K Number
K212506

Validate with FDA (Live)

Date Cleared
2021-11-07

(90 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large. The subject device is non-sterile.

AI/ML Overview

This document, K212506, is a 510(k) premarket notification for Medical Examination Gloves (Tested for Use with Chemotherapy). It details the acceptance criteria and the study proving the device meets these criteria.

Here's an analysis of the provided information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria and device performance are primarily based on the ASTM D6319 standard for medical examination gloves and ASTM D6978 for chemotherapy drug permeation.

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D6319Physical Dimensions TestLength (mm): S: ≥220; M/L/XL: ≥230Length: > 240/Pass (implies all sizes met or exceeded)
Width (mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10Width: S: 86-88 /Pass; M: 98-99/ Pass; L: 109-111/ Pass; XL: 115-117/ Pass
Thickness (mm): Finger: ≥0.05; Palm: ≥0.05Finger: 0.11-0.12/Pass; Palm: 0.09-0.10/Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125/Pass (0 defects out of 125 samples, meeting AQL 2.5)
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.15mg/Pass
ASTM D412Physical properties (Tensile Strength & Ultimate Elongation)Before Aging: Tensile Strength ≥14MPa; Ultimate Elongation ≥500%After Aging: Tensile Strength ≥14MPa; Ultimate Elongation ≥400%Before Aging: Tensile Strength 14.5-18.9/Pass; Ultimate Elongation 546-778/PassAfter Aging: Tensile Strength 14.4-16.1/Pass; Ultimate Elongation 579-699/Pass
ISO 10993-11Toxicity (Acute Systemic Toxicity)Non-acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivo / Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant/ Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer / Pass
ASTM D6978-05Chemotherapy Drug PermeationA set of specific minimum breakthrough detection times for various chemotherapy drugs (see table below). Note: Acceptance criteria for breakthrough times are implied by the provided values; these are the performance targets for the "tested for use with chemotherapy" claim.
Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes (Reported Performance)
-------------------------------------------------------------------------------------------------------------
Carmustine (BCNU)3.3 mg/ml (3,300 ppm)10.9 Minutes
Cisplatin1.0 mg/ml(1,000 ppm)> 240 Minutes
Cyclophosphamide (Cytoxan)20.0 mg/ml(20,000 ppm)> 240 Minutes
Dacarbazine (DTIC)10.0 mg/ml(10,000 ppm)> 240 Minutes
Doxorubicin HCL (Adriamycin)2.0 mg/ml(2,000 ppm)> 240 Minutes
Etoposide (Toposar)20.0 mg/ml(20,000 ppm)> 240 Minutes
Fluorouracil (Adrucil)50.0 mg/ml(50,000 ppm)> 240 Minutes
Paclitaxel (Taxol)6.0 mg/ml(6,000 ppm)> 240 Minutes
Thio TEPA10.0 mg/ml(10,000 ppm)15.2 Minutes

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • Watertightness Test (ASTM D5151): 125 samples (0/125/Pass).
    • Other physical performance tests (Dimensions, Physical properties, Powder Content): Specific sample sizes are not explicitly stated for all tests but are implied by the results provided (e.g., ranges for width/thickness indicate multiple measurements). These tests are typically performed on a statistically relevant sample size according to the respective ASTM standards.
    • Biocompatibility (ISO 10993 series): Sample sizes for the animal studies (toxicity, irritation, sensitization) are not detailed in this summary but would have followed the guidance of the ISO standards.
    • Chemotherapy Permeation (ASTM D6978): The permeation results are presented as single breakthrough times for each drug, implying a specific test protocol dictated by the standard, likely involving multiple replicates to ensure reproducibility, though the exact number isn't given.
  • Data Provenance: The data is derived from benchtop performance testing and biocompatibility testing. The manufacturer is based in Guangdong, China, which implies the testing was conducted there or by affiliated labs. The document does not specify if the data is retrospective or prospective; however, given it's a 510(k) submission, it's typically data collected specifically for the regulatory submission (i.e., prospective for this purpose).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This submission is for a Class I medical examination glove, which relies on standardized bench testing rather than interpretation of complex medical images or clinical outcomes that would require expert consensus (e.g., radiologists). Therefore, there were no human experts establishing "ground truth" in the diagnostic sense. The "ground truth" for the performance characteristics (e.g., tensile strength, breakthrough time) is defined by the validated test methods and their quantitative results.

