K Number
K212622

Validate with FDA (Live)

Device Name
Zyter RPM
Manufacturer
Date Cleared
2022-02-11

(177 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the Zyter RPM is to provide an interface with physiological patient monitoring systems to forward recorded device information to the patient's healthcare provider.

Zyter RPM does not alter the behavior of the primary medical devices and associated alert annunciations.

Zyter RPM is not intended to be used for diagnostic purposes. Zyter RPM is intended for use by professional clinical personnel and relies on proper use and operation of both the communication infrastructure in place at the use environment (e.g., healthcare facility or home setting) and the display interface used. Zyter RPM is a software product and cannot come into physical contact with patients.

Device Description

Zyter RPM is a Software as a Medical Device (SaMD) intended to provide healthcare professionals with supplemental information about events and alerts originating from medical devices, or other event or alert producing systems within a clinical or home setting. This device can route notifications including all or a subset of the event or alert information along with related contextual information to selected end points such as but not limited to mobile phones, or other computing or communications end points.

Zyter RPM is intended to provide remote central monitoring and display of information as recorded by multiple specifically compatible FDA cleared devices on a central screen and to provider alert and alert notifications on the screen and to other provider devices.

The Zyter RPM software application will receive, aggregate, process, distribute and display parameters, alerts, and events at locations other than at the patient for multiple patients.

Information provided by the Zyter RPM software is not intended for diagnosis or active patient monitoring where immediate action is required.

Receipt of notifications by the end point is not confirmed, and delivery to the end point is not guaranteed. The primary alert notification is the device or system producing the alert or event.

Zyter RPM is not intended to directly diagnose or treat patients or to be used in the prevention of a disease or condition, nor does Zyter RPM come into direct contact with patients.

Zyter RPM serves as a convenient user interface to integrate data and provide health care professionals the ability to access and monitor parameters of devices and data that have been cleared by the FDA. Zyter RPM does not affect the intended use, or alter the indications for use, for the cleared devices with which it is intended to function.

The Zyter RPM has two separate user interfaces: one is the provider web application and the second is the patient mobile application. The patient mobile application is used by patient-users along with their assigned medical devices to see their own health data and send device readings to their provider. The web application is used by healthcare providers to manage their patients, receive alerts and view their patient's medical data.

Zyter RPM is comprised of two main components:

  • Mobile App (for patient users)
  • Web App (for healthcare provider users)

The patient mobile application is used by patient-users along with their assigned medical devices to see their own health data and send device readings to their provider. The web application is used by healthcare providers to manage their patients, receive alerts and view their patient's medical data.

The patient-user is provided with a pre-configured Zyter Tablet for remote health monitoring. The patientuser may also choose to use their own smartphone and download the Zyter Mobile App. The patient is also assigned a number of health devices by their healthcare provider. The patient-user is responsible for connecting the health devices to the Zyter Tablet (or mobile phone). Pairing of the health devices with the tablet (or mobile phone) ensures wireless communication between both devices. Instructions for connecting health devices to the patient mobile app are included in the patient app guide. Zyter RPM mobile application also has a functionality that allows the patient to schedule a tele-visit with their healthcare provider. The patient may choose to take a self-assessment questionnaire found in the Zyter RPM prior to scheduling a visit.

Zyter RPM is a digital healthcare platform that automatically collects and routes data from the patient's pre-configured, compatible, at-home medical devices (blood pressure cuffs, glucometers, thermometers, digital weight scales, and smart watches) to the hospital office via a 4G cellular or Wireless connection.

The health devices are configured to a user in the Zyter user management using the Device Mac address. Zyter validation algorithm doesn't allow the duplicate Mac address entries. Zyter RPM provides patients with Tablets pre-installed with Zyter mobile application. Users connect with the wireless health devices through the mobile app. Zyter also supports BYOD model where patients use their own mobile devices to install the Zyter app from the App store and Google Play store to connect with the wireless devices.

