K Number
K212311

Validate with FDA (Live)

Date Cleared
2021-10-21

(90 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medical Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Device Description

The proposed device is Powder Free Medical Examination Gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Medical Examination Gloves, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Biocompatibility - Skin Sensitization (ISO 10993-10)Provided grades less than 1, otherwise sensitization.All grades are 0. All animals survived and no abnormal signs were observed during the study.
Biocompatibility - Skin Irritation (ISO 10993-10)If the primary irritation index is 0-0.4, the response category is Negligible. 0.5-1.9 means slight, 2-4.9 means moderate, 5-8 means severe.The primary irritation index is 0. The response of the proposed device was categorized as negligible under the test condition.
Biocompatibility - In Vitro Cytotoxicity (ISO 10993-5)Viability reduced to <70% of the blank indicates cytotoxic potential.Viability of 100% test article extract is 38.4%. (Indicating potential cytotoxicity to L-929 cells in the MTT method).
Biocompatibility - Systemic Toxicity (ISO 10993-11)Within the monitoring period (72 h), if the toxicosis response of testing group is not greater than that of control group, the testing sample is regarded as acceptable.No toxicosis response in testing group. (Meaning no potential acute systemic toxicity on ICR mice in the extraction method).
Residual Powder (ASTM D6124)Powder residue limit of 2.0 mg/glove.0.11 mg /glove.
Freedom from Holes (ASTM D5151)Samples number: 125 gloves. AQL: 2.5 (ISO 2859). Criterion ≤7 gloves for water leakage.No glove water leakage found.
Dimensions (ASTM D6319)S: width 80±10mm, Length ≥220 mmM: width 95±10mm, Length ≥230 mmL: width 110±10mm, Length ≥230 mmXL: width 120±10mm, Length ≥230 mmThickness: Finger ≥0.05 mm, Palm ≥0.05 mmS: width: 84-87 mm, Length 239-242 mmM: width 95-97 mm, Length 243-245 mmL: width 104-106 mm, Length 250-252 mmXL: width 114-117 mm, Length 244-247 mmThickness: Finger 0.09 mm, Palm 0.07 mm
Physical Properties - Before Aging (ASTM D6319)Tensile strength ≥ 14MPaUltimate Elongation ≥ 500%Tensile strength 14.06-20.59 MPaUltimate Elongation 552.680% - 652.080%
Physical Properties - After Accelerated Aging (ASTM D6319)Tensile strength ≥ 14MPaUltimate Elongation ≥ 400%Tensile strength 14.06-16.90MPaUltimate Elongation 508.43% - 646.33%

2. Sample Sizes and Data Provenance

  • Freedom from Holes (ASTM D5151): The sample size used for the test set was 125 gloves. The data provenance is not explicitly stated as retrospective or prospective, nor is the country of origin mentioned. However, typically such testing is prospective and conducted at the manufacturer's facility or a certified lab.
  • For other tests (Biocompatibility, Residual Powder, Dimensions, Physical Properties), specific sample sizes for each test are not provided in the summary. They generally fall under the methodologies prescribed by the respective ASTM/ISO standards.
  • Data Provenance: The document does not specify the country of origin for the data (e.g., test samples) or if the studies were retrospective or prospective. Given that this is a 510(k) summary for a premarket notification, the testing would generally be prospective to demonstrate compliance with standards for the device being submitted.

3. Number of Experts and Qualifications for Ground Truth

  • This device is a medical examination glove, which is a Class I device. The testing outlined focuses on physical properties, chemical properties (biocompatibility), and manufacturing quality (freedom from holes, powder content).
  • No "experts" in the traditional sense (e.g., radiologists) were used to establish ground truth in the context of image interpretation or complex clinical diagnoses. The ground truth for these tests is based on objective measurements and established scientific protocols (e.g., chemical analysis, physical property testing, biological assays) as defined by the referenced ISO and ASTM standards. The "truth" is determined by the outcome of these standardized tests against predefined limits.

4. Adjudication Method for the Test Set

  • Not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies, particularly for diagnostic devices where multiple human readers or a consensus of experts is needed to establish a "ground truth" for ambiguous cases.
  • For these gloves, the acceptance criteria are based on objective, quantifiable measurements or qualitative observations by trained laboratory personnel following standard operating procedures. There is no human interpretation or diagnostic "adjudication" necessary.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No. This type of study is not relevant for a medical examination glove. MRMC studies are used for diagnostic devices, particularly AI-powered ones, to compare the performance of human readers with and without AI assistance on a set of clinical cases.
  • Medical examination gloves are physical barriers, and their effectiveness is evaluated by their material properties, barrier integrity, and biocompatibility, not by their influence on human reading or diagnostic accuracy.

