: Nitrile Examination Gloves (Powder free, Blue) Common Name: NITRILE Patient Examination Gloves (Powder Free) Regulatory Information: Classification: I Product Code: LZA Regulation Number: 21 CFR 880.6250
{4}------------------------------------------------
Review Panel: General Hospital
Indication For Use Statement:
The Nitrile Examination Gloves (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
- Predicate Device Identification 5.
510(k) Number: K150340
Product Name: POWDER FREE Nitrile GLOVES (White, Cobalt Blue, Black, Ice Blue) Manufacturer: HEBEI HONGSEN PLASTICS TECHNOLOGY CO., LTD
- Device Description 6.
The proposed device, Nitrile Examination Gloves (Powder free, Blue) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size. The color of the proposed device is Blue.
The proposed device is provided non-sterile. The proposed device is made of Nitrile.
| Designation | Size | Tolerance | ||||
|---|---|---|---|---|---|---|
| XS | S | M | L | XL | ||
| Length, mm | 220 | 220 | 230 | 235 | 235 | min |
| Width, mm | 70 | 80 | 95 | 110 | 120 | ±10 |
| Thickness, mm: | ||||||
| Finger | 0.05 | min | ||||
| Palm | 0.05 | min | ||||
| Cuff | 0.05 | min |
Table 1 Device Size Specifications
| Table 2 Performance and Physical Specifications | |
|---|---|
| ------------------------------------------------- | -- |
| Before Aging | After Aging | Pinhole AQL | ||
|---|---|---|---|---|
| TensileStrength | UltimateElongation | TensileStrength | UltimateElongation | 2.5 |
| 14 MPa, min | 500 % min | 14 MPa, min | 400 % min |
{5}------------------------------------------------
The above data of size, performance, and physical specifications of proposed gloves meet all the current specifications listed in the ASTM standard D6319.
Comparison of Technological Characteristics 7.
| ITEM | Proposed Device(K213029) | Predicate Device (K150340) | Remark |
|---|---|---|---|
| Nitrile Examination Gloves (Powder free,Blue) | POWDER FREE Nitrile GLOVES (White,Cobalt Blue, Black, Ice Blue) | ||
| Product Code | LZA | LZA | SAME |
| Regulation No. | 21 CFR 880.6250 | 21 CFR 880.6250 | SAME |
| Class | I | I | SAME |
| Intended Use | The Nitrile Examination Gloves (Powderfree, Blue) is a disposable device intendedfor medical purposes that is worn on theexaminer's hands to prevent contaminationbetween patient and examiner. | The POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, Ice Blue) is adisposable device intended for medicalpurposes that is worn on the examiner'shands to prevent contamination betweenpatient and examiner. | SAME |
| Powdered orPowered free | Powdered free | Powdered free | SAME |
Table 1 General Comparison
{6}------------------------------------------------
| Proposed Device(K213029) | Size | |||||||
|---|---|---|---|---|---|---|---|---|
| Nitrile Examination Gloves(Powder free, Blue) | Designation | XS | S | M | L | XL | Tolerance | |
| Length, mm | 220 | 220 | 230 | 235 | 235 | min | ||
| Width, mm | 70 | 80 | 95 | 110 | 120 | ±10 | ||
| Thickness, mm: | ||||||||
| Finger | 0.05 | min | ||||||
| Palm | 0.05 | min | ||||||
| Cuff | 0.05 | min | ||||||
| Predicate Device (K150340) | Size | |||||||
| POWDER FREE NitrileGLOVES (White, Cobalt Blue,Black, Ice Blue) | Designation | XS | S | M | L | XL | Tolerance | |
| Length, mm | 230 | 230 | 230 | 230 | 230 | min | ||
| Width, mm | 70 | 80 | 95 | 110 | 120 | ±10 | ||
| Thickness, mm: | ||||||||
| Finger | 0.10-0.12 | ±0.03 | ||||||
| Palm | 0.08-0.10 | ±0.03 | ||||||
| Cuff | 0.06-0.09 | ±0.03 | ||||||
| Remark | Similar |
Table 2 Device Dimensions Comparison
Different Analysis:
The proposed device has different size specification to the predicate device, but all proposed devices are meet the specifications of ASTM D 6319.
So we consider this as the proposed device is similar to the predicate device.
| ITEM | Proposed Device(K213029)Nitrile Examination Gloves(Powder free, Blue) | Predicate Device (K150340)POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, IceBlue) | Remark | ||
|---|---|---|---|---|---|
| Colorant | Blue | White, Cobalt Blue, Black, Ice Blue | Different | ||
| Single Use | Yes | Yes | SAME | ||
| PhysicalProperties | BeforeAging | Tensile Strength | 14 MPa, min | 15 MPa, min | Different |
| UltimateElongation | 500 % min | 500 % min | SAME | ||
| AfterAging | TensileStrength | 14 MPa, min | 14 MPa, min | SAME |
Table 3 Performance Comparison
{7}------------------------------------------------
| UltimateElongation | 400 % min | 400 % min | SAME | ||
|---|---|---|---|---|---|
| Comply with ASTM D6319 | Comply with ASTM D6319 | SAME | |||
| Freedom from Holes | Be free from holes when tested inaccordance with ASTM D5151,AQL 2.5 | Be free from holes when tested inaccordance with ASTM D5151,AQL 1.5 | Different | ||
| Powder Content | Less than 2 mg per glove when testedin accordance with ASTM D6124 | Meet the requirements of ASTM6124 | SAME | ||
Different Analysis:
The proposed device has different color to the predicate device, this different may causes potential biocompatibility risk, for this risk we conducted the biocompatibility test according to the ISO 10993-5, and ISO 10993-10 the test results showed that the proposed devices with blue colorant did not induce skin irritation and showed no significant evidence of causing skin sensitization and no cytotoxic potential.