Zyter RPM supports two transmission protocols, device data transmission through Bluetooth technology where the Bluetooth device needs to be paired with the mobile app and the 4G/LTE devices where the device data is transmitted to Zyter via device cloud via APIs and no device pairing required.

Zyter RPM achieves EMR connectivity using on-prem or on-cloud solutions.

The vitals reading from the health devices are securely processed through device SDKs/APIs and they are accessed through the patient and provider RPM dashboard. The Zyter RPM Dashboard enables the monitoring of multiple patients with an information tab for each patient. All physiological patient data, plus alerts and notifications from each patient's RPM devices, go directly to the dashboard for physician review, as well as intervention via a telehealth virtual visit if a patient's readings are out of the normal range.

Physicians can also configure alerts for custom parameters of device readings for individual patients. Zyter RPM devices also send alerts to the physician dashboard, email and SMS, and a reminder to the patient's Zyter RPM app, if a device has been dormant for a pre-determined amount of time.

Physicians can set a Zyter RPM alert based on any patient device reading that is out of range. Depending on the device(s) the patient receives, alert ranges can be set to measure the following values:

  • BP (Systolic), mmHg
  • BP (Diastolic), mmHg
  • Pulse Oximeter, % Sp02
  • Temperature, Fº or Cº
  • Weight Scale, lbs
  • Blood Glucose, mmoI/L

Zyter RPM is compatible with the following medical devices and accessories:
BodyTrace Blood Pressure Monitor K131395
iHealth BP Monitor - iHealth Track (KN550BT) K160014
BodyTrace Weighing Scale N/A - Wellness device
iHealth Weighing Scale N/A - Wellness device
Smart Meter iGlucose Blood Glucose Monitor Starter Kit K161790
iHealth Blood Glucose - iHealth Gluco + (BG5s): K181070
iHealth Thermometer PT3 K202753
iHealth Air Pulse Oximeter - iHealth Air (PO3M): K131111
Nonin Pulse Oximeter K131021
Apple Health Kit N/A - Apple Health System
Samsung Android Tablet

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Zyter RPM device, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state specific pass/fail acceptance criteria in terms of numerical thresholds for device performance. Instead, it indicates that performance testing was conducted to demonstrate substantial equivalence through software verification and validation, and human factors testing.

The document discusses the types of performance testing performed. Here's a table summarizing those and the general reported outcome:

Acceptance Criteria CategoryReported Device Performance
Software Verification and ValidationConducted in compliance with: - ANSI/AAMI/IEC 62304:2006/A1:2016 Medical device software - Software life cycle processes - FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - FDA Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices - FDA Guidance for Industry and Food and Drug Administration Staff, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices - FDA Guidance for Industry and Food and Drug Administration Staff Postmarket Management of Cybersecurity in Medical Devices Conclusion: Demonstrated substantial equivalence.
Performance Bench Testing (Server Capacity)Testing conducted on the number of concurrent transactions (readings/alerts) APIs can handle within a stipulated interval. Transactions stimulated through auto scripts to test server capacity and identify when application performance begins to degrade. Implied successful performance within expected loads.
Human Factors TestingA summative human factors study was conducted to evaluate the usability of the subject device by the intended users, in compliance with ANSI/AAMI/EC 62366-1:2015/A1:2020 and FDA Guidance for Industry and Food and Drug Administration Staff Applying Human Factors and Usability Engineering to Medical Devices. Implied successful usability for intended users.
Risk AnalysisPerformed (listed under performance testing). Implied acceptable risk profile.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a sample size for any "test set" in the traditional sense of a clinical or imaging study. The testing described is primarily focused on software verification/validation, performance bench testing (server capacity), and human factors.

  • Software Verification and Validation: This typically involves testing various modules, functionalities, and integration points. A "sample size" isn't directly applicable in the same way as a clinical trial.
  • Performance Bench Testing: This involves simulating transactions. The document doesn't provide the number of simulated transactions or specific data provenance beyond general "auto scripts."
  • Human Factors Testing: While a summative study was conducted, the document does not state the number of participants (the "test set" for human factors) or their demographics.
  • Data Provenance: Not applicable as there is no mention of patient data being used for a test set to evaluate diagnostic or clinical performance.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

This information is not provided as the study described is not a diagnostic performance study requiring expert ground truth.

