(122 days)
The Dermoaroma Disposable Nitrile Examination Gloveare disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed device is Powder Free Dermoaroma Disposable Nitrile Examination Glove. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124,ASTM D5151, and ASTM D6319. The proposed device is non-sterile.
The provided document is a 510(k) Premarket Notification summary for the Dermoaroma Disposable Nitrile Examination Glove. This device is a Class I, reserved product, and the information presented focuses on demonstrating substantial equivalence to a predicate device, rather than proving a novel device's safety and effectiveness through a comprehensive clinical study.
Therefore, the document does not contain the kind of detailed information about acceptance criteria for an AI/algorithm-based device and its validation study that your request is looking for. There is no AI or algorithm involved in this specific medical device (disposable examination gloves).
Here's why the document doesn't fit your request and what information it does provide:
-
AI/Algorithm Focus: The request explicitly asks about an "AI device" or an "algorithm-based performance study" and its acceptance criteria. The device in this document is a physical product (nitrile gloves), not an AI or algorithm.
-
Clinical Study vs. Substantial Equivalence: The document explicitly states: "No clinical study is included in this submission." Instead, it relies on non-clinical performance testing and comparison to a legally marketed predicate device to demonstrate substantial equivalence.
-
Type of Performance Data: The performance data presented refers to physical and chemical properties of the gloves (e.g., length, width, thickness, tensile strength, elongation, freedom from holes, powder content, biocompatibility). These are not the types of metrics (e.g., sensitivity, specificity, AUC) or study designs (MRMC, standalone algorithm performance) relevant to an AI/algorithm.
To address your specific questions based on the provided document (and explain why most cannot be answered):
-
A table of acceptance criteria and the reported device performance:
- Context: This table relates to the physical and biological characteristics of the gloves, not AI performance metrics.
- Table 5: Summary of Non-Clinical Performance Testing provides this information.
- Acceptance Criteria Examples:
- Skin Sensitization Test: Grades less than 1.
- Skin Irritation Test: Primary irritation index 0-0.4 (Negligible).
- In Vitro Cytotoxicity: Viability reduced to <70% of blank indicates cytotoxic potential (acceptance is >70%).
- Systemic Toxicity: Toxicosis response of testing group not greater than control.
- Residual Powder: Powder residue limit of 2.0 mg.
- Freedom from Holes: AQL 2.5, Criterion ≤7 gloves for water leakage (for 125 samples).
- Dimensions (e.g., for Size S glove): width 80±10mm, Length ≥220 mm.
- Physical Properties (Before Aging): Tensile strength ≥ 14MPa, Ultimate Elongation ≥ 500%.
- Physical Properties (After Accelerated Aging): Tensile strength ≥ 14MPa, Ultimate Elongation ≥ 400%.
- Reported Device Performance Examples:
- Skin Sensitization Test: All grades are 0.
- Skin Irritation Test: Primary irritation index is 0 (Negligible).
- In Vitro Cytotoxicity: Viab.% of 100% test article extract is 20.8% (Indicates potential toxicity, but the submission argues this difference does not affect safety/efficacy when compared to predicate, likely due to the nature and contact of the device).
- Systemic Toxicity: No toxicosis response in testing group.
- Residual Powder: 0.13mg - 0.16 mg / glove.
- Freedom from Holes: No glove water leakage found.
- Dimensions (e.g., for Size S glove): width: 80-87 mm, Length 240-246 mm.
- Physical Properties (Before Aging): Tensile strength 15.3-19.8 MPa, Ultimate Elongation 508.480% - 669.787%.
- Physical Properties (After Accelerated Aging): Tensile strength 14.3-15.2MPa, Ultimate Elongation 435.176% - 542.968%.
- Acceptance Criteria Examples:
-
Sample sizes used for the test set and the data provenance:
- Sample Sizes (for physical/biocompatibility tests):
- For "Freedom from Holes" test (ASTM D5151), the sample size was 125 gloves.
- Other tests (e.g., dimensions, physical properties, biocompatibility) would have specific sample sizes per the referenced ASTM/ISO standards, but these are not explicitly listed in the summary beyond "125 gloves"for the holes test. Often, these standards specify minimum sample sizes for material testing.
