(169 days)
Lunit INSIGHT MMG is a radiological Computer-Assisted Detection and Diagnosis (CADe/x) software device based on an artificial intelligence algorithm intended to aid in the detection, and characterization of suspicious areas for breast cancer on mammograms from compatible FFDM systems. As an adjunctive tool, the device is intended to be viewed by interpreting physicians after completing their initial read.
It is not intended as a replacement for a complete physician's review or their clinical judgement that takes into account other relevant information from the image or patient history. Lunit INSIGHT MMG uses screening mammograms of the female population.
Lunit INSIGHT MMG is a radiological Computer-Assisted Detection and Diagnosis (CADe/x) software for aiding interpreting physicians with the detection, and characterization of suspicious areas for breast cancer on mammograms from compatible FFDM (full-field digital mammography) systems. The software applies an artificial intelligence algorithm for recognition of suspicious lesions, which are trained with large databases of biopsy proven examples of breast cancer, benign lesions and normal tissues.
Lunit INSIGHT MMG automatically analyzes the mammograms received from the client's image storage system (e.g., Picture Archiving and Communication System (PACS)) or other radiological imaging equipment. Following receipt of mammograms, the software device de-identifies the copies of images in DICOM format (.dcm) and then automatically analyzes each image and identifies and characterizes suspicious areas for breast cancer. The analysis result is converted into DICOM file and the result is saved within the designated storage location (e.g., PACS, x-ray system, etc.)
Lunit INSIGHT MMG processes mammograms and the output of the device can be viewed by interpreting physicians after completing their initial read. As an analysis result, the software device allows a visualization and quantitative estimation of the presence of a malignant lesion. The suspicious lesions are marked by a visualized map and an abnormality score, which reflects general likelihood of presence of malignancy, is presented for each lesion, as well as for each breast.
{
"acceptance_criteria_and_study_description": {
"acceptance_criteria_and_performance_table": [
{
"criterion": "Standalone Performance: ROC AUC",
"acceptance_criteria": "Statistical significance (p < 0.0001) and improvement compared to interpretation performance of radiologists when reading mammograms unaided.",
"reported_performance": "0.903 (95% CI: 0.889-0.917), p < 0.0001 (demonstrates improvement compared to unaided radiologists)"
},
{
"criterion": "Clinical Performance (MRMC Reader Study): Primary Endpoint - Average inter-test difference in ROC AUC (Test 2 vs. Test 1)",
"acceptance_criteria": "p-value of ROC AUC difference test < 5% and lower bound of two-sided 95% CI of AUC difference (Test 2 - Test 1) > 0",
"reported_performance": "0.051 (95% CI: 0.027-0.075), P=0.0001"
}
],
"test_set_sample_size_and_provenance": {
"standalone_performance_study": {
"sample_size": "2412 mammograms",
"data_provenance": "Collected using Hologic, GE Healthcare, and Siemens mammography equipment. Data is independent from dataset used for algorithm development and US pivotal reader study.",
"country_of_origin": "Not explicitly stated, but implies global or general collection from compatible equipment."
},
"clinical_testing_reader_study": {
"sample_size": "240 mammograms",
"data_provenance": "Collected using Hologic and GE Healthcare mammography equipment in the US. Retrospective study.",
"country_of_origin": "US"
}
},
"experts_for_ground_truth": {
"number_of_experts": "Not explicitly stated for establishing ground truth, but for the clinical reader study, 12 MQSA qualified reading panelists performed interpretations.",
"qualifications_of_experts": "For the clinical reader study, 'MQSA qualified reading panelists' were used. Specific years of experience are not mentioned."
},
"adjudication_method_for_test_set": {
"standalone_performance_study": "Not explicitly described, but implies comparison against 'reference standards'.",
"clinical_testing_reader_study": "Not explicitly described beyond the use of 12 MQSA qualified reading panelists. The process states that readers re-interpret cases with AI assistance after an initial unaided read, with randomized reading order to minimize bias."
