K Number
K212779

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-10-05

(399 days)

Product Code
Regulation Number
864.7750
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Coag-Sense Prothrombin Time (PT) / INR Monitoring System for Patient Self-Testing is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in INR units. It uses fresh capillary whole blood.

The Coag-Sense Prothrombin Time (PT) / INR Monitoring System for Patient Self-Testing is intended for use by properly selected and suitably trained patients or the order of the treating physician to monitor patients who are on anticoagulation therapy. Patients should be stabilized on warfarin-type (coumarin) anticoagulation therapy prior to self-testing.

The Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is not intended to be used for screening purposes.

The Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is intended to be used by a single person and should not be shared.

Device Description

The Coaq-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is a portable medical device for the measurement of the Prothrombin Time (PT) using fresh capillary whole blood obtained from a finger stick. The Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is a handheld device that directly detects clot formation. The System measures the PT of fresh capillary whole blood using micromechanical end point detection. The test is performed by inserting a test strip into the meter and applying a drop of blood to the sample receptacle of the disposable test strip. The test strip contains a rotating, spoked wheel that draws the sample into the reaction well after it is applied to the sample receptacle. The spokes rotate across the path of an infrared light beam and mix the liquid sample with the thromboplastin which is dried in the reaction well. When the sample clots, the clot is picked up by the spokes, interrupting the path of the infrared light beam that is detected by the meter.

The PT test and the result is displayed as International Normalized Ratio (INR). The result is date/time stamped and stored in the memory of the meter.

The device is powered by 4 AA batteries. This Coag-Sense Prothrombin Time (PT)//NR Monitoring System for Patient Self-Testing uses the exact same test and control strip as the predicate devices.

AI/ML Overview

The information provided describes a Special 510(k) submission for the Coag-Sense Prothrombin Time (PT) / INR Monitoring System for Patient Self-Testing (K212779). This submission focuses on modifications to the user interface (from touchscreen to button navigation) and power source, asserting substantial equivalence to its predicate device (K183255).

Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

Acceptance Criteria and Reported Device Performance

The document states that "All testing results met the pre-determined acceptance criteria that were established in the test protocols." However, the specific quantitative acceptance criteria for each test and the corresponding reported performance values are not explicitly listed in a detailed table format within the provided text. The text generally states that the device "demonstrated that it met the predetermined acceptance criteria and design specifications."

Based on the available text, a table can be constructed to show the types of tests performed and the general statement about meeting criteria, rather than specific numerical acceptance criteria and performance data.

Acceptance Criteria CategoryReported Device Performance
Functional Testing (Component Testing)Met predetermined acceptance criteria. Device performs as intended.
Functional Testing (Physical/Environmental)Met predetermined acceptance criteria. Device performs as intended.
Electrical Safety & EMC TestingMet predatory acceptance criteria. Device performs as intended, adhering to IEC 61010-1, IEC 61010-2-101, IEC 60606-1-2 standards.
Cybersecurity Evaluation & Risk ManagementNo significant detectable cybersecurity risks; acceptable overall risk.
Software Verification & ValidationPerformed as intended, met acceptance criteria, no negative impact on product performance or safety.
Usability Testing100% of professional and self-test users agreed/strongly agreed navigation features were easy to use, and overall setup/display was easy to use.
Side-by-Side Performance Comparison (with Predicate)No difference in results obtained on the modified meter compared to the predicate device.

Study Details:

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document does not specify a numerical sample size for the test set used in the functional bench testing, electrical safety/EMC testing, or software V&V. For the usability study, it mentions "All users" and "100% of professional and self-test users," but does not provide the number of users involved. For the side-by-side performance comparison, it mentions "identical test samples" but does not specify the number of samples.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be prospective in nature, as they were conducted specifically for this 510(k) submission to verify the modified device's performance.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided. The testing described (functional, electrical, software V&V, usability, and side-by-side performance) does not typically involve "experts" establishing a "ground truth" in the same way clinical diagnostic studies might (e.g., radiologists interpreting images). Instead, performance is assessed against established engineering specifications, regulatory standards, or predicate device results. For the usability study, users evaluated the device, but they are not described as "experts" establishing a "ground truth."
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This information is not provided and is generally not applicable to the types of engineering and performance verification tests described for this device. Adjudication methods are typically associated with clinical studies involving interpretation of results by multiple human readers.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. This device is a Prothrombin Time (PT) / INR Monitoring System, which measures a physiological parameter. It is an in vitro diagnostic device, not an AI-assisted diagnostic tool that would typically involve human readers interpreting results with or without AI assistance. The study described focuses on device performance and usability, not on improving human reader effectiveness.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • The core of the described studies focuses on the standalone performance of the modified device. The functional bench testing, electrical safety/EMC testing, and software verification/validation are all evaluations of the device itself (algorithm and hardware) without human interpretation affecting the primary measurement. The side-by-side comparison also assesses the device's measurement accuracy against a predicate. While there is a human user interface, the primary measurement output (PT/INR) is generated by the device's mechanism and algorithm. Therefore, yes, standalone performance was assessed in the sense that the device's ability to generate accurate PT/INR results was verified independently.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the various performance tests seems to be based on:
      • Predetermined acceptance criteria and design specifications: For functional, electrical safety, and software V&V.
      • Applicable electrical safety standards (e.g., IEC 61010-1): For electrical safety and EMC.
      • Predicate device results: For the side-by-side performance comparison, where the predicate device served as the established reference for "no difference" in results.
      • User feedback: For the usability study.
  7. The sample size for the training set:

    • Not applicable / Not explicitly provided. This submission is for modifications to an existing device and focuses on verification and validation of those changes. There is no mention of a "training set" in the context of machine learning model development. The device's underlying algorithm for detecting clot formation (micromechanical end point detection) is established and not part of a machine learning paradigm that would typically involve training sets.
  8. How the ground truth for the training set was established:

    • Not applicable. As a traditional in vitro diagnostic device, the concept of a "training set" and "ground truth establishment" for it as it relates to machine learning is not relevant to this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 5, 2022

CoaguSense, Inc. Robin Bush Consulting Director, Regulatory Affairs 48377 Fremont Blvd. Suite 113 Fremont, California 94538

Re: K212779

Trade/Device Name: Coag-Sense Prothrombin Time (PT) / INR Monitoring System for Patient Self-Testing Regulation Number: 21 CFR 864.7750 Regulation Name: Prothrombin Time Test Regulatory Class: Class II Product Code: GJS Dated: September 20, 2022 Received: September 01, 2021

Dear Robin Bush:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Min Wu Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212779

Device Name

Coag-Sense Prothrombin (PT) / INR Monitoring System for Patient Self-Testing

Indications for Use (Describe)

The Coag-Sense Prothrombin Time (PT) / INR Monitoring System for Patient Self-Testing is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in INR units. It uses fresh capillary whole blood.

The Coag-Sense Prothrombin Time (PT) / INR Monitoring System for Patient Self-Testing is intended for use by properly selected and suitably trained patients or the order of the treating physician to monitor patients who are on anticoagulation therapy. Patients should be stabilized on warfarin-type (coumarin) anticoagulation therapy prior to self-testing.

The Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is not intended to be used for screening purposes.

The Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is intended to be used by a single person and should not be shared.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for CoaguSense. The logo features the word "CoaguSense" in a stylized font, with the first part of the word, "Coagu", in red and the second part, "Sense", in black. To the left of the word is a red circle that is not fully closed, giving the impression of a ring or target.

Appendix 2

Date of Summary

September 28, 2022

Type of Submission

Special 510(k)

510(k) Applicant

CoaguSense Inc. 48377 Fremont Blvd, Suite 113 Fremont CA, 94538 Phone: 510-892-2379

Primary Contact

Robin Bush Consulting Director, Regulatory Affairs 510-908-5716 Email: rbushconsulting@earthlink.net

Device Overview

Trade Name:Coag-Sense Prothrombin Time (PT)/INR Monitoring Systemfor Patient Self-Testing
Common Name:PT/INR Test System
Classification Name:Prothrombin Time Test
Regulation Number:§864.7750
Product Code:GJS

Predicate Device

The predicate devices for this premarket submission:

Trade Name510(k) Submitter510(k)Number
Coag-Sense ProthrombinTime (PT)/INR SystemFor Patient Self-TestingCoaguSense Inc.K183255

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for CoaguSense. The logo features a red and black circular design on the left, with the word "CoaguSense" in black and red text to the right of the circle. The word "Coagu" is in red, while "Sense" is in black.

Device Description

The Coaq-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is a portable medical device for the measurement of the Prothrombin Time (PT) using fresh capillary whole blood obtained from a finger stick. The Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is a handheld device that directly detects clot formation. The System measures the PT of fresh capillary whole blood using micromechanical end point detection. The test is performed by inserting a test strip into the meter and applying a drop of blood to the sample receptacle of the disposable test strip. The test strip contains a rotating, spoked wheel that draws the sample into the reaction well after it is applied to the sample receptacle. The spokes rotate across the path of an infrared light beam and mix the liquid sample with the thromboplastin which is dried in the reaction well. When the sample clots, the clot is picked up by the spokes, interrupting the path of the infrared light beam that is detected by the meter.

The PT test and the result is displayed as International Normalized Ratio (INR). The result is date/time stamped and stored in the memory of the meter.

The device is powered by 4 AA batteries. This Coag-Sense Prothrombin Time (PT)//NR Monitoring System for Patient Self-Testing uses the exact same test and control strip as the predicate devices.