4. Adjudication Method for the Test Set

Not applicable. As described above, this involves quantitative bench testing and not subjective expert interpretation requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices where human readers interpret medical images with and without AI assistance. This device is a physical glove, and its performance is evaluated through laboratory bench tests.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This question is not applicable to this device. This is a physical non-diagnostic medical device. There is no algorithm or AI component involved, hence no "standalone" algorithm performance to evaluate. The device's performance is inherently "standalone" in that it performs its function (e.g., barrier to chemicals, physical strength) intrinsically.

7. The Type of Ground Truth Used

The ground truth used for this device's performance testing is based on:

  • Standardized Test Methods (Benchtop Data): Adherence to specific ASTM and ISO standards (e.g., ASTM D6319, ASTM D5151, ASTM D6124, ASTM D412, ASTM D6978, ISO 10993-10, ISO 10993-11). The results obtained from these tests are considered the objective ground truth for the glove's physical and chemical resistance properties.
  • Chemical Permeation Data: Direct quantitative measurement of breakthrough times for specific chemotherapy drugs, as determined by ASTM D6978-05.
  • Biocompatibility Data: Results from in vivo (for toxicity, irritation, sensitization) studies following ISO 10993 guidelines, which define the biological response to the material.

8. The Sample Size for the Training Set

Not applicable. This device is a physical product, not an AI/ML algorithm. Therefore, there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 7, 2021

Guangdong Gymda Medical Technology Co.,ltd Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161 Lujiazui East Rd.,Pudong Shanghai. 200120 China

Re: K212506

Trade/Device Name: Medical Examination Gloves (Tested for Use with Chemotherapy) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: July 30, 2021 Received: August 9, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212506

Device Name

Medical Examination Gloves (Tested for Use with Chemotherapy)

Indications for Use (Describe)

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes
Carmustine (BCNU)3.3 mg/ml (3,300 ppm)10.9 Minutes
Cisplatin1.0 mg/ml(1,000 ppm)> 240 Minutes
Cyclophosphamide (Cytoxan)20.0 mg/ml(20,000 ppm)> 240 Minutes
Dacarbazine (DTIC)10.0 mg/ml(10,000 ppm)> 240 Minutes
Doxorubicin HCL (Adriamycin)2.0 mg/ml(2,000 ppm)> 240 Minutes
Etoposide (Toposar)20.0 mg/ml(20,000 ppm)> 240 Minutes
Fluorouracil (Adrucil)50.0 mg/ml(50,000 ppm)> 240 Minutes
Paclitaxel (Taxol)6.0 mg/ml(6,000 ppm)> 240 Minutes
Thio TEPA10.0 mg/ml(10,000 ppm)15.2 Minutes

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 10.9 Minutes Thio-Tepa 10.0 mg/ml 15.2 Minutes

Warning: Please do not use with Carmustine (BCNU) and Thiotepa.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary-K212506

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

GUANGDONG GYMDA MEDICAL TECHNOLOGY CO.,LTD Name: Address: No.13, Quan'an Third Road, Phase 2 of High-tech Zone, Nanxiong City, Shaoguan City, Guangdong Province, 512400, China Contact: Olivia Chen Date of Preparation: Jul 30, 2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 Lujiazui East Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Medical Examination Gloves (Tested for Use with Chemotherapy) Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove S, M, L, XL Model(s):

3.0 Classification

Production code: LZA.LZC Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Ever Growth (Vietnam) Co., Ltd. Manufacturer:

Device: Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Orange Color

510(k) number: K190860

{4}------------------------------------------------

5.0 Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large. The subject device is non-sterile.

