K Number
K211927

Validate with FDA (Live)

Date Cleared
2021-10-12

(112 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The nitrile examination glove is intended to be worn on the hands of examiner's to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

Device Description

The propose device (Mode: NG100) is powder free nitrile examination gloves, provided as non-sterile and disposable device. The proposed devices are blue color and there are five sizes, includes XS (6.5"), S (7"), M (8"), L (8.5"), XL (9") for optional. The gloves are provided with blue color. The examination glove is smooth surface with textured fingertips and a rolled rim at the cuff edge.

The gloves are manufactured in accordance with the requirements of ASTM D6319-19 and Medical Glove Guidance Manual.

AI/ML Overview

This document describes the FDA 510(k) clearance for "Nitrile examination gloves" (K211927). The clearance is based on substantial equivalence to a predicate device (K203593). The device is a non-sterile, powder-free, single-use examination glove made of nitrile rubber.

The key acceptance criteria for this device are defined by a set of ASTM and ISO standards, which dictate the physical properties, freedom from holes, powder residue, and biocompatibility of the gloves.

Here's the breakdown of the information requested:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaStandardReported Device Performance (K211927)Predicate Device Performance (K203593)
Dimensions (Palm Width)ASTM D6319-19XS (70±10mm)S (80±10mm)M (95±10mm)L (110±10mm)XL (120±10mm)Small (80±10mm)Medium (95±10mm)Large (110±10mm)X large (120±10mm)
Dimensions (Length)ASTM D6319-19XS (220mm min)S (220mm min)M (230mm min)L (230mm min)XL (230mm min)S (220mm min)M (230mm min)L (230mm min)XL (230mm min)
ThicknessASTM D6319-19Palm: 0.05mm minFinger: 0.05mm minPalm: 0.05mm minFinger: 0.05mm min
Freedom from holesASTM D5151-06Free from holes when tested (Pass)Free from holes when tested (Pass)
Physical Properties (before aging)ASTM D6319-19Tensile Strength: 14MPa, minUltimate Elongation: 500% minTensile Strength: 14MPa, minUltimate Elongation: 500% min
Physical Properties (after aging)ASTM D6319-19Tensile Strength: 14MPa, minUltimate Elongation: 400% minTensile Strength: 14MPa, minUltimate Elongation: 400% min
Powder ResidualASTM D6124≤2.0 mg/gloves≤2.0 mg/gloves (implied by "Same" discussion)
Biocompatibility - Skin Sensitization TestISO 10993-10Not a sensitizer (Pass)Not a sensitizer (Pass)
Biocompatibility - Skin Irritation TestISO 10993-10Not an irritant (Pass)Not an irritant (Pass)
Biocompatibility - Cytotoxicity TestISO 10993-5Non-cytotoxic to L-929 cells (Pass)Cytotoxicity assessed via rationale (Pass)
Shelf-lifeNot specified in table3 years3 years

2. Sample size used for the test set and the data provenance

The document does not explicitly state the sample sizes for the individual non-clinical tests (e.g., number of gloves tested for tensile strength, number of samples for biocompatibility). The studies are non-clinical (laboratory testing) and based on internationally recognized standards (ASTM, ISO). The data provenance is from the manufacturing site in Guangdong, China, as indicated by the submitter's address. The tests are prospective as they were performed to verify the device met specifications.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This is not applicable as the "ground truth" for these tests isn't established by human experts in the sense of clinical diagnosis (e.g., radiologists). Instead, the performance is measured against objective, validated scientific standards (ASTM, ISO) and laboratory methods. The "experts" would be the qualified laboratory personnel performing the standardized tests, but their number and specific qualifications are not detailed in this submission.

4. Adjudication method for the test set

This is not applicable. The device performance is determined by meeting objective criteria defined in the referenced ASTM and ISO standards, not by an adjudication process typically used for human-reviewed data or clinical outcomes.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. Nitrile examination gloves are a physical medical device, not a diagnostic AI/imaging product. Therefore, no MRMC study, human reader improvement, or AI assistance is relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is a physical glove, not an algorithm or AI.

7. The type of ground truth used

The "ground truth" for this device is based on objective, standardized measurements and laboratory tests defined by ASTM and ISO standards. This includes:

  • Physical measurements: Dimensions (palm width, length, thickness) measured according to ASTM D6319-19 and ASTM D3767-03(2020).
  • Performance testing: Freedom from holes (ASTM D5151-19), tensile strength and ultimate elongation before and after aging (ASTM D6319-19, ASTM D412-16, ASTM D573-04(2019)), and powder residual (ASTM D6124-06(2017)).
  • Biocompatibility testing: Skin sensitization, skin irritation (ISO 10993-10: 2010), and in vitro cytotoxicity (ISO 10993-5: 2009).

8. The sample size for the training set

This is not applicable. There is no "training set" for physical medical devices like examination gloves. The testing is for validation of the final product specifications against established standards.

9. How the ground truth for the training set was established

This is not applicable. As there is no training set for this type of device, the concept of establishing ground truth for a training set does not apply.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".