6. Standalone (Algorithm Only) Performance Study

  • No. This concept is not applicable. A "standalone" study refers to the performance of an algorithm (AI) without any human interaction. Medical examination gloves do not involve algorithms or AI. The device is a physical product.

7. Type of Ground Truth Used

  • The ground truth used for the tests is derived from:
    • Objective measurement standards: For dimensions, physical properties (tensile strength, elongation), and powder content, the ground truth is an objective, quantitative measurement against the limits specified in ASTM D6319 and ASTM D6124.
    • Biological assay outcomes: For biocompatibility tests (irritation, sensitization, cytotoxicity, systemic toxicity), the ground truth is the biological response (or lack thereof) observed in in vitro or in vivo models, interpreted according to the criteria defined in ISO 10993 series.
    • Absence of defect: For "freedom from holes," the ground truth is the absence of water leakage, as determined by ASTM D5151.

8. Sample Size for the Training Set

  • Not applicable. Medical examination gloves are not "trained" in the sense of machine learning algorithms. There is no training set for this device. The physical gloves are manufactured and then tested to ensure they meet specified standards.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no training set for this device, the question of how its ground truth was established is irrelevant.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 21, 2021

Guangzhou Junda Gloves Co., Ltd Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room 608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K212311

Trade/Device Name: Medical Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: July 20, 2021 Received: July 23, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K212311

Device Name Medical Examination Gloves

Indications for Use (Describe)

The Medical Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K212311

This summary of 510(k) safety and effectiveness information is being submitted in accordance with 21 CFR 807.92.

1.0 submitter's information

Name: Guangzhou Junda Gloves Co., Ltd Address: No.38 Heting Fengwei Industrial Zone Renhe Town Baiyun District, Guangzhou,Guangdong,510470,China Phone Number: +86-20-37738661 Contact: Olivia Chen Date of Preparation: 2021.07.20

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device information

Trade name: Medical Examination Gloves Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL

3.0 Classification

Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate device information

Manufacturer: Ever Global (Vietnam) Enterprise Corp Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422

{4}------------------------------------------------

5.0 Intended use

The Medical Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands or finger to prevent contamination between patient and examiner.

6.0 Device description

The proposed device is Powder Free Medical Examination Gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

7.0 Summary comparing technological characteristics with predicate device

ItemProposed devicePredicated deviceRemark
510(k) numberK212311K171422
Product CodeLZALZASame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseThe Medical ExaminationGloves is a disposabledevice intended formedical purposes that isworn on the examiner'shands or finger to preventcontamination betweenpatient and examinerThe Nitrile Powder Freepatient examination gloveis a non-sterile disposabledevice intended formedical purpose that isworn on the examiner'shands or finger to preventcontamination betweenpatient and examinerSame
Powdered or Powered freePowdered freePowdered freeSame
Design FeatureambidextrousambidextrousSame
Labeling InformationSingle-use indication,powder free, device color,device name, glove sizeand quantity, MedicalExamination Gloves,Non-SterileSingle-use indication,powder free, device color,device name, glove sizeand quantity, DisposablePowder Free NitrileExamination Glove,Non-SterileSame

Table1-General Comparison

Table2 Device Dimensions Comparison

PredicateDesignationSizeTolerance
-----------------------------------------

{5}------------------------------------------------

Device(K171422)XSMLXL
Length, mm230230230230230min
Width, mm7585ત્ત્વની તે જેવાયું છે છે. આ ગામનાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલ105115– 5
Thickness, mm:Finger0.05
min
Palm0.05min
Proposed DeviceDesignationSizeTolerance
MLXL
Length, mm220230230230min
Width, mm80ત્ત્વની તે જેવાયું છે છે. આ ગામનાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલ110120+10
Thickness, mm:
Finger0.05min
Palm0.05min
RemarkAnalysis1

Analysis1: The sizes and tolerances of proposed device are different with those of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

Table3 Performance Comparison
ItemProposed devicePredicated deviceRemark
ColorantblueWhite/ Blue/ Black/ PinkAnalysis2
PhysicalPropertiesBeforeAgingTensileStrength14MPa, min14MPa, minSAME
UltimateElongation500%min500%minSAME
AfterAgingTensileStrength14MPa, min14MPa, minSAME
UltimateElongation400%min400%minSAME
Comply with ASTM D6319Comply with ASTM D6319Comply with ASTM D6319SAME
Freedom from HolesBe free from holeswhen tested inaccordance withASTMD5151AQL=2.5Be free from holes whentested in accordance withASTMD5151 AQL=2.5SAME
Powder Content0.11Meet the requirements ofASTM D6124SIMILAR

Analysis 2: The proposed device has different color to the predicate device, but all proposed devices are conducted the biocompatibility test, the test results shown that the color difference do not effect the safety of proposed device

