K Number
K212613

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-12-21

(125 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A Nitrile powder free examination glove is a disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between examiner and patient. This specially glove has also been tested with Chemotherapy drugs and the Opioid Fentanyl citrate.

Device Description

The K212613 gloves are produced as nitrile synthetic gloves. They are manufactured without natural rubber latex. The nitrile gloves are blue in color, disposable, single-use only, non-sterile, medical examination gloves. K212613 meets all the requirements of ASTM D6978-10(2019) Standard Specifications for Examination Gloves for Medical Applications. They have been tested for use with chemotherapy drugs and the opioid Fentanyl citrate, both of which were evaluated using ASTM D6378-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

AI/ML Overview

The provided document describes the acceptance criteria and performance of Powder Free Nitrile Examination Gloves, Non-sterile, Tested for Use with Chemotherapy Drugs and Opioid Fentanyl Citrate (Blue). The evaluation is based on non-clinical performance tests, not clinical studies involving human or animal subjects. Therefore, sections related to human reader studies, ground truth establishment by experts, and training set information are not applicable.

Here's the information extracted from the document:

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D6978-2005 (2019)Permeation by Chemotherapy DrugsMinimum breakthrough detection time > 240 minutes for most drugsAll Chemotherapy Drugs & the Opioid passed >240 min except Carmustine (13.0 min) & ThioTepa (37.9 min)
ASTM D412-2016Physical properties (Tensile strength)Minimum 14 MpaPass
ASTM D412-2016Physical properties (Elongation)Minimum 400%Pass
ASTM D5151-19Water-leak testAQL 1.5 (ISO 2859-1)Pass
ASTM D6124-06 (2017)Residual powder test< 2mg/glovePass
ISO 10993-10:2010 (2016)Dermal IrritationUnder conditions of this study, the subject device is non-irritatingPass
ISO 10993-10:2010 (2016)Skin SensitizationUnder conditions of this study, the subject device is non-sensitizingPass
ISO 10993-11: Tests for Systemic ToxicitySystemic Toxicity (Biocompatibility)Under the conditions of the Systemic Toxicity test, the predicate glove showed no adverse biological reaction.Passes (Subject device used ISO 10993-11 as an accepted alternative to the predicate's ISO 10993-5)
ISO 10993-5 (Cytotoxicity)Cytotoxicity (Biocompatibility)Not explicitly stated for subject device, but predicate passed at certain concentrationsNot directly tested for subject device; ISO 10993-11 Acute Systemic Toxicity was used as an accepted alternative

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes (e.g., number of gloves) used for each non-clinical test. The tests are standard industry tests (e.g., ASTM, ISO standards) typically performed by manufacturers and independent labs. The data provenance is from the manufacturer's testing (Koon Seng Sdn. Bhd. located in Malaysia) or accredited laboratories performing these standard tests. These are non-clinical, controlled laboratory tests, not retrospective or prospective clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This is not applicable as the evaluation is based on objective, non-clinical laboratory tests according to specified standards, not expert interpretation of outputs like in AI/imaging studies.

4. Adjudication method for the test set

This is not applicable as the evaluation is based on objective, non-clinical laboratory tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study was done. This device is an examination glove, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not applicable. The device is a physical product (gloves), not an algorithm. The tests performed are standalone evaluations of the glove's physical and chemical resistance properties.

7. The type of ground truth used

The "ground truth" for the performance criteria is defined by established international and national standards for medical gloves (e.g., ASTM D6978, ASTM D412, ASTM D5151, ASTM D6124, ISO 10993). These standards specify the test methodologies and acceptable limits for physical properties, chemical permeation, and biocompatibility.

8. The sample size for the training set

This is not applicable as this is not a machine learning or AI device.