October 12, 2021

Fitone Latex Products Co., Ltd. Guangdong % Stuart Situ Director Landlink Healthcare Technology (Shanghai) Co., Ltd. Room 1308, Baohua International Plaza, 555 West Guangzho Shanghai. 200072 China

Re: K211927

Trade/Device Name: Nitrile examination gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 7, 2021 Received: September 7, 2021

Dear Stuart Situ:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211927

Device Name Nitrile examination gloves

Indications for Use (Describe)

The nitrile examination glove is intended to be worn on the hands of examiner's to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

l. Submitter

Fitone Latex Products Co., Ltd. Guangdong No.5 Huitong road, Lingbei Industrial Zone, Suixi, 524338 Zhanjiang, Guangdong, China

Contact person: Christine Ou Position: Manager Tel.: 0759-7905808 E-mail: market-intl@fitonelatex.com

Preparation date: Jun.11, 2021

US Agent

Qihui Zhang ZYPPEL LLC 1337 Massachusetts Avenue #158 Arlington MA, MA US 02476

II. Proposed Device

Device Trade NameNitrile examination gloves
ModelNG100
Common name:Polymer Patient Examination Glove
Regulation Number:21 CFR 880.6250
Regulatory Class:Class I
Product code:LZA
Review PanelGeneral Hospital

III. Predicate Devices

510(k) Number:K203593
Trade name:Nitrile examination gloves
Common name:Polymer Patient Examination Glove
Classification:Class I
Product Code:LZA
ManufacturerGUANGDONG KINGFA SCI.&TECH.CO.,LTD.

IV. Device description

{4}------------------------------------------------

The propose device (Mode: NG100) is powder free nitrile examination gloves, provided as non-sterile and disposable device. The proposed devices are blue color and there are five sizes, includes XS (6.5"), S (7"), M (8"), L (8.5"), XL (9") for optional. The gloves are provided with blue color. The examination glove is smooth surface with textured fingertips and a rolled rim at the cuff edge.

The gloves are manufactured in accordance with the requirements of ASTM D6319-19 and Medical Glove Guidance Manual.

V. Indication for use

The nitrile examination glove is intended to be worn on the hands of examiner's to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

VI. Comparison of technological characteristics with the predicate devices

ItemStandardProposed device(K211927)Predicate device(K203593)Discussion
Product name/Nitrile ExaminationGlovesNitrile ExaminationGloves-
Product Code/LZALZASame
Regulation No./21 CFR 880.625021 CFR 880.6250Same
Classification/Class IClass ISame
Powder free/YesYesSame
Indication foruse/The nitrileexamination glove isintended to be wornon the hands ofexaminer's to preventcontaminationbetween patient andexaminer. This is asingle-use,powder-free,non-sterile device.The nitrileexamination glove isintended to be wornon the hands ofexaminer's to preventcontaminationbetween patient andexaminer. This is asingle-use,powder-free,non-sterile device.Same
Main Material/Nitrile rubberNitrile rubberSame
Color/BlueBlueSimilar*
Size/X-small, Medium, largeSmall, Large, X largeSimilar Dimensions meet ASTM D6319-19
Palm widthASTM D6319-19XS (70±10mm)S (80±10mm)M (95±10mm)L (110±10mm)XL (120±10mm)Small (80±10mm)Medium (95±10mm)Large (110±10mm)X large (120±10mm)Similar Dimensions meet ASTM D6319-19
LengthASTM D6319-19XS (220mm min)S (220mm min)M (230mm min)L (230mm min)XL (230mm min)S (220mm min)M (230mm min)L (230mm min)XL (230mm min)Similar Dimensions meet ASTM D6319-19
ThicknessASTM D6319-19Palm: 0.05mm minFinger: 0.05mm minPalm: 0.05mm minFinger: 0.05mm minSame
Freedom from holesASTM D5151-06The gloves shall be free from hole when tested in accordance with the method given in ASTM D5151The gloves shall be free from hole when tested in accordance with the method given in ASTM D5151Same
Physical Properties (before aging)ASTM D6319-19Tensile Strength:14MPa, minUltimate Elongation:500% minTensile Strength:14MPa, minUltimate Elongation:500% minSame
Physical Properties (after aging)ASTM D6319-19Tensile Strength:14MPa, minUltimate Elongation:400% minTensile Strength:14MPa, minUltimate Elongation:400% minSame
Powder residualASTM D6124≤2.0 mg/glovesSame
Sterility/Non-sterileNon-sterileSame
For single use/YesYesSame
Type of use/Over the counter useOver the counter useSame
Shelf-life/3 years3 yearsThe shelf-life testing was
performedthatdemonstratemeet theclaimedshelf-life.
Biocompatibility- SkinSensitizationTestISO10993-10Under the testcondition of study notan sensitizerUnder the testcondition of study notan sensitizerSame
Biocompatibility- Skin IrritationTestISO10993-10Under the testcondition of study notan irritantUnder the testcondition of study notan irritantSame
Biocompatibility- CytotoxicityTestISO10993-5Under the testconditions, the testarticle wasnon-cytotoxic to L-929cells.Cytotoxicity isassessed viarationale. Under thecondition of acutesystemic toxicity test,the test article did notshow acute systemictoxicity in vivo.Different

Table 1 Comparison of Nitrile examination gloves

{5}------------------------------------------------

{6}------------------------------------------------

*As above comparison, the difference in the dimensions and reference standard version of the subject and predicate device does not raise additional questions for safety and effectiveness of the device. The biocompatibility test and performance test of the subject devices have been performed on the final finished device. The test results shows pass the requirements.

VII. Non-Clinical Testing

Non clinical tests were conducted in accordance with following standards to verify that the proposed device met all design specifications.

  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for . Medical Application
  • ASTM D3767-03(2020), Practice for rubber-Measurement of Dimensions ●
  • ASTM D5151-19,Test Method for Detection of Holes in Medical Gloves ●
  • ASTM D6124-06(2017), Test Method for Residual Powder on Medical Gloves ●
  • ASTM D573-04(2019), Test Method for Rubber-Deterioration in an Air Oven ●
  • ASTM D412-16, Test Methods for Vulcanized Rubber and Thermoplastic ● Elastomers-Tension

{7}------------------------------------------------

  • ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For ● In Vitro Cytotoxicity
  • . ISO 10993-10: 2010, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

VIII. Clinical Testing

No clinical study is included in this submission.

IX. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Nitrile Patient Examination Glove is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K203593.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.