K Number
K212497

Validate with FDA (Live)

Date Cleared
2021-11-11

(94 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nitrile Examination Glove (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Device Description

The proposed device, Nitrile Examination Glove (Powder free, Blue) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size. The color of the proposed device is Blue.

The proposed device is not provided as sterilized The proposed device is made of Nitrile.

AI/ML Overview

This document is a 510(k) Pre-market Notification for a medical device, specifically a Nitrile Examination Glove. The purpose of this notification is to demonstrate substantial equivalence to a legally marketed predicate device. The information provided outlines the acceptance criteria and the results of non-clinical tests to prove the device meets these criteria.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D5151Testing for Freedom from holesFreedom from holesNo water leakage is inspected from 125 samples
ASTM D6124Determine the powder residue for powder free gloves<2.0 mg per gloveResidual Powder: Average 0.1 Mg
ASTM D412, ASTM D573Testing for Physical property characteristicsTensile Strength: 14 MPa min.Ultimate Elongation: 400% min.Tensile Strength: ≥ 14 MPa;Ultimate Elongation: ≥ 400%.
ASTM D3767Testing For physical dimensions specificationLength: 230 mm min. for all size (S, M, L, XL);Width: 80±10 mm for S; 95±10 mm for M; 110±10 mm for L; 120±10 mm for XL.Finger Thickness: ≥0.05 mm;Palm Thickness: ≥0.05 mm;Cuff Thickness: ≥0.05 mm.All acceptance criteria above meet the requirements in Table 2 Dimensions and Tolerances of ASTM D6319Length of Size S: ≥ 254 mm; Width of Size S: 88±1 (87-89) mm; Palm Thickness of Size S: ≥0.07 mm; Finger Thickness of Size S: ≥0.12 mm.Length of Size M: ≥ 251 mm; Width of Size M: 96±2 (95-98) mm; Palm Thickness of Size M: ≥0.07 mm; Finger Thickness of Size M: ≥0.12 mm.Length of Size L: ≥ 255 mm; Width of Size L: 110±1 (109-111) mm; Palm Thickness of Size L: ≥0.07 mm; Finger Thickness of Size L: ≥0.12 mm.Length of Size XL: ≥ 244mm; Width of Size XL: 113±2 (111-115) mm; Palm Thickness of Size XL: ≥0.07 mm; Finger Thickness of Size XL: ≥0.12 mm.
ISO 10993-11Evaluate the endpoint of systemic toxicity for biocompatibilityThe test article showed "negative" systemic toxicityUnder the conditions of the study, the test article showed "negative" systemic toxicity.
ISO 10993-10 (Irritation)Evaluate the endpoint of irritant for biocompatibilityThe response of the test article has no skin irritationUnder the experimental conditions, the test article has no skin irritation on rabbits.
ISO 10993-10 (Sensitization)Evaluate the endpoint of sensitization for biocompatibilityThe test article showed no evidence of causing delayed dermal contact sensitizationThe test article showed no evidence of causing delayed dermal contact sensitization in the guinea pig

2. Sample size used for the test set and the data provenance

  • ASTM D5151 (Freedom from holes): 125 samples.
  • Other tests (ASTM D6124, ASTM D412, ASTM D573, ASTM D3767, ISO 10993-11, ISO 10993-10): The specific sample sizes for these tests are not explicitly stated in the provided text beyond that they were sufficient to produce the reported results.
  • Data Provenance: The origin of the test data (e.g., country of origin, retrospective/prospective) is not explicitly stated. The tests were performed as "bench tests" to verify design specifications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable. The device is a medical glove, and the evaluation is based on objective physical, chemical, and biological performance tests against established standards, not on expert interpretations of data like in image analysis or diagnostic studies.

4. Adjudication method for the test set

This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving expert interpretation. The tests described are objective measurements against predefined acceptance criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This is a submission for a medical glove, not an AI-powered diagnostic device or a device requiring human reader interpretation. No MRMC study was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This is a medical glove. Standalone performance refers to the performance of an algorithm without human intervention, which is not relevant for this device.

7. The type of ground truth used

The "ground truth" for the performance evaluation of this medical glove is established by objective performance specifications defined in recognized international standards:

  • ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves for Medical Application)
  • ASTM D5151-19 (Standard Test Method for Detection of Holes in Medical Gloves)
  • ASTM D6124-17 (Standard Test Method for Residual Powder on Medical Gloves)
  • ISO 10993-10:2010 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization)
  • ISO 10993-11:2017 (Biological evaluation of medical devices - Part 11: Tests for systemic toxicity)

The device's performance is compared against the quantitative and qualitative requirements set forth in these standards.