9. How the ground truth for the training set was established

This is not applicable as this is not a machine learning or AI device. The "ground truth" for the performance of the gloves is established by adherence to the objective criteria and methods outlined in the mentioned ASTM and ISO standards for medical devices and materials.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

December 21, 2021

Koon Seng Sdn Bhd % Wava Truscott President Truscott MedSci Associates, LLC. 180 Burkemeade Ct. Roswell, Georgia 30075

Re: K212613

Trade/Device Name: Powder Free Nitrile Examination Gloves, Non-sterile,Tested For Use with Chemotherapy Drugs and Opioid Fentanyl Citrate (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QPO Dated: November 15, 2021 Received: November 22, 2021

Dear Wava Truscott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K212613

Device Name

Trade Name: Powder Free Nitrile Examination Gloves, Non-sterile, Tested for Use with Chemotherapy Drugs and the Opiod Fentanyl Citrate (Blue)

Indications for Use (Describe) A Nitrile powder free examination glove is a disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between examiner and patient. This specially glove has also been tested with Chemotherapy drugs and the Opioid Fentanyl citrate.

Chemotherapy Drugs and the opioid Fentanyl citrate were tested for breakthrough detection times:
--------------------------------------------------------------------------------------------------------
TESTCHEMOTHERAPY DRUGCONCENTRATIONTESTEDMINIMUM BREAKTHROUGHDETECTION TIME (MINUTES)
Carmustine (BCNU)3.3 mg/ml13.0
Cisplatin1.0 mg/mlNo Breakthrough Up To 240 minutes
Cyclophosphamide (Cyloxan)20.0 mg/mlNo Breakthrough Up To 240 minutes
Cytarabine100.0 mg/mlNo Breakthrough Up To 240 minutes
Dacarbazine (DTIC)10.0 mg/mlNo Breakthrough Up To 240 minutes
Doxorubicin Hydrochloride2.0 mg/mlNo Breakthrough Up To 240 minutes
Etoposide (Toposar)20.0 mg/mlNo Breakthrough Up To 240 minutes
Fluorouracil50.0 mg/mlNo Breakthrough Up To 240 minutes
Ifosfamide50.0 mg/mlNo Breakthrough Up To 240 minutes
Methotrexate25.0 mg/mlNo Breakthrough Up To 240 minutes
Mitomycin C0.5 mg/mlNo Breakthrough Up To 240 minutes
Mitoxantrone2.0 mg/mlNo Breakthrough Up To 240 minutes
Pacitaxel (Taxol)6.0 mg/mlNo Breakthrough Up To 240 minutes
Thiotepa10.0 mg/ml37.9
Vincristine Sulfate1.0 mg/mlNo Breakthrough Up To 240 minutes

Important: Carmustine and Thiolepa have extremely low minimal breakthrough times of 13.0 minutes and 37.9 minutes respectively. Warning: Do not use with Carmusline or Thiotepa.

TEST OPIOID DRUGCONCENTRATIONTESTEDMINIMUM BREAKTHROUGHDETECTION TIME (MINUTES)
Fentanyl Citrate (Injectable)100.0 mg/2mlNo Breakthrough up to 240 minutes

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo featuring two golden hands overlapping each other, set against a black background. The logo is encircled by a golden ring, and the text "Delivering Quality Excellence in Gloves" is written around the ring. The text is in a smaller font size and follows the curve of the circle.

QUALITY, TRANSPARENCY, PARTNER

OON SENG SDN. BHD = (107051-H)

(A member of the KS Group) Corporate Sales Office: 1321, 13th Floor, BlockA, Lift Lobby &, Damancara Intan No.1,
Jin. SS 20/27, 47400 Petaling Jaya, Selangor. Factory 1 : PTD 3274, Batu 41/2, Jalan Bakri, 84000 Muar, Johor, Factory 2 : PTD 16058, Kaw. Perindustrian Tangkak, 84900 Tangkak, Johor.

Email : koonseng@koonseng.com Website : www.koonseng.com Tel : 603- 7733 1388

Tel : 606-9868446 , 606-986 9257 Fax: 606-986 6908 Tel: 606-979 1778 , 606-978 2780 Fax: 606-979 1779

Image /page/3/Picture/6 description: The image contains several certification logos. There is an SMG logo, a CE logo, an ISO 22000 registered logo, and an ISO 13485 registered logo. These logos likely indicate that the product or company associated with the image meets certain standards for quality, safety, or environmental management.

