K Number
K193421

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-10-30

(326 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the Box View Smart Alarm Interface (SAI). Model SA-01 is to provide an interface with physiological patient monitoring systems to forward information associated to an alarm event to a designated display device(s). For medical, near real time alarms, the BoxView Smart Alarm Interface, Model SA-01 is intended to serve as a parallel, redundant, mechanism to inform healthcare professionals of particular medical alarm events. The BoxView Smart Alarm Interface, Model SA-01 does not alter the primary medical devices and associated alarm annunciations.

The BoxView Smart Alarm Interface. Model SA-01 is intended for use as a secondary alarm notification system. It does not replace the primary alarm function on the monitor.

The BoxView Smart Alarm Interface. Model SA-01 is not intended to be used for diagnostic purposes. The Box View Smart Alarm Interface, Model SA-01 is intended for use by professional clinical personnel and relies on proper use and operation of both the communication in place at the healthcare facility and the display devices used. The BoxView Smart Alarm Interface, Model SA-01 is a software product and cannot come into physical contact with patients.

Device Description

The BoxView Smart Alarm Interface (SAI), Model SA-01, is a Software Medical Device (SaMD) product intended to be located on-site in the hospital, or pre-configured off site in the 'cloud' utilizing a standard Linux operating system. The primary purpose of the BoxView Smart Alarm Interface (SAI), Model SA-01 is to act as a message integrator to forward patient monitor status and alarm event information originating from a patient monitoring network. Users receive interactive, time-critical information from patient monitoring devices directly via their display devices as text, alarms or data. The BoxView Smart Alarm Interface (SAI), Model SA-01 allows caregivers to be informed of their patient's alarm conditions when they are not in the patient vicinity.

The BoxView Smart Alarm Interface (SAI), Model SA-01 is an open system that is compatible with most smart phones or computers. The BoxView Smart Alarm Interface (SAI), Model SA-01 connects to the information sources through wired Ethernet connections which are part of the customer's infrastructure. The BoxView Smart Alarm Interface (SAI), Model SA-01 software acquires patient data from patient monitoring devices and allows the user to configure the system to determine which information, including alarm notifications, is delivered to which users communicators. The BoxView Smart Alarm Interface (SAI), Model SA-01 then formats the data for delivery to the display devices.

The BoxView Smart Alarm Interface (SAI), Model SA-01 system is designed to forward alarm event information as the alarms are recognized by the patient monitoring network. The system is also capable of being configured to periodically forward a patient's physiological data as well. All messaging activities are recorded by the BoxView Smart Alarm Interface (SAI), Model SA-01 providing real-time activity logging for audit trail records and reporting.

The BoxView Smart Alarm Interface (SAI), Model SA-01, system is a secondary alarm notification system. It does not replace the primary alarm function of the bedside monitor or telemetry monitoring system.

AI/ML Overview

The provided text describes the acceptance criteria and the study conducted for the BoxView Smart Alarm Interface (SAI), Model SA-01, a Software as a Medical Device (SaMD) that forwards information associated with alarm events from physiological patient monitoring systems to designated display devices.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Process and respond to at least 250 alarm signals per second.Smart Alarm Interface was able to correctly process and respond to 250 alarm signals per second over various time intervals.
Alarm messages filter correctly and notifications send to caregiver devices when applicable.The correct number of alarm signals were created in Smart Alarm Interface. For alarm signals that did match rules, start times were compared to validate the time periods that the alarm signals were received. Additionally, for each alarm signal created, a corresponding notification was created and a push notification was sent (validated by checking the "sendon" column in the notification table).
Error rate of 0.0% for alarm signal filtering and notification creation.At around 400 alarm signals a second, while throughput declined slightly and timestamps were delayed, the alarm signal filtering and notification creation was still handling correctly with an error rate of 0.0%.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: The test involved simulating 250 alert/alarm messages per second from 25 concurrent alert/alarm sources. The document also mentions testing "around 400 alarm signals a second," indicating a range of simulated loads. It's a simulated environment, not real patient data.
  • Data Provenance: The data used for testing was simulated. XprezzNet Simulator was used to simulate alarm messages from Spacelabs patient monitors. The testing was performed in a "simulated hospital network environment." This implies the data is synthetic and not from a specific country of origin, nor is it retrospective or prospective clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The ground truth for this device's performance testing was established by the expected behavior of the system based on its predetermined specifications (e.g., if an alarm signal matched a rule, a notification should be created and sent). It was a software functionality and performance test, not a diagnostic or clinical accuracy study requiring expert human interpretation or consensus.

