K Number
K212525

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-08-19

(373 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Advcare Vial Direct to Bag (VDB) Needle-free Admixture devices are intended for drug reconstitution and transfer of fluids from drug vials into the IV bag containing infusion solution and attaching an IV set to administer influsion liquid to patient. The devices are provided as single-use, sterile, and non-pyrogenic products. The VDB-13 & 20 has an integrated vial adaptor that is used to reconstitute and admix drugs from a drug vial into the solution IV bag. The devices are intended for use in healthcare environments by clinically trained healthcare professional providers to aid and support prescribed treatment and therapy.

Device Description

Advcare IV fluid transfer Vial Direct to Bag (VDB) VDB-13 & 20 Needle-free Admixture Devices are designed for drug reconstitution and transfer of fluids from drug vials into IV bag containing infusion solution. The VDB-13 & 20 serve as a fluid pathway for standard IV set to administer fluid to patient while minimizing exposure to drug aerosols and spills. The VDB-13 & 20 has an integrated vial adaptor that is used to reconstitute and admix drugs from a drug vial into the solution IV bag. Both IV fluid transfer system designs consist of a body, piercing spike, spike cover, IV port, and twist-off for the IV administration set. The VDB 13mm and 20mm has an integrated Vial Adaptor for access to the drug/solution vial. Both system transfer drug into an IV bag and to an external IV administration set to administer fluid to patient. The products do not contain any medicinal substances, and can be used with standard drug vials. It is intended for use in healthcare facilities by care-giver to aid and support prescribed treatment and therapy.

AI/ML Overview

The provided text describes the Advcare Vial Direct to Bag (VDB) Needle-free Admixture Devices (K212525). Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with specific quantitative thresholds. Instead, it lists the performance tests conducted, implying that the device was deemed acceptable if it successfully passed these tests. The "reported device performance" is essentially that the device met these requirements.

Acceptance Criteria CategorySpecific Tests ConductedReported Device Performance
Functionality & Design- Product Functionality According to IFU- Cap Detachment Force- Spike Penetration Force- Air Leakage Test- Vial Adaptor Tensile Detachment Force From Drug Vial- Vial Adaptor Torsional Breakage Test- Spike to IV Bag Port Tensile Detachment Force- IV Admin Spike to VDB-IV Port Attachment Force- Vial Attachment Force- Vial Detachment Force- Flange Breakage- Flow Rate- Coring of IV Bag Port- Coring and Fragmentation of Vial Stopper- Twist-Off Bonding To IV Spike- Leakage After Removal of Twist OFF- Leakage at VDB-IV Spike to IV Bag port following assembly of components- Leakage at VDB-IV Port to IV Admin Spike following assembly of components- Dimensional Measurements of the IV spike lumens- Dimensional Measurements of the Vial Adaptor Retention Feature"Testing was conducted to validate that the Advcare IV fluid transfer Vial Direct to Bag (VDB) VDB-13 & 20 Needle-free Admixture Devices met the applicable design and performance requirements." (Implies successful completion of all listed tests.)
Material Safety- Chemical Compatibility Testing"Biocompatibility testing demonstrated the materials are biocompatible."
Biocompatibility- Cytotoxicity- Sensitization- Intracutaneous Reactivity- Acute System Toxicity- Subacute System Toxicity- Pyrogenicity- Hemolysis- Particulate Test- Bacterial Endotoxin"The following biocompatibility tests has been successfully conducted."
Sterilization- Validation in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 10-6"The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014... The sterilization method provides a sterility assurance level (SAL) of 10-6."
Dose Accuracy- Dose Concentration Study- Dose Accuracy (Transfer of vial contents to the IV bag & to the administration set)Implied successful completion as part of "Performance Testing."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample sizes used for each individual test. It mentions "Testing was conducted to validate..." globally. The data provenance (country of origin, retrospective/prospective) is not explicitly stated for the performance testing. However, the manufacturer is Advcare Medical from New Taipei, Taiwan, suggesting the testing was likely conducted by or sponsored by the manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is a medical device for drug administration, not an AI or diagnostic imaging device. Therefore, the concept of "experts used to establish groundwork" for a test set in the way a diagnostic algorithm requires (e.g., radiologists interpreting images) is not applicable here. The "ground truth" for this device would be established by engineering specifications, material standards, and regulatory requirements, rather than expert interpretation of a diagnostic outcome.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as this is not a study requiring human adjudication for diagnostic discrepancies. The tests conducted are objective, primarily engineering and performance-based.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI diagnostic device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" in this context refers to established engineering standards, material specifications, and regulatory requirements (e.g., ISO standards for sterilization, biocompatibility, and general medical device performance). For instance, for "Air Leakage Test," the ground truth would be a defined permissible leakage rate, not an expert consensus on a diagnosis.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI-driven system requiring a training set.