},
"mrmc_comparative_effectiveness_study": {
"conducted": "Yes",
"effect_size": {
"roc_auc_improvement": "Average inter-test difference in ROC AUC between Test 2 (with CAD) and Test 1 (without CAD) was 0.051 (95% CI: 0.027-0.075), with statistical significance (P=0.0001). This indicates improved physician interpretation ability.",
"lcm_lroc_improvement": "0.094 (95% CI: 0.056 - 0.132)",
"lcm_roc_auc_improvement": "0.052 (95% CI: 0.026 - 0.079)",
"recall_rate_in_cancer_group_sensitivity_improvement": "5.97 (95% CI: 2.48 - 9.46)",
"recall_rate_in_non_cancer_group_1_specificity_improvement": "-1.46 (95% CI: -3.41 - 0.05)"
}
},
"standalone_performance_study_conducted": "Yes, two standalone performance analyses were done:\n1. A dedicated standalone performance study with 2412 mammograms, showing ROC AUC of 0.903.\n2. An additional standalone algorithm performance assessment using the 240 cases from the reader study, without reader intervention, yielding an ROC AUC of 0.863. This exceeded the AUC of every human reading panelist in the unaided (Test 1) scenario (unaided AUC = 0.754).",
"type_of_ground_truth_used": "Mainly implied as breast cancer detection based on 'reference standards' in the standalone study and by expert interpretation/consensus in the reader study. The device is trained with 'biopsy proven examples of breast cancer, benign lesions and normal tissues', suggesting pathology or clinical outcomes data as the ultimate ground truth source for training.",
"training_set_sample_size": "Not explicitly stated as a specific number, but described as 'large databases of biopsy proven examples of breast cancer, benign lesions and normal tissues'."
}
}
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Lunit Inc. % John Smith Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, District of Columbia 20004
February 7, 2022
Re: K211678
Trade/Device Name: Lunit INSIGHT MMG Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological computer assisted detection and diagnosis software Regulatory Class: Class II Product Code: QDQ
Dear John Smith:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 7, 2021. Specifically, FDA is updating this SE Letter due to a typographical error in the 510(k) summary as an administrative correction. Specifically the substantial equivalence table contains an additional column, which could trigger a misinterpretation for members of the public.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jessica Lamb, OHT7: Office of In Vitro Diagnostics and Radiological Health, 301-796-6167, Jessica.Lamb(@fda.hhs.gov.
Sincerely.
Jessica Lamb
Jessica Lamb Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Lunit Inc. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 Thirteenth Street NW WASHINGTON DC 20004
November 17, 2021
Re: K211678
Trade/Device Name: Lunit INSIGHT MMG Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological computer assisted detection and diagnosis software Regulatory Class: Class II Product Code: QDQ Dated: October 7, 2021 Received: October 7, 2021
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
510(k) Number (if known) K211678
Device Name
Lunit INSIGHT MMG Indications for Use (Describe)
Lunit INSIGHT MMG is a radiological Computer-Assisted Detection and Diagnosis (CADe/x) software device based on an artificial intelligence algorithm intended to aid in the detection, and characterization of suspicious areas for breast cancer on mammograms from compatible FFDM systems. As an adjunctive tool, the device is intended to be viewed by interpreting physicians after completing their initial read.
It is not intended as a replacement for a complete physician's review or their clinical judgement that takes into account other relevant information from the image or patient history. Lunit INSIGHT MMG uses screening mammograms of the female population.
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number
{4}------------------------------------------------
510(k) Summary
Lunit INSIGHT MMG (K211678)
This 510(k) summary of safety and effectiveness information is prepared in accordance with the requirements of 21 CFR §807.92.
Date Prepared: November 15, 2021
I. SUBMITTER
Manufacturer:
Lunit Inc. 15 Floor, 27 Teheran-ro 2-gil, Gangnam-gu, Seoul, Republic of Korea 06241 Phone +82-2-2138-0827 Fax +82-2-2135-5413
Contact Person:
Joohee Lee Regulatory Affairs Specialist
II. DEVICE
Name of Device: Lunit INSIGHT MMG Common or Usual Name: Lunit INSIGHT MMG Classification Name: Radiological Computer Assisted Detection/Diagnosis Software For Suspicious Lesions For Cancer Classification Regulation: 21 CFR 892.2090 Requlatory Class: II Product Code: QDQ
III. PREDICATE DEVICE
Predicate Device Manufacturer Name: Screenpoint Medical B.V. Device Trade Name: Transpara™ 510(k) Number: K192287
IV. DEVICE DESCRIPTION
Lunit INSIGHT MMG is a radiological Computer-Assisted Detection and Diagnosis (CADe/x) software for aiding interpreting physicians with the detection, and characterization of suspicious areas for breast cancer on mammograms from compatible FFDM (full-field digital mammography) systems. The software applies an artificial intelligence algorithm for recognition of suspicious lesions,
{5}------------------------------------------------
which are trained with large databases of biopsy proven examples of breast cancer, benign lesions and normal tissues.