Intended Use / Indications for Use

The Coaq-Sense® Prothrombin Time (PT) / INR Monitoring System for Patient Self-Testing is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in INR units. It uses fresh capillary whole blood.

The Coaq-Sense® Prothrombin Time (PT) / INR Monitoring System for Patient Self-Testing is intended for use by properly selected and suitably trained patients or their caregivers on the order of the treating physician to monitor patients who are on anticoagulation therapy. Patients should be stabilized on warfarin-type (coumarin) anticoagulation therapy prior to self-testing.

The Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is not intended to be used for screening purposes.

The Coag-Sense® Prothrombin Time (PT) / INR Monitoring System for Self-Testing is intended to be used by a single person and should not be shared.

Performance Data

This special 510(k) describes modifications to the user interface access (buttons to navigate the flow, compared to touchscreen in the predicate device). Functional bench testing, electrical safety and EMC testing was performed on the modified Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing, demonstrating that it met the predetermined acceptance criteria and design specifications.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for CoaguSense. The logo features a red circle that is not fully closed, with the word "CoaguSense" written in black and red letters. The word "Coagu" is in red, and the word "Sense" is in black.

  • Functional Testing .
    The following functional bench tests were performed:

  • Component Testing (including power consumption, battery, clock, temperature, o humidity, memory, communication)

  • Drop Test o

  • Vibration Testing о

  • o Transit Testing (Bulk)

  • Electrical Safety / Electromagnetic Compatibility Testing ●

Testing was performed per the applicable sections of the following electrical safety standards:

  • o IEC 61010-1:2017; Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
  • IEC 61010-2-101:2018; Safety requirements for electrical equipment for O measurement, control, and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
  • IEC 60606-1-2:2014, edition 4.1, 2020-09; Medical electrical equipment Part 1-2: o General requirements for basic safety and essential performance - Collateral standard: Electromaqnetic disturbances - Requirements and tests

Cybersecurity evaluation and risk management documentation was prepared according to FDA Guidance Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (October 2, 2014) and ANSI UL 2900-2-1, Standard for Safety, Software Cybersecurity for Network-Connectable Products, Part 2-1: Particular Requirements for Network Connectable Components of Healthcare and Wellness Systems.

The risk analysis concludes that the Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing has no significant detectable cybersecurity risks and has an acceptable overall risk.

All testing results met the pre-determined acceptance criteria that were established in the test protocols. Based on the testing, the Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing performs as intended, with no new questions of safety or effectiveness identified during testing.

Software Verification and Validation

Software verification and validation was performed on the functions of the software to ensure that the software performs as intended. Verification and Validation testing of the Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing demonstrated the software performed as intended, met acceptance criteria, and does not have a negative impact on product performance or product safety.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for CoaguSense. The logo features the word "CoaguSense" in a stylized font. The "Coagu" part of the word is in red, while the "Sense" part is in black. To the left of the word is a red and black circular design.

Usablity Testing

A usability study was conducted according to IEC 62366-1 and FDA quidance to verify the changes in user interface. All users agreed or strongly agreed that the navigation features were easy to use. 100% of professional and self-test users strongly agree or agree that overall, the set-up and display on the modified Coag-Sense meter is easy to use.

Testing In Support of Substantial Equivalence Determination

The design verification testing and design validation testing performed on the modified meter demonstrate the subject device meets defined product specification and intended use. The testing demonstrates that the subject device has comparable performance to the predicate device. The functional bench testing performed verifies that the subject Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing meets the specified performance specifications and thus, is substantially equivalent to the predicate device for home use.

Comparison to Predicate Device

No changes have been made in the test strips or chemistry. The mechanism of action to detect a clot remains the same. A side-by-side comparison study was performed with identical test samples to demonstrate that the performance of the modified Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is equivalent than the currently marketed predicate Coag-Sense PT/INR meter. Performance testing confirmed no difference in the results obtained on the modified meter when compared to the predicate device.

The subject Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is substantially equivalent to the cleared predicate (#K183255) with respect to indications for use for patient self-testing, intended use, and technological characteristics.

The differences between the subject device and the predicate devices do not raise any new or different issues of safety and effectiveness. Design verification testing and side by side performance testing confirmed that no new questions of safety or effectiveness were identified during testing, and that the subject Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing performs as intended.

Conclusion

Based on the comparative analysis, the intended use, principles of operation, performance characteristics and technological characteristics, the Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing does not introduce new risks or any new questions of safety and effectiveness. The minor differences in power source and user interface between the subject Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing and the predicate device raise no new issues of safety or effectiveness. Therefore, the Coag-Sense Prothrombin Time (PT)/INR Monitoring System for Patient Self-Testing is substantially equivalent to the predicate devices.

§ 864.7750 Prothrombin time test.

(a)
Identification. A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).(b)
Classification. Class II (performance standards).