4. Adjudication method for the test set

Not applicable. As noted above, this was a functional and performance test against predefined software rules and expected outputs, not a subjective assessment requiring human adjudication. The validation was done by comparing the number of sent simulated alarm signals to the number of alarm signals created in the database and checking the population of notification tables.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This device is a "Smart Alarm Interface," a middleware that forwards alarm information. It is not an AI-powered diagnostic tool that assists human readers in interpreting medical images or data. Therefore, an MRMC comparative effectiveness study involving human readers' improvement with or without AI assistance is not relevant to this device's function or the tests described.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the performance testing described is a standalone (algorithm only) test. The "Smart Alarm Interface" (SAI) operates as a software product within a simulated environment, processing simulated alarm signals and generating notifications based on its configured rules. The test specifically validates the SAI's ability to process and respond to alarm signals, filter them, and create/send notifications, without direct human intervention in the real-time processing of signals.

7. The type of ground truth used

The ground truth used was based on the predetermined specifications and expected behavior of the Smart Alarm Interface application after processing simulated alarm signals. Specifically:

  • Expected number of alarm signals created in the alarm database based on simulated input.
  • Correct filtering of alarm signals based on predefined rules.
  • Creation of corresponding notifications for matched alarm signals.
  • Population of the "sendon" column in the notification table at the expected time.

8. The sample size for the training set

Not applicable. The document describes performance testing for a software interface, not a machine learning model that requires a training set. The BoxView Smart Alarm Interface (SAI) is a message integrator and secondary alarm notification system, indicating it follows predefined logic and rules, rather than learning from data.

9. How the ground truth for the training set was established

Not applicable, as there was no training set for a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 30, 2020

BoxView, LLC % Robert Steurer Principal Consultant Steurer Consulting Group 800 Blue Quail Rd. Keller, Texas 76248

Re: K193421

Trade/Device Name: Box View Smart Alarm Interface (SAI), Model SA-01 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MSX Dated: July 28, 2020 Received: July 31, 2020

Dear Robert Steurer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193421

Device Name

BoxView Smart Alarm Interface (SAI), Model SA-01

Indications for Use (Describe)

The intended use of the Box View Smart Alarm Interface (SAI). Model SA-01 is to provide an interface with physiological patient monitoring systems to forward information associated to an alarm event to a designated display device(s). For medical, near real time alarms, the BoxView Smart Alarm Interface, Model SA-01 is intended to serve as a parallel, redundant, mechanism to inform healthcare professionals of particular medical alarm events. The BoxView Smart Alarm Interface, Model SA-01 does not alter the primary medical devices and associated alarm annunciations.

The BoxView Smart Alarm Interface. Model SA-01 is intended for use as a secondary alarm notification system. It does not replace the primary alarm function on the monitor.

The BoxView Smart Alarm Interface. Model SA-01 is not intended to be used for diagnostic purposes. The Box View Smart Alarm Interface, Model SA-01 is intended for use by professional clinical personnel and relies on proper use and operation of both the communication in place at the healthcare facility and the display devices used. The BoxView Smart Alarm Interface, Model SA-01 is a software product and cannot come into physical contact with patients.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
of this information collection, including suggestions for reducing this burden, to:The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspect
Food and Drug AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) StaffPRAStaff@fda.hhs.govDepartment of Health and Human Services
"An agency may not conduct or sponsor, and a person is not required to respond to. a collection ofinformation unless it displays a currently valid OMB number."
EORM EDA 3884 17/47)Page 1 of 1