Lunit INSIGHT MMG automatically analyzes the mammograms received from the client's image storage system (e.g., Picture Archiving and Communication System (PACS)) or other radiological imaging equipment. Following receipt of mammograms, the software device de-identifies the copies of images in DICOM format (.dcm) and then automatically analyzes each image and identifies and characterizes suspicious areas for breast cancer. The analysis result is converted into DICOM file and the result is saved within the designated storage location (e.g., PACS, x-ray system, etc.)
Lunit INSIGHT MMG processes mammograms and the output of the device can be viewed by interpreting physicians after completing their initial read. As an analysis result, the software device allows a visualization and quantitative estimation of the presence of a malignant lesion. The suspicious lesions are marked by a visualized map and an abnormality score, which reflects general likelihood of presence of malignancy, is presented for each lesion, as well as for each breast.
V. INDICATIONS FOR USE
Lunit INSIGHT MMG is a radiological Computer-Assisted Detection and Diagnosis (CADe/x) software device based on an artificial intelligence algorithm intended to aid in the detection, and characterization of suspicious areas for breast cancer on mammograms from compatible FFDM systems. As an adjunctive tool, the device is intended to be viewed by interpreting physicians affer completing their initial read.
lt is not intended as a replacement for a complete physician's review or their clinical judgement that takes into account other relevant information from the image or patient history. Lunit INSIGHT MMG uses screening mammograms of the female population.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERSTICS WITH THE PREDICATE DEVICE
The Lunit INSIGHT MMG and the predicate Transpara have the similar indications where both devices are intended to assist a clinical user in the detection, and characterization suspicious areas for breast cancer on mammography images. Also, both devices have intended user, and patient populations, and use the same imaging modality. The differences in the indications are with respect to the clinical workflow where the predicate device is intended to be used for concurrent read of images while the Lunit INSIGHT MMG is intended to be viewed by interpreting physicians after completing their initial read. Neither device is intended as a replacement for the review of a clinician or their clinical judgement and the patient management decisions are not made solely on the basis of analysis of the device. The minor difference in the indications with respect to the clinical workflow does not raise different questions of safety or effectiveness.
Lunit INSIGHT MMG and predicate share similar technological characteristics. Both devices are radiological computer assisted detection and diagnostic software based on an artificial intelligence algorithm that detect, identify and characterize findings based on features or information extracted from images. Both devices provide information about the presence and location of the findings to the user. As both devices process radiological images using similar machine learning techniques to
{6}------------------------------------------------
highlight abnormalities, the same types of safety and effectiveness questions are raised between devices. Both devices are support tools that provide relevant clinical data as an additional resource to the physician and will not replace the clinical expertise and judgement of the clinical user.
Both devices provide users with a numeric score indicating the likelihood of a presence of a malignant lesion however, the scoring system is different. Both devices provide a lesion score indicating likelihood of presence of malignancy. Lunit INSIGHT MMG analyzes per-breast abnormality score that takes the maximum scores from two views (i.e. CC, MLO). However, the predicate device analyzes the overall likelihood that cancer is present in a mammogram and categorizes the exams on a scale of 1 to 10 with increasing likelihood of cancer. The minor differences do not raise any different questions for its performance or safety; both devices provide clinically useful information to physicians who are reviewing a mammogram.
A comparison of the technological characteristics with the predicate device is summarized in Table 1.
