K Number
K211808

Validate with FDA (Live)

Date Cleared
2022-03-04

(266 days)

Product Code
Regulation Number
880.6250
Age Range
6 - 12
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The powder free patient examination gloves are disposable devices intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner. The device is for over the counter use.

Device Description

Medcare Powder-Free Blue Nitrile Patient Examination Gloves are non-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between a patient and an examiner. The gloves are nitrile, powder- free, ambidextrous, and blue-colored with a beaded cuff. The proposed device is offered in the following sizes: S, M, L, XL.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Medcare Powder-Free Blue Nitrile Patient Examination Gloves (K211808) based on the provided text, formatted to your request:

1. Table of Acceptance Criteria and Reported Device Performance

Test Method (Acceptance Criteria Defined By)PurposeAcceptance Criteria Reported asReported Device Performance
ASTM D6319-19Length of the glovesMin 230 mm for all sizes (±5mm)Small: Pass, Medium: Pass, Large: Pass, X-Large: Pass
ASTM D6319-19Width of the glovesSmall: 85mm, Medium: 95mm, Large: 105mm, X-Large: 115mmSmall: Pass, Medium: Pass, Large: Pass, X-Large: Pass
ASTM D6319-19Thickness of the glovesPalm 0.05 mm min, Finger 0.05 mm min for all sizesPalm: Pass for all sizes, Finger: Pass for all sizes
ASTM D6319-19Physical properties - Tensile Strength (Before Aging)14 Mpa Min for all sizesSmall: Pass, Medium: Pass, Large: Pass, X-large: Pass
ASTM D6319-19Physical properties - Tensile Strength (After Aging)14 Mpa Min for all sizes (at 70oC for 168 hrs or 100oC for 22 hrs)Small: Pass, Medium: Pass, Large: Pass, X-large: Pass
ASTM D6319-19Physical properties - Ultimate Elongation (Before Aging)500% Min for all SizeSmall: Pass, Medium: Pass, Large: Pass, X-large: Pass
ASTM D6319-19Physical properties - Ultimate Elongation (After Aging)400% Min for all sizesSmall: Pass, Medium: Pass, Large: Pass, X-large: Pass
ASTM D5151-19Detection of holes in medical glovesAQL 2.5Pass
ASTM D6124-06Residual powder in the gloves2 mg/Glove MaxSmall: Pass, Medium: Pass, Large: Pass, X-Large: Pass
ISO 10993-10:2010Dermal irritation in RabbitsUnder the condition of study not an irritantUnder the condition of study not an irritant
ISO 10993-10:2010Skin sensitization (Guinea Pig)Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
ISO 10993-5:2009In vitro cytotoxic potential (L-929 mouse fibroblasts)Under the conditions of study non cytotoxicUnder the conditions of the study non cytotoxic

Since the device is a Class I medical device (patient examination gloves), the studies described are non-clinical performance tests to demonstrate substantial equivalence to a predicate device, not typical clinical studies with human participants that would involve the following points you requested. Therefore, for most of the remaining points, the answer is "Not Applicable" or "Not Provided" in the context of this 510(k) submission.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample size: Not explicitly stated for each individual test, but the ASTM and ISO standards for material testing typically specify sample sizes.
  • Data provenance: Not explicitly stated whether retrospective or prospective. The manufacturer is Dongguan Grinvald Technology Co., Ltd. - China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. These are objective, quantitative material and biological tests performed according to standardized protocols (ASTM, ISO). "Ground truth" in the clinical sense with expert consensus is not relevant here. The "experts" would be the accredited laboratory personnel performing the tests according to the standard methods.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • None. Adjudication methods are typically used for subjective assessments or when there's disagreement among human readers in interpreting clinical data. These non-clinical tests have objective pass/fail criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is a non-clinical device (patient examination glove). AI assistance or human reader improvement studies are not relevant to its regulatory pathway.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Objective, standardized test results. The "ground truth" is established by the specified acceptance criteria within each ASTM or ISO standard, which are quantitative thresholds for physical properties and defined biological responses.

8. The sample size for the training set

  • Not Applicable. This device does not involve a training set in the machine learning context. The "training" for the device's manufacturing would be adherence to quality control procedures and standard operating procedures.

9. How the ground truth for the training set was established

  • Not Applicable. As above, no training set for an algorithm is involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".

