K Number
K212148

Validate with FDA (Live)

Date Cleared
2022-03-10

(244 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: 45.4 minutes and Thiotepa: 23.5 minutes. Warning: Do not use with Carmustine or Thiotepa.

Device Description

Microflex 73-743 Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs is a non-sterile, single use only, disposable, powder free examination gloves. Characteristics: - . Ambidextrous with beaded cuff and straight fingers - Finger-textured . - . Blue colored - Five (5) sizes – extra-small, small, medium, large, and extra-large - . Tested against chemotherapy drugs High levels of ozone will degrade rubber material of the glove; therefore, the glove should be protected from ozone in particular. The glove is designed to meet the specifications of ASTM D6977-19, Standard Specification for Polychloroprene Examination Gloves for Medical Application.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device: "Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs".

The information provided describes the product's acceptance criteria and the non-clinical testing performed to demonstrate that the device meets some of these criteria. However, it does not describe an AI/ML-driven medical device, nor does it detail a study proving that an AI/ML device meets acceptance criteria. Therefore, I am unable to answer the questions related to AI/ML device performance, ground truth, expert opinions, or MRMC studies.

Instead, I will extract the acceptance criteria and study information related to the glove's performance as described in the document.


Acceptance Criteria and Study for the Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs

This submission is for a physical medical device (examination glove), not an AI/ML system. Therefore, the "acceptance criteria" relate to the physical and chemical performance characteristics of the glove. The "study" refers to non-clinical laboratory testing.

1. Table of Acceptance Criteria and Reported Device Performance

The device performance is assessed against recognized standards, primarily ASTM D6977-19, ASTM D5151, ASTM D6124, and ASTM D6978-05 for chemotherapy drug permeation, as well as ISO 10993 series for biocompatibility.

Test Methodology / CharacteristicAcceptance Criteria (Standard)Reported Device Performance (Results)
Physical Characteristics:
Dimensions (ASTM D6977-19):Meets ASTM D6977-19 requirements for length, width and thickness
- LengthMinimum 230mmMinimum 240mm
- Palm Width (XS)$70 \pm 10$ mm$75 \pm 5$ mm
- Palm Width (S)$80 \pm 10$ mm$85 \pm 5$ mm
- Palm Width (M)$95 \pm 10$ mm$95 \pm 5$ mm
- Palm Width (L)$110 \pm 10$ mm$105 \pm 5$ mm
- Palm Width (XL)$120 \pm 10$ mm$115 \pm 5$ mm
- Thickness (Finger)Minimum 0.05 mmFinger - min 0.09 mm
- Thickness (Palm)Minimum 0.05 mmPalm - min 0.06 mm
- Thickness (Cuff)- (Not specified in standard for comparison)Cuff - min 0.05 mm
Physical Properties (ASTM D6977-19):Meets ASTM D6977-19 requirements for tensile strength and ultimate elongation before and after accelerated aging.Meets ASTM D6977-19 requirements for tensile strength and ultimate elongation before and after accelerated aging.
- Tensile Strength (Before Aging)Minimum 14 MPaMinimum 16 MPa
- Tensile Strength (After Aging)Minimum 14 MPaMinimum 14 MPa
- Ultimate Elongation (Before Aging)Minimum 500%Minimum 500%
- Ultimate Elongation (After Aging)Minimum 400%Minimum 400%
Freedom from Holes (ASTM D6977-19, ASTM D5151-06):AQL 2.5Meets or exceeds ASTM D6977-19 and ASTM D5151-06 requirements of AQL 2.5
Powder Residual (ASTM D6977-19, ASTM D6124-06):Not more than 2.0 mg/gloveMeets applicable requirement for powder free; ≤ 2 mg per glove
Biocompatibility:
- ISO Skin Irritation StudyISO10993-10:2010 (Not an irritant)Under the conditions of the study, not an irritant
- ISO Maximization Sensitization StudyISO 10993-10:2010 (Not a sensitizer)Under the conditions of the study, not a sensitizer
- ISO Acute Systemic ToxicityISO 10993-11: 2006 (No evidence of systemic toxicity)Under the conditions of the study, no evidence of systemic toxicity
- ISO In Vitro CytotoxicityISO 10993-5:2009 (Non-cytotoxic at specified dilutions)Under the conditions of the study, undiluted, 1:2, and 1:4 dilutions were cytotoxic (grade 4). Dilutions of 1:8 (grade 2), 1:16 (grade 1), 1:32 and 1:64 (grade 0) were non-cytotoxic. (This indicates cytotoxicity at high concentrations, which is a common finding for glove materials, and the acceptance is based on non-cytotoxicity at lower, clinically relevant dilutions for biocompatibility).
Chemotherapy Permeation (ASTM D6978-05):Times as per standard/previous testing (predicate device, if applicable)
- Carmustine - 3.3 mg/ml- (Comparison to predicate: 47.9 minutes)45.4 minutes
- Cisplatin - 1.0 mg/ml- (Not explicitly in predicate comparison table)>240 minutes
- Cyclophosphamide - 20.0 mg/ml>240 minutes>240 minutes
- Dacarbazine - 10.0 mg/ml- (Not explicitly in predicate comparison table)>240 minutes
- Doxorubicin HCl - 2.0 mg/ml>240 minutes>240 minutes
- Etoposide - 20.0 mg/ml>240 minutes>240 minutes
- Fluorouracil - 50.0 mg/ml>240 minutes>240 minutes
- Methotrexate - 25.0 mg/ml>240 minutes>240 minutes
- Paclitaxel - 6.0 mg/ml>240 minutes>240 minutes
- Thiotepa - 10.0 mg/ml>240 minutes (from predicate)23.5 minutes (Note: This is significantly lower than the predicate for the same drug, leading to a warning for its use with this glove.)
- Vincristine Sulfate - 1.0 mg/ml>240 minutes>240 minutes