6.0 Indication for Use

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs

Chemotherapy DrugConcentrationBreakthrough DetectionTime in Minutes
Carmustine (BCNU)3.3 mg/ml (3,300 ppm)10.9
Cisplatin1.0 mg/ml(1,000 ppm)> 240
Cyclophosphamide(Cytoxan)20.0 mg/ml(20,000 ppm)> 240
Dacarbazine (DTIC)10.0 mg/ml(10,000 ppm)> 240
Doxorubicin HCl2.0 mg/ml(2,000 ppm)> 240
Etoposide20.0 mg/ml(20,000 ppm)> 240
Fluorouracil50.0 mg/ml(50,000 ppm)> 240
Paclitaxel6.0 mg/ml(6,000 ppm)> 240
Thio Tepa10.0 mg/ml(10,000 ppm)15.2

Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 10.9 Minutes; Thio Tepa 10.0 mg/ml 15.2 Minutes.

Warning: Please do not use with Carmustine (BCNU) and Thiotepa.

7.0 Technological Characteristic Comparison Table

Table1-General Comparison

ItemSubject Device(K212506)Predicate Device(K190860)Remark
Product CodeLZA,LZCLZA,LZCSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseA patient examinationgloves is a disposabledevice intended for medicalpurpose that is worn on theexaminer's hand orfingers to preventcontamination betweenpatient and examiner. Inaddition, these gloves weretested for use withchemotherapy drugs inaccordance with ASTMD6978-05 StandardPractice for Assessment ofMedical gloves toPermeation byChemotherapy Drugs.A patient examinationgloves is a disposabledevice intended for medicalpurpose that is worn on theexaminer's hand orfingers to preventcontamination betweenpatient and examiner. Inaddition, these gloves weretested for use withchemotherapy drugs inaccordance with ASTMD6978-05 StandardPractice for Assessment ofMedical gloves toPermeation byChemotherapy Drugs.Same
Powdered or PoweredfreePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
SterilityNon-SterileNon-SterileSame
Labeling InformationSingle-use indication,powder free, device color,device name, glove size andquantity,Non-Sterile,a statement of standardASTM D6978-05compliance and asummary of the testingresults.Single-use indication,powder free, device color,device name, glove sizeand quantity, Non-Sterile, astatement of standardASTM D6978-05compliance and asummary of the testingresults.Same
Dimensions(mm)Length:S:≥220;M/L/XL: ≥230;Width:S: 80±10;M: 95±10;L: 110±10;XL: 120±10Length:XS/S/M/L/XL: ≥230;Width:XS:70±10;S: 80±10;M: 95±10;L: 110±10;XL: 120±10SimilarAnalysis1
Thickness(mm)Finger: ≥0.05;Palm: ≥0.05Finger: ≥0.05;Palm: ≥0.05Same

{5}------------------------------------------------

{6}------------------------------------------------

Analysis2
PhysicalPropertiesTensileStrength14MPa, minTensileStrength14MPa, minSame
BeforeAgingUltimateElongation500% minUltimateElongation500% minSame
AfterAgingTensileStrength14MPa, minTensileStrength14MPa, minSame
UltimateElongation400%minUltimateElongation400%minSame
Freedom from HolesBe free from holes when tested in accordance withASTMD5151 AQL=2.5Be free from holes when tested in accordance withASTMD5151 AQL=2.5Same
Powder Content0.15 mg per glove, Meet the requirements ofASTM D6124Meet the requirements ofASTM D6124SimilarAnalysis3
BiocompatibilityISO 10993-10;Under the conditions of the study, not an irritantor a sensitizerISO 10993-5Under conditions of thestudy, device extract iscytotoxicISO 10993-11;Under thecondition of acutesystemic toxicity test,the test article did notshow acute systemictoxicity in vivo.ISO 10993-10;Under the conditions of the study, not an irritantor a sensitizerISO 10993-5Under conditions of thestudy, device extract isnot cytotoxicN.A.Same//
Carmustine (BCNU) 3.3mg/ml: 10.9 MinutesCarmustine (BCNU) 3.3mg/ml:White:11.8 Minutes;Orange:31.6MinutesSimilarAnalysis 4
Cisplatin 1.0 mg/ml: > 240MinutesCisplatin 1.0 mg/ml: > 240MinutesSame

