(223 days)
ZOOM Image Enhancement System is an image processing software that can be used for image enhancement in MRI images. Enhanced images will be sent to PACS server and exist in conjunction to the original images.
ZOOM Image Enhancement System is an image processing software that can be used for image enhancement in MRI images. ZOOM image enhancement software implements a noise reduction algorithm using wavelets and image guided filtering. Original images are decomposed into different wavelet sub bands and noise in each band is soft threshold. De-noised images are reconstructed from softthresholded images using inverse wavelet transform. The software, which is installed on a remote computer, receives DICOM images from MRI host computer, automatically processes the received images and sends the enhanced images to a PACS server. Enhanced images exist in conjunction to the original images.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
| Requirement Specification | Reported Performance |
|---|---|
| For spin echo large phantom protocol in ACR quality control 2015, the software shall increase SNR by at-least 10% in slice 7 ACR data without compromising high contrast resolution in slice 1 and low contrast resolution in slice 11 | Pass |
| For fast spin echo sequences with slice thickness in the range 2-5mm and in-plane resolution in the range 0.6-1.4 mm, the software shall increase SNR by at-least 10% in slice 7 ACR data without compromising high contrast resolution in slice 1 and low contrast resolution in slice 11 | Pass |
| For gradient echo sequences with slice thickness in the range 2-5mm and in-plane resolution in the range 0.6-1.4 mm, the software shall increase SNR by at-least 10% in slice 7 ACR data without compromising high contrast resolution in slice 1 and low contrast resolution in slice 11 | Pass |
| For spin echo small phantom protocol in ACR quality control 2015, the software shall increase SNR by at-least 10% in slice 7 ACR data without compromising high contrast resolution in slice 1 and low contrast resolution in slice 11 | Pass |
Study Information
1. Sample size used for the test set and the data provenance:
* Sample Size: A total of 64 data sets were acquired using an ACR MRI phantom.
* Data Provenance: The data was acquired using various MRI systems: GE 1.5T Excite, Siemens Avanto 1.5T, Philips Intera 1.5T, and Toshiba Titan 1.5T. The data appears to be prospectively collected as part of a validation study for the device. The country of origin is not explicitly stated, but the use of the ACR (American College of Radiology) phantom suggests a North American context.
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
* The document does not mention human experts or their qualifications for establishing ground truth in this performance study. The ground truth appears to be based on objective phantom measurements (Signal-to-Noise Ratio and contrast resolution) as defined by the ACR quality control protocols.
3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
* None. The performance test relies on objective measurements from the ACR MRI phantom using defined protocols, not expert adjudication of images.
4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
* No. A MRMC comparative effectiveness study involving human readers and AI assistance was not conducted or described in this document. The study focuses solely on the objective performance of the image enhancement software itself.
5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
* Yes. The described study is a standalone performance test of the ZOOM software (algorithm only). It evaluates the software's ability to enhance SNR and maintain resolution based on phantom data, without human interaction in the evaluation process.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
* The ground truth is based on objective measurements derived from an ACR MRI phantom protocol. Specifically, it uses ACR quality control guidelines to measure signal-to-noise ratio (SNR) and high/low contrast resolution.
7. The sample size for the training set:
* The document does not specify the sample size used for the training set of the ZOOM software. It only describes the validation/test set.
8. How the ground truth for the training set was established:
* The document does not provide information on how the ground truth for the training set was established.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Zetta Medical Technologies, LLC. Main Ghazal President 1313 Ensell Road LAKE ZURICH IL 60047
April 24, 2018
Re: K172768
Trade/Device Name: ZOOM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: LLZ Dated: March 14, 2018 Received: March 23, 2018
Dear Main Ghazal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRHs Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information regulations, please Device about labeling see Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K172768
Device Name ZOOM
Indications for Use (Describe)
ZOOM Image Enhancement System is an image processing software that can be used for image enhancement in MRI images. Enhanced images will be sent to PACS server and exist in conjunction to the original images.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirement of Titles 21 CFR §807.87 and 807.92.
-
- Applicant & Submitted By: Zetta Medical Technologies, LLC. 1313 Ensell Road, Lake Zurich, IL 60047 Phone: (847) 550-9990 Fax: (847) 550-9994 Contact Person: Main M. Ghazal, President Date Prepared: March 14th 2018
2. Identification of the Device:
Trade Name: ZOOM Common Name: Image Enhancement System Classification Name: Image Processing System, Radiological (21 CFR, 892.2050, LLZ) Regulatory Description: Picture Archiving and Communications System
3. Predicate Device:
Context Vision Sharp View Image Enhancement System, K024028
4. Indications for Use:
ZOOM Image Enhancement System is an image processing software that can be used for image enhancement in MRI images. Enhanced images will be sent to PACS server and exist in conjunction to the original images.
