K Number
K212532

Validate with FDA (Live)

Date Cleared
2021-11-11

(92 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Disposable Powder Free Nitrile Exam Glove (Tested For Use With Chemotherapy Drugs) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Summarized in Table below, the proposed device was tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugConcentration (mg/mL)Breakthrough time (minutes)
Carmustine (BCNU)3.3 mg/mL6.2min
Cisplatin1.0 mg/mL>240min
Cyclophosphamide (Cytoxan)20.0 mg/mL>240min
Dacarbazine (DTIC)10.0 mg/mL>240min
Doxorubicin HCL (Adriamycin)2.0 mg/mL>240min
Etoposide (Toposar)20.0 mg/mL>240min
Fluorouracil (Adrucil)50.0 mg/mL>240min
Paclitaxel (Taxol)6.0 mg/mL>240min
ThioTEPA10.0 mg/mL8.4min
Methotrexate25.0 mg/mL>240min
Vincristine sulfate1.0 mg/mL>240min

Do not use with Carmustine or ThioTEPA.

Device Description

The Disposable Powder Free Nitrile Exam Gloves (Tested For Use With Chemotherapy Drugs) are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue color, powder free, nitrile ambidextrous gloves. The gloves are offered in sizes small, medium, large, extra large packed in a paper box. The gloves are designed and manufactured in accordance with the ASTM D6319-19 standard and are tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019).

AI/ML Overview

The provided text describes the acceptance criteria and performance of "Disposable Powder Free Nitrile Exam Gloves (Tested For Use With Chemotherapy Drugs)". This is a medical device, specifically examination gloves, and the studies performed are non-clinical, primarily physical and chemical resistance tests. There is no mention of a study involving AI, human readers, or medical image analysis.

Here's a breakdown of the requested information based on the provided document:

1. Table of acceptance criteria and the reported device performance

Test Method (Purpose)Acceptance CriteriaReported Device Performance
Dimensions (width & thickness) (To evaluate the physical dimension of the glove)Width: 70mm min. Length: 220mm min. Palm: 0.05mm min. Finger: 0.05mm min.Pass 76mm min width 229mm min length Palm: 0.07mm min. Finger: 0.08mm min.
Physical properties (To evaluate tensile strength and ultimate elongation before and after aging)Tensile Strength: Before Aging ≥ 14 MPa, min. After Aging ≥ 14 MPa, min. Elongation: Before Aging 500%, min. After Aging 400%, min.Pass Tensile Strength: Before Aging 15.08 MPa, min. After Aging 14.87 MPa, min. Elongation: Before Aging 570%, min. After Aging 529%, min.
Freedom from holes (To detect holes in the gloves)No leakage at sampling level of G-1, AQL 2.5Pass No leakage, 80 of 80 passed of each size
Residual Powder (To detect powder residue in the glove)<2mg per glovePass Max. 0.52 mg per glove
Chemotherapy Drugs Tested with Minimum Breakthrough Detection Time (To test the resistance of medical gloves to permeation by chemotherapy drugs)For the following drugs, greater than 240 minutes breakthrough time: Cisplatin (1.0 mg/mL) Cyclophosphamide (Cytoxan) (20.0 mg/mL) Dacarbazine (DTIC) (10.0 mg/mL) Doxorubicin HCL (Adriamycin) (2.0 mg/mL) Etoposide (Toposar) (20.0 mg/mL) Fluorouracil (Adrucil) (50.0 mg/mL) Paclitaxel (Taxol) (6.0 mg/mL) Methotrexate (25.0 mg/mL) Vincristine sulfate (1.0 mg/mL) Note: The device is not for use with Carmustine or ThioTEPA due to shorter breakthrough times (6.2 min and 8.4 min respectively, as explicitly stated on page 3).Pass Cisplatin (1.0 mg/mL): >240min Cyclophosphamide (Cytoxan) (20.0 mg/mL): >240min Dacarbazine (DTIC) (10.0 mg/mL): >240min Doxorubicin HCL (Adriamycin) (2.0 mg/mL): >240min Etoposide (Toposar) (20.0 mg/mL): >240min Fluorouracil (Adrucil) (50.0 mg/mL): >240min Paclitaxel (Taxol) (6.0 mg/mL): >240min Methotrexate (25.0 mg/mL): >240min Vincristine sulfate (1.0 mg/mL): >240min The document explicitly states "Do not use with Carmustine or ThioTEPA." which aligns with the observed breakthrough times being below the >240min threshold for these two drugs (Carmustine: 6.2min, ThioTEPA: 8.4min).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Physical Tests (Freedom from holes): 80 of 80 gloves passed for each size. (Specific sample sizes for other physical dimensions and properties are not explicitly stated, but the tests were performed per ASTM D6319-19 which would specify sample requirements).
  • Sample Size for Chemotherapy Permeation Test: Not explicitly stated, but the test was conducted according to ASTM D6978-05 (Reapproved 2019), which would define the sample size requirements.
  • Data Provenance: The tests were conducted to support the 510(k) submission, implying prospective testing specific to this device. No country of origin for the data itself is mentioned, but the manufacturer is based in China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This information is not applicable to this type of device and study. The "ground truth" here is defined by objective physical and chemical testing standards (ASTM standards), not by expert consensus or interpretation of medical data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • This information is not applicable. Adjudication methods are typically used in studies involving human interpretation or clinical outcomes. The tests performed are objective, laboratory-based measurements against established standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This information is not applicable as this is not a study involving AI or human readers for diagnostic purposes. It's a non-clinical performance evaluation of gloves.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This information is not applicable. This is a physical device, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • The "ground truth" is established by objective, standardized laboratory test methods and their defined acceptance criteria as detailed in ASTM D6319-19 for physical properties and ASTM D6978-05 (Reapproved 2019) for chemotherapy drug permeation. There is no expert consensus, pathology, or outcomes data involved in these specific tests.