{3}------------------------------------------------

Date Prepared:October 29, 2020
Submitter:BoxView, LLC14001 McAuley Blvd., Suite 220Oklahoma City, OK 73134
Submitter Contact:BoxView, LLC14001 McAuley Blvd., Suite 220Oklahoma City, OK 73134Mr. Cody GuestBoxView, LLCPhone: +1 (405) 740-5785Fax:Email: cody.guest@boxview.com
ApplicationCorrespondent:Bob SteurerPrincipal ConsultantSteurer Consulting Group800 Blue Quail Rd.Keller, TX 76248steurerbob@gmail.com+1 (425) 358-1072
Manufacturing Site:BoxView, LLC14001 McAuley Blvd., Suite 220Oklahoma City, OK 73134
Trade Name:BoxView Smart Alarm Interface (SAI), Model SA-01
Common Name:Network and Communication Middleware
Classification Name:System, Network and Communication, Physiological Monitors
PrimaryClassificationRegulation:21 CFR §870.2300
Primary ProductCode:MSX
Predicate Device:Ascom Mobile Monitoring Gateway (MMG), K111215

{4}------------------------------------------------

SubstantiallyEquivalent Devices:CompanyPredicate 510(k)NumberPredicateManufacturer / Model
AscomK111215Ascom Mobile MonitoringGateway (MMG)
Device Description:The BoxView Smart Alarm Interface (SAI), Model SA-01, is a SoftwareMedical Device (SaMD) product intended to be located on-site in thehospital, or pre-configured off site in the 'cloud' utilizing a standard Linuxoperating system. The primary purpose of the BoxView Smart AlarmInterface (SAI), Model SA-01 is to act as a message integrator to forwardpatient monitor status and alarm event information originating from apatient monitoring network. Users receive interactive, time-criticalinformation from patient monitoring devices directly via their displaydevices as text, alarms or data. The BoxView Smart Alarm Interface (SAI),Model SA-01 allows caregivers to be informed of their patient's alarmconditions when they are not in the patient vicinity.The BoxView Smart Alarm Interface (SAI), Model SA-01 is an open systemthat is compatible with most smart phones or computers. The BoxViewSmart Alarm Interface (SAI), Model SA-01 connects to the informationsources through wired Ethernet connections which are part of thecustomer's infrastructure. The BoxView Smart Alarm Interface (SAI),Model SA-01 software acquires patient data from patient monitoringdevices and allows the user to configure the system to determine whichinformation, including alarm notifications, is delivered to which userscommunicators. The BoxView Smart Alarm Interface (SAI), Model SA-01then formats the data for delivery to the display devices.The BoxView Smart Alarm Interface (SAI), Model SA-01 system is designed to forward alarm event information as the alarms are recognized by thepatient monitoring network. The system is also capable of beingconfigured to periodically forward a patient's physiological data as well.All messaging activities are recorded by the BoxView Smart AlarmInterface (SAI), Model SA-01 providing real-time activity logging for audittrail records and reporting.The BoxView Smart Alarm Interface (SAI), Model SA-01, system is asecondary alarm notification system. It does not replace the primaryalarm function of the bedside monitor or telemetry monitoring system.

{5}------------------------------------------------

Indications/Intended The intended use of the BoxView Smart Alarm Interface (SAI), Model SA-Use: 01 is to provide an interface with physiological patient monitoring systems to forward information associated to an alarm event to a designated display device(s).

For medical, near real time alarms, the BoxView Smart Alarm Interface, Model SA-01 is intended to serve as a parallel, redundant, mechanism to inform healthcare professionals of particular medical alarm events. The BoxView Smart Alarm Interface, Model SA-01 does not alter the behavior of the primary medical devices and associated alarm annunciations.

The BoxView Smart Alarm Interface, Model SA-01 is intended for use as a secondary alarm notification system. It does not replace the primary alarm function on the monitor.

The BoxView Smart Alarm Interface, Model SA-01 is not intended to be used for diagnostic purposes. The BoxView Smart Alarm Interface, Model SA-01 is intended for use by professional clinical personnel and relies on proper use and operation of both the communication infrastructure in place at the healthcare facility and the display devices used. The BoxView Smart Alarm Interface, Model SA-01 is a software product and cannot come into physical contact with patients.

Summary of Substantial Equivalence:

The predicate device is intended to interface to GE Healthcare patient monitoring network to forward alarm signals from patient monitors and telemetry transmitters to specified caregivers which is the same as the BoxView SAI which is designed to interface to the Spacelabs patient monitoring network using protocol provided from Spacelabs. Both the predicate device and the BoxView SAI provide a secondary means of alarm annunciation to a mobile device or computer. Neither the predicate nor the BoxView SAI alters the context of the alarm signal, the behavior of the monitoring system, and neither is intended for diagnostic use. The differences in intended use are not critical to the safety and effectiveness of the device when used per labeling instructions.

