K Number
K212470

Validate with FDA (Live)

Device Name
iRAD
Date Cleared
2021-10-20

(75 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

iRAD Image Enhancement System is an image processing software than can be used for image enhancement of MRI, CT, and X-Ray images. Enhanced images will be sent to PACS server and exist in conjunction to the original images.

Device Description

iRAD v1.0 Image Enhancement System is a medical image enhancement software, i.e., a Software as a Medical Device (SaMD), that can be used to enhance images of MRI, CT and X-Ray. iRAD takes as input DICOM [Digital Imaging and Communications in Medicine] files of MRI, CT, and X-Ray images, and produces an enhanced output of the same file, in DICOM format that can be sent to a PACS server. The objective is to enhance the DICOM files that are obscured and not clearly visible, to be more visible, sharper, and clearer through the iRAD image enhancement process. The iRAD image enhancement is done by the implementation of an image enhancement algorithm.

iRAD is intended to be used by medical doctors, radiologists and clinicians in hospitals, radiology centers and clinics, as well as by medical universities and research intuitions. The system allows selection of input DICOM images from PACS servers. DICOM images are sent to the iRAD image enhancement server, where they are processed and sent back to the PACS server after enhancement. The enhanced and original images exist in conjunction and can be compared. The system provides the user with a set of adjustable parameters through which to control the degree of contrast and strength enhancement and noise suppression.

iRAD implements a modified contrast limited adaptive histogram equalization algorithm to improve the visibility of the image and it uses the iRAD guided filter to reduce noise. The original image is deconstructed into overlapping rectangular components. The equalization and matching algorithm is executed in overlapping rectangular regions, resulting in several level-of-detail layers. The enhanced and denoised image is reconstructed using the level-of-detail layers based on user-controlled parameters of noise suppression and detail enhancement.

AI/ML Overview

Here's the breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

Key Takeaway Regarding the Study: The provided document describes performance testing for an image enhancement system (iRAD). This system aims to improve the visibility, sharpness, and clarity of medical images (MRI, CT, X-Ray) by increasing contrast and reducing noise. It's crucial to understand that this is not a diagnostic AI system that classifies or detects pathologies. Instead, it's a tool to improve image quality for human interpretation. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are focused on the technical performance of the image enhancement rather than diagnostic accuracy comparisons.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria Category/MetricReported Device Performance (iRAD)
Image Enhancement Effectiveness
Contrast Ratio ImprovementIncreased in all test cases.
Entropy ImprovementIncreased in all test cases.
Noise Reduction
Signal-to-Noise Ratio (SNR) ImprovementIncreased significantly in all test cases (mathematical phantoms); improved by at least 50% without degradation in MRI, CT, and X-Ray scans.

Note: The document describes these as "test cases" that "passed successfully" or "passed." These are implicit acceptance criteria based on the device's intended function to enhance images. Specific quantitative thresholds for "increase" or "improvement" are not explicitly stated as numerical acceptance limits in this summary, but the successful passing of these tests implies that the observed improvements met the internal criteria.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Mathematical Derenzo Phantom: Used for controlled testing of contrast ratio, entropy, and SNR improvement. (Quantitative count of phantoms not specified, but referred to as "the mathematical Derenzo Phantom" and "two mathematical phantoms").
  • X-Ray Scans: A collection of 82 lower resolution and 21 high resolution X-Ray scans (total 103) processed.
  • MRI Scans: 100 MRI scans processed.
  • CT Scans: CT scans were used for SNR improvement testing (quantitative count not specified, but implied to be part of the "all images tested").
  • Camera Scans: 38 camera scans processed (though the device is for medical images, these were used in one phase of testing).

Data Provenance: The document does not specify the country of origin for the X-Ray, MRI, CT, or camera scans. It also does not explicitly state whether the data was retrospective or prospective. Given the nature of performance testing for an image enhancement algorithm, it's highly likely to be retrospective data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The concept of "ground truth" for a diagnostic task (e.g., presence/absence of a disease) is not directly applicable here, as iRAD is an image enhancement system, not a diagnostic AI. The "ground truth" in this context refers to the ideal enhanced image or the known properties of the phantom.