2. Sample Size Used for the Test Set and the Data Provenance

The document does not specify exact sample sizes for each test beyond indicating the tests were conducted. However, standards like ASTM and ISO have prescribed sample sizes and methodologies. For instance, for chemotherapy drug permeation testing as per ASTM D6978-05, a minimum of three specimens are tested per drug (indicated by "Average Minimum Breakthrough Detection Time (Sample 1,2,3)").

The data provenance is not explicitly stated as retrospective or prospective or by country of origin, but it originates from non-clinical laboratory testing performed to meet recognized international standards. This type of testing is typically prospective, conducted under controlled laboratory conditions.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This is not applicable as this is a non-clinical device that relies on standardized laboratory testing rather than human expert interpretation of data.

4. Adjudication Method for the Test Set

Not applicable. The testing is based on objective measurements and adherence to specified standard protocols, not expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML device.

7. The type of ground truth used

The "ground truth" for the performance claims of this medical glove consists of measurements obtained according to established and recognized international laboratory standards (e.g., ASTM, ISO). For example:

  • Physical dimensions and properties are measured directly.
  • Freedom from holes is determined by a water leak test.
  • Powder residual is measured gravimetrically.
  • Biocompatibility is assessed through in vitro and in vivo biological tests with specific endpoints (e.g., cell viability for cytotoxicity, skin reactions for irritation/sensitization).
  • Chemotherapy drug permeation is measured using highly sensitive analytical methods to detect breakthrough time.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device; there is no "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable. This is not an AI/ML device; there is no "training set."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

March 10, 2022

Ansell Healthcare Products LLC Donald Cronk Associate Director 2301 Robb Drive Reno, Nevada 89523

Re: K212148

Trade/Device Name: Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: February 28, 2022 Received: March 1, 2022

Dear Donald Cronk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

encies. You must comply with all the Act's

Page 2

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K212148

Device Name

Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs

Indications for Use (Describe)

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: 45.4 minutes and Thiotepa: 23.5 minutes. Warning: Do not use with Carmustine or Thiotepa.

