K Number
K212467

Validate with FDA (Live)

Date Cleared
2021-11-04

(90 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Device Description

The subject device is powder free nitrile patient examination gloves. The subject device is blue color. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The subject device is non-sterile.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: Disposable Nitrile Examination Glove

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodologyAcceptance CriteriaReported Device Performance
ASTM D6319
Physical Dimensions Test
Length(mm)XS/S: ≥220; M/L/XL: ≥230>230 (for all sizes)
Width (mm)XS: 70±10; S: 80±10; M: 95±10; L: 105±10; XL: 115±10XS: 77-80; S: 87-88; M: 95-98; L: 106-107; XL: 115-117
Thickness (mm)Finger: ≥0.05; Palm: ≥0.05Finger: 0.11; Palm: 0.07
ASTM D5151
Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125 leaks
ASTM D6124
Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.11mg
ASTM D412
Physical Properties - Before Aging
Tensile Strength≥14MPa15.1-26.2
Ultimate Elongation≥500%503-634
Physical Properties - After Aging
Tensile Strength≥14MPa14.6-21.8
Ultimate Elongation≥400%503-620
ISO 10993-5
CytotoxicityNon-cytotoxicNot show potential toxicity to L-929 cells
ISO 10993-10
IrritationNon-irritatingNot an irritant
SensitizationNon-sensitizingNot a sensitizer

All reported device performances indicate that the device "Pass"ed the respective acceptance criteria.

2. Sample size used for the test set and the data provenance

The document states:

  • For Watertightness Test (ASTM D5151): "0/125 leaks" suggests a sample size of 125 gloves were tested for holes.
  • For other tests like physical dimensions, thickness, and physical properties (tensile strength, elongation), specific sample sizes are not explicitly stated. However, the standards (ASTM D6319, ASTM D412) typically specify sampling plans.
  • Data Provenance: The tests were conducted by the manufacturer, Inner Mongolia Boming Medical Supplies Co., Ltd., which is based in Inner Mongolia, China. The data would therefore be retrospective testing performed on manufactured batches of the device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This device is a disposable nitrile examination glove, and the testing involved evaluating its physical and biocompatibility characteristics against standardized performance metrics rather than interpretative medical imaging or diagnostic results requiring expert consensus.

4. Adjudication method for the test set

Not applicable. As noted above, the evaluation of the device performance against physical and material standards does not involve adjudication by experts. The results are objective measurements or observations (e.g., absence of leaks, tensile strength values).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a disposable nitrile examination glove, not an AI-powered diagnostic or assistive tool. MRMC studies are used for evaluating AI performance in clinical tasks, which is irrelevant here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical medical device, not an algorithm or AI system.

7. The type of ground truth used

The ground truth used for this device's evaluation is defined by established international and national standards for medical gloves:

  • ASTM D6319 (Standard Specification for Nitrile Examination Gloves for Medical Application)
  • ASTM D5151 (Standard Test Method for Detection of Holes in Medical Gloves)
  • ASTM D6124 (Standard Test Method for Residual Powder on Medical Gloves)
  • ASTM D412 (Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—Tension) - Implied for physical properties like tensile strength and elongation.
  • ISO 10993-5 (Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity)
  • ISO 10993-10 (Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization)

These standards provide the pre-defined, objective criteria (e.g., minimum tensile strength, maximum allowable powder content, AQL for watertightness, non-cytotoxicity, non-irritation, non-sensitization) against which the device's performance is measured.

8. The sample size for the training set

Not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI models. This device's performance is evaluated through direct physical and biological testing.

9. How the ground truth for the training set was established

Not applicable. There is no training set for this type of device. The "ground truth" (acceptance criteria) for evaluating the finished product's performance is established by the referenced consensus standards (ASTM, ISO).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

November 4, 2021

Inner Mongolia Boming Medical Supplies Co., Ltd. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161,East Lujiazui Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K212467

Trade/Device Name: Disposable Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: July 30, 2021 Received: August 6, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212467

Device Name Disposable Nitrile Examination Glove

Indications for Use (Describe)

The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K212467

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information _

Name:Inner Mongolia Boming Medical Supplies Co., Ltd.
Address:Room 326, New Industrial Park Management Committee Office,Tumet Right Banner, Baotou City, Inner Mongolia AutonomousRegion, China
Contact:Guo Hua
Tel:+86-13485097856

Designated Submission Correspondent

Contact:Mr. Boyle Wang
Address:Shanghai Truthful Information Technology Co., Ltd.Room 1801, No. 161 East Lu Jiazui Rd., Pudong, Shanghai200120 China
Tel:+86-21-50313932
Email:info@truthful.com.cn