9. How the ground truth for the training set was established

Not applicable. This is a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

August 19, 2022

Advcare Medical Meng Tan CEO No. 36, Sinsing Street, Shulin District New Taipei, 23877 Taiwan

Re: K212525

Trade/Device Name: Advcare Vial Direct to Bag (VDB) Needle-free Admixture Devices Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: LHI Dated: August 5, 2021 Received: August 11, 2021

Dear Meng Tan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Wolloscheck, Ph.D. For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212525

Device Name

Advcare Vial Direct to Bag (VDB) Needle-Free Admixture Devices

Indications for Use (Describe)

Advcare Vial Direct to Bag (VDB) Needle-free Admixture devices are intended for drug reconstitution and transfer of fluids from drug vials into the IV bag containing infusion solution and attaching an IV set to administer influsion liquid to patient. The devices are provided as single-use, sterile, and non-pyrogenic products. The VDB-13 & 20 has an integrated vial adaptor that is used to reconstitute and admix drugs from a drug vial into the solution IV bag. The devices are intended for use in healthcare environments by clinically trained healthcare professional providers to aid and support prescribed treatment and therapy.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K212525 510(k) Summary

Submitter:Advcare MedicalNo. 36 Sinsing StreetShulin District, New Taipei City, 23877
Application Correspondent:Telephone Number:Fax Number:E-mail:Date Prepared:Meng T. Tan+886-2-3501-7479+886-2-8970-1448sales@advcaremed.comAugust 17, 2022
Trade Name:Advcare Direct to Bag (VDB) Needle-freeAdmixture Devices
Common or Usual Name:I.V. Fluid Transfer Set
Regulation Medical Specialty:General Hospital
Regulation Name:Intravascular Administration Set
Regulation Number:21 CFR 880.5440
Product Code:LHI
Device Class:2
Predicate Device:K201415 Vial2Bag-Advanced

Device Indications For Use:

Advcare Vial Direct to Bag (VDB) Needle-free Admixture devices are intended for drug reconstitution and transfer of fluids from drug vials into the IV bag containing infusion solution and attaching an IV set to administer infusion liquid to patient. The devices are provided as single-use, sterile, and non-pyrogenic products. The VDB-13 & 20 has an integrated vial adaptor that is used to reconstitute and admix drugs from a drug vial into the solution IV bag. The devices are intended for use in healthcare environments by clinically trained healthcare professional providers to aid and support prescribed treatment and therapy.

Device Description:

{4}------------------------------------------------

Advcare IV fluid transfer Vial Direct to Bag (VDB) VDB-13 & 20 Needle-free Admixture Devices are designed for drug reconstitution and transfer of fluids from drug vials into IV bag containing infusion solution. The VDB-13 & 20 serve as a fluid pathway for standard IV set to administer fluid to patient while minimizing exposure to drug aerosols and spills. The VDB-13 & 20 has an integrated vial adaptor that is used to reconstitute and admix drugs from a drug vial into the solution IV bag.

Both IV fluid transfer system designs consist of a body, piercing spike, spike cover, IV port, and twist-off for the IV administration set. The VDB 13mm and 20mm has an integrated Vial Adaptor for access to the drug/solution vial. Both system transfer drug into an IV bag and to an external IV administration set to administer fluid to patient. The products do not contain any medicinal substances, and can be used with standard drug vials. It is intended for use in healthcare facilities by care-giver to aid and support prescribed treatment and therapy.

Technological Characteristics and Substantial Equivalence:

A technological comparison table is provided below that compares the subject devices and predicate device.

Predicate K201415Advcare VDB 13 & 20Equivalence to Predicate
Intended useThe Vial2BagAdvanced intendeduse is to allow theadministration of afluid from an IVcontainer to apatient's vascularsystem through aneedle or catheterinserted into avein. TheVial2bag-Advanced has20mm vial adaptorto access infusionsystem.Advcare IV fluid transferVial Direct to Bag(VDB) 13 & 20 intendeduse is to allow theadministration of a fluidfrom an IV container to apatient's vascular systemthrough a needle orcatheter inserted into avein. The Advcare IVfluid transfer Vial Directto Bag (VDB) Deviceshas 13mm and 20mmvial adaptor to accessinfusion system.Comment 1
Target usersPharmacist orother healthcareprofessionalsPharmacist or otherhealthcare professionalsSame
UseEnvironmentHospital, ClinicsHospital, ClinicsSame

{5}------------------------------------------------

Page3of5
----------------
Interaction withother devicesFor all uses, devicewill connect to anIV solutioncontainer and aprimaryadministration set.Based on IFU.For all uses, device willconnect to an IV solutioncontainer and a primaryadministration set. Basedon IFUSame
Vial size20mm20mm and 13 mmComment 1
IV bag sizes50ml,100m,250mlAll standard IVbagsComment 2
SterilizationEtoEtoSame
Prescription useRx onlyRx onlySame
Piercing SpikePlastic-Single LumenPlastic-SingleLumenSame
MaterialsMedical GradePlasticsMedical GradePlasticsComment 3

