K Number
K212832

Validate with FDA (Live)

Date Cleared
2021-12-14

(98 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

As an NMES device, indications are for the following conditions:

-Relaxation of muscle spasms

-Retardation or prevention of disuse atrophy

-Increased local blood circulation

-Re-Educating muscles

-Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis

-Maintaining or increasing range of motion.

As a TENS device, indications are for the following conditions: -Symptomatic relief and management of chronic intractable pain -Adjunctive treatment for post-surgical and post-traumatic acute pain

Device Description

TimeWaver Frequency is an electrical stimulation device that supports the treatment of specific medical conditions, using currents in the microampere range with different frequencies. It can be applied to different areas of the body and is a multifunctional electrotherapy device with two stimulation channels and two treatment modes that deliver neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulations (TENS).

The TimeWaver Frequency device is a prescription device intended to be used by healthcare providers in a healthcare setting and is not approved for use in a home environment. The use of the device on a particular patient is solely the decision of the medical professional, who is responsible for having the proper training and exercising his or her informed medical judgement.

The principals of electrotherapy correspond to the process observed during voluntary muscle contraction. NMES pulses stimulate motor points of target muscle contraction. This can help reeducate and strengthen muscles following injury or surgery. TENS blocks the pain signal sent from the affected area on nerve pathways.

The TimeWaver Frequency device provides 6 treatment programs for NEMS and two treatment programs for TENS. The NMES programs are for relaxation of muscle spasms, disuse atrophy, blood circulations, re-educating muscles, stimulation of calf muscles, and range of motion. The TENS programs are for chronic pain and acute pain. In all modes adhesive electrodes are used to deliver the stimulation.

The TimeWaver Frequency is controlled by an application software program. A personal computer or a notebook is necessary to install the application software and is not provided with the TimeWaver Frequency system. The TimeWaver Frequency software must be used as a stand-alone application on its own computer, which is not used for other software applications. The user interface has several sections. The title bar contains the current program version, the lower status bar shows if hardware components are correctly connected, and the menu bar allows selection of the section to work within. The software requires setting up a new client to include relevant patient information. Treatments are conducted by selecting the appropriate treatment from the Programs module.

The TimeWaver Frequency device operates using an external power supply and/or an internal battery. It is supplied with instructions for use, the TimeWaver Frequency device, power supply, USB cable, electrode connector cables, 50mm x 50mm and 50mm x 90 mm adhesive electrodes.

AI/ML Overview

Sorry, but the provided text from the FDA 510(k) submission for the TimeWaver Frequency device ({0} through {14}) describes a substantial equivalence determination based on non-clinical testing and comparison to a predicate device, rather than a study with acceptance criteria and reported device performance in terms of clinical outcomes or diagnostic accuracy.

The document states:

  • "TimeWaver has completed comprehensive design verification testing, electrical safety and electromagnetic compatibility testing, software verification and validation to ensure that the TimeWaver Frequency device performs as intended." (Page {5})
  • "All necessary non-clinical testing was conducted on the TimeWaver Frequency device to confirm that the device performs as intended." (Page {13})
  • The enumerated nonclinical tests include electrical safety and electromagnetic compatibility, biocompatibility verification, and software verification and validation. (Page {14})

There is no information within the provided text about:

  • Acceptance criteria related to clinical performance (e.g., sensitivity, specificity, accuracy).
  • A study specifically measuring device performance against such criteria.
  • Sample sizes for a clinical test set or data provenance.
  • Number of experts, their qualifications, or adjudication methods for ground truth.
  • A multi-reader multi-case (MRMC) comparative effectiveness study or related effect sizes.
  • A standalone algorithm performance study.
  • The type of ground truth used (expert consensus, pathology, outcomes data).
  • Sample size for a clinical training set or how its ground truth was established.

The FDA 510(k) clearance process for this type of device (Powered Muscle Stimulator and Transcutaneous Electrical Nerve Stimulator) often relies on demonstrating substantial equivalence to a legally marketed predicate device through engineering and performance bench testing, rather than new clinical trials demonstrating efficacy or safety if the indications for use and technological characteristics are similar and do not raise new questions of safety or effectiveness.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 14, 2021

TimeWaver Production GmbH % Douglas Herrington Principal Consultant Herrington Consulting LLC 2885 Sanford Ave SW 43083 Grandville, Michigan 49418

Re: K212832

Trade/Device Name: TimeWaver Frequency Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: IPF. GZJ Dated: August 2, 2021 Received: September 7, 2021

Dear Douglas Herrington:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CDR Jitendra Virani, MS Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212832

Device Name TimeWaver Frequency

Indications for Use (Describe)

As an NMES device, indications are for the following conditions:

-Relaxation of muscle spasms

-Retardation or prevention of disuse atrophy

-Increased local blood circulation

-Re-Educating muscles

-Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis

-Maintaining or increasing range of motion.