Date of Preparation: Oct.9,2021

2.0 Device Information

Trade name:Disposable Nitrile Examination Glove
Common name:Patient Examination Gloves
Classification name:Non-powdered patient examination glove
Model(s):XS, S, M, L, XL
Production code:LZA
Regulation number:21CFR880.6250
Classification:Class I
Panel:General Hospital

3.0 Predicate Device Information

Manufacturer:Ever Global (Vietnam) Enterprise Corp
Device:Disposable Powder Free Nitrile Examination Glove, White/Blue/ Black/ Pink Color
510(k) number:K171422

{4}------------------------------------------------

4.0 Indication for Use

The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

5.0 Device Description

The subject device is powder free nitrile patient examination gloves. The subject device is blue color. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The subject device is non-sterile.

ItemSubject Device(K212467)Predicate Device(K171422)
Product CodeLZALZA
Regulation No.21CFR880.625021CFR880.6250
ClassII
Intended UseThe Disposable NitrileExamination Glove is a non-sterile disposable deviceintended for medicalpurposes that is worn on theexaminer's hands or finger toprevent contaminationbetween patient andexaminer.The Nitrile Powder Freepatient examination glove is anon-sterile disposable deviceintended for medical purposesthat is worn on the examiner'shands or finger to preventcontamination between patientand examiner.
MaterialNitrileNitrile
Powdered orPowered freePowdered freePowdered free
Design FeatureAmbidextrousAmbidextrous
ColorantBlueWhite/ Blue/ Black/ Pink
Labeling InformationSingle-use indication,powder free, device color,device name, glove size andquantity,Non-SterileSingle-use indication, powderfree, device color, devicename, glove size and quantity,Non-Sterile
Dimensions(mm)Length:XS/S:\u2265220;M/L/XL: \u2265230;Width:XS: 70\u00b110;Length:XS/S:\u2265220; M: \u2265235;L/XL: \u2265245Width:XS: 75\u00b15;

6.0 Technological Characteristic Comparison Table

{5}------------------------------------------------

S: 80±10;S: 85±5;
M: 95±10;M: 95±5;
L: 110±10;L: 105±5;
XL: 120±10XL: 115±5
Thickness(mm)Finger: ≥0.05;Finger: ≥0.05;
Palm: ≥0.05Palm: ≥0.05
PhysicalPropertiesBeforeAgingTensileStrength14MPa, minTensileStrength14MPa, min
UltimateElongation500% minUltimateElongation500% min
AfterAgingTensileStrength14MPa, minTensileStrength14MPa, min
UltimateElongation400%minUltimateElongation400%min
Freedom fromHolesBe free from holes whentested in accordance withASTMD5151 AQL=2.5Be free from holes when testedin accordance withASTMD5151 AQL=2.5
Meet the requirements ofASTM D6124 <2.0mgMeet the requirements ofASTM D6124
BiocompatibilityISO 10993-10;Under the conditions of thestudy, not an irritant or asensitizerISO 10993-5Under conditions of thestudy, device extract is notcytotoxicComply withISO10993-10/

Analysis: The physical dimensions are little different with that of the predicate, but they all meet the requirements of ASTM D6319.

7.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

{6}------------------------------------------------

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

TestMethodologyPurposeAcceptance CriteriaResults
ASTM D6319PhysicalDimensionsTestLength(mm):XS/S:≥220;M/L/XL: ≥230;Length(mm):>230
Width:Width(mm):
XS: 70±10;XS: 77-80;
S: 80±10;S: 87-88
M: 95±10;M: 95-98
L: 105±10;L: 106-107
XL: 115±10XL: 115-117
Pass
Thickness (mm) :Finger: ≥0.05Palm: ≥0.05Finger: 0.11Palm: 0.07Pass
ASTM D5151WatertightnessTest forDetection ofHolesMeet the requirements ofASTM D5151 AQL 2.50/125 leaksPass
ASTM D6124PowderContentMeet the requirements ofASTM D6124 < 2.0mg0.11mgPass
ASTM D412PhysicalpropertiesTensile15.1-26.2
Before Strength ≥14MPaPass
Aging UltimateElongation ≥500%503-634Pass
Tensile14.6-21.8
After Strength ≥14MPaAging UltimateElongation ≥400%Pass503-620Pass
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditionsof the study, didnot show potentialtoxicity to L-929cells
Pass
ISO 10993-10IrritationNon-irritatingUnder theconditions of thestudy, not anirritant.Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditionsof the study, not asensitizer.Pass

Table 2: Summary of Non-clinical Testing Table

{7}------------------------------------------------

8.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

9.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Disposable Nitrile Examination Glove, is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.