K Number
K180057

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-03-09

(60 days)

Product Code
Regulation Number
876.4620
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The double loop ureteral stents are used for:

  • drainage of the upper urinary tract over fistulas or ureteral obstacles
  • healing of the ureter
Device Description

Vortek® Double Loop Ureteral Stents in this submission and both predicates. Vortek® Hydro-Coated Double Loop Ureteral Stents and Vortek® Double Loop Ureteral Stents, are implantable autostatic catheters used to maintain urine drainage and to allow for ureteral healing. They are inserted during a surgical procedure using either an antegrade or a retrograde technique or by open surgery typically using a guidewire and a pusher.
Stents are supplied in diameters between 4.8 and 8 Fr and lengths between 12 and 30 cm long. The stents can have either both tips open (0/0) or the vesical tip open and renal tip closed (0/c).
Vortek® Double Loop Ureteral Stents in this submission and the predicates are supplied in kits which include the following components:

  • A double-loop ureteral stent
  • A pusher
  • A guide wire
    And in some kits:
  • A ureteric catheter
    The Vortek® Double Loop Ureteral Stent and accessories included in the kits are supplied sterile via ethylene oxide. The ureteral stent and each accessory are packaged and sterilized separately prior to being combined in the kit.
AI/ML Overview

This document is a 510(k) Pre-Market Notification from the FDA for a medical device called "Vortek® Double Loop Ureteral Stents." It describes the device, its intended use, comparison to predicate devices, and the testing conducted to demonstrate substantial equivalence.

However, the provided text does not contain information about an AI/ML-based medical device or its acceptance criteria, performance studies, or ground truth establishment. The document focuses on a traditional medical device (a ureteral stent) and its non-clinical performance data (biocompatibility, mechanical testing, sterilization, packaging).

Therefore, I cannot extract the requested information regarding AI/ML device acceptance criteria and study details from this document. The questions you've asked are relevant for AI/ML device submissions, but this specific FDA document does not pertain to such a device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left, and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

March 9, 2018

Coloplast Corp Cori L. Ragan Regulatory Affairs Manager 1601 West River Road North Minneapolis, MN 55411

Re: K180057

Trade/Device Name: Vortek® Double Loop Ureteral Stents Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: January 5, 2018 Received: January 8, 2018

Dear Cori L. Ragan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, sans-serif font. The text is black and appears to be centered on a white background. The letters are clear and easy to read, with consistent spacing between them.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K180057

Device Name

Vortek® Double Loop Ureteral Stents

Indications for Use (Describe)

The double loop ureteral stents are used for:

  • drainage of the upper urinary tract over fistulas or ureteral obstacles
  • healing of the ureter

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2020

See PRA Statement below.

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

7. 510(k) SUMMARY

I.SUBMITTER510(K) Owner's Name:Coloplast A/S
Legal Manufacturer Address:Holtedam 13050 Humlebaek, Denmark
Phone/Fax/Email:Phone: (612) 597-5106Email: usclr@coloplast.com
Name of Contact Person:Cori L. RaganRegulatory Affairs Manager
Address/Contact:1601 West River RoadMinneapolis, MN 55411
Date Prepared:5 January 2018
II.DEVICETrade or Proprietary Name:Vortek® Double Loop Ureteral Stents
Common or Usual Name:Vortek® Double Loop Ureteral Stents
Classification Name:Stent, Ureteral(21CFR section 876.4620)Product Code: FADDevice Class: 2

III. PREDICATE DEVICE

Vortek® Hydro-Coated Double Loop Ureteral Stents (K170362) and Vortek® Double Loop Ureteral Stents (K881744 and K981591).

This predicate has not been subject to a design-related recall. No reference devices were used in this submission.

DEVICE DESCRIPTION IV.

Vortek® Double Loop Ureteral Stents in this submission and both predicates. Vortek® Hydro-Coated Double Loop Ureteral Stents and Vortek® Double Loop Ureteral Stents, are implantable autostatic catheters used to maintain urine drainage and to allow for ureteral healing. They are inserted during a surgical procedure using either an antegrade or a retrograde technique or by open surgery typically using a guidewire and a pusher.