Tested Chemotherapy drug & ConcentrationAverage Breakthrough Detection Time (Minutes)
Carmustine - 3.3 mg/ml45.4
Cisplatin - 1.0 mg/ml>240
Cyclophosphamide - 20.0 mg/ml>240
Dacarbazine - 10.0 mg/ml>240
Doxorubicin HCl - 2.0 mg/ml>240
Etoposide - 20.0 mg/ml>240
Fluorouracil - 50.0 mg/ml>240
Methotrexate - 25.0 mg/ml>240
Paclitaxel - 6.0 mg/ml>240
Thiotepa - 10.0 mg/ml23.5
Vincristine Sulfate - 1.0 mg/ml>240

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510K Summary

510(k) Number: K212148

Submitter:

Ansell Healthcare Products LLC. 2301 Robb Drive Reno, NV 89523

Contact Person(s):

Don Cronk Associate Director, Regulatory Affairs Phone: (775) 470-7106 Email: don.cronk@ansell.com

Jacob Ramirez Senior Coordinator, Regulatory Affairs Phone: (775) 624-8118 Email: jacob.ramirez@ansell.com

Date Prepared:

2/23/2022

Name of the Device:

Trade Names:Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue)Tested for Use with Chemotherapy Drugs
Common Name:Patient Examination Glove
Classification Name:Patient Examination Glove
Classification Regulation:21 CFR 880.6250
Device Class:I
Product Code:LZA, LZC
Classification Panel:Non-powdered patient examination glove

Legally Marketed Predicate Device:

Company:Ansell Healthcare Products LLC
Trade Names:Microflex® Nitrile Patient Examination Gloves Blue Colored Tested forUse with Chemotherapy Drugs and Fentanyl Citrate
510(k) Number:K210401
Device Class:Class I
Product Code:LZA, LZC, QDO
Device Name:Patient Examination Glove (21 CFR 880.6250)

{4}------------------------------------------------

Device Description:

Microflex 73-743 Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs is a non-sterile, single use only, disposable, powder free examination gloves.

Characteristics:

  • . Ambidextrous with beaded cuff and straight fingers
  • Finger-textured .
  • . Blue colored
  • Five (5) sizes – extra-small, small, medium, large, and extra-large
  • . Tested against chemotherapy drugs

High levels of ozone will degrade rubber material of the glove; therefore, the glove should be protected from ozone in particular.

The glove is designed to meet the specifications of ASTM D6977-19, Standard Specification for Polychloroprene Examination Gloves for Medical Application.

Indications for Use Statement:

Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: Carmustine: 45.4 minutes and Thiotepa: 23.5 minutes. Warning: Do not use with Carmustine or Thiotepa.

TEST CHEMOTHERAPY DRUG ANDCONCENTRATIONAVERAGE MINIMUM BREAKTHROUGH DETECTIONTIME(Sample 1,2,3) (Minutes)
Carmustine (3.3 mg/ml)45.4
Cisplatin (1.0 mg/ml)>240
Cyclophosphamide (20.0 mg/ml)>240
Dacarbazine (10.0 mg/ml)>240
Doxorubicin HCl (2.0 mg/ml)>240
Etoposide (20.0 mg/ml)>240
Fluorouracil (50.0 mg/ml)>240
Methotrexate (25.0 mg/ml)>240
Paclitaxel (6.0 mg/ml)>240
Thiotepa (10.0 mg/ml)23.5
Vincristine Sulfate (1.0 mg/ml)>240

Tested chemotherapy drugs are as follows:

{5}------------------------------------------------

Technological Characteristics:

PredicateSubject DeviceComparison
Trade nameMicroflex® Nitrile PatientExamination Gloves Blue ColoredTested for Use with ChemotherapyDrugs and Fentanyl CitrateNon-Sterile Powder-FreePolychloroprene ExaminationGlove (Blue) Tested for Use withChemotherapy DrugsDifferent
Product OwnerAnsell HealthcareAnsell HealthcareSame
Product CodeLZA, LZC, QDOLZA, LZCDifferent
Regulation Number21 CFR 880.625021 CFR 880.6250Same
Regulatory ClassIISame
Regulation NamePatient Examination GlovePatient Examination GloveSame
Indications for useA powder-free patientexamination glove is a disposabledevice intended for medicalpurposes that is worn on theexaminer's hand to preventcontamination between patientand examinerA powder-free patientexamination glove is a disposabledevice intended for medicalpurposes that is worn on theexaminer's hand to preventcontamination between patientand examinerSame
MaterialCompositionSynthetic nitrile rubberPolychloropreneDifferent
CoatingPolyacrylic polymer inner coatingto aid donningPolyacrylic polymer inner coatingto aid donningSame
DesignNon-sterileNon-sterileSame
Single useSingle useSame
Powder-freePowder-freeSame
AmbidextrousAmbidextrousSame
Beaded cuffBeaded cuffSame
ColorBlue, Green, and BlackBlueDifferent
Performance
a. DimensionsMeets ASTM D6319-10 requirementsMeets ASTM D6977-19 requirementsSimilar
b. Physical PropertiesMeets ASTM D6319-10 requirementsMeets ASTM D6977-19 requirementsSimilar
c. Freedom from holesMeets ASTM D6319-10 requirements of GI, AQL 2.5Meets ASTM D6977-19 requirements of GI, AQL 2.5Similar
d. Powder ResidualMeets ASTM D6319-10 requirements; Not more than 2.0mg/gloveMeets ASTM D6977-19 requirements; Not more than 2.0mg/gloveSimilar
e. SterilityNon-sterileNon-sterileSame
OzoneAvoid OzoneAvoid OzoneSame
BiocompatibilityPasses Primary Skin Irritation Test and Dermal Sensitization Test and Acute Systemic Toxicity Test and In Vitro Cytotoxicity TestPasses Primary Skin Irritation Test and Dermal Sensitization Test and Acute Systemic Toxicity Test and In Vitro Cytotoxicity TestSame
Chemotherapy ClaimMicroflex® Nitrile Patient Examination Gloves Blue Colored Tested for Use with Chemotherapy Drugs and Fentanyl CitrateA powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drug has an extremely low permeation time: Carmustine: 47.9 minutes. Warning; Do not use with Carmustine.Tested chemotherapy drugs/concentration & Average Minimum Breakthrough Detection Time (Minutes) are as follows:Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy DrugsA powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The glove was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment for Medical Gloves to Permeation by Chemotherapy Drugs. Please note that the following drugs have extremely low permeation times: Carmustine: 45.4 minutes and Thiotepa: 23.5 minutes. Warning; Do not use with Carmustine or Thiotepa.Tested chemotherapy drugs/concentration & Average Minimum Breakthrough Detection Time (Minutes) are as follows:Different
Carmustine (3.3 mg/ml) $47.9$Carmustine (3.3 mg/ml) $45.4$Different
Cyclophosphamide (20.0 mg/ml) $>240$Cyclophosphamide (20.0 mg/ml) $>240$Same
Doxorubicin HCI (2.0 mg/ml) $>240$Doxorubicin HCI (2.0 mg/ml) $>240$Same
Etoposide (20.0 mg/ml) $>240$Etoposide (20.0 mg/ml) $>240$Same
Fluorouracil (50.0 mg/ml)>240Fluorouracil (50.0 mg/ml)>240Same
Methotrexate (25.0 mg/ml)>240Methotrexate (25.0 mg/ml)>240Same
Paclitaxel (6.0 mg/ml)>240Paclitaxel (6.0 mg/ml)>240Same
Thiotepa (10.0 mg/ml)>240Thiotepa (10.0 mg/ml)23.5Different
Vincristine Sulfate (1.0 mg/ml)>240Vincristine Sulfate (1.0 mg/ml)>240Same
Cisplatin (3.3 mg/ml)>240Different
Dacarbazine (10.0 mg/ml)>240Different
Tested hazardous drug & AverageMinimum Breakthrough Detection Time(Minutes) are as follows:Different
Fentanyl Citrate Injection(100 mcg/2 ml)>240

