K Number
K211911

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-03-10

(262 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For short-term (up to 30 days) percutaneous drainage of the upper urinary tract in adult patients.

Device Description

The Kolibri® Percutaneous Nephrostomy Catheters are available in two configurations based on the tip configuration termed either J or Malecot. The distal end of the J catheter resembles a loop, and the distal end of the Malecot catheter has 4-wings which form an open basket to hold the catheter in place. J and Malecot catheters are single lumen catheters with outer diameters ranging from 6 Fr to 20 Fr. The J and Malecot catheters are manufactured from the material branded Vortek by Coloplast. The catheters are supplied either individually or as part of a kit. The following components are included depending on the specific configuration:

  • · Stylet
  • · Silicone sleeve
  • · Stopcock
  • Female Luer
  • Urine bag connector
  • Guidewire (kits only)
  • · Dilator set(s) (kits only)
  • · Chiba needles (kits only)
    The dilators. Chiba Needles, and quidewires are all packaged separately. The J and Malecot catheters are supplied sterile via ethylene oxide for single use.

Kolibri Direct Puncture Set
The Direct Puncture Set consists of an 8 Fr or 10 Fr J catheter with a silicone sleeve, stopcock, urine bag connector, transparent reinforced tube, stainless-steel trocar needle, and stainlesssteel stylet. The Direct Puncture Set is supplied sterile via ethylene oxide for single use.

AI/ML Overview

The provided text is a 510(k) summary for the Kolibri® Percutaneous Nephrostomy Catheters, Kolibri® Percutaneous Nephrostomy Kits, and Kolibri® Direct Puncture Sets. This document is a premarket notification to the FDA, demonstrating that the new device is substantially equivalent to a legally marketed predicate device.

However, the document does not describe an acceptance criteria table or a study that proves the device meets specific performance acceptance criteria related to a diagnostic or AI-driven medical device. Instead, it outlines the device's technical characteristics, its comparison to a predicate device, and a list of biocompatibility, sterilization, and bench tests performed to demonstrate substantial equivalence for a mechanical catheter device.

Therefore, I cannot extract the information required for your request in the format you specified, as the input document does not contain data related to a clinical study or performance metrics for a diagnostic or AI-based device. The device described is a medical instrument (catheter), and the "performance data" refers to its physical and material properties, not diagnostic accuracy, sensitivity, specificity, or AI performance.

If you have a document that describes the acceptance criteria and study for a diagnostic or AI-driven medical device, please provide that, and I will do my best to extract the requested information.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 10, 2022

Coloplast Corp Angela Kilian Head of Regulatory, North America 1601 West River Road North Minneapolis, MN 55411

Re: K211911

Trade/Device Name: Kolibri® Percutaneous Nephrostomy Catheters, Kolibri® Percutaneous Nephrostomy Kits, Kolibri® Direct Puncture Sets Regulatory Class: Unclassified Product Code: LJE Dated: February 11, 2022 Received: February 11, 2022

Dear Angela Kilian:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211911

Device Name

Kolibri® Percutaneous Nephrostomy Catheters, Kolibri® Percutaneous Nephrostomy Kits , Kolibri® Direct Puncture Sets

Indications for Use (Describe)

For short-term (up to 30 days) percutaneous drainage of the upper urinary tract in adult patients.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (k) Summary

1. Submitter Information

510 (k) submitterColoplast A/S
AddressHoltedam 13050 HumlebaekDenmark
Contact PersonAngela KilianHead of Regulatory Affairs, North America1601 West River Road, Minneapolis, MN 55411Office : 612-422-7956Email : usaby@coloplast.com

Preparation date March 09, 2022

2. Device Name

Trade Name of the DeviceKolibri® Percutaneous Nephrostomy Catheters
Kolibri® Percutaneous Nephrostomy Kits
Kolibri® Direct Puncture Sets
Common NamePercutaneous Nephrostomy Catheter
Classification NameUnclassified
Classification RegulationUnclassified
Device ClassUnclassified
PanelGastroenterology/Urology
Product CodeLJE

3. Predicate and Reference Devices

Predicate DeviceReference Device
OwnerCook IncorporatedColoplast Corp
Trade Name of theDeviceUniversa Percutaneous DrainageCatheterVortek Double LoopUreteral Stent
Classification NameUnclassifiedStent, Ureteral
RegulationClassificationUnclassified21 CFR 876.4620
Device ClassUnclassifiedII
PanelGastroenterology/UrologyGastroenterology/Urology
Product CodeFEW, KOB, LJEFAD
510(k) NumberK140085K180057