K212613: 510(k) Summary

1.0Submitter
Name:Koon Seng, Sdn. Bhd.
Address:PTD 16058, Kawasan Perindustrian Tangkak,84900 Tangkak, Johor, Malaysia
Contact:Yeo Pai Shuang
Phone No.:+603-7733 1388
Fax No.:+603-7733 6499
Date of Preparation:June 29, 2021
2.0Name of the Device
Common Name:Nitrile Powder Free Examination Glove
Trade Name:Powder Free Nitrile Examination Gloves, Non-sterile,Tested for Usewith Chemotherapy Drugs and the Opioid Fentanyl Citrate (Blue)
Classification Name:Patient Examination Glove Specialty (21CFR 880.6250)
Product Codes:LZA, LZC, OPJ, QDO
Device Class:Class I
3.0Predicate: Identification of Legally Marketed Device
Predicate 510(k) No.:K192954
Trade Name:Blue Colored, Powder Free Nitrile Examination Gloves Tested for Use withChemotherapy Drugs and Fentanyl Citrate
Owner:Comfort Rubber Gloves Industries Sdn. Bhd.
Product Codes:LZA, LZC, QDO
Device Class:Class I

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Koon Seng. Above the name is the text "Quality - Transparency - Partnership". Below the name is the text "Since 1987".

4.0 Description of Subject Device

The K212613 gloves are produced as nitrile synthetic gloves. They are manufactured without natural rubber latex. The nitrile gloves are blue in color, disposable, single-use only, non-sterile, medical examination gloves. K212613 meets all the requirements of ASTM D6978-10(2019) Standard Specifications for Examination Gloves for Medical Applications. They have been tested for use with chemotherapy drugs and the opioid Fentanyl citrate, both of which were evaluated using ASTM D6378-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

5.0 Indications for Use

A Nitrile powder free examination glove is a disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between examiner and patient. This specialty glove has also been tested for use with Chemotherapy drugs and the Opioid Fentanyl citrate.

Test Chemotherapy DrugConcentrationMinimum Breakthrough Detection Time(minutes)
Carmustine (BCNU)3.3 mg/mL13 min.
Cisplatin1.0 mg/mL>240 min.
Cyclophosphamide (Cytoxan)20.0 mg/mL>240 min.
Cytarabine100 mg/mL>240 min.
Dacarbazine10.0 mg/mL>240 min.
Doxorubicin2.0 mg/mL>240 min.
Etoposide20.0 mg/mL>240 min.
Fluorouracil50.0 mg/mL>240 min.
Ifosfamide50.0 mg/mL>240 min.
Methotrexate25.0 mg/mL>240 min.
Mitomycin C0.5 mg/mL>240 min.
Mitoxantrone2.0 mg/mL>240 min.
Paclitaxel6.0 mg/mL>240 min.
ThioTepa10.0 mg/mL37.9 min
Vincristine Sulfate1.0 mg/mL>240 min.
Important: Carmustine and ThioTepa have extremely low minimum breakthrough times of13.0 minutes and 37.9 minutes respectively.
Warning: Do not use with Carmustine or ThioTepa.
Test Opioid DrugConcentrationMinimum Breakthrough Detection Time(minutes)
Fentanyl Citrate Injection100mcg/2mL>240 min.
Please Note: Glove used for protection against possible Chemotherapy Drug exposureshould be selected specifically for the type of drugs used. Users should review drug labelingor material safety data sheets for the drugs being used to determine an adequate level ofprotection.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Koon Seng. The logo includes the words "Quality • Transparency • Partnership" at the top. Below that is the name "KOON SENG" in large, bold letters. Finally, the words "Since 1987" are at the bottom of the logo.

Technological Characteristic Comparison:

Shown below is a comparison of The Powder Free Nitrile Examination Gloves, Non-sterile, Tested for Use with Chemotherapy Drugs and Opioid Fentanyl Citrate (Blue) and Blue Colored, Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.

