K Number
K212842

Validate with FDA (Live)

Date Cleared
2022-01-18

(133 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Microvolume Luer Access Device is a valve intended for the aspiration, injection, or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector. The Microvolume Luer Access Device may be used with power injectors at a maximum pressure of 400 psi and a maximum flow rate of 10ml/sec.

Device Description

The Microvolume Luer Access Device (LAD) consists of a body, piston, spike/nut, and a male luer lock cover. The Microvolume LAD is a neutral displacement needleless connector intended to provide needle-free access to IV gravity sets, pump sets and extension sets for the administration of IV fluids and blood. The Microvolume LAD may be used with a power injector. The Microvolume LAD is individually packaged and supplied as a sterile, nonpyrogenic, single use disposable device.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device (Microvolume Luer Access Device) and its comparison to a predicate device, focusing on demonstrating substantial equivalence. However, it does not contain information about the acceptance criteria and study proving device performance as requested in the prompt. The document states that the device was evaluated using "Functional performance bench testing" and "No clinical was performed" because the device does not require clinical studies to demonstrate substantial equivalence.

Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor details about sample size (test or training set), data provenance, number or qualifications of experts, adjudication methods, MRMC studies, standalone performance, or how ground truth was established, as these aspects relate to clinical studies or a more detailed performance evaluation that is not present in this 510(k) summary.

The document lists several standards and tests performed, but these are part of the bench testing to show the device meets certain functional requirements and not specifically "acceptance criteria" against which a study directly proves performance in the context of clinical outcomes or diagnostic accuracy.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 18, 2021

B. Braun Medical Inc. Kimberly Smith Sr. Regulatory Affairs Specialist 901 Marcon Blvd. Allentown, Pennsylvania 18109-9341

Re: K212842

Trade/Device Name: Microvolume Luer Access Device Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: December 15, 2021 Received: December 17, 2021

Dear Kimberly Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gang Peng For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212842

Device Name Microvolume Luer Access Device

Indications for Use (Describe)

The Microvolume Luer Access Device is a valve intended for the aspiration, or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector. The Microvolume Luer Access Device may be used with power injectors at a maximum pressure of 400 psi and a maximum flow rate of 10ml/sec.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K212842

I. Submitter Information

Name:B. Braun Medical Inc.
Address:901 Marcon BoulevardAllentown, PA 18109-9341
Contact Person:Kimberly SmithSr. Regulatory Affairs Specialist
Telephone Number:(610) 596-2326
Fax Number:(610) 266-4962
Emailkim.smith@bbraunusa.com
Date Prepared:January 13, 2022

II. Device Name and Classification

Device Trade Name:Microvolume Luer Access Device
Common Name:Luer Access Device, Needleless Connector; Needle-freeInjection Site; Needle-free Luer Access Device; Needle-freeconnector; Valve
Classification Name:Intravascular administration set; §21 CFR 880.5440
Regulatory ClassClass II
Product Code:FPA

III. Predicate Device

Device Trade Name:Caresite Luer Access Device
Common Name:Needleless Connector; Needle-free Injection Site; Needle-free Luer Access Device
Classification Name:Intravascular administration set; §21 CFR 880.5440
Regulatory ClassClass II
Product Code:FPA
510(k) Number:K140311, B. Braun Medical Inc.
Reference Device Trade Name:Clave Connector
Classification Name:Intravascular administration set; §21 CFR 880.5440
510(k) Number:K970855, ICU Medical, Inc.

{4}------------------------------------------------

IV. Device Description

The Microvolume Luer Access Device (LAD) consists of a body, piston, spike/nut, and a male luer lock cover. The Microvolume LAD is a neutral displacement needleless connector intended to provide needle-free access to IV gravity sets, pump sets and extension sets for the administration of IV fluids and blood. The Microvolume LAD may be used with a power injector. The Microvolume LAD is individually packaged and supplied as a sterile, nonpyrogenic, single use disposable device.

V. Indications for Use / Intended Use

Indications for Use

The Microvolume Luer Access Device is a valve intended for the aspiration, injection, or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector. The Microvolume Luer Access Device may be used with power injectors at a maximum pressure of 400 psi and a maximum flow rate of 10ml/sec.