- Data Provenance: The tests were conducted by the manufacturer, ChuZhou Harmony Gloves Medical Technology Co.,Ltd, based in AnHui Province, China. The document does not specify if these were retrospective or prospective studies in the general sense, but rather standard performance evaluations of manufacturing batches.
- Sample Sizes (for physical/biocompatibility tests):
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This device does not involve expert-established ground truth like an AI diagnostic algorithm. Ground truth here is based on measurements against established engineering and biological standards (e.g., "0.13mg - 0.16 mg / glove" for powder content as measured by ASTM D6124).
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, as there's no diagnostic component requiring human interpretation and consensus.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is for a non-AI device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is for a non-AI device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The ground truth for this device's performance relies on established international standards and their specified methodologies for measuring physical, chemical, and biological properties (e.g., ASTM D6319 for dimensions and physical properties, ASTM D5151 for freedom from holes, ISO 10993 for biocompatibility).
-
The sample size for the training set: Not applicable. This device does not use an AI model that would require a training set.
-
How the ground truth for the training set was established: Not applicable. This device does not use an AI model or a training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 27, 2021
ChuZhou Harmony Gloves Medical Technology Co.,Ltd Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room 608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China
Re: K212716
Trade/Device Name: Dermoaroma Disposable Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: November 19, 2021 Received: November 29, 2021
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K212716
Device Name
Dermoaroma Disposable Nitrile Examination Glove
Indications for Use (Describe)
The Dermoaroma Disposable Nitrile Examination Gloveare disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with 21 CFR 807.92(a)(1).
1.0 Submitter's information
Name: ChuZhou Harmony Gloves Medical Technology Co.,Ltd Address: Technology Road 6, branching stream economic development zone, LaiAn Town,Chuzhou City ,AnHui Province,China Phone Number: +86-18689321491 Contact: WeiHong Ou Date of Preparation: December 7, 2021
Designated Submission Correspondent
Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Tel: +86-21-50313932 Email: Info@truthful.com.cn
2.0 Device information
Trade name: Dermoaroma Disposable Nitrile Examination Glove Patient Examination Gloves Common name: Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL
3.0 Classification
Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital
4.0 Predicate device information
Manufacturer: Jiangxi Surefine Medical Co., Ltd. Blue Nitrile Exam Gloves Device: 510(k) number: K211341
{4}------------------------------------------------
5.0 Indications for Use
The Dermoaroma Disposable Nitrile Examination Glove are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.
6.0 Device description
The proposed device is Powder Free Dermoaroma Disposable Nitrile Examination Glove. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124,ASTM D5151, and ASTM D6319. The proposed device is non-sterile.
7.0 Summary comparing technological characteristics with predicate device
| Item | Proposed device | Predicated device | Remark |
|---|---|---|---|
| 510(k) number | K212716 | K211341 | |
| Product Code | LZA | LZA | Same |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same |
| Class | I | I | Same |
| Indications for Use | The Dermoaroma Disposable Nitrile Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | The Blue Nitrile Exam Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | Same |
| Powdered or Powered free | Powdered free | Powdered free | Same |
| Design Feature | ambidextrous | ambidextrous | Same |
| Labeling Information | Single-use indication, powder free, device color, device name, glove size and quantity, Dermoaroma Disposable Nitrile Examination Glove, Non-Sterile | Single-use indication, powder free, device color, device name, glove size and quantity, Blue Nitrile Exam Gloves, Non-Sterile | Same |
Table1-General Comparison
Table2 Device Dimensions Comparison
| Predicate | Designation | Size | Tolerance | |||
|---|---|---|---|---|---|---|
| Device(K211341) | S | M | L | XL |
{5}------------------------------------------------
| Length, mm | 220 | 230 | 230 | 230 | min | |
|---|---|---|---|---|---|---|
| Width, mm | 80 | 95 | 110 | 120 | ±10 | |
| Thickness, mm: | ||||||
| Finger | 0.05 | min | ||||
| Palm | 0.05 | min | ||||
| Proposed Device | Designation | Size | Tolerance | |||
| S | M | L | XL | |||
| Length, mm | 220 | 230 | 230 | 230 | min | |
| Width, mm | 80 | 95 | 110 | 120 | ±10 | |
| Thickness, mm: | ||||||
| Finger | 0.05 | min | ||||
| Palm | 0.05 | min | ||||
| Remark | Same |
Analysis1: The proposed device has same sizes to the predicate device, and all proposed devices are conducted the properties test, the test results shown that the sizes comply with the requirements of standard ASTM D6319-19, Standard Specification For Dermoaroma Disposable Nitrile Examination Glove For Medical Application.