{6}------------------------------------------------

{7}------------------------------------------------

The subject device meets the applicable requirements for patient examination gloves regarding dimensions and sizes, physical properties, freedom from holes, and powder residues as found in the following standards: ASTM D6977, ASTM D5151 and ASTM D6124.

Non-Clinical Testing:

Non-Sterile Powder-Free Polychloroprene Examination Glove (Blue) Tested for Use with Chemotherapy Drugs has the following technological characteristics as compared to ASTM or equivalent standards:

Test MethodologyPurposeAcceptance CriteriaResults
Physical Characteristics:
Dimensions:Standard Specificationfor PolychloropreneExamination Gloves forMedical ApplicationASTM D6977-19Meets ASTM D6977-19requirements for length, width andthickness
LengthMinimum 230mmMinimum 240mm
Palm width (mm)
Size — XS$70 \pm 10$$75 \pm 5$
Size — S$80 \pm 10$$85 \pm 5$

{8}------------------------------------------------

Size - M$95 \pm 10$$95 \pm 5$
Size - L$110\pm 10$$105 \pm 5$
Size - XL$120 \pm 10$$115 \pm 5$
Thickness (mm) - single-wall
Fingerminimum 0.05Finger - min 0.09
Palmminimum 0.05Palm - min 0.06
Cuff-Cuff - min 0.05
Physical Properties:Standard Specificationfor PolychloropreneExamination Gloves forMedical ApplicationMeets ASTM D6977-19requirements for tensile strengthand ultimate elongation before andafter accelerated aging:
Tensile Strength
Before Agingminimum 14 MPaminimum 16 MPa
After Agingminimum 14 MPaminimum 14 MPa
Ultimate Elongation
Before Agingminimum 500%minimum 500%
After Agingminimum 400%minimum 400%
Freedom from holesStandard Test Methodfor Detection of Holesin Medical GlovesASTM D6977-19ASTM D5151-06Meets or exceeds ASTM D6977-19and ASTM D5151-06 requirementsof AQL 2.5
Powder ResidualStandard Test Methodfor Residual Powder onMedical GlovesASTM D6977-19ASTM D6124-06Meets applicable requirement forpowder free; ≤ 2 mg per glove
Biocompatibility:
ISO Skin IrritationStudyBiological Evaluation ofMedical Devices – Part10: Tests for Irritationand Skin SensitizationISO10993-10:2010Under the conditions of the study,not an irritant
ISO MaximizationSensitization StudyBiological Evaluation ofMedical Devices – Part10: Tests for Irritationand Skin SensitizationISO 10993-10:2010Under the conditions of the study,not a sensitizer
ISO Acute SystemicToxicityBiological Evaluation ofMedical Devices – Part11: Tests for SystemicToxicityISO 10993-11: 2006Under the conditions of the study,no evidence of systemic toxicity
ISO In VitroCytotoxicityBiological Evaluation ofMedical Devices – Part5: Tests for In VitroCytotoxicityISO 10993-5:2009Under the conditions of the studyundiluted, 1:2, and 1:4 dilutionswere cytotoxic (grade 4). Dilutionsof 1:8 (grade 2), 1:16 (grade 1),1:32 and 1:64 (grade 0) were non-cytotoxic

{9}------------------------------------------------

The subject device passes biological reactivity testing for skin irritation, dermal sensitization, acute systemic toxicity, and in vitro cytotoxicity; in accordance with the ISO 10993 series of standards.

A clinical study was not required for the subject or predicate device.

Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.