K Number
K212789

Validate with FDA (Live)

Date Cleared
2021-12-01

(91 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The qloves are offered in four sizes: extra-small, small, medium, large, and extra-large.

AI/ML Overview

This document describes the premarket notification (510(k)) for Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs). The core of the submission is to demonstrate substantial equivalence to a legally marketed predicate device (K190860). The "device" in this context refers to the Nitrile Examination Gloves. The acceptance criteria and the study that proves the device meets them are based on non-clinical (bench) testing, as no clinical testing was performed or required for this device type.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally defined by compliance with specific ASTM and ISO standards, and the reported performance are the test results demonstrating this compliance.

Test Methodology / StandardPurposeAcceptance CriteriaReported Device Performance
Biocompatibility
ISO 10993-10:2010 (Irritation & Sensitization)Assesses possible contact hazards from chemicals released from medical devices, which may produce skin and mucosal irritation, eye irritation, or skin sensitization.Skin Sensitization Test: Provided grades less than 1, otherwise sensitization. Skin Irritation Test: If the primary irritation index is 0-0.4, the response category is Negligible. 0.5-1.9 means slight, 2-4.9 means moderate, 5-8 means severe.Skin Sensitization Test: All grades are 0. All animals survived and no abnormal signs were observed. Skin Irritation Test: The primary irritation index is 0. The response of the proposed device was categorized as negligible.
ISO 10993-5:2009 (Cytotoxicity)Describes test methods to assess the in vitro cytotoxicity of medical devices.The viability (%) of the 100% extract of the test article is the final result, and if viability is reduced to <70% of the blank, it has cytotoxic potential.Viability (%) of 100% test article extract is 86.5%. This means the proposed device has no potential toxicity to L-929 in the MTT method.
Physical Performance
ASTM D6124-06 (Residual Powder)Determines the amount of residual powder (or filter-retained mass) found on medical gloves.Powder residue limit of 2.0 mg.0.1 mg / glove.
ASTM D5151-19 (Holes)Covers the detection of holes in medical gloves.Samples number: 125 gloves. AQL: 2.5 (ISO 2859). Criterion $\leq$ 7 gloves for water leakage.No glove water leakage found.
ASTM D6319-19 (Dimensions & Physical Properties)Covers certain requirements for nitrile rubber gloves used in conducting medical examinations and diagnostic and therapeutic procedures.Dimensions: XS: width 70 $\pm$ 10mm, Length $\geq$ 220 mm S: width 80 $\pm$ 10mm, Length $\geq$ 220 mm M: width 95 $\pm$ 10mm, Length $\geq$ 230 mm L: width 110 $\pm$ 10mm, Length $\geq$ 230 mm XL: width 120 $\pm$ 10mm, Length $\geq$ 230 mm Thickness: Finger $\geq$ 0.05 mm, Palm $\geq$ 0.05 mm Physical properties: Before aging: Tensile strength $\geq$ 14MPa, Ultimate Elongation $\geq$ 500% After Accelerated Aging: Tensile strength $\geq$ 14MPa, Ultimate Elongation $\geq$ 400%Dimensions: XS: width: 75-76 mm, Length 234-243 mm S: width: 86-87 mm, Length 240-248 mm M: width 98-99 mm, Length 246-256 mm L: width 109-111 mm, Length 250-258 mm XL: width 115-118mm, Length 253-255mm Thickness: Finger: 0.091-0.127 mm (range across sizes) Palm: 0.071-0.094 mm (range across sizes) Physical properties: Before aging: Tensile strength 21.6-29.5 MPa, Ultimate Elongation 536% - 571% After Accelerated Aging: Tensile strength 24.1-31.0 MPa, Ultimate Elongation 479% - 519%
Chemotherapy Drug Permeation
ASTM D6978-05 (Reapproved 2019)Assesses resistance of medical gloves to permeation by chemotherapy drugs. Minimum breakthrough detection times for various chemotherapy drugs.Carmustine (BCNU) 3.3 mg/ml: > 15.1 Minutes Cisplatin 1.0 mg/ml: > 240 Minutes Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes Carboplatin 10.0 mg/ml: > 240 Minutes Doxorubicin HCl 2.0 mg/ml: > 240 Minutes Etoposide 20.0 mg/ml: > 240 Minutes Fluorouracil 50.0 mg/ml: > 240 Minutes Paclitaxel 6.0 mg/ml: > 240 Minutes Thio Tepa 10.0 mg/ml: > 32.8 MinutesCarmustine (BCNU) 3.3 mg/ml: 15.1 minutes (individual results: 15.1, 16.4, 16.6) Cisplatin 1.0 mg/ml: > 240 Minutes Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes Carboplatin 10.0 mg/ml: > 240 Minutes Doxorubicin HCl 2.0 mg/ml: > 240 Minutes Etoposide 20.0 mg/ml: > 240 Minutes Fluorouracil 50.0 mg/ml: > 240 Minutes Paclitaxel 6.0 mg/ml: > 240 Minutes Thio Tepa 10.0 mg/ml: 32.8 minutes (individual results: 48.6, 32.8, 46.1) (Note: A warning is included for Carmustine (BCNU) and Thiotepa due to low permeation times.)