4. Adjudication Method for the Test Set

This information is not provided as the study described is not a diagnostic performance study requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The Zyter RPM is a remote patient monitoring software that provides an interface for physiological data and alerts; it is not intended for diagnostic purposes and does not involve human readers interpreting medical images or data in a diagnostic context where AI assistance would be measured for improving diagnostic accuracy. The device's purpose is to aggregate and route data and alerts.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

The performance testing described (Software V&V, Performance Bench Testing) can be considered "standalone" in the sense that it evaluates the software's functionality, reliability, and capacity on its own. However, the device's intended use is inherently connected to human users (healthcare providers and patients), and human factors testing was conducted to assess usability with human interaction. The device itself is an "algorithm only" software from the perspective that it processes and presents data, but its utility is in its interaction with connected devices and human users. There isn't a "diagnostic algorithm" performance being measured in isolation.

7. The Type of Ground Truth Used

The concept of "ground truth" as typically used in diagnostic performance studies (e.g., pathology, outcomes data) is not applicable here. The Zyter RPM is not a diagnostic device. The "truth" being verified in its testing relates to:

  • Software Functionality: Does the software perform its coded tasks correctly? (Verified through V&V).
  • Data Aggregation and Routing Accuracy: Does it correctly receive, process, distribute, and display parameters/alerts as intended? (Verified through V&V and bench testing).
  • System Capacity: Can it handle the expected load? (Verified through performance bench testing).
  • Usability: Is it intuitive and safe for its intended users? (Verified through human factors testing).

8. The Sample Size for the Training Set

The document does not mention a training set. This is expected because the Zyter RPM appears to be a rule-based or conventional software system for data aggregation and alert management, rather than an AI/Machine Learning model that would require a "training set" of data for learning patterns to make predictions or classifications.

9. How the Ground Truth for the Training Set Was Established

Since there is no mention of a training set, this information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.

February 11, 2022

Zyter, Inc. % Mary Vater 510(k) Consultant Medical Device Academy, Inc 345 Lincoln Hill Rd. Shrewsbury, Vermont 05738

Re: K212622

Trade/Device Name: Zyter RPM Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX Dated: August 17, 2021 Received: August 18, 2021

Dear Mary Vater:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212622

Device Name Zyter RPM

Indications for Use (Describe)

The intended use of the Zyter RPM is to provide an interface with physiological patient monitoring systems to forward recorded device information to the patient's healthcare provider.

Zyter RPM does not alter the behavior of the primary medical devices and associated alert annunciations.

Zyter RPM is not intended to be used for diagnostic purposes. Zyter RPM is intended for use by professional clinical personnel and relies on proper use and operation of both the communication infrastructure in place at the use environment (e.g., healthcare facility or home setting) and the display interface used. Zyter RPM is a software product and cannot come into physical contact with patients.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

l. SUBMITTER Zyter, Inc. 2600 Tower Oaks Blvd. Suite 700 Rockville, MD 20852 Tel: +1.301.355.7760 Fax: N/A

Contact Person: Mary Vater, 510(k) Consultant Date Prepared: January 6, 2022

II. DEVICE
510(K) Number:K212622
Name of Device:Zyter RPM
Classification Name:Cardiac Monitor (including cardiotachometer)
Regulation:21 CFR §870.2300
Regulatory Class:Class II
Product Classification Code:MSX
III. PREDICATE DEVICE
Primary Predicate Manufacturer:BoxView, LLC
Primary Predicate Trade Name:BoxView Smart Alarm Interface (SAI), Model SA-01
Primary Predicate 510(k):K193421
Secondary Predicate Manufacturer:EARLYSENSE LTD.
Secondary Predicate Trade Name:EarlySense Central Display Station (CDS)

K151006

No reference devices were used in this submission.