As a TENS device, indications are for the following conditions: -Symptomatic relief and management of chronic intractable pain -Adjunctive treatment for post-surgical and post-traumatic acute pain

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

[As required by 21 CFR 807.92]

General Information

Date: December 13, 2021

Type of 510(k) Submission: Traditional 510(k) Notification

Sponsor & Manufacturing Address:

TimeWaver Production GmbH Schloss Kränzlin, Darritzer Straße 6 16818 Kränzlin, Germany Phone: +49 3391 40022-11 FAX: +49 3391 40022-99 Registration Number: 3009411292

Official Contact Name

Douglas Herrington Principal Consultant Herrington Consulting LLC 2885 Sanford Ave, SW#43083 Grandville, MI 49418-1342 Telephone: 1.248.369.5564 Fax: 1.877.881.4412 E-mail: dgls herrington@yahoo.com

Regulatory Information

Device Name: TimeWaver Frequency Model Number: REF 0002

Panel Code: 89 Physical Medicine Classification: Class II Product Code(s) and Regulation Numbers:

Product CodeRegulations NumberRegulation Name
IPF890.5850Powered MuscleStimulatorPrimary Code
GZJ882.5890Transcutaneouselectrical nervestimulatorSecondary Code

Predicate Device 510k Number: K150413. Manufacturer: CyMedica Orthopedics Model: QB1 Powered Muscle Stimulator System (NMES) & Transcutaneous Electrical Nerve Stimulator (TENS) QB-1000 Classification: Class II

{4}------------------------------------------------

DEVICE DESCRIPTION [807.92(a)(4)]

TimeWaver Frequency is an electrical stimulation device that supports the treatment of specific medical conditions, using currents in the microampere range with different frequencies. It can be applied to different areas of the body and is a multifunctional electrotherapy device with two stimulation channels and two treatment modes that deliver neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulations (TENS).

The TimeWaver Frequency device is a prescription device intended to be used by healthcare providers in a healthcare setting and is not approved for use in a home environment. The use of the device on a particular patient is solely the decision of the medical professional, who is responsible for having the proper training and exercising his or her informed medical judgement.

The principals of electrotherapy correspond to the process observed during voluntary muscle contraction. NMES pulses stimulate motor points of target muscle contraction. This can help reeducate and strengthen muscles following injury or surgery. TENS blocks the pain signal sent from the affected area on nerve pathways.

The TimeWaver Frequency device provides 6 treatment programs for NEMS and two treatment programs for TENS. The NMES programs are for relaxation of muscle spasms, disuse atrophy, blood circulations, re-educating muscles, stimulation of calf muscles, and range of motion. The TENS programs are for chronic pain and acute pain. In all modes adhesive electrodes are used to deliver the stimulation.

The TimeWaver Frequency is controlled by an application software program. A personal computer or a notebook is necessary to install the application software and is not provided with the TimeWaver Frequency system. The TimeWaver Frequency software must be used as a stand-alone application on its own computer, which is not used for other software applications. The user interface has several sections. The title bar contains the current program version, the lower status bar shows if hardware components are correctly connected, and the menu bar allows selection of the section to work within. The software requires setting up a new client to include relevant patient information. Treatments are conducted by selecting the appropriate treatment from the Programs module.

The TimeWaver Frequency device operates using an external power supply and/or an internal battery. It is supplied with instructions for use, the TimeWaver Frequency device, power supply, USB cable, electrode connector cables, 50mm x 50mm and 50mm x 90 mm adhesive electrodes.

Indication for Use Statement

As an NMES device, indications are for the following conditions:

    1. -Relaxation of muscle spasms
    1. Retardation or prevention of disuse atrophy
    1. Increased local blood circulation
    1. -Re-Educating muscles
    1. -Immediate post-surgical stimulation of leg and arm muscles
    1. Maintaining or increasing range of motion.