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size for Test Set:
    • Freedom from Holes (ASTM D5151-19): 125 gloves.
    • Chemotherapy Drug Permeation (ASTM D6978-05): The table provides individual breakthrough detection times for some drugs, indicating multiple samples were tested (e.g., three reported times for Carmustine and Thio Tepa). Specific sample sizes for each drug are not explicitly stated beyond these examples.
    • Other tests (Biocompatibility, Physical Properties, Residual Powder): Specific sample sizes are not detailed in the provided text, but testing was conducted "in accordance with" the listed standards, implying standard-appropriate sample sizes were used.
  • Data Provenance: This is a 510(k) submission for a medical device (nitrile examination gloves). The testing is non-clinical (bench testing), meaning it was performed in a laboratory setting to assess the physical and chemical properties of the gloves. The country of origin of the data is not explicitly stated for each test, but the submitting company is 5R MED Instruments (CHENGDU) Co., Ltd. located in China. The testing would have been conducted in a controlled lab environment. The studies are prospective tests performed on the manufactured device to ensure it meets the required standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. For a device like examination gloves, "ground truth" is established by laboratory measurements against defined standards (e.g., strength, thickness, permeation time), not by expert human interpretation of medical images or clinical outcomes. The tests are objective, quantifiable measurements.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies where multiple human readers interpret data (like medical images) and their discrepancies need to be resolved. For non-clinical bench testing, the results are quantitative and directly measured by calibrated equipment and documented procedures.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No. MRMC studies are used to evaluate the diagnostic performance of medical imaging devices, often comparing human readers with and without AI assistance. This submission is for examination gloves, which do not involve such diagnostic tasks.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. "Standalone performance" typically refers to the performance of an AI algorithm without human involvement. This device is not an AI algorithm. Its performance is demonstrated through physical and chemical property testing.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is objective, measurable physical properties and chemical resistance, defined by international standards:

  • Standard-defined thresholds: For properties like tensile strength, elongation, freedom from holes (AQL), residual powder, dimensions (length, width, thickness), and breakthrough times for chemotherapy drugs.
  • Laboratory measurements: Results obtained from standardized test methods (ASTM, ISO).
  • Biocompatibility assessments: Based on established criteria for irritation, sensitization, and cytotoxicity from biological assays.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML model. There is no "training set" in the context of examination gloves.

9. How the Ground Truth for the Training Set was Established

Not applicable. As there is no training set for an AI/ML model, there is no ground truth established for it. The ground truth for the device's performance is based on its physical, chemical, and biological properties measured against established standards during non-clinical testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 1, 2021

5R MED Instruments (CHENGDU) Co., Ltd. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161 Lujiazui East Rd., Pudong Shanghai. 200120 China

Re: K212789

Trade/Device Name: Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: August 27, 2021 Received: September 1, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number K212789

Device Name

Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs)

Indications for Use (Describe)

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes
Carmustine (BCNU)3.3 mg/ml (3,300 ppm)15.1(15.1, 16.4, 16.6) Minutes
Cisplatin1.0 mg/ml(1,000 ppm)> 240 Minutes
Cyclophosphamide (Cytoxan)20.0 mg/ml(20,000 ppm)> 240 Minutes
Carboplatin10.0 mg/ml(10,000 ppm)> 240 Minutes
Doxorubicin HCl2.0 mg/ml(2,000 ppm)> 240 Minutes
Etoposide20.0 mg/ml(20,000 ppm)> 240 Minutes
Fluorouracil50.0 mg/ml(50,000 ppm)>240 Minutes
Paclitaxel6.0 mg/ml(6,000 ppm)>240 Minutes
Thio Tepa10.0 mg/ml(10,000 ppm)32.8 (48.6, 32.8, 46.1) Minutes

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 15.1 Minutes (min.); Thio Tepa 10.0 mg/ml 32.8 Minutes (min.).