The predicate device is configured on hardware provided by Ascom while the BoxView SAI can be configured on local hardware that meets minimum requirements specified by BoxView and can be configured as a cloud-based system. The hardware specifications from BoxView reflect newer technology than that offered by Ascom in 2011/2012. This does

{6}------------------------------------------------

not alter the intended use of the BoxView SAI but provides additional flexibility for users. This difference does not affect safety or effectiveness of the BoxView SAI when hardware is used as specified.

Technology Comparison:

The BoxView Smart Alarm Interface, Model SA-01 employs the same technological characteristics as the predicate device.

CharacteristicPredicate Device –Ascom MMGSmart Alarm InterfaceModel SA-01
Alarm InterfaceTransports alarm signalsto display devicesprovided by Ascom orthird-party mobiledevice companies.Transports alarm signalsto display devicesprovided by compatiblethird-party mobiledevice companiesincluding Ascom
ConfigurationsWeb browser-basedapplication configuredto send alarmnotifications to specificusersWeb browser-basedapplication configuredto send alarmnotifications to specificusers
HardwareRequirementsServer: (Elise EmbeddedLinux Server) Appliancewith Microsoft® InternetExplorer® 6.0 or laterbrowser.Sun™ Java™ RuntimeEnvironment (JRE) 6 orlater; Network10 baseT or 100 baseT(10 MB or 100 MB)Minimum Specs forServer: 64-bit(x64) processor, twocore; 8GB Ram; 40GBHard Drive; Linux OS -CentOS 8 or Later(CentOS is a version ofLinux); Google Chrome54.0 or later browser;Network - 100MB.
CommunicationsOperationWirelesscommunicationsplatform, WiFi, pagingtechnology, hospitalinfrastructure.Wirelesscommunicationsplatform, WiFi, pagingtechnology, hospitalinfrastructure.
AdministrationUser ID and Password required to enter the configuration managerUser ID and Password required to enter the configuration manager
Connection to NetworkMessaging integration to hardware on-site via wired Ethernet connectionsMessage integration to hardware on-site via wired Ethernet connections; or Cloud based application connected through site-to-site internet connection
LoggingStores messages, alarms, faults, input/output activities.Stores messages, alarms, faults, input/output activities.
User InterfaceAccept or reject messages if busy – automatically redirect to another caregiver.Accept or reject messages if busy – automatically redirect to another caregiver.
Alarm ConfigurationFilter to provide specific alarm signals to specific caregiversFilter to provide specific alarm signals to specific caregivers

{7}------------------------------------------------

{8}------------------------------------------------

Performance Testing Summary

Non-Clinical Bench Performance Testing Information

Test Report Summaries

    1. Test(s) Performed
      The bench performance test for Smart Alarm Interface involves simulating alert/alarm messages from concurrent alert/alarm sources to validate that below and at server capacity the alarm messages filter correctly and that notifications send to caregiver devices when applicable. Another bench performance test will involve simulating alert/alarm messages over server capacity to identify when the application performance begins to degrade.
    1. Objective(s) of the Test(s)
      The first objective of the test is to confirm that Smart Alarm Interface can process and respond to 250 alert/alarm messages per second from 25 concurrent alert/alarm sources, this results to each alert/alarm source sending 10 alerts/alarm signals a second.

The second objective is to determine the number of alert/alarm messages and concurrent alert/alarms sources in which the application performance starts to worsen. Performance test is also used to help identify any bottle necks or areas of the code that have the potential to be better optimized. These performance tests are to be executed before every major software release to confirm that the application is performing per requirements.