March 4, 2022 Dongguan Grinvald Technology Co., Ltd. % Aristotle Nafpliotis Regulatory Affairs Consultant/Engineer mdi Consultants, Inc. 55 Northern Blvd Great Neck, New York 11021

Re: K211808

Trade/Device Name: Medcare Powder-Free Blue Nitrile Patient Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: January 31, 2022 Received: February 1, 2022

Dear Aristotle Nafpliotis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K211808

Device Name

Medcare Powder-Free Blue Nitrile Patient Examination Gloves

Indications for Use (Describe)

The powder free patient examination gloves are disposable devices intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner. The device is for over the counter use.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a stylized, curved checkmark symbol. The checkmark is rendered in a gradient of blue, with darker shades suggesting depth and a three-dimensional effect. The design is simple and modern, with smooth lines that create a sense of fluidity and motion.

Dongguan Grinvald Technology CO., LTD. - China

510(k) SUMMARY

The assigned 5l0(k) number is: K211808.

1. Submitter's Identification:

Date Summary Prepared: 4 March 2022

Dongguan Grinvald Technology Co.,Ltd. 401, Building3 No.4 of Guangming New Village 2 Road, Dongcheng , Dongguan City, Guangdong, China 523000

Contact: Dr. Anna Roxana Grinvald Email : anna@gts-hk.net Phone number 86 0769 23326470 Mobile phone no. 86 136 2261 7084

    1. Name of the Device: Medcare Powder-Free Blue Nitrile Patient Examination Gloves
      Model #: MD0120

Regulation Number: 21 CFR 880.6250

Regulation Name: Non-powdered patient examination glove

Regulatory Class: Class 1

Product Code: LZA

3. Information for the 510(k) Cleared Device (Predicate Device):

510(K) Number: K200326

Trade Name: Powder Free Nitrile Examination Glove (Aqua green)

Device Name: Nitrile Patient Examination Gloves

Device Classification Name: Patient Examination Gloves

Device Class: Class I

Product code LZA

{4}------------------------------------------------

4. Device Description:

Medcare Powder-Free Blue Nitrile Patient Examination Gloves are non-sterile, single use, disposable gloves intended for medical purposes to be worn on the hands of examiners to prevent contamination between a patient and an examiner. The gloves are nitrile, powder- free, ambidextrous, and blue-colored with a beaded cuff. The proposed device is offered in the following sizes:

Module numberSize
MD0120SS
MD0120MM
MD0120LL
MD0120XLXL

Indications for Use: 5.

The powder free patient examination gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner. The device is for over the counter use.

Comparison to the 510(k) Cleared Devices (Predicate Devices): 6.

Comparison Chart

ItemAcceptance CriteriaSubject DevicesMedcare Powder-FreeBlue Nitrile PatientExamination GlovesK211808Predicate DevicePowder Free NitrileExamination Glove(Aqua green)K200326Same,Similar orDifferent
Product CodeLZALZALZASame
IntendedUse/Indicationsfor Use---The powder free patientexamination gloves aredisposable devicesintended for medicalpurposes that are wornThe Powder FreeNitrile ExaminationGlove (Aqua Green)is a disposable deviceintended for MedicalSimilar*
on the examiner's handto preventcontamination betweenpatient and examiner.The device is for overthe counter use.purpose that is wornon the examiner'shands or finger topreventcontamination patientand examiner.
Color---BlueAqua GreenDifferent*
MaterialsASTMD6910-10(Reapproved2015)NitrileNitrileSame
Sterility---Non-sterileNon-sterileSame
Single Use---YesYesSame
Dimensions (mm)Overall LengthMin 230mm Width(±5mm)Size S = 85mmSize M= 95mmSize L = 105mmSize XL = 115mmThickness at PalmMin; 0.05 mmThickness at FingerTipMin 0.05 mmMeets ASTM D6319-10Meets ASTM D6319-10Same
PhysicalpropertiesBefore AgeingTensileStrength (MPa)Min 14minUltimate Elongation(%)= 500min After Agingat 70oC for 168 hrs@ 100oC for 22 hrsTensile Strength(MPa)Meets ASTM D6319-10Meets ASTM D6319-10Same
= 14minUltimate Elongation(%)= 400min
Freedom frompinholesAQL 2.5Inspection Level G-1Meets ASTM D5151-06Meets ASTM D5151-06Same
Residual Powder< 2.0 mg/pcMeets ASTM D6124-06Meets ASTM D6124-06Same
Biocompatibility
CytotoxicityISO 10993-5Tests for in vitrocytotoxicity ofmedical devicesNon-cytotoxicPassPassSame
IrritationISO 10993-10Tests for irritationand skinsensitizationNon-IrritatingPassPassSame
SensitizationISO 10993-10Tests for irritationand skinsensitizationNon-sensitizingPassPassSame

{5}------------------------------------------------

{6}------------------------------------------------

*The items marked different and similar only differ in color and grammatical structure. These differences do not affect the overall substantial equivalence comparison as the former can be shown in the biocompatibility reports.