{5}------------------------------------------------

Treatment ProgramPulse ShapeDurationFrequencyPulse WidthDuty CycleWork CycleRelaxation TimeContraction TimeRest TimeIndication Numbers
NMES Post-OpMonophasic20 min50 pps5 ms25 %13 s10 s3 s2 cycles3.4 s1, 2, 3, 4, 5, 6
NMES StrengthMonophasic20 min50 pps5 ms25 %13 s10 s1 s5 cycles1.4 s1, 2, 3, 4, 5, 6

As a TENS device, indications are for the following conditions:

    1. Symptomatic relief and management of chronic intractable pain
    1. Adjunctive treatment for post-surgical and post-traumatic acute pain
TENS Pain ManagementBiphasic. symmetrical20 min 100 pps1 ms20%Continuous1 4 msставите седание в седание в сединание в седение в седение в сединание в седение в седение в сединание в сединание в седение в сединание в сединание в сединание в сединание вવી છ

Comparison of Technological Characteristics with the Predicate Device

Considering technological characteristics, the TimeWaver Frequency device is similar to the predicate with only minor differences.

The TimeWaver Frequency and CyMedica QB-1 devices are both electrotherapy devices with the two treatment modes of neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS). They are both prescription devices intended for use under the direction of a medical provider and have the same indications for use. In addition, both products have the same product codes IPF (CFR 890.5850) and GZJ (CFR 882.5890).

In general, the TimeWaver Frequency output parameters are substantially the same as the output parameters for CyMedica OB-1. For example, the maximum current density (mA/cm2) for TimeWaver Frequency @ 500 ohms is 0.16, while the CyMedica QB-1 is 0.27. The maximum average power density (mW/cm2) @500 ohms is 2.0 for TimeWaver Frequency, whereas the CyMedica QB-1 maximum average power density @500 ohms is 1.0. The TimeWaver Frequency and CyMedica OB-1 both use pulsed monophasic and symmetric biphasic to deliver treatment programs.

TimeWaver has completed comprehensive design verification testing, electrical safety and electromagnetic compatibility testing, software verification and validation to ensure that the TimeWaver Frequency device performs as intended. The patient-contacting material has been verified to demonstrate acceptable biocompatibility. The electrodes and cables comply with 21 CFR 898, so they do not raise new questions of safety or effectiveness.

The minor differences in labeling and technological characteristics between the proposed device and the predicate device have been evaluated and determined to not raise different questions of safety or effectiveness. As such, the proposed TimeWaver Frequency is substantially equivalent to the predicate device.

The following table summarizes the specifications and features of the proposed TimeWaver Frequency device and the predicate device.

{6}------------------------------------------------

Substantial Equivalence Summary

Basic Unit Characteristics
510(k) NumberCurrent SubmissionK150413Comparison
Device Name/ModelTimeWaver Frequency, REF 0002QB1 & QB-1000, (NEMS) & (TENS)N/A
Indications for UseAs an NMES device, indications are for the following conditions:-Relaxation of muscle spasms-Retardation or prevention of disuse atrophy-Increased local blood circulation-Re-Educating muscles-Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis-Maintaining or increasing range of motion.As a TENS device, indications are for the following conditions:-Symptomatic relief and management of chronic intractable pain-Adjunctive treatment for post-surgical and post-traumatic acute painThe QB1 Powered Muscle Stimulator System (NMES) &Transcutaneous Electrical Nerve Stimulator System (TENS); QB-1000 is a multifunctional electrotherapy device with two treatment modes that allow for neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS).Indications for Use:As an NMES device, indications are for the following conditions:- Relaxation of muscle spasms- Retardation or prevention of disuse atrophy- Increasing local blood circulationSame