Different Analysis:
The proposed device has different Tensile Strength before aging specification to the predicate device, but all proposed device meets the specification requirements of ASTM D 6319.
Different Analysis:
The proposed device has different AQL to the predicate device, but all proposed device meets the specification requirements of ASTM D 6319.
| ITEM | Proposed Device (K213029)Nitrile Examination Gloves(Powder free, Blue) | Predicate Device (K150340)POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, IceBlue) | Remark | |
|---|---|---|---|---|
| Material | Nitrile | Nitrile | SAME | |
| Biocompatibility | Irritation | Under the conditions of the study,not an irritant | Under the conditions of the study,not an irritant | SAME |
| Sensitization | Under conditions of the study, not asensitizer. | Under conditions of the study, not asensitizer. | SAME | |
| Cytotoxic | Under the conditions of the study,there was no evidence of cytotoxicpotential from the extract. | Not Available | Different | |
| Label and Labeling | Meet FDA's Requirements | Meet FDA's Requirements | SAME |
Table 4 Safety Comparison
Different Analysis:
The proposed device has conducted the acute cytotoxic testing, and the test results shown that there was
{8}------------------------------------------------
no evidence of cytotoxic potential.
8. Summary of Non-Clinical Testing
Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6124-17, Standard Test Method for Residual Powder on Medical Gloves.
ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
| Test Method | Purpose | Acceptance Criteria | Results |
|---|---|---|---|
| ASTM D5151 | Testing for Freedomfrom holes | Freedom from holesAQL 2.5 | No water leakage is inspected form 200 samples |
| ASTM D6124 | Determine the powderresidue for powder freegloves | <2.0 mg per glove | Residual Powder of Size XS: Average 0.28 mg; |
| Residual Powder of Size S: Average 0.31 mg; | |||
| Residual Powder of Size M: Average 0.34 mg; | |||
| Residual Powder of Size L: Average 0.31 mg; | |||
| Residual Powder of Size XL: Average 0.40mg; | |||
| ASTM D412ASTM D573 | Testing for Physicalproperty characteristics | Before AgingTensile Strength: 14 MPa min.Ultimate Elongation: 500% min. | Before AgingTensile Strength: ≥ 19MPa;Ultimate Elongation: ≥ 500%. |
| Before AgingTensile Strength: 14 MPa min.Ultimate Elongation: 400% min. | After AgingTensile Strength: ≥ 18 MPa;Ultimate Elongation: ≥ 473%. | ||
| ASTM D412ASTM D3767 | Testing For physicaldimensions specification | Length: 220 mm min. for size (XS, S);Length: 230 mm min. for size (M);Length: 235 mm min. for size (L, XL)Width: 70±10 mm for XS; 80±10 mmfor S; 95±10 mm for M; 110±10 mm forL; 120±10 mm for XL.Cuff Thickness: ≥0.05 mm;Finger Thickness: ≥0.05 mm;Palm Thickness: ≥0.05 mm;All acceptance criteria above meet therequirements in Table 1 Dimensions andTolerances of ASTM D6319 | Length of Size XS: ≥ 223 mm;Width of Size XS: 75±2 (73-77) mm;Cuff Thickness of Size XS: ≥0.06 mm;Palm Thickness of Size XS: ≥0.07 mm;Finger Thickness of Size XS: ≥0.10 mm.Length of Size S: ≥ 224mm;Width of Size S: 85±2 (85-87) mm;Cuff Thickness of Size S: ≥0.06 mm;Palm Thickness of Size S: ≥0.07 mm;Finger Thickness of Size S: ≥0.10 mm.Length of Size M: ≥ 233 mm; |
Table 5 Performance Test Results Summary
{9}------------------------------------------------
| Width of Size M: 95±3 (95-98) mm; | |||
|---|---|---|---|
| Cuff Thickness of Size M: ≥0.06 mm; | |||
| Palm Thickness of Size M: ≥0.07 mm; | |||
| Finger Thickness of Size M: ≥0.10 mm. | |||
| Length of Size L: ≥ 236mm; | |||
| Width of Size L: 115±4 (115-119) mm; | |||
| Cuff Thickness of Size L: ≥0.06 mm; | |||
| Palm Thickness of Size L: ≥0.07 mm; | |||
| Finger Thickness of Size L: ≥0.10 mm. | |||
| Length of Size XL: ≥ 236 mm; | |||
| Width of Size XL: 125±2 (115-117) mm; | |||
| Cuff Thickness of Size XL: ≥0.06 mm; | |||
| Palm Thickness of Size XL: ≥0.07 mm; | |||
| Finger Thickness of Size XL: ≥0.10 mm. | |||
| ISO 10993-5 | Evaluate the endpoint ofCytotoxicity forbiocompatibility | The test article showed no evidence ofcytotoxic potential from the extract. | The test article showed no evidence ofsystemic toxicity from the extract. |
| ISO 10993-10 | Evaluate the endpoint ofirritant forbiocompatibility | The test article showed no irritation onthe skin. | The test article showed no irritation on theskin. |
| Evaluate the endpoint ofsensitization forbiocompatibility | The test article showed no evidence ofcausing delayed dermal contactsensitization | The test article showed no evidence of causingdelayed dermal contact sensitization in theguinea pig. |
9. Summary of Clinical Testing Clinical Testing is not applicable.
10. Conclusion
Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, POWDER FREE Nitrile GLOVES (White, Cobalt Blue, Black, Ice Blue) cleared under K150340.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.