7. Summary of Non-Clinical Tests Performed:

The following National and International Standards were utilized for testing the subject device:

• ASTM D6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application

{7}------------------------------------------------

• ASTM D6124-06 (Reaffirmation 2011) Standard Test Method for Residual Powder on Medical Gloves

• ASTM D5151-06 (Reapproved 2011) Standard Test Method for Detection of Holes in Medical Gloves

· ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

• ISO 10993-05:2009 Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity

Test MethodPurposeAcceptance CriteriaResult
ASTM D6319-19Standard Specificationfor Nitrile ExaminationGloves for MedicalApplicationTo determine thelength of the glovesMin 230 mm for allsizes (±5mm)Small: PassMedium: PassLarge: PassX-Large: Pass
ASTM D6319-19Standard Specificationfor Nitrile ExaminationGloves for MedicalApplicationTo determine thewidth of the glovesSmall: 85mmMedium: 95mmLarge: 105mmX-Large: 115mmSmall: PassMedium: PassLarge: PassX-Large: Pass
Test MethodPurposeAcceptance CriteriaResult
ASTM D6319-19Standard Specificationfor NitrileExamination Glovesfor MedicalApplicationTo determinethe length ofthe glovesPalm 0.05 mm minFinger 0.05 mm minfor all sizesSizeSmallMediumLargeX-largePalmPassFingerPass
ASTM D6319-19Standard Specificationphysical properties-for NitrileExamination GlovesUltimate Elongationfor MedicalTo Determinethe physicalpropertiesTensilestrengthBefore AgeingTensile Strength14Mpa Min for allsizesSizeSmallMediumPassPass

{8}------------------------------------------------

ApplicationAfter AgeingLarge
Tensile StrengthX-large
14Mpa Min for allsize
To Determinethe physicalpropertiesUltimateElongationBefore AgeingSizeBeforeageingAfterageing
UltimateElongation
Small
500% Min for allSizeMediumPassPass
After AgeingLarge
UltimateX-large
Elongation 400%Min for all sizes
Test MethodPurposeAcceptanceCriteriaResult
ASTM D5151-19Standard Test Methodfor detection of holesin medical glovesTo determine theholes in the glovesAQL 2.5Pass
ASTM D6124-06Standard Test Methodfor ResidualPowder on MedicalGlovesTo determine theresidual powder inthe gloves2 Mg/GloveMaxSizeSmallMediumLargeX-Large
ResidualPowderContentPass
Test MethodPurposeAcceptanceCriteriaResult
ISO 10993-10:2010Biological Evaluation ofTo determine thepotential of the materialunder test to produceUnder thecondition of studynot an irritantUnder thecondition of studynot an irritant

{9}------------------------------------------------

Medical Devices Test forIrritation and SkinSensitization. Test donefor irritation.dermal irritation inRabbits
ISO10993-10:2010Biological Evaluation ofMedical Devices Test forIrritation and SkinSensitization. Test doneSkin sensitization.To determine the skinsensitization potential ofthe material both in termsof induction andelicitation in Guinea Pig.Under theconditions of thestudy not asensitizerUnder theconditions of thestudy not asensitizer
ISO 10993-5:2009biological evaluation ofmedical devices - part 5,tests for in vitrocytotoxicity.To evaluate the in vitrocytotoxic potential of thetest item (both inner andouter surface) Extractsin L-929 mousefibroblasts cells usingelution method.Under theconditions of studynon cytotoxicUnder theconditions of thestudy noncytotoxic.

8. Discussion of Clinical Tests Performed:

Not applicable – Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

9. Conclusions:

The conclusion drawn from the nonclinical tests demonstrates that the subject device, the Medcare Powder-Free Blue Nitrile Patient Examination Gloves are as safe, effective, and perform as well as or better than the legally marketed predicate device cleared under K200326.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.