5. Device Description:
ZOOM Image Enhancement System is an image processing software that can be used for image enhancement in MRI images. ZOOM image enhancement software implements a noise reduction algorithm using wavelets and image guided filtering. Original images are decomposed into different wavelet sub bands and noise in each band is soft threshold. De-noised images are reconstructed from softthresholded images using inverse wavelet transform. The software, which is installed on a remote computer, receives DICOM images from MRI host computer, automatically processes the received images and sends the enhanced images to a PACS server. Enhanced images exist in conjunction to the original images.
{4}------------------------------------------------
6. Substantial Equivalence Table
The subject device ZOOM is substantially equivalent to the predicate device, Sharp View Image Enhancement System. The main difference is that Sharp View Enhancement Image transfer/storage/enhancement system where as ZOOM (as of 7/27/2017) is strictly an MRI image enhancement software. Detailed differences between ZOOM and Sharp view systems are listed in Table-1.
| Characteristics | ZOOM | Sharp View ImageEnhancement System |
|---|---|---|
| Indications forUse | ZOOM ImageEnhancement System isan image processingsoftware that can be usedfor image enhancementin MRI images.Enhanced images will besent to PACS server andexist in conjunction tothe original images. | The Image EnhancementSystem is intended for use bya qualified/trainedtechnologist for transfer,storage, enhancement andviewing of multi-modalityimages. |
| Computer | PC or PC Compatible | PC compatible |
| OperatingSystem | Windows 7 | Windows 98, NT 4.0, 2000 andXP |
| Storage | Is not a primary imagestorage system.However, processedimages are archived onlocal hard drive | Hard disk or any compatiblePC method: Optical, CDROM,Tape |
| Image ProcessingHardware | Intel i3 processor, 4GBRAM, 500GB Hard drive | Javelin (PCI-bus) or Similar |
| Software core | ZOOM ImageEnhancement Software(Zetta's own trademark) | GOP® Enhancement software(The GOP trademark is theproperty of Context Vision) |
| Image Input | DICOM | DICOM |
| Image output | DICOM | DICOM |
Table -1: ZOOM vs Sharp View Image Enhancement System
{5}------------------------------------------------
7. Performance Testing:
ZOOM has been designed, verified and validated in compliance with FDA 21 CFR Part 820 requirements. The device has been validated through the use of ACR MRI PHANTOM. A total of 64 data sets were acquired using spin echo, fast spin echo and gradient echo based sequences from the following systems: GE 1.5T Excite, Siemens Avanto 1.5T, Philips Intera 1.5T and Toshiba Titan 1.5T. Parameters (slice thickness, field of view, matrix dimensions and number of averages) that effect the signal to noise ratio (SNR) were varied while acquiring the 64 data sets. These data sets were processed using ZOOM software for image enhancement and results are compared between original and processed images. Performance test results indicate that the ZOOM software improves SNR by at least 10% without degrading high and low contrast resolutions.
| Requirement specification | Verified on systems | Result |
|---|---|---|
| For spin echo large phantom protocol inACR quality control 2015, the softwareshall increase SNR by at-least 10% in slice7 ACR data without compromising highcontrast resolution in slice 1 and lowcontrast resolution in slice 11 | GE 1.5T Excite Siemens Avanto 1.5T Philips Intera 1.5T Toshiba Titan 1.5T | Pass |
| For fast spin echo sequences with slicethickness in the range 2-5mm and in-plane resolution in the range 0.6-1.4 mm,the software shall increase SNR by at-least10% in slice 7 ACR data withoutcompromising high contrast resolution inslice 1 and low contrast resolution in slice11 | GE 1.5T Excite Siemens Avanto 1.5T Philips Intera 1.5T Toshiba Titan 1.5T | Pass |
| For gradient echo sequences with slicethickness in the range 2-5mm and in-plane resolution in the range 0.6-1.4 mm,the software shall increase SNR by at-least10% in slice 7 ACR data withoutcompromising high contrast resolution inslice 1 and low contrast resolution in slice11 | GE 1.5T Excite Siemens Avanto 1.5T Philips Intera 1.5T Toshiba Titan 1.5T | Pass |
| For spin echo small phantom protocol inACR quality control 2015, the softwareshall increase SNR by at-least 10% in slice7 ACR data without compromising high | GE 1.5T Excite Siemens Avanto 1.5T Philips Intera 1.5T Toshiba Titan 1.5T | Pass |
Table - 2: High level performance test results
{6}------------------------------------------------
| contrast resolution in slice 1 and low |
|---|
| contrast resolution in slice 11 |
8. Safety and Effectiveness:
Based on the ZOOM software performance test results and incorporated risk minimization methods in design, Zetta Medical Technologies concludes that this device is substantially equivalent to the predicate device.
9. Conclusion:
ZOOM is an image enhancement software which has similar indications for use as predicate device. The main difference is that the predicate device is a multimodality image transfer/storage/enhancement system where as ZOOM (as of 7/27/2017) is strictly an MRI image enhancement system. Performance test results and incorporated risk minimization methods demonstrate that ZOOM is as safe and effective as predicate device.
N/A