8. The sample size for the training set

  • This information is not applicable. There is no "training set" as this is not a machine learning or AI-based device.

9. How the ground truth for the training set was established

  • This information is not applicable. Since there is no training set mentioned, there's no ground truth for it.

In summary: The provided document details the non-clinical performance evaluation of disposable nitrile exam gloves against recognized industry standards (ASTM D6319-19 and ASTM D6978-05) to assess their physical properties and resistance to chemotherapy drugs. The evaluation focuses on objective measurements rather than clinical or AI-driven assessments.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 11, 2021

CRDLight Optoelectronic Technology Co., Ltd Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China

Re: K212532

Trade/Device Name: Disposable Powder Free Nitrile Exam Gloves (Tested For Use With Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: August 11, 2021 Received: August 11, 2021

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212532

Device Name

Disposable Powder Free Nitrile Exam Gloves (Tested For Use With Chemotherapy Drugs)

Indications for Use (Describe)

Disposable Powder Free Nitrile Exam Glove (Tested For Use With Chemotherapy Drugs) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Summarized in Table below, the proposed device was tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugConcentration (mg/mL)Breakthrough time (minutes)
Carmustine (BCNU)3.3 mg/mL6.2min
Cisplatin1.0 mg/mL>240min
Cyclophosphamide (Cytoxan)20.0 mg/mL>240min
Dacarbazine (DTIC)10.0 mg/mL>240min
Doxorubicin HCL (Adriamycin)2.0 mg/mL>240min
Etoposide (Toposar)20.0 mg/mL>240min
Fluorouracil (Adrucil)50.0 mg/mL>240min
Paclitaxel (Taxol)6.0 mg/mL>240min
ThioTEPA10.0 mg/mL8.4min
Methotrexate25.0 mg/mL>240min
Vincristine sulfate1.0 mg/mL>240min

Do not use with Carmustine or ThioTEPA.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

K212532

(As requirement by 21 CFR 807.92)

Date prepared: 2021-11-10

A. Applicant:

Name: CRDLIGHT OPTOELECTRONIC TECHNOLOGY CO., LTD Address: Floor 1-5 BULIDING NO.7 & FLOOR 1-4 BUILDING NO.5 NO.18 XINYI ROAD, JIANGHAI DISTRICT, JIANGMEN GUANGDONG, CHINA Contact: Fishy liang Title: General Manager Tel: +86-13924689685 Email: fishy@gdyanyang.com

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

B. Device:

Trade Name: Disposable Powder Free Nitrile Exam Gloves (Tested For Use With Chemotherapy Drugs) Common Name: Non-powdered patient examination glove

Regulatory Information

Classification Name: Medical Gloves with Chemotherapy Labeling Claims – Test For Use with Chemotherapy Drugs Classification: Class I Product code: LZA, LZC Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

C. Predicate device:

K200960

Medline Nitrile Powder-Free Dark Blue Examination Gloves (Tested for use with Chemotherapy Drugs) Medline Industries, Inc.

{4}------------------------------------------------

D. Indications for use of the device:

Disposable Powder Free Nitrile Exam Glove (Tested For Use With Chemotherapy Drugs) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Summarized in Table 1 below, the proposed device was tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugConcentration (mg/mL)Breakthrough time (minutes)
Carmustine (BCNU)3.3 mg/mL6.2min
Cisplatin1.0 mg/mL>240min
Cyclophosphamide (Cytoxan)20.0 mg/mL>240min
Dacarbazine (DTIC)10.0 mg/mL>240min
Doxorubicin HCL (Adriamycin)2.0 mg/mL>240min
Etoposide (Toposar)20.0 mg/mL>240min
Fluorouracil (Adrucil)50.0 mg/mL>240min
Paclitaxel (Taxol)6.0 mg/mL>240min
ThioTEPA10.0 mg/mL8.4min
Methotrexate25.0 mg/mL>240min
Vincristine sulfate1.0 mg/mL>240min

Do not use with Carmustine or ThioTEPA.