Stents are supplied in diameters between 4.8 and 8 Fr and lengths between 12 and 30 cm long. The stents can have either both tips open (0/0) or the vesical tip open and renal tip closed (0/c).

{4}------------------------------------------------

Vortek® Double Loop Ureteral Stents in this submission and the predicates are supplied in kits which include the following components:

  • A double-loop ureteral stent ।
  • । A pusher
  • A guide wire ।

And in some kits:

  • | A ureteric catheter
    The Vortek® Double Loop Ureteral Stent and accessories included in the kits are supplied sterile via ethylene oxide. The ureteral stent and each accessory are packaged and sterilized separately prior to being combined in the kit.

V. INDICATIONS FOR USE

Vortek® Double Loop Ureteral Stents in this submission and the predicates Vortek® Hydro-Coated Double Loop Ureteral Stents and Vortek® Double Loop Ureteral Stents have the same intended use:

  • Drainage of the upper urinary tract over fistulas or ureteral obstacles
  • Healing of the ureter ●

Both the subject and predicate device are intended to be used in both adult and pediatric populations and remain in place for up to six months.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Vortek® Double Loop Ureteral Stents are substantially equivalent in performance, design and materials to the Vortek® Hydro-Coated Double Loop Ureteral Stents, cleared under premarket notification number K170362 and Vortek® Double Loop Ureteral Stents cleared under K881744 and K981591. All devices are tubes made from the same radiopaque, polymeric materials with loops at both the renal and vesical ends to hold the stent in place. Both devices use guidewires and pushers or direct surgical placement to be implanted for up to six months. All of the devices are sold as kits with the same accessories intended to facilitate implantation. The only difference between the subject Vortek® Double Loop Ureteral Stent and the primary predicate, Vortek® Hydro-Coated Double Loop Ureteral Stent is that the predicate device includes a lubricious coating to decrease friction during the implant procedure.

PERFORMANCE DATA VII.

The following performance data was provided in support of the substantial equivalence determination.

{5}------------------------------------------------

Biocompatibility Testing

Biocompatibility testing was conducted based upon ISO 10993-1 (2009): Biological evaluation of medical devices – Part 1: "Evaluation and testing within a risk management process" and FDA Guidance for Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" - Guidance for Industry and Food and Drug Administration Staff - June 16, 2016. A comprehensive regimen of testing for both the Vortek® Double Loop Ureteral Stent and the accessories included:

  • Cytotoxicity ●
  • Sensitization
  • Irritation ●
  • Acute Systemic toxicity ●
  • Material-mediated pyrogenicity
  • Sub-chronic toxicity
  • Genotoxicity
  • . Implantation
  • Chronic Toxicity ●
  • Extractable and Leachable evaluation ●
  • Toxicological Risk evaluation ●

Mechanical Testing

Mechanical testing was completed using the FDA guidance document "Guidance for the content of premarket notifications for ureteral stents" for reference:

  • Flow Rate
  • Elongation / Yield and tensile strength
  • Curl Strength
  • Shelf life evaluation
  • Visual and dimensional testing ●
  • . Radiopacity testing
  • Suture testing
  • Magnetic resonance compatibility testing ●

Sterilization

The Vortek® Double Loop Ureteral Stents and accessories are sterilized using ethylene oxide in a validated cycle demonstrating a microbial assurance level of 10-6.

Packaging and Distribution

The Vortek® Double Loop Ureteral Stent was subjected to distribution testing and verification testing to demonstrate that the product and package would be undamaged throughout the product life and maintain the device sterility.

{6}------------------------------------------------

No animal studies or clinical testing were provided to support substantial equivalence between the subject and predicate devices.

VIII. CONCLUSIONS

The Vortek® Double Loop Ureteral stent and related accessories have been demonstrated to be substantially equivalent to the two predicates, Vortek® Hydro-Coated Double Loop Ureteral Stents and the previously submitted Vortek® Double Loop Ureteral stent based on the nonclinical data provided, the same intended use, patient population, implant duration, materials, biocompatibility, kit composition, and technological characteristics.

§ 876.4620 Ureteral stent.

(a)
Identification. A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.(b)
Classification. Class II (performance standards).