Glove Comparison Table":

SubjectStandardsKoon Seng SubjectDevice510(k): K212613Predicate Device:510(k): K192954Comparison K212613Subject glove toPredicate K192954
ManufacturersNAKoon Seng Sdn. Bhd.Comfort Rubber GlovesIndustries Sdn. Bhd.Different
Trade NameNAPowder Free NitrileExamination Gloves, Non-Sterile, Tested for Use withChemotherapy Drugs and theOpiod Fentanyl Citrate (Blue)Blue Colored, Powder FreeNitrile Examination GloveTested for Use withChemotherapy Drugs andFentanyl CitrateDifferent
CommonNameASTM D6319 -10(2019)Nitrile Powder free PatientExamination GloveNitrile PatientExamination GloveSimilar
Device Class21 CFR880.6250Class 1Class 1Same
Base MaterialNANitrileNitrileSame
ColorNABlueBlueSame
Product Codes21 CFR880.6250LZA, LZC, OPJ, QDOLZA, LZC, QDOSimilar OPJ is new: denotestested with Chemotherapydrugs to further define theLZC specialty designation
Sizes offeredASTM D3767Extra SmallSmallMediumLargeExtra LargeExtra SmallSmallMediumLargeExtra LargeSame
Dimensions:ASTM D6319 -10(2019)MeetsMeetsSame
Sterile v.Non-SterileNANon-SterileNon-SterileSame
Prescription orOTCNAOTCOTCSame
Single Use-DisposableNASingle use-DisposableSingle use-DisposableSame
Thickness:Finger & palmASTM D6319 -10(2019)MeetsMeetsSame
Before agingPhysicalPropertiesASTM D412ASTM D6319 -10(2019)MeetsMeetsSame
After agingPhysicalPropertiesASTM D412ASTM D6319MeetsMeetsSame
Freedom fromHolesASTM D5151ASTM D6319Meets AQL 1.5Meets AQL 1.5Same
SubjectStandardsKoon Seng SubjectDevice510(k): K212613Predicate Device:510(k): K192954Comparison K212613Subject glove toPredicate K192954
Powder-FreeASTM D6124< 2mg/gloveMeetsMeetsSame
Intended Use(part ofIndications forUse)A Nitrile powder freeexamination glove is adisposable device intendedfor medical purposes, wornon the examiner's hand orfinger to preventcontamination betweenexaminer and patient. Thisspecialty glove has also beentested for use withChemotherapy drugs and theOpioid Fentanyl citrate.A patient examinationglove is a disposabledevice intended formedical purpose that isworn on the examiner'shand to preventcontamination betweenpatient and examiner.Similar
Indications forUseChemotherapyASTM D6978Chemotherapy drugs:All >240min BreakthroughTimes except Carmustine &ThioTEPA• Cisplatin, 1mg/ml• Cyclophosphamide(Cytoxan), 20mg/ml• Cytarabine, 100mg/ml• Dacarbazine, 10mg/ml• Doxorubicin HCL, 2mg/ml• Etoposide, 20mg/ml• Fluorouracil, 50mg/ml• Ifosfamide, 50mg/ml• Methotrexate, 25mg/ml• Mitomycin C, 0.5mg/ml• Mitoxantrone, 2mg/ml• Paclitaxel, 6mg/ml• Vincristine Sulfate,mg/ml• Carmustine (BCNU), 3.3mg/ml:Breakthrough: 13 min.• ThioTepa, 10mg/mlBreakthrough: 37.9 min.Chemotherapy drugs:All>240min BreakthroughTimes except Carmustine &ThioTEPA•Cisplatin 1mg/ml• Cyclophosphamide(Cytoxan) 20mg/ml• Dacarbazine (DTIC) 10mg/ml• Doxorubicin HCL, 2mg/ml• Hydrochloride 2mg/ml• Etoposide (Toposar)20mg/ml• FluorouracilS0mg/ml• Paclitaxel (Taxol)6mg/ml• Carmustine (BCNU) 3.3mg/ml:Breakthrough: 18.2 min.• ThioTepa (THT) 10mg/ml:Breakthrough: 57.3 min.Similar
SubjectStandardsKoon Seng SubjectDevice510(k): K212613Predicate Device:510(k): K192954Comparison K212613Subject glove toPredicate K192954
Cautions &WarningsImportant: Carmustine andThioTEPA have extremely lowminimal breakthrough times of13.0 minutes and37.9 minutes respectively.Warning: Do not use withCarmustine or ThioTepaPlease note that thefollowing drugs haveextremely low permeationtimes: Carmustine (BCNU):18.2minutes and Thiotepa:57.3 minutes. Warning:Do not use withCarmustine.Similar: except Subjectdevice added ThioTepa tothe " Warning: Do not usewith..." listing. Technically,it is not required to do soas the cut-off to requirethe warning is below 30minutes.
Indications forUse: FentanylASTM D6978Fentanyl Citrate (Injectable)100mg/2ml:Minimumbreakthrough time: >240Fentanyl Citrate 100mg/2ml:Minimum breakthrough time:>240Similar: Fentanyl Citrateconcentration & diluentsame. US DEA requiredtesting street injectableterm by test lab in testresults
BiocompatibilityISO 10993-11: Testsfor SystemicToxicityPassesUnder the conditions of TheSystemic Toxicity test thepredicate glove showed noadverse biological reaction.(Used Accepted Alternative:ISO 10993-5 (Cytotoxicity)Similar: Used Differentalternativebiocompatibilitytest method: Bothacceptable
ISO 10993-5Cytotoxicity(Used Accepted Alternative:ISO 10993-11 Acute SystemicToxicity)Passed at: Exhibits severecytotoxicity reactivity at100%, and 66% extractconcentrations and nocytotoxicity reactivity at44%, 30%, 20% and 15%extract concentrationsunder the condition of thistest.Similar: Used DifferentAccepted AlternativeBiocompatibility test
ISO 10993-10SkinIrritationPassesUnder conditions of thisstudy, subject glove is non-irritatingPassesUnder conditions of thestudy, the device is non-irritatingSame
ISO 10993-10DermalSensitizationPassesUnder conditions of thisstudy, subject glove is anon-sensitizerPassesUnder conditions of thestudy, the device is non-sensitizationSame