E. Device Description:

The Disposable Powder Free Nitrile Exam Gloves (Tested For Use With Chemotherapy Drugs) are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue color, powder free, nitrile ambidextrous gloves. The gloves are offered in sizes small, medium, large, extra large packed in a paper box. The gloves are designed and manufactured in accordance with the ASTM D6319-19 standard and are tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019).

F. Summary of Technological Characteristics

Table 2 Comparison of Proposed and Predicate Devices

DeviceProposed DevicePredicate DeviceResult
510K #K212532K200960-
ManufacturerCRDLIGHT OPTOELECTRONIC TECHNOLOGYCO., LTDMedline Industries, Inc.-

{5}------------------------------------------------

CRDLIGHT OPTOELECTRONIC TECHNOLOGY CO., LTD Floor 1-5 BULIDING NO.7 & FLOOR 1-4 BUILDING NO.5 NO.18 XINYI ROAD, JIANGHAI DISTRICT, JIANGMEN GUANGDONG, CHINA

Product NameDisposable Powder Free Nitrile Exam Gloves(Tested For Use With Chemotherapy Drugs)Medline Nitrile Powder-Free DarkBlue Examination Gloves (Tested foruse with Chemotherapy Drugs)Similar
Product CodeLZA, LZCLZA, LZCSame
RegulationNumber21 CFR 880.625021 CFR 880.6250Same
Indications foruseDisposable Powder Free Nitrile Exam Glove isa disposable device intended for medicalpurposes that is worn on the examiner's handto prevent contamination between patientand examiner. These gloves were tested foruse with chemotherapy drugs.A patient examination glove is adisposable device intended for medicalpurposes that is worn on theexaminer's hand to preventcontamination between patient andexaminer. These gloves were tested foruse with chemotherapy drugs.Same
DesignconfigurationsBlueDard blueSimilar
MaterialNitrileNitrileSame
SizeS, M, L, XLS, M, L, XL, XXLSimilar
Contact durationLimited ≤ 24 hoursLimited ≤ 24 hoursSame
OTC useYesYesSame
SterilityNon-sterileNon-sterileSame
UseSinge useSingle useSame
DimensionsComplies with: ASTM D6319-19Complies with: ASTM D6319-10Similar
Dimensions(thickness)Complies with: ASTM D6319-19Palm – 0.05mm min.Finger – 0.05mm min.Complies with: ASTM D6319-10Palm - 0.05mm min.Finger - 0.05mm min.Similar
PhysicalpropertiesComplies with: ASTM D6319-19Tensile Strength:Before Aging ≥14 MPa, min.After Aging ≥14 MPa, min.Elongation:Before Aging 500%, min.After Aging 400%, min.Complies with: ASTM D6319-10Tensile Strength:Before Aging ≥14 MPa, min.After Aging ≥14 MPa, min.Elongation:Before Aging 500%, min.After Aging 400%, min.Similar
Freedom fromholesComplies with: ASTM D6319-19 and ASTMD5151-19G-1, AQL 2.5Complies with: ASTM D6319-10 andASTM D5151-06G-1, AQL 1.5Different
Powder freeYesYesSame
Residual PowderComplies with: ASTM D6319-19<2mg per gloveComplies with: ASTM D6319-10<2mg per gloveSimilar
BiocompatibilityComplies with ISO 10993-10:Not a skin irritantNot a skin sensitizerComplies with AAMI/ANSI/ISO10993-10:Not a skin irritantSimilar

{6}------------------------------------------------

CRDLIGHT OPTOELECTRONIC TECHNOLOGY CO., LTD Floor 1-5 BULIDING NO.7 & FLOOR 1-4 BUILDING NO.5 NO.18 XINYI ROAD. JIANGHAI DISTRICT, JIANGMEN GUANGDONG, CHINA