DEVICE DESCRIPTION IV.

Secondary Predicate 510(k):

Zyter RPM is a Software as a Medical Device (SaMD) intended to provide healthcare professionals with supplemental information about events and alerts originating from medical devices, or other event or alert producing systems within a clinical or home setting. This device can route notifications including all or a subset of the event or alert information along with related contextual information to selected end points such as but not limited to mobile phones, or other computing or communications end points.

Zyter RPM is intended to provide remote central monitoring and display of information as recorded by multiple specifically compatible FDA cleared devices on a central screen and to provider alert and alert notifications on the screen and to other provider devices.

The Zyter RPM software application will receive, aggregate, process, distribute and display parameters, alerts, and events at locations other than at the patient for multiple patients.

Information provided by the Zyter RPM software is not intended for diagnosis or active patient monitoring where immediate action is required.

Receipt of notifications by the end point is not confirmed, and delivery to the end point is not guaranteed. The primary alert notification is the device or system producing the alert or event.

{4}------------------------------------------------

Zyter RPM is not intended to directly diagnose or treat patients or to be used in the prevention of a disease or condition, nor does Zyter RPM come into direct contact with patients.

Zyter RPM serves as a convenient user interface to integrate data and provide health care professionals the ability to access and monitor parameters of devices and data that have been cleared by the FDA. Zyter RPM does not affect the intended use, or alter the indications for use, for the cleared devices with which it is intended to function.

The Zyter RPM has two separate user interfaces: one is the provider web application and the second is the patient mobile application. The patient mobile application is used by patient-users along with their assigned medical devices to see their own health data and send device readings to their provider. The web application is used by healthcare providers to manage their patients, receive alerts and view their patient's medical data.

Zyter RPM:

Zyter RPM is comprised of two main components:

  • Mobile App (for patient users) ●
  • . Web App (for healthcare provider users)

The patient mobile application is used by patient-users along with their assigned medical devices to see their own health data and send device readings to their provider. The web application is used by healthcare providers to manage their patients, receive alerts and view their patient's medical data.

The patient-user is provided with a pre-configured Zyter Tablet for remote health monitoring. The patientuser may also choose to use their own smartphone and download the Zyter Mobile App. The patient is also assigned a number of health devices by their healthcare provider. The patient-user is responsible for connecting the health devices to the Zyter Tablet (or mobile phone). Pairing of the health devices with the tablet (or mobile phone) ensures wireless communication between both devices. Instructions for connecting health devices to the patient mobile app are included in the patient app guide. Zyter RPM mobile application also has a functionality that allows the patient to schedule a tele-visit with their healthcare provider. The patient may choose to take a self-assessment questionnaire found in the Zyter RPM prior to scheduling a visit.

Zyter RPM is a digital healthcare platform that automatically collects and routes data from the patient's pre-configured, compatible, at-home medical devices (blood pressure cuffs, glucometers, thermometers, digital weight scales, and smart watches) to the hospital office via a 4G cellular or Wireless connection.

The health devices are configured to a user in the Zyter user management using the Device Mac address. Zyter validation algorithm doesn't allow the duplicate Mac address entries. Zyter RPM provides patients with Tablets pre-installed with Zyter mobile application. Users connect with the wireless health devices through the mobile app. Zyter also supports BYOD model where patients use their own mobile devices to install the Zyter app from the App store and Google Play store to connect with the wireless devices.

Zyter RPM supports two transmission protocols, device data transmission through Bluetooth technology where the Bluetooth device needs to be paired with the mobile app and the 4G/LTE devices where the device data is transmitted to Zyter via device cloud via APIs and no device pairing required.

Zyter RPM achieves EMR connectivity using on-prem or on-cloud solutions.

{5}------------------------------------------------

The vitals reading from the health devices are securely processed through device SDKs/APIs and they are accessed through the patient and provider RPM dashboard. The Zyter RPM Dashboard enables the monitoring of multiple patients with an information tab for each patient. All physiological patient data, plus alerts and notifications from each patient's RPM devices, go directly to the dashboard for physician review, as well as intervention via a telehealth virtual visit if a patient's readings are out of the normal range.