8. The sample size for the training set

This information is not applicable. A "training set" refers to data used to train machine learning models. This device is a physical product (a medical glove) and does not involve machine learning or AI.

9. How the ground truth for the training set was established

This information is not applicable for the same reasons as #8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

November 11, 2021

Jiangsu Jinlian Medical Technology Co., Ltd % Ray Wang General Manager Beijin Believe-Med Technology Service Co., Ltd Rm.912, Building#15, XiYueHui, No.5, YiHe North Rd. FangShan District Beigin, Beijing 102401 China

Re: K212497

Trade/Device Name: Nitrile Examination Glove (Powder free, Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: August 5, 2021 Received: August 9, 2021

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K212497

Device Name

Nitrile Examination Glove (Powder free, Blue)

Indications for Use (Describe)

The Nitrile Examination Glove (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K212497

This 510(k) Summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

The assigned 510(k) Number: K212497

    1. Date of Preparation: 08/05/2021
    1. Sponsor

Jiangsu Jinlian Medical Technology Co., Ltd

No.14 Factory Building, 1st Floor, West of Century Avenue, Suqian Economic and Technological Development Zone, Suqian City, Jiangsu Province, China,223800 Contact Person: Wu Min Position: General Manager Tel: +86-18641219728 Fax: +86-527-82860533 Email: 247001296@qq.com

  • Submission Correspondent 3.

Beijing Believe-Med Technology Service Co., Ltd.

Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China,102401 Contact Person: Ray Wang Position: General Manager Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang(@believe-med.com

    1. Proposed Device Identification
      Trade Name: Nitrile Examination Glove (Powder free, Blue) Common Name: NITRILE Patient Examination Gloves (Powder Free) Regulatory Information: Classification: I Product Code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

{4}------------------------------------------------

Indication For Use Statement:

The Nitrile Examination Glove (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

  • Predicate Device Identification 5.
    510(k) Number: K150340 Product Name: POWDER FREE Nitrile GLOVES (White, Cobalt Blue, Black, Ice Blue) Manufacturer: HEBEI HONGSEN PLASTICS TECHNOLOGY CO., LTD

  • Device Description 6.
    The proposed device, Nitrile Examination Glove (Powder free, Blue) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size. The color of the proposed device is Blue.

The proposed device is not provided as sterilized The proposed device is made of Nitrile.

SizeModelCuffThickness(mm)PalmThickness(mm)FingerThickness(mm)Width(mm)Length(mm)Color
S$\geq 0.05$$\geq 0.05$$\geq 0.05$80±10
M$\geq 0.05$$\geq 0.05$$\geq 0.05$95±10$\geq 230$Blue
L$\geq 0.05$$\geq 0.05$$\geq 0.05$110±10
XL$\geq 0.05$$\geq 0.05$$\geq 0.05$120±10

Table 1 Device Size Specifications

Table 2 Performance and Physical Specifications

Before AgingAfter AgingPinhole AQL
Tensile StrengthUltimate ElongationTensile StrengthUltimate Elongation1.5
14 MPa, min500 % min14 MPa, min400 % min

The above data of size, performance, and physical specifications of proposed gloves meet all the current specifications listed in the ASTM standard D6319.

{5}------------------------------------------------

7. Technological Characteristics Comparison Table

ITEMProposed DevicePredicate Device (K150340)Comparison
ITEMNitrile Examination Glove (Powder free,Blue)POWDER FREE Nitrile GLOVES (White,Cobalt Blue, Black, Ice Blue)SAME
Product CodeLZALZASAME
Regulation No.21 CFR 880.625021 CFR 880.6250SAME
ClassIISAME
Intended UseThe Nitrile Examination Glove (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.The POWDER FREE Nitrile GLOVES (White, Cobalt Blue, Black, Ice Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.SAME
Powdered orPowered freePowdered freePowdered freeSAME

Table 1 General Comparison

{6}------------------------------------------------

Proposed DeviceDesignationSizeTolerance
Nitrile Examination Glove(Powder free, Purple-Blue,Blue)Length, mm230230230230min
Width, mm8095110120±10
Thickness, mm:
Finger0.05min
Palm0.05min
Cuff0.05min
Predicate Device (K150340)DesignationXSSMLXLTolerance
POWDER FREE NitrileGLOVES (White, Cobalt Blue,Black, Ice Blue)Length, mm230230230230230min
Width, mm708095110120±10
Thickness, mm:
Finger0.10-0.12±0.03
Palm0.08-0.10±0.03
Cuff0.06-0.09±0.03
RemarkAnalysis 1

Table 2 Device Dimensions Comparison

Analysis 1:

The proposed device has different size specification to the predicate device, but all proposed devices are meet the specifications of ASTM D 6319.

So we consider this as the proposed device is same with the predicate device.