ISO 10993-05: cytotoxicity potentialNot a skin sensitizerSimilar
ISO 10993-11:AAMI/ANSI/ISO 10993-05ISO 10993-11:Non-Toxic
Non-Toxic
ChemotherapyDrugs Testedwith MinimumBreakthroughDetection Timeas tested perASTM D6978Chemotherapy DrugConcentration(mg/mL)Breakthrough time(minutes)Chemotherapy DrugConcentrationBreakthrough time
Carmustine(BCNU)3.3 mg/mL6.2minCarmustine(BCNU)3.3 mg/ml(3,300 ppm)12.4 minutes
Cisplatin1.0 mg/mL>240minCisplatin1.0 mg/ml(1,000 ppm)>240 minutes
Cyclophosphamide(Cytoxan)20.0 mg/mL>240minCyclophosphamide(Cytoxan)20.0 mg/ml(20,000 ppm)>240 minutes
Dacarbazine (DTIC)10.0 mg/mL>240minDacarbazine(DTIC)10.0 mg/ml(10,000 ppm)>240 minutes
Doxorubicin HCL(Adriamycin)2.0 mg/mL>240minDoxorubicinHydrochloride2.0 mg/ml(2,000 ppm)>240 minutes
Etoposide(Toposar)20.0 mg/mL>240minEtoposide(Toposar)20.0 mg/ml(20,000 ppm)>240 minutes
Fluorouracil(Adrucil)50.0 mg/mL>240minFluorouracil50.0 mg/ml(50,000 ppm)>240 minutes
Paclitaxel (Taxol)6.0 mg/mL>240minMethotrexate25 mg/ml(25,000 ppm)>240 minutes
ThioTEPA10.0 mg/mL8.4minMitomycin C0.5 mg/ml (500 ppm)>240 minutes
Methotrexate25.0 mg/mL>240minPaclitaxel (Taxol)6.0 mg/ml(6,000 ppm)>240 minutes
Vincristine sulfate1.0 mg/mL>240minThio Tepa10.0 mg/ml(10,000 ppm)27.4 minutes
Vincristine Sulfate(Oncovin)1.0 mg/ml(1,000 ppm)>240 minutes
Do not use with Carmustine or ThioTEPA.Do Not Use with Carmustine or Thiotepa

G. Summary of Non-Clinical Testing

> Biocompatibility

Biocompatibility Testing according to ISO 10993-1:2018, the nature of body contact for the subject device is Surface Device category, Skin Contact and duration of contact is A-Limited (≤24h). The following tests for the subject device were conducted to evaluated the biocompatibility of Disposable Powder Free Nitrile Exam Gloves:

  • ISO 10993-10: Primary Skin Irritation
  • ISO 10993-10: Dermal Sensitization
  • · ISO 10993-05: Cytotoxicity
  • · ISO 10993-11: Systemic Toxicity

A Performance Testing

Physical performance testing of the proposed device were conducted as per ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs.

To summarize, the performance testing of the subject device were conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
  • ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves

{7}------------------------------------------------

CRDLIGHT OPTOELECTRONIC TECHNOLOGY CO., LTD Floor 1-5 BULIDING NO.7 & FLOOR 1-4 BUILDING NO.5 NO.18 XINYI ROAD, JIANGHAI DISTRICT, JIANGMEN GUANGDONG, CHINA

  • ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
    • ASTM D6978-05 (Reapproved 2019)Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs
Test MethodPurposeAcceptance CriteriaResults
Dimensions (width)(thickness)The purpose of thetest is to evaluatethe physicaldimension of thegloveWidth 70mm minLength 220mm minPalm – 0.05mm min.Finger–0.05mm min.Pass76mm min width229mm min lengthPassPalm – 0.07mm min.Finger–0.08mm min
Physical propertiesThe purpose of thetest is to evaluatethe tensile strengthand ultimateelongation beforeand after agingTensile Strength:Before Aging ≥ 14 MPa, min.After Aging ≥ 14 MPa, min.Elongation:Before Aging 500%, min.After Aging 400%, min.PassTensile Strength:Before Aging 15.08 MPa, min.After Aging 14.87 MPa, min.Elongation:Before Aging 570%, min.After Aging 529%, min.
Freedom from holesThe purpose of thetest is to detectholes in the glovesNo leakage atsampling level of G-1,AQL 2.5PassNo leakage, 80 of 80 passed of each size
Residual PowderThe purpose of thetest is to detect thepowder residue inthe glove<2mg per glovePassMax. 0.52 mg per glove
ChemotherapyDrugs Tested withMinimumBreakthroughDetection Time astested per ASTMD6978The purpose of thetest is to test theresistance ofmedical gloves topermeation bychemotherapy drugs>240minPass Chemotherapy Drug Concentration (mg/mL) Breakthrough time (minutes) Cisplatin 1.0 mg/mL >240min Cyclophosphamide (Cytoxan) 20.0 mg/mL >240min Dacarbazine (DTIC) 10.0 mg/mL >240min Doxorubicin HCL (Adriamycin) 2.0 mg/mL >240min Etoposide (Toposar) 20.0 mg/mL >240min Fluorouracil (Adrucil) 50.0 mg/mL >240min Paclitaxel (Taxol) 6.0 mg/mL >240min Methotrexate 25.0 mg/mL >240min Vincristine sulfate 1.0 mg/mL >240min

H. Clinical Test Conclusion

{8}------------------------------------------------

No clinical study is included in this submission.

l. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Disposable Powder Free Nitrile Exam Glove (Tested For Use With Chemotherapy Drugs) is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K200960.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.