Intended Use

The intended use of the Microvolume Luer Access Device is to provide a sterile needle-free fluid pathway for the administration or aspiration of blood, IV therapy, or medications as prescribed by the physician. Fluids are administered to the patient through a catheter or cannula. The device can be utilized as a conduit between two devices, in addition to being used to deliver contrast media via power injection. This is a general use device that may be used for any patient population with consideration given to the adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.

{5}------------------------------------------------

VI. Comparison of Technological Characteristics with the Predicate Device

The similarities and differences of the technological characteristics between the predicate and the subject device are summarized below.

Similarities:

  • · The subject Microvolume Luer Access Device and the predicate device are identical in their intended use.
  • · Mode of fluid delivery is the same for both the predicate and subject devices. Both devices are used for gravity/pump administration.
  • · Conditions of use are the same for both the predicate and subject devices. Both devices are used by the same type of healthcare professionals under the same conditions of use.
  • · Sterilization method, SAL Level and sterilization cycle are the same for both the predicate and subject devices.
  • · Interfaces are similar between the subject device and predicate device. Both devices contain threaded female and male ports intended for use with associated luers.
  • · Mechanical and Performance Specifications are similar between the subject device and predicate device. The predicate device performance testing and referenced standards were utilized as a guide to establish the testing for the subject device. The subject device meets all the standards that were met by the predicate. Additional standards were considered for performance testing of the subject device.

Differences:

  • · Principle of operation is different but achieves the same intended use. The difference does not raise different questions of safety and effectiveness and is evaluated by design verification testing.
  • · Physical characteristics and dimensions are different but do not impact the intended use. The differences do not raise different questions on safety and effectiveness and is evaluated by performance testing.
  • · Displacement Volume is different but does not impact the intended use. The difference does not raise different questions on safety and effectiveness and is evaluated by performance testing.