  • For Phantoms: Mathematical phantoms inherently have a known "ground truth" for their properties (e.g., known contrast, known signal, known noise levels). No human experts are needed to establish this.
  • For Clinical Images: The "ground truth" for enhancement is assessed by the measured improvement in objective metrics like contrast ratio, entropy, and SNR, rather than expert labels of image content. The document does not mention human readers or experts for grading image quality or establishing a "ground truth" for the test sets used to measure these quantitative improvements.

4. Adjudication Method for the Test Set

Not applicable, as the evaluation relies on quantitative, objective metrics for image enhancement (contrast ratio, entropy, SNR) calculated algorithmically, rather than subjective human assessment where adjudication might be needed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not described in the provided text. This type of study is typically performed for diagnostic AI systems to show how AI assistance impacts human reader performance (e.g., changes in sensitivity, specificity, reading time). Since iRAD is an image enhancement tool, its primary evaluation is on the objective improvement of image characteristics, not directly on diagnostic accuracy.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, the testing described is primarily standalone algorithm performance. The device's ability to improve contrast ratio, entropy, and SNR values was measured objectively, without direct human interaction or diagnostic performance evaluation.

7. The Type of Ground Truth Used

  • Mathematical/Computational Ground Truth: For the phantom studies, the ground truth of the image properties (contrast, signal, noise) is known conceptually from the phantom design specification.
  • Objective Metric-Based Ground Truth: For the clinical image sets, the "ground truth" for success is measured by the objective improvement in quantitative metrics (contrast ratio, entropy, SNR) calculated before and after processing by the iRAD algorithm. There is no "pathology" or "outcomes data" ground truth involved, as the device does not perform diagnosis.

8. The Sample Size for the Training Set

The document does not specify a sample size for a training set. Given that iRAD implements a "modified contrast limited adaptive histogram equalization algorithm" and "iRAD guided filter" (Section 5.6), which are described as algorithmic processes rather than deep learning models requiring large-scale data training, it's possible that a traditional "training set" in the machine learning sense was not used, or if it was, its size is not disclosed. The description points to a rules-based or filter-based image processing approach.

9. How the Ground Truth for the Training Set was Established

As no training set (in the machine learning sense) is explicitly mentioned, the method for establishing its ground truth is also not described. The algorithms appear to be designed based on image processing principles rather than being learned from labeled data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

Claritas HealthTech Pte Ltd % Ms. Devika Dutt COO 20A Tanjong Pagar Road Singapore 088443 SINGAPORE

October 20, 2021

Re: K212470

Trade/Device Name: iRAD Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: September 7, 2021 Received: September 13, 2021

Dear Ms. Dutt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212470

Device Name

iRAD

Indications for Use (Describe)

iRAD Image Enhancement System is an image processing software than can be used for image enhancement of MRI, CT, and X-Ray images. Enhanced images will be sent to PACS server and exist in conjunction to the original images.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human ServicesFood and Drug AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) StaffPRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 20 of 332

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

5. 510(k) Summary - K212470

The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:

5.1 Submitter Information

Company:Devika Dutt
COO
Claritas HealthTech Pte Ltd
20A Tanjong Pagar Road
Singapore, Singapore 088443 Singapore
Telephone: 6597951921
Fax: N/A
d.d@claritasco.com
Contact:Devika Dutt
COO
Claritas HealthTech Pte Ltd
20A Tanjong Pagar Road
Singapore, Singapore 088443 Singapore
Telephone: 6597951921
Fax: N/A
d.d@claritasco.com
Date Summary Prepared:August 3, 2021
5.2 Name of the Device
Trade Name:iRAD
Common Name:Image Enhancement System
Classification Name:Processing System, Radiological (21 CFR, 892.205LLZ)
Review Panel:Radiology (RA)
Regulation:892.2050
Class:Class II
Product Code:LLZ

5.3 Equivalence Claimed to Predicate Device

The iRAD is equivalent to the ZOOM (K172768), manufactured by Zetta Medical Technologies, LL.