Physicians can also configure alerts for custom parameters of device readings for individual patients. Zyter RPM devices also send alerts to the physician dashboard, email and SMS, and a reminder to the patient's Zyter RPM app, if a device has been dormant for a pre-determined amount of time.

Physicians can set a Zyter RPM alert based on any patient device reading that is out of range. Depending on the device(s) the patient receives, alert ranges can be set to measure the following values:

  • BP (Systolic), mmHg .
  • . BP (Diastolic), mmHg
  • Pulse Oximeter, % Sp02
  • . Temperature, Fº or Cº
  • . Weight Scale, lbs
  • . Blood Glucose, mmoI/L

Zyter RPM is compatible with the following medical devices and accessories:

Accessory510(k) Number
BodyTrace Blood Pressure MonitorK131395
iHealth BP Monitor - iHealth Track (KN550BT)K160014
BodyTrace Weighing ScaleN/A - Wellness device
iHealth Weighing ScaleN/A - Wellness device
Smart Meter iGlucose Blood Glucose Monitor Starter KitK161790
iHealth Blood Glucose - iHealth Gluco + (BG5s):K181070
iHealth Thermometer PT3K202753
iHealth Air Pulse Oximeter - iHealth Air (PO3M):K131111
Nonin Pulse OximeterK131021
Apple Health KitN/A - Apple Health System

Zyter Component Architecture:

Image /page/5/Figure/14 description: The image shows a diagram of a healthcare network. The diagram is divided into five sections: Connected Devices and Health Parameters (3.1), Connected and Local Gateway (3.2), Platform Infrastructure (3.3), Healthcare Network (3.4), and RPM (3.5). The Connected Devices and Health Parameters section includes weight, glucose levels, heart rate, and blood pressure. The Connected and Local Gateway section includes iOS and Android devices and a Health IoT Gateway with security, connectivity, compliance, device management, data exchange, and data management.

{6}------------------------------------------------

V. INDICATIONS FOR USE

The intended use of the Zyter RPM is to provide an interface with physiological patient monitoring systems to forward recorded device information to the patient's healthcare provider.

Zyter RPM does not alter the behavior of the primary medical devices and associated alert annunciations.

Zyter RPM is not intended to be used for diagnostic purposes. Zyter RPM is intended for use by professional clinical personnel and relies on proper use and operation of both the communication infrastructure in place at the use environment (e.g., healthcare facility or home setting) and the display interface used. Zyter RPM is a software product and cannot come into physical contact with patients.

{7}------------------------------------------------

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVCE

The following characteristics were compared between the predicate devices in order to demonstrate substantial equivalence:

Zyter RPMBoxView Smart Alarm Interface (SAI) – K193421EarlySense Central Display Station (CDS) – K151006
RegulationNumber21 CFR 870.230021 CFR 870.230021 CFR 870.2300
RegulationNameCardiac Monitor (IncludingCardiac Monitor (IncludingCardiac Monitor (Including
Cardiotachometer and Rate Alarm)Cardiotachometer and Rate Alarm)Cardiotachometer and Rate Alarm)
Primary ProductCodeMSXMSXMSX
Indications forUseThe intended use of the Zyter RPM is toprovide an interface with physiologicalpatient monitoring systems to forwardrecorded device information to thepatient's healthcare provider.Zyter RPM does not alter the behavior ofthe primary medical devices andassociated alert annunciations.Zyter RPM is not intended to be used fordiagnostic purposes. Zyter RPM isintended for use by professional clinicalpersonnel and relies on proper use andoperation of both the communicationinfrastructure in place at the useenvironment (e.g., healthcare facility orhome setting) and the display interfaceused. Zyter RPM is a software productand cannot come into physical contactwith patients.The intended use of the BoxView SmartAlarm Interface (SAI), Model SA-01 isto provide an interface withphysiological patient monitoring systemsto forward information associated to analarm event to a designated displaydevice(s).For medical, near real time alarms, theBoxView Smart Alarm Interface, ModelSA-01 is intended to serve as a parallel,redundant, mechanism to informhealthcare professionals of particularmedical alarm events. The BoxViewSmart Alarm Interface, Model SA-01does not alter the behaviorof the primary medical devices andassociated alarm annunciations.The BoxView Smart Alarm Interface,Model SA-01 is intended for use as asecondary alarm notification system. Itdoes not replace the primary alarmfunction on the monitor.The Early Sense Central Display Systemis intended to provide remote centralmonitoring and display of information asrecorded by multiple EarlySense bedsideunits, on a central remote screen. Thesystem can be used in hospitals orhospital type and clinic environment.