Table 3 Performance Comparison
ITEMProposed DeviceNitrile Examination Glove(Powder free, Blue)Predicate Device (K150340)POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, Ice Blue)Comparison
ColorantBlueWhite, Cobalt Blue, Black, Ice BlueAnalysis 2
PhysicalPropertiesBeforeAgingTensileStrength14 MPa, min15 MPa, minAnalysis 3
UltimateElongation500 % min500 % minSAME
AfterAgingTensileStrength14 MPa, min14 MPa, minSAME
Ultimate400 % min400 % minSAME

{7}------------------------------------------------

Elongation
Comply with ASTM D6319Comply with ASTM D6319SAME
Freedom from HolesBe free from holes when tested inaccordance with ASTM D5151Be free from holes when tested inaccordance with ASTM D5151SAME
Powder ContentLess than 2 mg per glove when testedin accordance with ASTM D6124Meet the requirements of ASTM6319SAME

Analysis 2:

The proposed device has different color to the predicate device

Analysis 3:

The proposed device has different Tensile Strength before aging specification to the predicate device

ITEMProposed DeviceNitrile Examination Glove(Powder free, Blue)Predicate Device (K150340)POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, IceBlue)Comparison
MaterialNitrileNitrileSAME
BiocompatibilityIrritationUnder the conditions of the study,not an irritantUnder the conditions of the study,not an irritantSAME
SensitizationUnder conditions of the study, not asensitizer.Under conditions of the study, not asensitizer.SAME
SystemictoxicityUnder conditions of the study, not asystemic toxicity.Under conditions of the study, not asystemic toxicity.SAME
Label and LabelingMeet FDA's RequirementsMeet FDA's RequirementsSAME

Table 4 Safety Comparison

Summary of Non-Clinical Tests 8.

Bench tests were conducted to verify that the proposed device met all design specifications found in the standards and test methods described below. The test results demonstrated that the proposed device complies with the following standards:

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6124-17, Standard Test Method for Residual Powder on Medical Gloves.

ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

{8}------------------------------------------------

Table 5 Performance Test Results Summary

Test MethodPurposeAcceptance CriteriaResults
ASTM D5151Testing for Freedomfrom holesFreedom from holesNo water leakage is inspected form 125samples
ASTM D6124Determine the powderresidue for powder freegloves<2.0 mg per gloveResidual Powder: Average 0. 1 Mg;
ASTM D412ASTM D573Testing for Physicalproperty characteristicsTensile Strength: 14 MPa min.Ultimate Elongation: 400% min.Tensile Strength: ≥ 14 MPa;Ultimate Elongation: ≥ 400%.
ASTM D3767Testing For physicaldimensions specificationLength: 230 mm min. for all size (S, M,L, XL);80±10 mm for S; 95±10 mm for M;110±10 mm for L; 120±10 mm for XL.Finger Thickness: ≥0.05 mm;Palm Thickness: ≥0.05 mm;All acceptance criteria above meet therequirements in Table 2 Dimensions andTolerances of ASTM D6319Length of Size S: ≥ 254 mm;Width of Size S: 88±1 (87-89) mm;Palm Thickness of Size XL: ≥0.07 mm;Finger Thickness of Size XL: ≥0.12 mm.Length of Size M: ≥ 251 mm;Width of Size M: 96±2 (95-98) mm;Palm Thickness of Size XL: ≥0.07 mm;Finger Thickness of Size XL: ≥0.12 mm..Length of Size L: ≥ 255 mm;Width of Size L: 110±1 (109-111) mm;Palm Thickness of Size XL: ≥0.07 mm;Finger Thickness of Size XL: ≥0.12 mm.Length of Size XL: ≥ 244mm;Width of Size XL: 113±2 (111-115) mm;Palm Thickness of Size XL: ≥0.07 mm;Finger Thickness of Size XL: ≥0.12 mm.
ISO 10993-11Evaluate the endpoint ofsystemic toxicity forbiocompatibilityThe test article showed "negative"systemic toxicityUnder the conditions of the study, the testarticle showed "negative" systemic toxicity.
ISO 10993-10Evaluate the endpoint ofirritant forbiocompatibilityThe response of the test article has noskin irritationUnder the experimental conditions, the testarticle has no skin irritation on rabbits.
Evaluate the endpoint ofsensitization forbiocompatibilityThe test article showed no evidence ofcausing delayed dermal contactsensitizationThe test article showed no evidence of causingdelayed dermal contact sensitization in theguinea pig

{9}------------------------------------------------

9. Summary of Clinical Test

No clinical study is included in this submission.

10. Conclusion

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, POWDER FREE Nitrile GLOVES (White, Cobalt Blue, Black, Ice Blue) cleared under K150340.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.