K Number
K212401

Validate with FDA (Live)

Date Cleared
2021-10-25

(84 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Device Description

The subject device is powder free nitrile patient examination gloves. The subject device is blue color and has 5 models of XS,S, M, L, XL. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The subject device is non-sterile.

AI/ML Overview

The provided text describes the acceptance criteria and the results of non-clinical testing for a medical device, specifically a "Disposable Nitrile Examination Glove." It does not contain information about studies involving AI, human readers, or clinical trials.

Here's a breakdown of the requested information based on the provided text, focusing on the available non-clinical data:

1. A table of acceptance criteria and the reported device performance

Test MethodologyPurposeAcceptance CriteriaReported Device Performance
ASTM D6319Physical Dimensions TestLength (mm):XS/S: ≥220;M/L/XL: ≥230;Width:XS: 70±10;S: 80±10;M: 95±10;L: 105±10;XL: 115±10Thickness (mm):Finger: ≥0.05Palm: ≥0.05Length (mm): >230Width (mm):XS: 72-74;S: 80-83M: 95-98L: 110-114XL: 118-121PassThickness (mm):XS (Finger: 0.07-0.10, Palm: 0.08-0.10)S (Finger: 0.08-0.11, Palm: 0.08-0.11)M (Finger: 0.08-0.12, Palm: 0.07-0.11)L (Finger: 0.08-0.12, Palm: 0.08-0.11)XL (Finger: 0.08-0.12, Palm: 0.08-0.12)Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.5XS: 2/125 leaksS: 0/125 leaksM: 0/125 leaksL: 1/125 leaksXL: 0/125 leaksPass
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mgXS: 0.05mgS: 0.06mgM: 0.06mgL: 0.07mgXL: 0.09mgPass
ASTM D412Physical Properties (Before Aging)Tensile Strength: ≥14MPaUltimate Elongation: ≥500%Tensile Strength:XS: 15.4-17.3S: 15.3-16.9M: 15.4-17.3L: 15.4-17.6XL: 15.3-17.1PassUltimate Elongation:XS: 524-569S: 525-568M: 525-567L: 527-566XL: 520-570Pass
ASTM D412Physical Properties (After Aging)Tensile Strength: ≥14MPaUltimate Elongation: ≥400%Results for "After Aging" for Tensile Strength and Ultimate Elongation are specified in the acceptance criteria, but specific numerical performance results are not explicitly provided in the "Results" column of the table. However, since there is a "Pass" next to the "ultimate Elongation" line, it can be inferred that these too met the criteria.
ISO 10993-5CytotoxicityNon-cytotoxicUnder conditions of the study, did not show potential toxicity to L-929 cells.Pass
ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant.Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer.Pass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Watertightness (ASTM D5151): 125 gloves were tested for each size (XS, S, M, L, XL).
  • Other tests (Physical Dimensions, Powder Content, Physical Properties, Biocompatibility): The document does not specify the exact sample sizes for these tests, but it indicates that tests were performed for each glove size (XS, S, M, L, XL).
  • Data Provenance: The tests were conducted to demonstrate compliance with international (ISO) and US (ASTM) standards. The manufacturer is based in China. The data's specific origin beyond that is not detailed in the provided text, and it appears to be prospective testing conducted specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is Not Applicable. The study described is a non-clinical device testing study, relying on standardized test methods (ASTM, ISO) rather than expert interpretation of medical images or patient data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is Not Applicable. Adjudication methods are typically relevant for studies involving human interpretation or subjective assessments, which are not present in this non-clinical testing. The adherence to standard methodologies (e.g., AQL for ASTM D5151) implies a defined pass/fail criterion rather than expert adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is Not Applicable. The provided document describes non-clinical testing of a physical medical device (examination gloves) and does not involve AI, human readers, or comparative effectiveness studies of that nature.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is Not Applicable. The device is a physical examination glove, not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is established by the defined parameters and measurement techniques of the referenced international and national standards (ASTM D6319, ASTM D5151, ASTM D6124, ASTM D412, ISO 10993-5, ISO 10993-10). For example, for "Watertightness," the ground truth is a quantitative measurement of leaks based on ASTM D5151 and an AQL of 2.5. For "Cytotoxicity," it's the observed cellular response in an in-vitro test according to ISO 10993-5.

8. The sample size for the training set

This section is Not Applicable. The provided document describes non-clinical performance testing of a manufactured product, not machine learning model development. Therefore, there is no "training set."

9. How the ground truth for the training set was established

This section is Not Applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 25, 2021

Inner Mongolia Cureguard Medical Technology Co.,Ltd. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161, East Lujiazui Rd., Pudong Shanghai, Shanghai 200120 China

Re: K212401

Trade/Device Name: Disposable Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: July 16, 2021 Received: August 2, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212401

Device Name Disposable Nitrile Examination Glove

Indications for Use (Describe)

The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K212401)

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Inner Mongolia Cureguard Medical Technology Co.,Ltd. Name: Address: Room 326, Management Committee of New Industrial Park, Tumote youqi. Baotou. Inner Monqolia Autonomous Region 014100. China. Phone Number: +86-13485097856 Contact: Guo Hua Date of Preparation: Oct.22,2021

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Disposable Nitrile Examination Glove Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): XS,S, M, L, XL

3.0 Classification

Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Ever Global (Vietnam) Enterprise Corp Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422

{4}------------------------------------------------

5.0 Indication for Use

The Disposable Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

6.0 Device Description

The subject device is powder free nitrile patient examination gloves. The subject device is blue color and has 5 models of XS,S, M, L, XL. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The subject device is non-sterile.