{8}------------------------------------------------

Zyter, Inc.
---------------
The BoxView Smart Alarm Interface, Model SA-01 is not intended to be used for diagnostic purposes. The BoxView Smart Alarm Interface, Model SA-01 is intended for use by professional clinical personnel and relies on proper use and operation of both the communication infrastructure in place at the healthcare facility and the display devices used. The BoxView Smart Alarm Interface, Model SA-01 is a software product and cannot come into physical contact with patients.
Rx or OTCRx onlyRx onlyRx only
Intended UserHealthcare professionals and lay-usersHealthcare professionalsHealthcare professionals
Devices itcommunicateswithThe Zyter RPM supported devicesinclude:The BoxView SAI provides an interfacewith physiological patient monitoringsystems. Specific medical devices orparameters the primary predicatesupports are not referenced in theirpublicly available 510(k) summary.The secondary predicate communicateswith EarlySense Bedside monitoringdevices (K131379 and K120465). Thetransmitted information from BedsideUnit to CDS and backwards includesalert information and physiologicalparameters (such as patient in / out ofbed status, heart rate, respiration rate,motion rate and SpO2, if monitored atbedside unit, as well as room and bednumber, etc.).
Blood pressureMonitoring BodyTrace Blood PressureMonitor iHealth BP Monitor Weighing scale BodyTrace Weighing Scale iHealth Weighing Scale Glucometer Smart Meter iGlucose BloodGlucose Monitor Starter Kit iHealth Blood Glucose Thermometer iHealth Thermometer PT3 Pulse oximeter iHealth Air Pulse Oximeter Nonin Pulse Oximeter Health kit & wearables Apple Health Kit Tablet
o Samsung Android Tablet
Alert/AlarmInterfaceTransports alert signals to displaydevices provided by compatible third-party mobile device companiesTransports alarm signals to displaydevices provided by compatible third-party mobile device companiesTransports alarm signals fromEarlySense Bedside monitoring devicesto a central remote screen (EarlySenseCDS).
ConfigurationsWeb browser-based applicationconfigured to send alert notifications tospecificUsers. Parameters can be adjusted, e.g.,change settable parameters like alertthresholds depending on patient.Web browser-based applicationconfigured to send alert notifications tospecificusersSoftware application is installed onstandard off-the-shelf PC computer withcomputer screen. Bedside unit'sparameters can be adjusted, e.g., changesettable parameters like alert thresholds.
HardwareRequirementsProcessor and RAM requirements:Processor - Single-core 1Ghz or higher(Minimum); 4GB RAM.Operating Systems:Ubuntu 16.04macOS X with macOS 10.9, Windows10, Windows 8 or 8.1Browser:Chrome 63, FireFox 57Network:Broadband wired or wireless (3G or4G/LTE)High-level hardware components neededinclude: Standard PC with monitor,keyboard, mouse, screenMinimum Specs forServer: 64-bit(x64) processor, twocore; 8GB Ram; 40GBHard Drive; Linux OS -CentOS 8 or Later(CentOS is a version ofLinux); Google Chrome54.0 or later browser;Network - 100MB.High-level hardware components neededinclude: Standard PC with monitor,keyboard, mouse, screen
CommunicationsOperationCommunication performed throughHTTPS/WSS protocol, Wirelesscommunications platform, WiFi, hospitalinfrastructure and BLE or LTE fordevicesWireless communications platform,WiFi, paging technology, hospitalinfrastructure.Communication performed throughTCP/IP protocol, wirelesscommunication, wired, externalcommunication devices e.g. pagers
Administration
User ID and Password required to enterthe configuration managerUser ID and Password required to enterthe configuration managerNot referenced in publicly available510(k).
Connection toNetworkCloud based application connectedthrough site to-site internet connectionMessage integration to hardware on-sitevia wired Ethernet connections; or Cloudbased application connected through siteto-site internet connectionCommunication performed throughTCP/IP protocol, via standard wired orwireless LAN
LoggingZyter RPM stores messages (encrypted),alerts, RPM Notes, Device readings.Stores messages, alarms, faults,input/output activities.Not referenced in publicly available510(k).
User InterfaceCloud based application with the userdashboard, where users can view alertmessages and reading values, add notesand set parameters, such as alertthresholds.Accept or reject messages if busy –automatically redirect toanother caregiver.Users can access user interface ofindividual bedside units via the CDS'sscreen and view or adjust bedside unit'sparameters, e.g. change settableparameters, such as alert thresholds.Remote view of the CDS screen from atablet or additional PC computer is alsopossible.
AlarmConfigurationFilter to provide specific alert signals tospecific providersFilter to provide specific alarm signals tospecific caregiversNot referenced in publicly available510(k).
PerformanceTestingPerformance testing of the Zyter RPMincluded:1. Risk Analysis2. Software verification andvalidation3. Performance bench testingPerformance testing done on number ofconcurrent transaction (readings fromdevice or alerts generated from areading) APIs can handle within astipulated interval. The transactions arestimulated through auto scripts to the testthe server capacity to identify when theapplication performance begins todegradeThe bench performance test for SmartAlarm Interface involves simulatingalert/alarm messages from concurrentalert/alarm sources to validate that belowand at server capacity the alarmmessages filter correctly and thatnotifications send to caregiver deviceswhen applicable. Another benchperformance test will involve simulatingalert/alarm messages over servercapacity to identify when the applicationperformance begins to degrade.Performance testing of the EarlySenseCDS included:4. Risk Analysis5. Software verification andvalidation6. Performance bench testing