VII. PERFORMANCE DATA
Non-clinical performance Data
The Lunit INSIGHT MMG is considered as a "moderate" level of concern, since a failure or a latent design flaw of the software device could result in Minor injury, either to a patient or to a user of the device. Based on results of verification and validation tests, Lunit INSIGHT MMG is effective in the detection of suspicious lesion for breast cancer at an appropriate safety level in mammograms. Verification tests which consisted of software unit testing and system testing were successfully completed to assure that the software application satisfied the software requirements. Validation testing determining standalone performance of the algorithms in Lunit INSIGHT MMG was also completed successfully. Validation and verification test results demonstrated that Lunit INSIGHT MMG is as safe and effective as the predicate device.
{7}------------------------------------------------
Standalone Performance Testing
Standalone Performance Study of the Lunit INSIGHT MMG assessed the performance of software algorithm for breast cancer detection in screening mammography.
Total of 2412 mammograms were collected using Hologic, GE Healthcare, and Siemens mammography equipment. The standalone performance of Lunit INSIGHT MMG for breast cancer detection in screening mammography is examined as comparing the reference standards and the analysis results of the device. The dataset used in the study is independent from the dataset used for development of the algorithm and the US pivotal reader study.
The primary endpoint of the standalone performance measured the Lunit INSIGHT MMG performance compared to radiologist performance alone. ROC AUC in the standalone performance analysis was 0.903 (95% Cl: 0.889-0.917) with statistical significance (p < 0.0001), which demonstrates improvement compared to the interpretation performance of radiologists when reading mammograms unaided. The secondary exploratory endpoints are two types of LROC AUC, sensitivity and specificity and these results are as follow:
[Standalone Analysis Results]
| ROC AUC [95% CI] | 0.903 [0.889, 0.917] |
|---|---|
| *p < 0.0001 | |
| Type I LROC AUC [95% CI] | 0.781 [0.751, 0.812] |
| Type II LROC AUC [95% CI] | 0.792 [0.763, 0.822] |
| Sensitivity (%) [95% CI] | 85.74 [82.95, 88.53] |
| Specificity (%) [95% Cl] | 75.62 [73.64, 77.60] |
Clinical Testing - Reader Study
Clinical Performance Assessment was conducted to evaluate effectiveness of Lunit INSIGHT MMG in the assistance of detection and diagnosis of breast cancer during screening mammography interpretation. A retrospective, multi-case (MRMC) study was conducted comparing the reading panel's interpretation performance with and without the assistance of the Lunit INSIGHT MMG during the screening mammography interpretation.
Total of 240 mammograms were collected using Hologic and GE Healthcare mammography equipment in the US. During the study, 12 MQSA qualified reading panelists performed their interpretation independently using a setting similar to a screening procedure in the U.S. Reading panelists interpret cases without CAD assistance first then re-interpret of the same cases with Test 1 reading results in assistance of the Lunit INSIGHT MMG. To minimize any potential bias, the reading order for each reading panelists was randomized.
The primary objective of the reader study is to determine that the diagnostic ability of radiologist with CAD assistance is superior to without CAD assistance. Radiologist performance was assessed with an inter-test comparison of measuring FBR (Forced Bl-RADS) ROC AUC between reader's interpretation and device-assisted interpretation for the detection of malignant lesions in the study. Success criteria of the study was defined as the p-value of the ROC AUC difference test is less than
{8}------------------------------------------------
the significance level of 5% and the lower bound of two-sided 95% Cl (Confidence Interval) of the AUC difference (Test 2-Test 1) is above 0. Secondary exploratory endpoints of the reader study assessed inter-test difference in LCM (level of confidence of malignancy) LROC AUC, LCM ROC AUC, and recall rate in cancer and non-cancer group (sensitivity and 1-specificity).
For the primary endpoint, the average inter-test difference in ROC AUC between Test 2 and Test 1 was 0.051 (95% Cl: 0.027-0.075) with statistical significance (P=0.0001), which indicates that physician's interpretation ability for breast cancer in mammograms is improved in Test 2 (with CAD) from Test 1 (without CAD).