{4}------------------------------------------------

4. Device Description

Kolibri® Percutaneous Nephrostomy Catheters and Kits

The Kolibri® Percutaneous Nephrostomy Catheters are available in two configurations based on the tip configuration termed either J or Malecot. The distal end of the J catheter resembles a loop, and the distal end of the Malecot catheter has 4-wings which form an open basket to hold the catheter in place. J and Malecot catheters are single lumen catheters with outer diameters ranging from 6 Fr to 20 Fr. The J and Malecot catheters are manufactured from the material branded Vortek by Coloplast. The catheters are supplied either individually or as part of a kit. The following components are included depending on the specific configuration:

  • · Stylet
  • · Silicone sleeve
  • · Stopcock
  • Female Luer
  • Urine bag connector
  • Guidewire (kits only)
  • · Dilator set(s) (kits only)
  • · Chiba needles (kits only)

The dilators. Chiba Needles, and quidewires are all packaged separately. The J and Malecot catheters are supplied sterile via ethylene oxide for single use.

Kolibri Direct Puncture Set

The Direct Puncture Set consists of an 8 Fr or 10 Fr J catheter with a silicone sleeve, stopcock, urine bag connector, transparent reinforced tube, stainless-steel trocar needle, and stainlesssteel stylet. The Direct Puncture Set is supplied sterile via ethylene oxide for single use.

5. Indication For Use

For short-term (up to 30 days) percutaneous drainage of the upper urinary tract in adult patients.

6. Comparison of the Technological Characteristics with Predicate and Reference Devices

Device &PredicateDevice(s):K211911K140085 (PredicateDevice)K180057(ReferenceDevice)
General DeviceCharacteristics
Indication ForUseFor short-term (up to 30days) percutaneousdrainage of the upperurinary tract in adultpatientsThe UniversaLoop/Malecot DrainageCatheter Set is intendedto provide percutaneousurine drainage from thegenitourinary system.The double loop ureteralstents are used for:drainage of the upperurinary tract over fistulasor ureteral obstacles andfor healing of the ureterSimilar to thepredicate
Device TypeCatheter, NephrostomyCatheter, NephrostomyStent, UreteralSame as the

{5}------------------------------------------------

Product Codes: LJEProduct Codes: LJEProduct Code: FADpredicateDevice
DeviceComponentsKolibri® PercutaneousNephrostomy Catheterincludes either a J orMalecot catheter, andsome combination ofsilicone sleeve,stopcock, Luer, urinebag connectorand stylet in somemodels.The UniversaPercutaneous DrainageCatheter Sets availableas a loop catheter or aMalecot catheter.Suprapubic sets includesdrainage catheter,connecting tube,straightening stylets,trocar needle withobturator, hollow needle,wire guide, dilators,silicone retention discwith pull tie, or one-waystopcock.The Vortek Double LoopUreteral Stents aresupplied in kits, whichcontain a double-loopureteral stent, asteerable pusher or anon-steerable pusherwith clamp and aguidewire.Similar to thepredicate
Catheter tubeMaterialPolyether block amidePolyurethanePolyether block amideDifferent fromthe predicatebut similar toreferencedevice
Length29 - 36 cm15 - 30 cm12 - 30 cmDifferent fromthe predicate
CatheterFrench Size06 Fr. to 14 Fr. (J)8 Fr. to 20 Fr.(Malecot)06 to 14 Fr. (Loop)08 to 24 Fr. (Malecot)CHF 4.8,6,7,8 FR.Similar to thepredicate
TipConfigurationJ and MalecotLoop and MalecotDouble loopSimilar to thepredicate
Duration of Useup to 30 days andcumulative use up to 90days4 weeks6 monthsSimilar to thepredicate
IntendedpatientPopulationAdults onlyAdults onlyAdults and pediatric(infants, andadolescents) populationSimilar to thepredicate

As evidenced by the above table, both the subject and the predicate devices have the same intended use, but the subject and predicate devices have different technological characteristics. However, performance testing was conducted on the subject catheters, and it was established that the differences in technological characteristics between the subject and the predicate does not raise different questions of safety or effectiveness.

7. Performance Data

Below is a list of the tests that have been performed and successfully completed for the subject catheters:

  • Biocompatibility testing according to ISO 10993-1:2018 and FDA Guidance "Use of ● International Standard ISO 10993-1" (2016)
  • Ethylene oxide residual testing according to ISO 10993-7 ●
  • EO sterilization according to ISO 11135:2014 ●
  • Bacterial Endotoxins testing according to ANSI/AAMI ST72:2019 and USP <85>

The following list of bench testing was conducted on the subject catheters using established methods to determine the substantial equivalence with the predicate:

{6}------------------------------------------------

  • Visual inspection
  • Compatibility Testing .
  • Dimensional verification ●
  • Flow Rate ●
  • Loop strength
  • Tensile testing including tip, body, loop, and connector/body junction. .
  • . Simulated use
  • Implant simulation
  • Shelf Life/Expiration date ●

All predetermined acceptance criteria were met.

8. Conclusion

Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate.

N/A