{6}------------------------------------------------

Proposed DevicePredicate DeviceComparison
Microvolume Luer Access(K140311)
DeviceCaresite Luer Access Device
ManufacturerB. Braun Medical Inc.B. Braun Medical Inc.Same
Intended UseThe intended use of theThe intended use of theSame
Microvolume Luer AccessCaresite Luer Access Device is
Device is to provide a sterileto provide a sterile needle-free
needle-free fluid pathway forfluid pathway for the
the administration oradministration or aspiration of
aspiration of blood, IVblood, IV therapy, or
therapy, or medications asmedications as prescribed by
prescribed by the physician.the physician. Fluids are
Fluids are administered to theadministered to the patient
patient through a catheter orthrough a catheter or cannula.
cannula. The device can beThe device can be utilized as a
utilized as a conduit betweenconduit between two devices, in
two devices, in addition toaddition to being used to
being used to deliver contrastdeliver contrast media via
media via power injection.power injection. This device
This device may be used formay be used for any patient
any patient population withpopulation with consideration
consideration given to thegiven to the adequacy of
adequacy of vascularvascular anatomy and
anatomy and appropriatenessappropriateness for the solution
for the solution being infusedbeing infused and duration of
and duration of therapy.therapy.
Indications for UseThe Microvolume LuerThe Caresite Luer AccessDifferent
Access Device is a valveDevice is a valve intended forDifference in
intended for the aspiration,the aspiration, injection, orpower injector
injection, or gravity/pumpgravity/pump flow of IV fluidsmaximum flow rate
flow of IV fluids and bloodand blood upon insertion of afalls within the
upon insertion of a male luermale luer connector. Theindications for use
connector. The MicrovolumeCaresite Luer Access Deviceof the predicate
Luer Access Device may bemay be used with powerdevice, therefore;
used with power injectors at ainjectors at a maximumthe two devices
maximum pressure of 400 psiand a maximum flow rate ofpressure of 400 psi and amaximum flow rate ofhave equivalentindications for use.
10ml/sec.15ml/sec.
Mode of Fluid DeliveryGravity/Pump AdministrationGravity/Pump AdministrationSame
Components of Device3-piece assembly containing:3-piece assembly containing:Same
1. Body1. Body
2. Piston2. Piston
3. Luer Nut3. Luer Nut
4. Male luer lock vented4. Male luer lock vented cover
cover
Summary of nonclinicaltests for determinationof substantialequivalence• Flow Rate & CapacityAttributes• Seal and Snap Tests• Durability Testing• Seal Tests - PostDurability• Power Injection• 7 Day IPA & CHGCompatibility• IPA Exposure, InfusateCompatibility• Duration & ConnectorRemoval• ISO 80369-7 ComplianceTesting• Associate Male ID Testing• 7 Day Disinfection CapCompatibility with SwabCap, Curos Jet, andCareCap. Disinfection CapValve Activation• Particulate Contamination• Cap Assessment after ShipTest• Mechanical Hemolysis• Microbial Ingress Testing• Tensile Testing• Real-Time Aging• Displacement Volume• Flushability• Ship Testing• Flow Rate & CapacityAttributes• Seal and Weld Tests• Durability Testing• Seal Tests - Post Durability• Power Injection• 4 Day IPA & CHGCompatibility• IPA Exposure, InfusateCompatibility• Duration• ISO 594 Compliance Testing• Associate Device Testing• Mechanical Hemolysis• Microbial Ingress Testing• Flushability• Ship TestingDifferentThe additionalnonclinical tests donot raise differentquestions on safetyand effectiveness,as demonstrated byperformancetesting.
Physical CharacteristicsMale Luer ID = 0.0673 in.Male Luer OD = 0.1565in.w/0.060 in/in taperFemale ID = 0.1669in.Priming Volume = 0.0345 mL,averageResidual Volume = 0.052 mL,averageMale Luer ID = 0.080 in.Male Luer OD = 0.1565 in.w/0.060 in/in taperFemale ID = 0.1669 in.Priming Volume = 0.22 mL,averageResidual Volume = 0.20 mL,averageDifferentThe physicalcharacteristics donot raise differentquestions on safetyand effectiveness, asdemonstrated byperformance testing.
Materials• Body - Copolyester• Piston - Silicone Elastomer,Self-lubricating• Luer Nut - ABS• Male Luer Lock VentedCover - Polyethylene• Body - Polycarbonate• Piston - Silicone Elastomer,Self-lubricating• Luer Nut - Polycarbonate• Male Luer Lock Vented Cover- PolyethyleneDifferentThe difference inmaterials do notraise differentquestions on safetyand effectiveness asdemonstrated byperformance testingandbiocompatibility.
Patient Contactcategory/durationExternally Communicating,Blood Path Indirectprolonged exposureExternally Communicating,Blood Path Indirectprolonged exposureSame
BiocompatibilityConforms to ISO 10993-1Conforms to ISO 10993-1Same

{7}------------------------------------------------

{8}------------------------------------------------

Performance Data VII.

Performance Testing Bench

Functional performance bench testing was conducted to demonstrate that the Microvolume Luer Access Device performs as intended. No clinical was performed as this device does not require clinical studies to demonstrate substantial equivalence with the predicate device. The following nonclinical performance tests were conducted in support of the substantial equivalence determination:

StandardTest Performed
ISO 8536-4:2019Flow Rate & Capacity Attributes Seal and Snap Tests Seal Test Seal Tests - Post Durability Particulate Contamination Associate Male ID Testing Tensile Testing
ISO 8536-8:2015Seal and Snap Tests Seal Test Seal Tests - Post Durability Associate Male ID Testing Tensile Testing
ISO 8536-9:2015Tensile Testing
ISO 8536-10:2015Seal and Snap Tests Seal Test Seal Tests - Post Durability Associate Male ID Testing Tensile Testing
ISO 10993-4:2017Mechanical Hemolysis
ISO 10993-12:2021
ISO 80369-7:2021Dimensional requirements for Luer connectors Fluid leakage Sub-atmospheric pressure air leakage Stress cracking Resistance to separation from axial load Resistance to separation from unscrewing Resistance to overriding
ISO 11607-1:2020ASTM F1980-16ASTM F1886M-16ASTM D4169-16ASTM D4332-14ASTM F88/F88M-15ASTM F1929-15ASTM F2096-11ASTM F2638-18Sterile Barrier System Validation