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Summary of Non-Clinical Performance Tests conducted

Test MethodPurposeAcceptance CriteriaResults
ASTM D6978-2005(2019)Permeation byChemotherapy DrugsMinimum detection time >240minuteAll Chemotherapy Drugs & theOpioid Passed >240min. exceptCarmustine & ThioTepa
ASTM D412-2016ASTM D573-04(2019)Physical propertiesTensile strength(min=14Mpa); Elongation (min400%)Pass
ASTM D5151-19Water-leak testAQL 1.5 (ISO 2859-1)Pass
ASTM D6124-06(2017)Residual powder test<2mg/glovePass
ISO 10993-10:2010(2016)Dermal IrritationUnder conditions of this study,the subject device is non-irritatingPass
ISO 10993-10:2010(2016)Skin SensitizationUnder conditions of this study,the subject device is non-sensitizingPass

Non-clinical performance testing method full Titles:

  • ASTM D412 Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— Tension O
  • O ASTM D573 Test Method for Rubber—Deterioration in an Air Oven x ASTM D3578 Specificationfor Rubber Examination Gloves
  • ASTM D6319 Standard Specification for Nitrile Examination Gloves for Medical Application O
  • ASTM D5151 Test Method for Detection of Holes in Medical Gloves O
  • O ASTM D6124 Test Method for Residual Powder on Medical Gloves
  • O ISO 2859 Sampling Procedures and Tables for Inspection by Attributes Test results show that under the conditions of the testing, there is no difference in physical attributes between the proposed device and the predicate device.
  • O ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs

Biocompatibility Testing was performed utilizing: ISO 10993 Biological Evaluation of Medical Devices

  • ISO 10993 Part 10: Tests for Irritation and Sensitization. Both Skin Irritation and Dermal O Magnuson/Kligman Sensitization performed.
  • ISO 10993 Part 11: Tests for assessment of Systemic Toxicity O

No human clinical or animal performance testing was performed.

Conclusion: The conclusions drawn from comparing the physical attributes and the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as, or better than, the legally marketed predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.