| Item | Proposed device | Predicated device | Remark | |||
|---|---|---|---|---|---|---|
| Colorant | blue | Blue | SAME | |||
| PhysicalProperties | BeforeAging | TensileStrength | 14MPa, min | 14MPa, min | SAME | |
| UltimateElongation | 500%min | 500%min | SAME | |||
| AfterAging | TensileStrength | 14MPa, min | 14MPa, min | SAME | ||
| UltimateElongation | 400%min | 400%min | SAME | |||
| Comply with ASTM D6319 | Comply with ASTM D6319 | Comply with ASTM D6319 | SAME | |||
| Freedom from Holes | Be free from holeswhen tested inaccordance withASTMD5151AQL=2.5 | Be free from holes whentested in accordance withASTMD5151 AQL=2.5 | SAME | |||
| Powder Content | 0.13mg -0.16 mg | 0.03 mg | SIMILARAnalysis2 |
Table3 Performance Comparison
Analysis2: The proposed device has different powder content to the predicate device, but the proposed device is conducted the biocompatibility and performance tests, the test results shown that the difference does not affect the safety and efficacy of proposed device.
{6}------------------------------------------------
| Item | Proposed device | Predicated device | Remark | |
|---|---|---|---|---|
| Material | Nitrile | Nitrile | SAME | |
| Biocompatibility | Irritation | Under the conditions of the study, not an irritant | Comply with ISO10993-10 | SAME |
| Sensitization | Under conditions of the study, not a sensitizer. | |||
| Cytotoxicity | Under the conditions of the study, the device is potentially cytotoxic | Comply with ISO10993-5 | SIMILAR | |
| Systemic toxicity | Under the conditions of the study, the device does not elicit a systemic toxicity response in the model animal. |
Biocompatibility Testing Comparison Table4
8.0 Summary of Non-Clinical Performance Testing
The following performance data has been provided to demonstrate that the subject device meets the acceptance criteria in the standard.
| No. | Name of the TestMethodology / Standard | Purpose | Acceptance Criteria | Results |
|---|---|---|---|---|
| 1 | ISO 10993-10:2010Biological Evaluation OfMedical Devices - Part10: Tests For IrritationAnd Skin Sensitization. | This part of ISO10993 assessespossible contacthazards fromchemicalsreleased frommedical devices, | Skin SensitizationTest:provided grades less than 1,otherwisesensitization. | All grades are 0.All animals were survived and noabnormal signs were observedduring the study. |
| 2 | which mayproduce skin andmucosal irritation,eye irritation orskin sensitization. | Skin Irritation Test:If the primary irritationindex is 0-0,4, theresponse category isNegligible.0,5-1,9 means slight2-4,9 means moderate5-8 means severe | The primary irritation index is 0.The response of the proposeddevice was categorized asnegligible under the testcondition | |
| 3 | ISO 10993-5:2009Biological Evaluation OfMedical Devices - Part5: Tests For In VitroCytotoxicity | This part of ISO10993 describestest methods toassess the in vitrocytotoxicity ofmedical devices. | The viab.% of the100% extract of thetest article is the finalresult, and if viability isreduced to <70% ofthe blank, it hascytotoxic potential. | Viab.% of 100% test articleextract is 20.8%It means the proposed devicehave potential toxicity to L-929 inthe MTT method |
| 4 | ISO 10993-11: 2017Biological evaluation ofmedical devices — Part11: Tests for systemictoxicity | To evaluate thepotential formedical devicematerials to causeadverse systemicreactions. | Within the monitoringperiod (72 h), if thetoxicosis response oftesting group is notgreater than that ofcontrol group, thetesting sample isregarded asacceptable. | No toxicosis response in testinggroup.It means the testarticle has no potential acutesystem toxicity on ICR mice inthe extraction method. |