[Primary Endpoint Results of the Pivotal Reader Study]
| Test1 | Test2 | Test2-Test1 | P-value | |
|---|---|---|---|---|
| ROC AUC (95%CI) | 0.754(0.702, 0.807) | 0.805(0.759, 0.852) | 0.051(0.027, 0.075) | 0.0001 |
All secondary exploratory endpoints have shown higher performance in assistance of the device, the inter-test differences for each assessment are as follow: LCM LROC 0.094 (95% C1: 0.056 - 0.132), LCM ROC AUC 0.052 (95% Cl: 0.026 - 0.079), recall rate in cancer group (sensitivity) 5.97(95% Cl: 2.48 - 9.46), and recall rate in non-cancer group (1-specificity) -1.46 (95% Cl: -3.41 - 0.05). Results of the primary and secondary analysis of the clinical testing demonstrate improvement for breast cancer detection and diagnosis using Lunit INSHGT MMG in mammograms.
Separated from the standalone performance test, an additional assessment of standalone algorithm performance was conducted with same cases from the reader study, using the Lunit INIGHT MMG without reader intervention as a sole means of reading mammograms compared to the software output. ROC AUC in the standalone performance analysis is 0.863 (95% Cl: 0.818 - 0.909) which exceeds the results of every reading panelist in Test 1 (CAD unaided).
{9}------------------------------------------------
Image /page/9/Figure/0 description: The image is a plot of three ROC curves, showing the performance of different models. The x-axis is the false positive rate (FPF), and the y-axis is the true positive rate (TPF). The red curve represents the Standalone Lunit MMG model, which has an AUC of 0.863. The green curve represents the Unaided model, which has an AUC of 0.754, and the blue curve represents the With Lunit MMG model, which has an AUC of 0.805.
[Plot Comparison of the standalone performance, CAD unaided and aided reader interpretation]
VIII. CONCLUSIONS
The non-clinical performance data and clinical data showed that Lunit INSIGHT MMG is as safe and effective as the predicate device. In particular, the effectiveness of the Lunit INSIGHT MMG in assistance of detection and diagnosis of breast cancer during screening mammography interpretation has been verified in the US population by confirming that the improvement for breast cancer detection and diagnosis before and after use of the Lunit INSIGHT MMG. In addition, the study results lead a positive expectation in an aspect of the clinical benefit that false positives and false negatives are reduced in the screening mammograms with the use of the Lunit INSIGHT MMG. In a conclusion, with all the clinical evidence achieved from the clinical study, the Lunit INSIGHT MMG can be considered to perform effectively in assistance of detection and diagnosis of breast cancer.
The Lunit INSIGHT MMG has the similar intended uses, indications, technological characteristics, and principles of operation as its predicate device. Both devices detect and characterize findings in radiological breast images and provide information about the presence, location, and characteristics of the findings to the user in a similar way. The minor differences in indications do not alter the intended of the device and do not affect its safety and effectiveness. In addition, the minor technological differences between the Lunit INSIGHT MMG and its predicate devices raise no new questions of safety or effectiveness. Performance data demonstrate that the Lunit INSIGHT MMG is as safe and effective as the predicate Transpara. Thus, the Lunit INSIGHT MMG is substantially equivalent.
{10}------------------------------------------------
| TechnicalCharacteristics | Subject DeviceLunit INSIGHT MMG | Predicate DeviceTranspara™ (K192287) |
|---|---|---|
| ClassificationRegulation | 21 CFR 892.2090Radiological Computer AssistedDetection and Diagnosis software | 21 CFR 892.2090Radiological Computer AssistedDetection and Diagnosis software |
| Medical DeviceClassification | Class II | Class II |
| Product Code | QDQ | QDQ |
| Level ofConcern | Moderate | Moderate |
| Intended Use | A reading aid for physiciansinterpreting screening FFDMacquired with compatiblemammography systems, to identifyfindings and assess their level ofsuspicion. | A reading aid for physiciansinterpreting screening FFDMacquired with compatiblemammography systems, to identifyfindings and assess their level ofsuspicion. |
| Target patientpopulation | Women undergoing FFDM screeningmammography | Women undergoing FFDM screeningmammography |
| Target userpopulation | Physicians interpreting FFDMscreening mammograms | Physicians interpreting FFDMscreening mammograms |