{7}------------------------------------------------

ManufacturerTimeWaver Production GmbHCyMedica Orthopedics IncN/A
Power Source:Power Supply Adapter/BatteryBatteryDifferent, but does notraise different questionsof safety andeffectiveness
- Battery
- Number11
- Size3200 mAh, 3.7 V3.7V
- TypeLithium PolymerLithium Ion
- Power Supply AdapterDifferent, but does not
- Input100 - 240 V AC, 50 - 60 Hz, 0.5 AN/Araise different questions
- Output9 V DC, 2.0 AN/Aof safety andeffectiveness
- Method of Line Current IsolationPower Supply AdapterN/ADifferent, but does notraise different questionsof safety andeffectiveness
- Patient Leakage Current (uA)Same
- normal Condition, uA<100μAPass
- Single Fault Condition, uA<100μΑPass
Number of Output Modes22Same
Number of Output Channels22Same
- Synchronous or AlternatingAlternatingUnknownDifferent, but does notraise different questionsof safety andeffectiveness
- Method of Channel Isolationgalvanic isolationUnknownDifferent, but does notraise different questionsof safety andeffectiveness
Regulated Current or RegulatedVoltagebothbothSame
Software/Firmware/MicroprocessorControl?yesyesSame
Automatic Overload TripyesyesSame
Automatic No Load TripyesyesSame
Automatic Shut OffYes, PC-softwareYes, PC-softwareSame
User Override ControlYes, PC-softwareUnknownDifferent, but does notraise different questionsof safety andeffectiveness
Indicator Display
On/Off Status?yesyesSame
Low Battery?yesyesSame
Voltage/Current Level?yesUnknownDifferent, but does notraise different questionsof safety andeffectiveness
Compliance with VoluntarystandardsIEC 60601-1IEC 60601-1-2IEC 60601-1-6IEC 60601-2-10ISO 10993-1: 2009ISO 10993-5IEC 60601-1IEC 60606-1-2IEC 60601-2-10IEC 60601-1-6IEC 60601-1-11IEC 62366: 2007ISO 10993-1: 2009ISO 10993-5: 2009ISO 10993-10: 2010ISO 14971: 2007Different, but does notraise different questionsof safety andeffectiveness

{8}------------------------------------------------

{9}------------------------------------------------

510(k) Summary: K212832

Output SpecificationComparison
510(k) NumberCurrent SubmissionK150413
Product Code (s)IPF, GZJIPF (NMES), GZJ (TENS)N/A
Device NameTimeWaver FrequencyQB1 & QB1000N/A
ManufacturerTimeWaver Production GmbHCyMedica Orthopedics IncN/A
Program Modes/NamepulsedmonophasicsymmetricbiphasicpulsedmonophasicsymmetricbiphasicNEMSPost-OpNEMSStrengthTENSN/A
Output Mode (voltage or Current)CurrentCurrentVoltageVoltageBothBothBothDifferent,but doesnot raisedifferentquestionsof safetyandeffectiveness
WaveformPulsedMonophasicSymmetricBiphasicPulsedMonophasicSymmetricBiphasicPulsedMonophasicPulsedMonophasicSymmetricBiphasicSame
ShaperectanglerectanglerectanglerectangleComplexComplexComplexDifferent,but doesnot raise
differentquestionsof safetyandeffectiveness
Max Output Voltage (V) +/- 20%
500 Ohms0.250.105.02.03.43.40.180Different,
2k Ohms1.20.45.02.06.16.10.190but doesnot raisedifferentquestionsof safetyandeffectiveness
10k Ohms12.02.05.02.08.58.50.200
Max Output Current (mA) +/- 20%
500 Ohms0.50.210.04.06.86.80.36Different,
2k Ohms0.50.22.51.03.03.00.10but doesnot raise
10k Ohms0.50.20.50.20.90.90.02differentquestionsof safetyandeffectivenessss
Frequency, Hz50100501005050100Same
Duration of primary (depolarizing)phase (u sec)500010005000100050005000N/A(continuous)Different,but doesnot raisedifferentquestionsof safetyand
effectiveness
Pulse Duration, uSec5000100050001000500050001000Same
Net Charge uC per pulse2.50(symmetricBiphasic)500(symmetricBiphasic)1261260(symmetricBiphasic)Different,but doesnot raisedifferentquestionsof safetyandeffectiveness
Maximum Phase Charge (uC), 500Ohms2.50.250412612643Different,but doesnot raisedifferentquestionsof safetyandeffectiveness
Maximum Current Density,mA/cm² @ 500 Ohms0.0200.0080.4000.1600.2700.2700.014Different,but doesnot raisedifferentquestionsof safetyandeffectiveness
Maximum Average Power Density[mW/cm²] @ 500 Ohms0.0050.0012.0000.3201.0001.0000.0026Different,but doesnot raisedifferentquestionsof safetyandeffectiveness

{10}------------------------------------------------

{11}------------------------------------------------

Page 9 of 12

{12}------------------------------------------------

of safetyandeffectiveness
Electrode Sizes
50 x 50 mm, 25 cm²25.86 cm²Different, but does not
50 x 90 mm, 45 cm²51.61 cm²raise different questions
51.61 cm²of safety andeffectiveness