| 5 | ASTM D6124-06(Reapproved 2017),Standard Test Methodfor Residual Powder onMedical Gloves | This standard isdesigned todetermine theamount ofresidualpowder (orfilter-retainedmass) foundon medical gloves | powder residue limit of2.0 mg | 0.13mg -0.16 mg /glove |
| 6 | ASTMD5151-06(Reapproved2015), Standard TestMethod for Detection ofHoles in Medical Gloves. | This test methodcovers thedetection of holesinmedical gloves. | Samples number: 125glovesAQL: 2.5 (ISO 2859)Criterion ≤7 glovesfor water leakage | no glove water leakage found |
| 7 | ASTMD6319-10(Reapproved2015),StandardSpecification For NitrileExamination Gloves ForMedical Application. | This specificationcovers certainrequirements fornitrile rubbergloves used inconductingmedicalexaminations anddiagnostic andtherapeuticprocedures. | Sterility: no needFreedom from holes:pl. Refer to No. 5 intable 5Dimensions:S: width 80±10mmLength ≥220 mmM: width 95±10mmLength ≥230 mmL: width 110±10mmLength ≥230 mm | N.A.Please refer to No. 5 in table 5Lot no.:202100505Dimensions:S: width: 80-87 mmLength 240-246 mmM: width 93-98 mmLength 240-245 mmL: width 103-110 mmLength 249-252 mmXL: width 118-123 mm |
Table 5 Summary of Non-Clinical Performance Testing
{7}------------------------------------------------
{8}------------------------------------------------
| XL: width 120±10mm | Length 265-270 mm | ||
|---|---|---|---|
| Length ≥230 mm | Thickness: | ||
| Thickness: | Finger 0.10-0.12 mm | ||
| Finger ≥0.05 mm | Palm 0.07-0.10 mm | ||
| Palm ≥0.05 mm | |||
| Physical properties: | Physical properties: | ||
| Before aging | Before aging | ||
| Tensile strength ≥ 14MPa | Tensile strength 15.3-19.8 MPa | ||
| Ultimate Elongation ≥ 500% | Ultimate Elongation 508.480% - 669.787% | ||
| After Accelerated Aging | After Accelerated Aging | ||
| Tensile strength ≥ 14MPa | Tensile strength 14.3-15.2MPa | ||
| Ultimate Elongation ≥ 400% | Ultimate Elongation 435.176% - 542.968% | ||
| Powder-free Residue: pl. Refer to No. 4 in table 5 | |||
| Powder-free Residue: pl. Refer to No. 4 in table 5 | Lot no.:202100520 | ||
| Dimensions: | |||
| S: width: 80-85 mmLength 245-251 mm | |||
| M: width 89-96 mmLength 240-245 mm | |||
| L: width 105-112 mmLength 249-253 mm | |||
| XL: width 118-125 mmLength 259-264 mm | |||
| Thickness: | |||
| Finger 0.10-0.12 mm | |||
| Palm 0.08-0.10 mm | |||
| Physical properties: | |||
| Before aging | |||
| Tensile strength 14.9-19.8 MPa | |||
| Ultimate Elongation 508.480% - 669.787% | |||
| After Accelerated Aging | |||
| Tensile strength 14.3-15.2MPa | |||
| Ultimate Elongation 503.493% - 644.640% | |||
| Powder-free Residue:pl. Refer to No. 4 in table 5 | |||
| Lot no.:202100601 | |||
| Dimensions: | |||
| S: width: 82-90 mmLength 241-245 mm | |||
| M: width 95-103 mmLength 240-245 mm | |||
| L: width 103-112 mmLength 248-255 mm | |||
| XL: width 114-120 mmLength 258-265 mm | |||
| Thickness:Finger 0.09-0.12 mmPalm 0.08-0.09 mm | |||
| Physical properties:Before agingTensile strength 15.1-19.9 MPaUltimate Elongation 521.280% -665.453% | |||
| After Accelerated AgingTensile strength 14.1-15.9MPaUltimate Elongation 430.518% -513.693% | |||
| Powder-free Residue:pl. Refer to No. 5 in table 5 |
{9}------------------------------------------------
9. Summary of Clinical Performance Test
No clinical study is included in this submission.
10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.