    1. Test Methods
    • a. Test Environment
      • a. XprezzNet Simulator Used to simulate alarm messages from Spacelabs patient monitors. (XprezzNet is the name Spacelabs has designated for their network communication)
      • b. Amazon Web Services A cloud computing service that hosts the docker container(s) that run the Smart Alarm Interface application and Postgres database.
      • c. Mobile Devices iOS and Android communicator devices.
        • i. iPhone XR iOS 13.4.1
        • ii. iPhone 8 iOS 12.3.1
        • iii. iPhone XR iOS 13.5.1
        • iv. iPhone X iOS 13.1
        • iPhone XS iOS 13.3.1 V.
        • vi. Samsung Galaxy A10e Android 9
        • vii. Google Pixel 4 Android 10b
    • b. JMeter is used to execute requests with the XprezzNet simulate alarm sources that send alert/alarm messages to validate that Smart Alarm Interface filters alarm signals properly and creates or updates alarm signals and notifications. A set number of alarm signals will be sent from a set number of alarm sources over different intervals of time to simulate and validate the applications performance under different load conditions. The alarm signals are sent to the alarm source end point and injected into the application to begin the filtering process where corresponding alarm signals and/or notifications are created if they match predefined rules. If a notification is created, then a push notification is expected to send to a mobile device. The performance test is executed in a simulated hospital network environment.

{9}------------------------------------------------

4. Pass/Fail Criteria

The pass criteria of the performance test are that Smart Alarm Interface can process and respond to at least 250 alarm signals a second. If Smart Alarm Interface is not able to meet the pass criteria or if it processes an alarm signal incorrectly the test is considered a failure and the deviation will be corrected immediately.

5. Results Summary

The acceptance criteria were met for multiple iterations of the tests performed. Smart Alarm Interface was able to correctly process and respond to 250 alarm signals a second over various time intervals. The results were validated by comparing the number of sent simulated alarm signals that match a rule in Smart Alarm Interface to the number of alarm signals created in the alarm database. Several simulated alert signals intentionally do not match a rule so we know Smart Alarm Interface can properly match alarm signals to rules under high use. During the simulation, the correct number of alarm signals were created in Smart Alarm Interface. For alarm signals that did match rules, start times were compared to validate the time periods that the alarm signals were received. Additionally, the results were validated by making sure that for each alarm signal created in Smart Alarm Interface had a corresponding notification that was created and that a push notification was sent by checking that the "sendon" column in the notification table was populated with a value that matched the expected time period.

At around 400 alarm signals a second, throughput started to decline slightly with alarm signal and notification timestamps started to be delayed slightly, but the alarm signal filtering and notification creation was still handling correctly with an error rate of 0.0%. It is unclear if the decline in throughput is due to the application performance or to the test setup.

6. Discussion/Conclusions

The performance test used simulated alarm signals from alarm sources where the alarm signals were injected directly into the Smart Alarm Interface application. It is plausible that results may vary in production with physical alarm source devices. The objective of this test was to validate the speed at which Smart Alarm Interface can process alarm signals and create/push corresponding notifications. This test is not intended to validate the performance of the alarm source or the notification messaging system. Based on the test results Smart Alarm Interface can successfully process the recommended server capacity of 250 alarm signals per second from 25 concurrent alarm signal sources and potential to process above that under extremely high usage.

{10}------------------------------------------------

Sterilization andShelf-lifeThe BoxView Smart Alarm Interface (SAI) is a Software as a Medical Device(SaMD) product and is not provided sterile. The BoxView Smart AlarmInterface (SAI) does not have a shelf-life.Therefore, this section is not applicable
BiocompatibilityThe BoxView Smart Alarm Interface (SAI) is a Software as a Medical Device(SaMD) product and does not directly nor indirectly contact the patient.Therefore, this section is not applicable.
Software TestingThe BoxView Smart Alarm Interface (SAI) was designed and developedaccording to the BoxView internal software development process. TheBoxView Smart Alarm Interface (SAI) was tested using verification andvalidation methods, the results of which indicate the Smart AlarmInterface (SAI) complies with its specifications.
Electrical SafetyThe BoxView Smart Alarm Interface (SAI) is a Software as a Medical Device(SaMD) product. Therefore, this section is not applicable.
ElectromagneticCompatibilityTestingThe BoxView Smart Alarm Interface (SAI) is a Software as a Medical Device(SaMD) product designed to perform its intended purpose without beingpart of a hardware medical device. Therefore, this section is notapplicable.
Performance Testing– BenchThe BoxView Smart Alarm Interface (SAI) was tested for performance inaccordance with its predetermined specifications.
The results of the performance testing show the BoxView Smart AlarmInterface (SAI) complies with these specifications.
ConclusionA comparison of the predicate device and a review of the testing resultsshows that the BoxView Smart Alarm Interface (SAI) is substantiallyequivalent to the predicate device.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).