| Design | Software-only device | Software-only device |
| Indication forUse | Lunit INSIGHT MMG is a radiologicalComputer-Assisted Detection andDiagnosis (CADe/x) software devicebased on an artificial intelligencealgorithm intended to aid in thedetection, localization, and | The ScreenPoint Transpara™ system is intended for use as aconcurrent reading aid for physiciansinterpreting screening mammogramsfrom compatible FFDM system, toidentify regions suspicious for breast |
| TechnicalCharacteristics | Subject DeviceLunit INSIGHT MMG | Predicate DeviceTranspara™ (K192287) |
| characterization of suspicious areasfor breast cancer on mammogramsfrom compatible FFDM systems. Asan adjunctive tool, the device isintended to be viewed by interpretingphysicians after completing theirinitial read. It is not intended as areplacement for a completephysician's review or their clinicaljudgement that takes into accountother relevant information from theimage or patient history. The LunitINSIGHT MMG uses screeningmammograms of the femalepopulation. | cancer and assess their likelihood ofmalignancy. Output of the deviceincludes marks placed on suspicioussoft tissue lesions and suspiciouscalcifications; region-based scores,displayed upon the physician's query,indicating the likelihood that cancer ispresent in specific regions; and anoverall score indicating the likelihoodthat cancer is present on themammogram. Patient managementdecisions should not be made solelyon the basis of analysis byTranspara™™. | |
| Device outputin case ofpositivedetection | Output of the device includes markswith visualized map such as heatmapplaced on suspicious lesions forbreast cancer; lesion score,indicating the likelihood of presenceof the malignancy in specific regions;and an abnormality score indicatingthe likelihood of the presence ofmalignancy per-breast in themammogram. The per-breastabnormality score takes maximumscore from two views (i.e., CC, MLO)of a breast which is the maximumlesion score in the breast. | Output of the device includes marksin outlines placed on suspicious softtissue lesions and suspiciouscalcifications; region-based scores,displayed upon the physician's query,indicating the likelihood that cancer ispresent in specific regions; and anoverall score indicating the likelihoodthat cancer is present on themammogram. |
| Score | Finding level: | Finding level: |
| TechnicalCharacteristics | Subject DeviceLunit INSIGHT MMG | Predicate DeviceTranspara™ (K192287) |
| Score 1-100 indicating likelihood ofpresence of malignancy (from lowsuspicion to high suspicion). | Continuous score 1-100 indicatingthe level of suspicion of malignancy(from low suspicion to highsuspicion). | |
| Breast level:Score 1-100 indicating the level ofsuspicious of malignancy.The maximum scores from two view(i.e., CC, MLO) of a breast which isthe maximum lesion score in thebreast | Breast level: None | |
| Exam level: None | Exam level: 10-point scale scoreindicative of higher frequency ofcancer positive | |
| Performance | Reader study:• 240 cases• 12 radiologistscf. Reading time was not measuredin this study. | Reader study:• 240 cases• 14 radiologistsReading time:• 146 seconds (unaidedsession) |
| • 149 seconds (withTranspara™) | ||
| ROC AUC:radiologists AUC (unaided) = 0.754radiologists AUC (aided) = 0.805Standalone Performance Test:• 2412 cases | AUC:• radiologists AUC (unaided) =0.866• radiologists AUC (aided) =0.886 | |
| TechnicalCharacteristics | Subject DeviceLunit INSIGHT MMG | Predicate DeviceTranspara™ (K192287) |
| ROC AUC = 0.903 | standalone AUC = 0.887 | |
| Image SourceModality | FFDM | FFDM |
| Fundamentalscientifictechnology | In Lunit INSIGHT MMG, a range ofmedical image processing andmachine learning techniques areimplemented. The system includes'deep learning' algorithm applied toimages for recognition of suspiciouslesions for breast cancer. Themachine learning components aretrained to detect suspicious lesionsfor breast cancer with largedatabases of biopsy-proven cases ofbreast cancer, benign lesions andnormal tissues. | In Transpara, a range of medicalimage processing and machinelearning techniques are implemented.The system includes 'deep learning'algorithm applied to images forrecognition of suspicious lesions forbreast cancer. The machine learningcomponents are trained to detectsuspicious lesions for breast cancerwith large databases of biopsy-proven cases of breast cancer,benign lesions and normal tissues. |
Table 1 – Comparison Between Subject and Predicate Devices
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
N/A