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Sterilization & Shelf-life Testing

Not Applicable (Standalone Software)

Biocompatibility Testing

Not Applicable (Standalone Software)

Electrical safety and electromagnetic compatibility (EMC)

Not Applicable (Passive Device)

Software Verification and Validation Testing

Software verification and validation testing was conducted in compliance with ANSI/AAMI/IEC 62304:2006/A1:2016 Medical device software -Software life cycle processes, and the following FDA Guidance documents:

  • Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
  • । Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices
  • Guidance for Industry and Food and Drug Administration Staff, Content of Premarket — Submissions for Management of Cybersecurity in Medical Devices
  • Guidance for Industry and Food and Drug Administration Staff Postmarket | Management of Cybersecurity in Medical Devices

Human Factors Testing

A summative human factors study was conducted to evaluated the usability of the subject device by the intended users. The usability studies were conducted in compliance with ANSI/AAMI/EC 62366-1:2015/A1:2020, and the following FDA Guidance document:

  • Guidance for Industry and Food and Drug Administration Staff Applying Human Factors and Usability Engineering to Medical Devices

Animal Study

Animal performance testing was not required to demonstrate substantial equivalence of the device.

Human Clinical Performance Testing

Clinical testing was not required to demonstrate the substantial equivalence of the device.

VIII. CONCLUSIONS

The Zyter RPM is substantially equivalent to the identified predicates in terms of design features, fundamental scientific technology, and intended use. Substantial equivalence has been demonstrated with non-clinical performance testing. The verification and validation results provided in this 510(k) premarket notification demonstrate that the subject device, Zyter RPM, is substantially equivalent to the identified predicate devices.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).