{6}------------------------------------------------

ItemProposed devicePredicated deviceRemark
MaterialNitrileNitrileSAME
BiocompatibilityIrritationUnder the conditions of the study,not an irritantComply withISO10993-10SAME
SensitizationUnder conditions of the study, not asensitizer.
CytotoxicityUnder the conditions of the study,the device is potentially cytotoxicComply withISO10993-5Analysis3
SystemictoxicityUnder the conditions of the study,the device does not elicit a systemictoxicity response in the modelanimal.Complies with ISO10993-11 Third edition2017-09
Label and LabelingMeet FDA's RequirementMeet FDA'sRequirementSAME

Table4 Safety Comparison

Analysis3: The proposed device is potentially cytotoxic, but all proposed devices are conducted the systemic toxicity test, the test results show that the proposed device is safe.

8.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications and acceptance criteria. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ISO 10993-11 Third edition 2017-09, Biological evaluation of medical devices -Part 11: Tests for systemic toxicity.

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19,Standard Specification For Nitrile Patient Examination Gloves For Medical Application.

Table 5 Summary of Non-Clinical Performance Testing

No.Name of the TestPurposeAcceptance CriteriaResults
Methodology / Standard
1ISO 10993-10:2010Biological Evaluation OfMedical Devices - Part10: Tests For IrritationAnd Skin Sensitization.This part of ISO 10993 assesses possible contact hazards from chemicals released from medical devices, which may produce skin and mucosal irritation, eye irritation or skin sensitization.Skin Sensitization Test:provided grades less than 1, otherwise sensitization.All grades are 0.All animals were survived and no abnormal signs were observed during the study.
2Skin Irritation Test:If the primary irritation index is 0-0.4, the response category is Negligible.0.5-1.9 means slight2-4.9 means moderate5-8 means severeThe primary irritation index is 0.The response of the proposed device was categorized as negligible under the test condition
3ISO 10993-5:2009Biological Evaluation OfMedical Devices - Part5: Tests For In VitroCytotoxicityThis part of ISO 10993 describes test methods to assess the in vitro cytotoxicity of medical devices.The viab.% of the 100% extract of the test article is the final result, and if viability is reduced to <70% of the blank, it has cytotoxic potential.Viab.% of 100% test article extract is 38.4%It means the proposed device have potential toxicity to L-929 in the MTT method
4ISO 10993-11: 2017Biological evaluation of medical devices — Part11: Tests for systemic toxicityTo evaluate the potential for medical device materials to cause adverse systemic reactions.Within the monitoring period (72 h), if the toxicosis response of testing group is not greater than that of control group, the testing sample is regarded as acceptable.No toxicosis response in testing group.It means the test article has no potential acute system toxicity on ICR mice in the extraction method.
5ASTM D6124-06(Reapproved 2017),Standard Test Methodfor Residual Powder onMedical GlovesThis standard is designed to determine the amount of residual powder (or filter-retained mass) found on medical glovespowder residue limit of 2.0 mg0.11 mg /glove
6ASTMD5151-06(Reapproved2015), Standard TestMethod for Detection ofHoles in Medical Gloves.This test methodcovers thedetection of holesinmedical gloves.Samples number: 125glovesAQL: 2.5 (ISO 2859)Criterion ≤7 glovesfor water leakageno glove water leakage found
7ASTMD6319-10(Reapproved2015),StandardSpecification For NitrileExamination Gloves ForMedical Application.This specificationcovers certainrequirements fornitrile rubbergloves used inconductingmedicalexaminations anddiagnostic andtherapeuticprocedures.Sterility: no needFreedom from holes:Dimensions:S: width 80±10mmLength ≥220 mmM: width 95±10mmLength ≥230 mmL: width 110±10mmLength ≥230 mmXL: width 120±10mmLength ≥230 mmThickness:Finger ≥0.05 mmPalm ≥0.05 mmN.A.Dimensions:S: width: 84-87 mmLength 239-242 mmM: width 95-97 mmLength 243-245 mmL: width 104-106 mmLength 250-252 mmXL: width 114-117 mmLength 244-247 mmThickness:Finger 0.09 mmPalm 0.07 mm
Physical properties:Before agingTensile strength ≥14MPaUltimate Elongation ≥500%After AcceleratedAgingTensile strength ≥14MPaUltimate Elongation ≥400%Powder-free Residue:pl. Refer to No. 4 intable 5Physical properties:Before agingTensile strength 14.06-20.59MPaUltimate Elongation 552.680% -652.080%After Accelerated AgingTensile strength 14.06-16.90MPaUltimate Elongation 508.43% -646.33%Powder-free Residue:pl. Refer to No. 5 in table 5

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

9.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.