Warning: Please do not use with Carmustine (BCNU) and Thiotepa.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K212789

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

5R MED Instruments (CHENGDU) Co., Ltd. Name: Address: Building 9, Section 1, No.618 of West Kelin Road, Chengdu Cross-Straits Technology Industry Development Zone, Wenjiang District, Chengdu, Sichuan Province, 611130, China Mr. Pan Yuzhang Contact: Date of Preparation: 2021.08.12

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 Lujiazui East Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Nitrile Examination Gloves (Tested for Use with Chemotherapy Trade name: Drugs) Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): XS, S, M, L, XL

3.0 Classification

Production code: LZA, LZC Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Ever Growth (Vietnam) Co., Ltd.

Device: Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Disposable Powder Free

{4}------------------------------------------------

Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Orange Color

510(k) number: K190860

5.0 Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The qloves are offered in four sizes: extra-small, small, medium, large, and extra-large.

6.0 Indication for Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs

Chemotherapy DrugConcentrationBreakthrough DetectionTime in Minutes
Carmustine (BCNU)3.3 mg/ml (3,300 ppm)15.1(15.1,16.4,16.6)
Carboplatin10.0 mg/ml(10,000 ppm)> 240
Cyclophosphamide(Cytoxan)20.0 mg/ml(20,000 ppm)> 240
Cisplatin1.0 mg/ml(1,000 ppm)> 240
Doxorubicin HCI2.0 mg/ml(2,000 ppm)> 240
Etoposide20.0 mg/ml(20,000 ppm)> 240
Fluorouracil50.0 mg/ml(50,000 ppm)> 240
Paclitaxel6.0 mg/ml(6,000 ppm)> 240
Thio Tepa10.0 mg/ml(10,000 ppm)32.8(48.6, 32.8, 46.1)

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 15.1 Minutes (min.); Thio Tepa 10.0 mg/ml 32.8 Minutes (min.).

Warning: Please do not use with Carmustine (BCNU) and Thiotepa.

7.0 Technological Characteristic Comparison Table

{5}------------------------------------------------

Subject DevicePredicate Device
ItemK212789(K190860)Remark
Product CodeLZA,LZCLZA,LZCSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseA patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.Same
Powdered orPowered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
SterilityNon-SterileNon-SterileSame
Labeling InformationSingle-use indication, powder free, device color, device name, glove size and quantity,Non-Sterile, a statement of standard ASTM D6978-05 compliance and a summary of the testing results.Single-use indication, powder free, device color, device name, glove size and quantity, Non-Sterile, a statement of standard ASTM D6978-05 compliance and a summary of the testing results.Similar

Table1-General Comparison

Table2 Device Dimensions Comparison
PredicateDevice(K190860)DesignationSizeTolerance
XSSMLXL
Length, mm230230230230230min
Width, mm708095110120±10
Thickness, mm:

Table2 Device Dimensions Comparison

{6}------------------------------------------------

Finger0.05min
Palm0.05min
DesignationSize
XSSMLXLTolerance
Subject DeviceLength, mm220220230230230min
Width, mm708095110120±10
Thickness, mm:
Finger0.05min
Palm0.05min
RemarkDifferent

Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319-19,so the differences do not raise any new safety or performance questions.