{7}------------------------------------------------

Chemotherapy DrugsTested with MinimumBreakthroughDetection Time asTested per ASTM D6978Cyclophosphamide(Cytoxan) 20.0 mg/ml: >240 MinutesCyclophosphamide(Cytoxan)20.0 mg/ml: > 240 MinutesSame
Dacarbazine (DTIC) 10.0mg/ml: > 240 MinutesDacarbazine (DTIC) 10.0mg/ml:>240 MinutesSame
Doxorubicin HCl 2.0 mg/ml:> 240 MinutesDoxorubicin Hydrochloride2.0mg/ml: >240 MinutesSame
Etoposide 20.0 mg/ml: >240 MinutesEtoposide (Toposar) 20.0mg/ml: >240 MinutesSame
Fluorouracil 50.0 mg/ml: >240 MinutesFluorouracil 50.0 mg/ml: >240 MinutesSame
Paclitaxel 6.0 mg/ml: >240MinutesPaclitaxel (Taxol) 6.0mg/ml:>240 MinutesSame
Thio Tepa 10.0 mg/ml: 15.2MinutesThio-Tepa 10.0 mg/ml:White:16.9 Minutes;Orange: 72.5 MinutesSimilarAnalysis 5

Analysis 1:

The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

Analysis 2:

The color of the subject device is different with that of the predicate. The subject device was evaluated according to ISO 10993-1 standards, and there were no risks identified.

Analysis 3:

Powder Content of subject device is similar with that of the predicate, because the predicate did not publish the exact results of the powder content. But they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

Analysis 4:

And Breakthrough detection times of Carmustine (BCNU) and Thio Tepa of subject device are different with those of the predicate. The Chemotherapy Labeling Claims has clearly defined on the labeling. So it does not raise any new safety or performance questions.

{8}------------------------------------------------

8.0 Summary of Non-Clinical Testing

Biocompatibility Testing

The biocompatibility evaluation for Medical Examination Gloves (Tested for Use with Chemotherapy) was conducted in accordance with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization.

ISO 10993-11 Third edition 2017-09, Biological evaluation of medical devices -Part 11: Tests for systemic toxicity

Performance Testing (Bench)

Physical performance qualities of the proposed device were evaluated per ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs.

In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • ASTM D6124-06 (Reapproved 2017), Standard Test Method for -Residual Powder on Medical Gloves
  • -ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
  • ASTM D 6978-05 (Reapproved 2019) ,Standard Practice for -Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Table 2 - Summary of non-clinical performance testing

{9}------------------------------------------------

TestMethodPurposeAcceptance CriteriaResults
ASTMD6319PhysicalDimensionsTestLength(mm):S:≥220;M/L/XL:≥230;Length:> 240/Pass;
Width(mm):S: 80±10;Width:S: 86-88 /Pass
M: 95±10;M: 98-99/ Pass
L: 110±10;L: 109-111/ Pass
XL: 120±10XL:115-117/ Pass
Thickness (mm):Finger: ≥0.05Palm: ≥0.05Finger: 0.11-0.12/Pass
Palm: 0.09-0.10/Pass
ASTMD5151WatertightnessTest forDetection ofHolesMeet the requirements of ASTM D5151AQL 2.50/125/Pass
ASTMD6124PowderContentMeet the requirements of ASTM D6124 <2.0mg0.15mg/Pass;
ASTMD412PhysicalpropertiesBeforeAgingTensileStrength≥14MPa14.5-18.9/Pass;
UltimateElongation≥500%546-778/Pass;
AfterAgingTensileStrength≥14MPa14.4-16.1/Pass;
UltimateElongation≥400%579-699/Pass;
ISO10993-11ToxicityNon- acute systemictoxicityUnder conditions ofthe study, did notshow acute systemictoxicity in vivo / Pass
ISO10993-10IrritationNon-irritatingUnder the conditionsof the study, not anirritant/ Pass
ISO10993-10SensitizationNon-sensitizingUnder conditions ofthe study, not asensitizer / Pass

{10}------------------------------------------------

9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Medical Examination Gloves (Tested for Use with Chemotherapy) is as safe, as effective, and performs as well as or better than the legally marketed predicated device under K190860.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.