{13}------------------------------------------------

[As required by 21 CFR 807.92]

PREDICATE DEVICE(S) [807.92(a)(3)]

The TimeWaver Frequency device is substantially equivalent to the predicate device, the CyMedica Orthopedics QB1 powered Muscle Stimulator (NEMS) & Transcutaneous Electrical Nerve Stimulator (TENS) with regard to product labeling, intended use, performance testing, technological and safety characteristics.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES, [807.92(a)(6)]

Regarding technological characteristics, the TimeWaver Frequency device is similar to the predicate's functions with only minor technological differences.

The CyMedical QB-1 product is targeted at knee rehabilitation, while the TimeWaver Frequency is more traditional in its approach to NMES and TENS treatment leaving specific treatment options to the judgement of a medical professional. While the treatment areas are different the output parameters of both devices are quite similar.

The CyMedica device provides 3 electrodes, where the TimeWaver Frequency provides 2 electrodes, which is more common in NEMS and TENS devices, and allows the device to be used in multiple locations. The combined surface area of the CyMedica electrodes is 129 cm² and the TimeWaver Frequency is 90 cm². This does not pose new questions of safety or efficacy as the maximum power density and maximum current density is similar to the CyMedica device.

The CyMedica device is used in conjunction with a conductive garment to facilitate electrode placement for knee treatment. As a traditional NEMS/TENS device TimeWaver Frequency does not limit treatment application to the knee area. TimeWaver Frequency use is limited to health care professionals, where CyMedica allows home use, so limiting electrode placement to what is allowable with the garment is not required. This does not pose new questions of safety or efficacy.

The CyMedica device has only limited treatment options available to the user that are activated by simple on product buttons for Post Op or Strength and intensity level adjustment for knee or thigh treatment. The TimeWaver Frequency uses software installed on a PC to control the treatment applications at the direction of the medical professional. As the maximum outputs of the TimeWaver Frequency device are controlled by the software, this does not pose any new questions of safety or efficacy.

TimeWaver has completed comprehensive design verification testing, electrical safety and electromagnetic compatibility testing, software verification and validation to ensure that the TimeWaver Frequency device performs as intended. The TimeWaver Frequency also passed all testing requirements for electrical safety and EMC, and the device adhesive electrode has been previously cleared by FDA. The minor differences in labeling and technological characteristics between the proposed device and the predicate device have been evaluated and determined to not raise different questions of safety or effectiveness. As such, the proposed TimeWaver Frequency is substantially equivalent to the predicate device.

Performance Data [807.92(b)]

All necessary non-clinical testing was conducted on the TimeWaver Frequency device to confirm that the device performs as intended.

Nonclinical Testing Summary:

The nonclinical, bench testing included performance verification to confirm acceptable performance of device features and functions. Other nonclinical safety testing included:

{14}------------------------------------------------

  • · Electrical safety and electromagnetic compatibility testing
  • · Biocompatibility verification
  • · Software verification and validation

The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the TimeWaver Frequency meet the established specifications.

Performance Data: Electromagnetic Compatibility and Electrical Safety

The TimeWaver Frequency device has been thoroughly tested against applicable EN and IEC standards by a third party and found to be compliant with the applicable sections. It was functionally tested and found to be compliant with the specification. The applicable electrical and safety standards met are:

IEC 60601-1:Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC 60601-1-6Medical electrical equipment Part 1-6 General requirements for safety - Collateral Standard: Usability
IEC 60601-2-10:Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
ANSI/AAMI NS4:2013/(R)2017:Transcutaneous electrical nerve stimulators

Software Validation

The TimeWaver Frequency device software was validated following FDA 's Guidance for the Content of Premarket Submissions for Software contained in Medical Devices.

The applicable software validation standards met are:

EN ISO 13485: 2012 Medical devices -- Quality management systems -- Requirements for regulatory purposes EN ISO 14971: 2012 Medical devices - Application of risk management to medical devices EN 62304: 2006 Medical device software -- Software life cycle processes

Biocompatibility Data

Biocompatibility testing is not required because the primary patient contact surface is through the adhesive electrode which was previously cleared by FDA under K091163.

Conclusion

Based upon the intended use and technical information provided in this pre-market notification, the TimeWaver Frequency device has been shown to be substantially equivalent to the currently marketed predicate device.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).