ItemSubject devicePredicate device(K190860)Remark
ColorantBlueWhite, OrangeDifferent 1
PhysicalPropertiesBeforeAgingTensileStrength14MPa, min14MPa, minSame
UltimateElongation500% min500% minSame
AfterAgingTensileStrength14MPa, min14MPa, minSame
UltimateElongation400%min400%minSame
Comply with ASTM D6319Comply with ASTMD6319Same
Freedom from HolesBe free from holes when testedin accordance withASTMD5151 AQL=2.5Be free from holeswhen tested inaccordance withASTMD5151 AQL=2.5Same
Powder Content0.1mg per glove, Meet therequirements of ASTM D6124Meet the requirementsof ASTM D6124Same
Chemotherapy DrugsTested with MinimumBreakthroughDetection Time asTested per ASTM D6978Carmustine (BCNU) 3.3mg/ml: 15.1 Minutes (min.)Carmustine (BCNU)3.3 mg/ml:White:11.8 Minutes;Orange:31.6MinutesSimilar
Cisplatin 1.0 mg/ml: > 240MinutesCisplatin 1.0 mg/ml:>240 MinutesSame
Cyclophosphamide (Cytoxan)20.0 mg/ml: > 240 MinutesCyclophosphamide(Cytoxan)20.0 mg/ml: > 240Same

Table3 Performance Comparison

{7}------------------------------------------------

Minutes
Carboplatin 10.0 mg/ml: > 240MinutesDacarbazine (DTIC)10.0 mg/ml:>240 MinutesDifferent 2
Doxorubicin HCI 2.0 mg/ml: >240 MinutesDoxorubicinHydrochloride 2.0mg/ml: >240 MinutesSame
Etoposide 20.0 mg/ml: > 240MinutesEtoposide (Toposar)20.0mg/ml: >240 MinutesSame
Fluorouracil 50.0 mg/ml: >240 MinutesFluorouracil 50.0mg/ml: >240MinutesSame
Paclitaxel 6.0 mg/ml: >240MinutesPaclitaxel (Taxol) 6.0mg/ml:>240 MinutesSame
Thio Tepa 10.0 mg/ml: 32.8Minutes (min.)Thio-Tepa 10.0 mg/ml:White: 16.9 Minutes;Orange: 72.5 MinutesSimilar

Analysis:

Different 1: The color of the subject device is different of that of the predicate. Biocompatibility testing was successfully completed for the subject device, demonstrating that any color differences do not affect the safety of the proposed device.

Different 2: The chemotherapy drug is different with that of the predicate, but they all meet the requirements of ASTM D6978-05(2019),so the differences do not raise any new safety or performance questions.

Table4 Safety Comparison
ItemSubject devicePredicated device (K190860)Remark
MaterialNitrileNitrileSame
BiocompatibilityIrritation (ISO 10993-10:2010BiologicalEvaluation ofMedical Devices -Part 10: Tests ForIrritation And SkinSensitization)Under the conditions ofthe study, notan irritantComply withISO10993-10Same
Sensitization (ISO10993-10:2010Under conditions of