5.4 Predicate Device Information

{4}------------------------------------------------

Trade name: ZOOM

Manufacturer: Zetta Medical Technologies, LLC .. 1313 Ensell Road, Lake Zurich, IL 60047

Regulation Number: 21 CFR 892.2050

Regulation Name: System, Image Processing, Radiological

Device Class: Class II

Product Code: LLZ

510(k) Number: K172768

510(k) Clearance Date: April 24, 2018

5.5 Indications for Use

iRAD Image Enhancement System is an image processing software than can be used for image enhancement of MRI, CT, and X-Ray images. Enhanced images will be sent to PACS server and exist in conjunction to the original images.

5.6 Device Description

iRAD v1.0 Image Enhancement System is a medical image enhancement software, i.e., a Software as a Medical Device (SaMD), that can be used to enhance images of MRI, CT and X-Ray. iRAD takes as input DICOM [Digital Imaging and Communications in Medicine] files of MRI, CT, and X-Ray images, and produces an enhanced output of the same file, in DICOM format that can be sent to a PACS server. The objective is to enhance the DICOM files that are obscured and not clearly visible, to be more visible, sharper, and clearer through the iRAD image enhancement process. The iRAD image enhancement is done by the implementation of an image enhancement algorithm.

iRAD is intended to be used by medical doctors, radiologists and clinicians in hospitals, radiology centers and clinics, as well as by medical universities and research intuitions. The system allows selection of input DICOM images from PACS servers. DICOM images are sent to the iRAD image enhancement server, where they are processed and sent back to the PACS server after enhancement. The enhanced and original images exist in conjunction and can be compared. The system provides the user with a set of adjustable parameters through which to control the degree of contrast and strength enhancement and noise suppression.

iRAD implements a modified contrast limited adaptive histogram equalization algorithm to improve the visibility of the image and it uses the iRAD guided filter to reduce noise. The original image is deconstructed into overlapping rectangular components. The equalization and matching algorithm is executed in overlapping rectangular regions, resulting in several level-of-detail layers. The enhanced and denoised image is reconstructed using the level-of-detail layers based on user-controlled parameters of noise suppression and detail enhancement.

5.7 Substantial Equivalence Comparison of Technological Characteristics

The subject device iRAD is substantially equivalent to the predicate device ZOOM. Both predicate and subject device are image enhancement systems that process images for enhancement. Both devices have similar

{5}------------------------------------------------

technological characteristics as both implement image enhancement algorithms as their core technology. The difference is that the predicate device, ZOOM applies its image enhancement system only to MRI images, whereas the subject device, iRAD applies its image enhancement system to MRI images and to CT and X-Rays images. The method and process of enhancement that iRAD uses for MRI images is the same as the method and process of enhancement that iRAD uses for CT and X-Ray images. Verification, validation and testing demonstrate that the difference in number of modalities do not raise new questions of safety or effectiveness for the subject device. The table below shows the similarities and differences between the technological characteristics of the two devices.