{9}------------------------------------------------

B. Braun Medical Inc. Microvolume Luer Access Device Traditional 510(k) Premarket Notification

USP <161>ANSI/AAMI ST72:2019● Pyrogen
ISO 11135:2014/AMD 1:2018● Sterilization Validation
ISO 10993-7:2008/AMD 1:2019● EO Residuals
B. Braun Medical Inc. deviceperformance test methods● Durability● Microbial Ingress● Displacement Volume● Power Injection● 7 Day IPA & CHG Compatibility● IPA Exposure, Infusate Compatibility● Duration & Connector Removal● Disinfection Cap 7-Day Compatibility with Swab Cap, Curos Jet, and CareCap. Disinfection Cap Valve Activation● Cap Assessment after Ship Test● Flushability
USP <788>● Particulate Contamination

Biocompatibility Testing

The Microvolume Luer Access Device was evaluated according to ISO 10993-1:2018. The following biocompatibility testing was performed with the reference standard utilized:

Test PerformedStandard
Cytotoxicity Study Using the ISO Elution MethodISO 10993-5:2009
Guinea Pig Maximization Sensitization TestISO 10993-10:2010
Intracutaneous Study in RabbitsISO 10993-10:2010
Acute Systemic Toxicity Study in MiceISO 10993-11:2017
Subacute Toxicity Study in RatsISO 10993-11:2017
Hemocompatibility - HemolysisISO 10993-4:2017
Partial Thromboplastin Time (PTT) AssayISO 10993-4:2017
Heparinized Blood Platelet and Leukocyte Count AssayISO 10993-4:2017
Complement Activation AssayISO 10993-4:2017
Material Mediated Pyrogenicity in RabbitsISO 10993-11:2017
FTIR - Confirmation of MaterialsASTM E1252
Section 8; Annex B - Chemical RequirementsISO 8536-4:2020

{10}------------------------------------------------

VIII. Substantial Equivalence

Intended Use/Indications for Use - Discussion of differences and similarities

The intended use of the proposed and the predicate device are identical. The indications for use of the proposed and the predicate device are equivalent and do not create a new intended use:

  • · Both devices are valves intended for the administration of blood. IV therapy, or medications as prescribed by the physician.
  • Both devices can be utilized as a conduit between two devices, in addition to being used to deliver contrast media via power injection.
  • · Both devices are general used device and may be used for any patient population with consideration given to the adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.
  • · The indications for use of the proposed device fall within the intended use of the predicate device and, therefore, the two devices have the same intended use.

The differences in the Indications for Use do not affect safety and effectiveness and do not alter the intended use of the proposed device:

  • · The Indications for Use of the proposed device allows for use with a power injector at a maximum pressure of 400 psi and a maximum flow rate of 10ml/sec and the predicate allows for use with a power injector at a maximum pressure of 400 psi and a maximum flow rate of 15ml/sec.
  • · Both devices are used by the same type of healthcare professionals under the same conditions of use.
  • · Both devices have the same fundamental scientific technology.

Technological Characteristics - Discussion of Differences

Based on the performance data, the Microvolume Luer Access Device was found to have a safety and effectiveness profile similar to the predicate device. The technological characteristics of the proposed device are substantially equivalent to the predicate device as compared in sections VI and VII above.

Conclusion on Substantial Equivalence

The proposed Microvolume Luer Access Device has the same intended use and equivalent Indications for Use as the predicate device. The proposed device has the same mode of fluid delivery as the predicate. The proposed device has similar technological characteristics to the predicate, meets the same relevant performance specifications as the predicate, and the descriptive and performance information provided within this premarket notification demonstrates that:

  • any differences do not raise different questions of safety and effectiveness than that of the . predicate device; and
  • the proposed device is as safe and effective as the legally marketed predicate device. .

The Microvolume Luer Access Device is substantially equivalent to the predicate device.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.