Table4 Safety Comparison

{8}------------------------------------------------

BiologicalEvaluationofMedical Devices -Part 10: Tests ForIrritation And SkinSensitization)the study, not asensitizer.
Cytotoxicity (ISO10993-5:2009BiologicalEvaluationofMedical Devices -Part 5: Tests For InVitro Cytotoxicity)Underconditions ofthe study, thedevice extractis not cytotoxicUnderconditions ofthe study, thedevice extractis not cytotoxicSame
TestMethodology /StandardPurposeAcceptance CriteriaResults
ISO 10993-10:2010BiologicalEvaluation OfMedical Devices- Part 10: TestsFor Irritation AndSkinSensitization.This part of ISO10993 assessespossible contacthazards fromchemicals releasedfrom medicaldevices,which may produceskin and mucosalirritation, eyeirritation or skinsensitization.Skin Sensitization Test:providedgrades less than 1,otherwise sensitization.Skin Irritation Test:If the primary irritationindex is 0-0,4, theresponse category isNegligible.0,5-1,9 means slight2-4,9 means moderate5-8 means severeAll grades are 0.All animals were survivedand no abnormal signs wereobserved during the study.The primary irritation indexis 0.The response of theproposed device wascategorized as negligibleunder the test condition
ISO 10993-5:2009 BiologicalEvaluation OfMedical Devices- Part 5: TestsFor In VitroCytotoxicityThis part of ISO10993 describestest methods toassess the in vitrocytotoxicity ofmedical devices.The viab.% of the 100%extract of the test articleis the final result, and ifviability is reduced to<70% of the blank, ithas cytotoxic potential.Viab.% of 100% test articleextract is 86.5%It means the proposeddevice have no potentialtoxicity to L-929 in the MTTmethod
ASTM D6124-06(Reapproved2017), StandardTest Method forResidual Powderon MedicalGlovesThis standard isdesigned todetermine theamount ofresidualpowder (or filter-retained mass)foundpowder residue limit of2.0 mg0.1 mg /glove
on medical gloves
ASTM D5151-19Standard TestMethod forDetection ofHoles in MedicalGlovesThis test methodcovers thedetection of holesinmedical gloves.Samples number: 125glovesAQL: 2.5 (ISO 2859)Criterion $\leq$ 7 gloves forwater leakageno glove water leakagefound
ASTM D6319-19StandardSpecification forNitrileThis specificationcovers certainrequirements fornitrile rubber glovesSterility: no needFreedom from holes: pl.Refer to table 3N.A.
ExaminationGloves forMedicalApplicationused in conductingmedicalexaminations anddiagnostic andtherapeuticprocedures.Dimensions:XS: width 70 $\pm$ 10mmLength $\geq$ 220 mmS: width 80 $\pm$ 10mmLength $\geq$ 220 mmM: width 95 $\pm$ 10mmDimensions:XS: width: 75-76 mmLength 234-243 mmThickness:Finger 0.091-0.100 mmPalm 0.071-0.079 mm
Length $\geq$ 230 mmL: width 110 $\pm$ 10mmLength $\geq$ 230 mmXL: width 120 $\pm$ 10mmLength $\geq$ 230 mmThickness:Finger $\geq$ 0.05 mmS: width: 86-87 mmLength 240-248 mmThickness:Finger 0.098-0.114 mmPalm 0.084-0.094mm
Palm $\geq$ 0.05 mmPhysical properties:Before agingTensile strength $\geq$14MPaM: width 98-99 mmLength 246-256 mmThickness:Finger 0.106-0.124 mmPalm 0.077-0.093 mm
Ultimate Elongation $\geq$500%After Accelerated AgingTensile strength $\geq$14MPaUltimate Elongation $\geq$L: width 109-111 mmLength 250-258 mmThickness:Finger 0.111-0.127 mmPalm 0.080-0.088 mm
400%Powder-free Residue:pl. Refer to table 3XL: width 115-118mmLength 253-255mmThickness:Finger 0.112-0.124 mm
Palm 0.081-0.093 mm
Physical properties:Before agingTensile strength 21.6-29.5 MPaUltimate Elongation 536% - 571%After Accelerated AgingTensile strength 24.1-31.0 MPaUltimate Elongation 479% - 519%
Powder-free Residue:pl. Refer to table 3
ChemotherapyDrugsTested withMinimumBreakthroughDetection TimeCarmustine (BCNU) 3.3 mg/ml: 15.1 Minutes (min.)
ASTM D6978Cisplatin 1.0 mg/ml: > 240 Minutes
Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes
Carboplatin 10.0 mg/ml: > 240 Minutes
Doxorubicin HCl 2.0 mg/ml: > 240 Minutes
Etoposide 20.0 mg/ml: > 240 Minutes
Fluorouracil 50.0 mg/ml: > 240 Minutes
Paclitaxel 6.0 mg/ml: > 240 Minutes
Thio Tepa 10.0 mg/ml: 32.8 Minutes (min.)

8.0 Summary of Non-Clinical Testing

Biocompatibility Testing

The biocompatibility evaluation for Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs) was conducted in accordance with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

Performance Testing (Bench)

Physical performance qualities of the proposed device were evaluated per ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs.

In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

{9}------------------------------------------------

  • ASTM D6124-06 (Reapproved 2017), Standard Test Method for - Residual Powder on Medical Gloves
  • ASTM D5151-19, Standard Test Method for Detection of Holes in -Medical Gloves.
  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
  • ASTM D6978-05 (Reapproved 2019) ,Standard Practice for -Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

{10}------------------------------------------------

{11}------------------------------------------------

9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs)

{12}------------------------------------------------

is as safe, as effective, and performs as well as or better than the legally marketed predicated device under K190860.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.