CharacteristicsPredicate DeviceZOOM [K172768]Subject DeviceiRAD
Device ClassClass IISame
Product CodeLLZSame
Intended UseImage enhancement system which is animage processing software for imageenhancement.Same
Indications for UseZOOM Image Enhancement System isan image processing software that canbe used for image enhancement in MRIimages.Enhanced images will be sent to PACSserver and exist in conjunction to theoriginal images.iRAD Image Enhancement System is animage processing software that can beused for image enhancement in MRI,CT, and X-Ray images.Enhanced images will be sent to PACSserver and exist in conjunction to theoriginal images.
ModalitiesMRIMRI, CT, X-Ray
Physical CharacteristicsSoftware that operates on off-the-shelfhardwareSame
ComputerPC or PC CompatibleSame
DICOM standardComplianceThe software processes DICOMcompliant image dataSame
Operating SystemWindowsWindows / Linux
Image ProcessingHardwareIntel i3 processor, 4GB RAM, 500GBHard driveIntel i7 processor, Nvidia GPU
Image InputDICOMSame
CharacteristicsPredicate DeviceZOOM [K172768]Subject DeviceiRAD
Image OutputDICOMSame
User InterfaceThe software works on radiologyworkstation.The software works on radiologyworkstation, both offline and online.
Core TechnologyImage Enhancement AlgorithmSame
Software CoreZOOM Image Enhancement Algorithm(Zetta's own trademark)iRAD Image Enhancement Algorithm(Claritas HealthTech's own trademark)
Comparisonwithoriginal ImageThe enhanced image can be comparedto the original DICOM image as bothexist in conjunction.Same
WorkflowThe software, which is installed on aremote computer, receives DICOMimages, processes the received imagesand sends the enhanced images back toPACS server.Same

5.8 Technological Characteristics Comparison Table

{6}------------------------------------------------

Summary of Technological Characteristics Comparison Table

As per the table above the two devices are technologically similar. Verification, validation, and performance testing demonstrates the differences in modalities do not raise new questions of safety and effectiveness.

5.9 Performance Testing

iRAD has been designed, verified and validated according to the software development plan which is in compliance with IEC 62304:2006 requirements. Safety and performance has been evaluated and verified in accordance with software specification to ensure performance meets specified requirements and the requirement of the FDA guidance document, titled, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

iRAD is an image processing software which has the objective of emphasizing small and dim features that otherwise could have been missed. To make dim features visible, the contrast is improved, which can be quantified by calculating the RMS contrast ratio. The information visible in the enhanced image is increased, which can be quantified with the entropy, and the noise is reduced which is measured with the signal to noise ratio improvement.

The verification, validation and the performance evaluation of iRAD had three phases:

I. Using the mathematical Derenzo Phantom, the contrast ratio and entropy improvement were examined, in a scenario where the ground truth is available, and the noise is varying in a controlled way. The iRAD software

{7}------------------------------------------------

was also tested in a situation where the noise is greater than the just visible difference in the true signal. The contrast ratio and the entropy have been increased by iRAD in each of the test cases have passed successfully.

  1. A collection of 82 lower resolution and 21 high resolution X-Ray, 100 MRI, and 38 camera scans, haven been processed by iRAD to test the improvement in the contrast ratios and the entropy values as compared to the original images. As per the processed test results, the measured contrast ratio and the entropy values have been increased by iRAD in each of the test cases. This test passed.

  2. The signal-to-noise ratio (SNR) improvement of iRAD has been analyzed and tested in the following methods:

a. Using two mathematical phantoms. As per test results the SNR is increased significantly by iRAD in all test cases. This test passed.

b. SNR improvement was also tested in MRI, CT and X-Ray scans. Based on test results it was concluded that the iRAD algorithm improves the SNR in all images tested by at least 50% without any degradation. This test passed.

5.10 Safety and Effectiveness

Based on the iRAD software performance test results and incorporated risk minimisation methods in design, Claritas HealthTech Pte. Ltd. concludes that this device is substantially equivalent to the predicate device.

5.11 Conclusion

iRAD is an image enhancement software which has similar intended use and indications for use as the predicate device. The difference is that the predicate device is a Single Modality image enhancement system where as iRAD is a Multi Modality image enhancement system. The two devices have similar technological characteristics: both predicate device and subject device use image enhancement algorithms as their core technology; both algorithms use image based guided filtering and reconstruction, and both methods have optimized parameters to ensure robustness of the algorithm. Performance test results and incorporated risk minimization methods demonstrate that iRAD is as safe and effective as the predicate device. This 510(k) submission includes information on the iRAD technological characteristics, as well as performance data and verification and validation activities demonstrating that iRAD is as safe and effective as the predicate, and does not raise different questions of safety and effectiveness.

N/A