7.0 Technological Characteristic Comparison Table

ItemSubject Device(K212401)Predicate Device(K171422)Remark
Product CodeLZALZASame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Indication for Use/Intended UseThe Disposable NitrileExamination Glove is a non-sterile disposable deviceintended for medicalpurposes that is worn on theexaminer's hands or finger toprevent contaminationbetween patient andexaminer.The Disposable Powder FreeNitrile Examination Glove,White/ Blue/ Black/ Pink Coloris a disposable deviceintended for medical purposesthat is worn on the examiner'shands to preventcontamination between patientand examiner.Same
MaterialNitrileNitrileSame
Powdered orPowered freePowder freePowder freeSame
Design FeatureAmbidextrousAmbidextrousSame
ColorantBlueWhite/ Blue/ Black/ PinkSimilar
Labeling InformationSingle-use indication,powder free, device color,device name, glove size andquantity, Non-SterileSingle-use indication, powderfree, device color, devicename, glove size and quantity,Non-SterileSame
Dimensions(mm)Length:XS/S: $\geq$ 220;M/L/XL: $\geq$ 230;Width:XS: 70 $\pm$ 10:Length:XS/S: $\geq$ 220; M: $\geq$ 235;L/XL: $\geq$ 245Width:XS: 75 $\pm$ 5:Similar

{5}------------------------------------------------

S: 80±10;M: 95±10;L: 105±10;XL: 115±10S: 85±5;M: 95±5;L: 105±5;XL: 115±5
Thickness(mm)Finger: ≥0.05;Palm: ≥0.05Finger: ≥0.05;Palm: ≥0.05Same
TensileStrengthTensileStrengthSame
BeforeAging14MPa, min14MPa, min
UltimateElongationUltimateElongationSame
PhysicalProperties500% min500% min
TensileStrengthTensileStrengthSame
AfterAging14MPa, min14MPa, min
UltimateElongationUltimateElongationSame
400%min400%min
Freedom fromHolesBe free from holes whentested in accordance withASTM D5151 AQL=2.5Be free from holes when testedin accordance withASTMD5151 AQL=2.5Same
Powder ContentMeet the requirements ofASTM D6124 $<$ 2.0mgMeet the requirements ofASTM D6124Same
BiocompatibilityISO 10993-10;Under the conditions of thestudy, not an irritant or asensitizerComply withISO10993-10Same
ISO 10993-5
Under conditions of thestudy, device extract is notcytotoxic/Similar

Analysis:

The color(blue) of the subject device is different with those (white/ blue/ black/ pink) of the predicate device, biocompatibility test has been performed on subject device and the test result can meet the requirements of ISO 10993 standards.

The physical dimensions are little different with that of the predicate, but they all meet the requirements of ASTM D6319.

Therefore, the differences will not raise any safety and effectiveness issues.

8.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device

{6}------------------------------------------------

complies with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

TestMethodologyPurposeAcceptance CriteriaResults
ASTM D6319PhysicalDimensionsTestLength(mm):XS/S: ≥220;M/L/XL: ≥230;Width:XS: 70±10;S: 80±10;M: 95±10;L: 105±10;XL: 115±10Thickness (mm) :Finger: ≥0.05Palm: ≥0.05Length(mm):>230Width(mm):XS: 72-74;S: 80-83M: 95-98L: 110-114XL: 118-121PassXS:Finger: 0.07-0.10Palm: 0.08-0.10S:Finger: 0.08-0.11Palm: 0.08-0.11M:Finger: 0.08-0.12Palm: 0.07-0.11L:Finger: 0.08-0.12Palm: 0.08-0.11XL:Finger: 0.08-0.12

Table 2: Summary of Non-clinical Testing Table

{7}------------------------------------------------

Palm: 0.08-0.12
Pass
ASTM D5151WatertightnessTest forDetection ofHolesMeet the requirements ofASTM D5151 AQL 2.5XS:2/125 leaksS:0/125 leaksM:0/125 leaksL: 1/125 leaksXL: 0/125 leaksPass
ASTM D6124PowderContentMeet the requirements ofASTM D6124 < 2.0mgXS:0.05mgS:0.06mgM:0.06mgL:0.07mgXL:0.09mgPass
ASTM D412PhysicalpropertiesBeforeAgingTensileStrength≥14MPaXS:15.4-17.3S:15.3-16.9M: 15.4-17.3L:15.4-17.6XL:15.3-17.1Pass
UltimateElongation≥500%XS:524-569S:525-568M: 525-567L:527-566XL:520-570Pass
AfterAgingTensileStrength≥14MPa
UltimateElongation≥400%

{8}------------------------------------------------

ISO 10993-5CytotoxicityNon-cytotoxicUnder conditionsof the study, didnot show potentialtoxicity to L-929cells.Pass
ISO 10993-10IrritationNon-irritatingUnder theconditions of thestudy, not anirritant.Pass
ISO 10993-10SensitizationNon-sensitizingUnder conditionsof the study, not asensitizer.Pass

9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device, Disposable Nitrile Examination Glove, is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.