Discussion of differences in technological characteristics:

Comment 1- Advcare IV fluid transfer Vial Direct to Bag (VDB) Devices has two designs, 13mm vial adaptor and 20mm vial adaptor designs. The predicate device only has a 20mm vial adaptor. The VDB-13 is to accommodate smaller drug vial sizes. Performance testing confirmed that the VDB- 13mm vial adaptor design does not raise different questions of safety and effectiveness. The materials, function, and principal of operation are very similar as the primary predicate 20mm vial adaptor.

Comment 2 - the predicate device is for use with 50ml, 100ml and 250ml IV solution bags. The Advcare IV fluid transfer Vial Direct to Bag (VDB) Devices are for use with all standard IV solution bags. Standard IV solution bags available on the market utilize a standard spike connector and this is consistent for all size standard solution bags. The Advcare IV fluid transfer Vial Direct to Bag (VDB) Devices utilize a standard bag spike that is compatible with all size standard IV solution bags and as such, will not raise any new safety or efficacy questions.

COMPARATIVE SAFETY AND PERFORMANCE: The Advcare fluid transfer Vial Direct to Bag (VDB) 13 & 20 uses the same technique and parameters as the predicate device for reconstituting and transferring medication. There are no differences in technology. Therefore, the technology used in the Adveare fluid transfer Vial Direct to Bag (VDB) 13 & 20 does not affect the comparative safety and performance of the device.

Comment 3- Biocompatibility testing demonstrated the materials are biocompatible.

PERFORMANCE DATA:

The following non-clinical data were provided in support of the substantial equivalence determination.

{6}------------------------------------------------

Performance Testing

Testing was conducted to validate that the Advcare IV fluid transfer Vial Direct to Bag (VDB) VDB-13 & 20 Needle-free Admixture Devices met the applicable design and performance requirements throughout its shelf life, conformity to applicable standards, and demonstrate substantial equivalent to the predicate device.

Conformity to Standard:

ISO 8536-4: 2019

Additional Performance Testing to internal specifications:

  • Product Functionality According to IFU
  • Cap Detachment Force
  • Spike Penetration Force ●
  • Air Leakage Test ●
  • Vial Adaptor Tensile Detachment Force From Drug Vial ●
  • Vial Adaptor Torsional Breakage Test ●
  • Spike to IV Bag Port Tensile Detachment Force ●
  • IV Admin Spike to VDB-IV Port Attachment Force ●
  • Vial Attachment Force
  • Vial Detachment Force
  • Flange Breakage
  • Flow Rate ●
  • Coring of IV Bag Port
  • Coring and Fragmentation of Vial Stopper
  • Twist-Off Bonding To IV Spike
  • Leakage After Removal of Twist OFF
  • Leakage at VDB-IV Spike to IV Bag port following assembly of components ●
  • Leakage at VDB-IV Port to IV Admin Spike following assembly of components ●
  • Dimensional Measurements of the IV spike lumens
  • Dimensional Measurements of the Vial Adaptor Retention Feature. ●
  • Chemical Compatibility Testing. ●
  • Dose Concentration Study
  • Dose Accuracy (Transfer of vial contents to the IV bag & to the administration set)

Biocompatibility

Accordance to ISO 10993-1, the Advcare IV fluid transfer Vial Direct to Bag (VDB) VDB-13 & 20 Needle-free Admixture Devices are classified as an externally communicating device with prolonged contact duration (>24 hours to 30 days) and blood path indirect contact. The finished device's patient contacting parts were tested in accordance with the tests recommended in the FDA Guidance: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk

{7}------------------------------------------------

management process." The following biocompatibility tests has been successfully conducted.

  • Cytotoxicity
  • Sensitization
  • Intracutaneous Reactivity ●
  • Acute System Toxicity ●
  • Subacute System Toxicity ●
  • Pyrogenicity
  • Hemolysis ●
  • Particulate Test ●
  • Bacterial Endotoxin ●

Sterilization

The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014. Sterilization of health-care products-Ethylene oxide-Requirements for the development, validation and routine control of a sterilization process for medical devices. The sterilization method provides a sterility assurance level (SAL) of 10-6.

CLINICAL DATA:

No clinical trial was performed

CONCLUSION:

In summary, the Advcare IV fluid transfer Vial Direct to Bag (VDB) VDB-13 & 20 Needlefree Admixture Devices subject of this Premarket Notification is considered to be substantially equivalent in its intended use, technological, material, and performance to the predicate device, Vial2bag: Advanced (K201415).

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.