K Number
K192287

Validate with FDA (Live)

Device Name
Transpara
Date Cleared
2019-12-10

(110 days)

Product Code
Regulation Number
892.2090
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ScreenPoint Transpara™ system is intended for use as a concurrent reading aid for physicians interpreting screening mammograms from compatible FFDM systems, to identify regions suspicious for breast cancer and assess their likelihood of malignancy. Output of the device includes marks placed on suspicious soft tissue lesions and suspicious calcifications; region-based scores, displayed upon the physician's query, indicating the likelihood that cancer is present in specific regions; and an overall score indicating the likelihood that cancer is present on the mammogram. Patient management decisions should not be made solely on the basis of analysis by Transpara™.

Device Description

Transpara™ is a software-only device for aiding radiologists with the detection and diagnosis of breast cancer in mammograms. The product consists of a processing server and an optional viewer. The software applies algorithms for recognition of suspicious calcifications and soft tissue lesions, which are trained with large databases of biopsy proven examples of breast cancer, benign lesions and normal tissue. Processing results of Transpara™ can be transmitted to external destinations, such as medical imaging workstations or archives, using the DICOM mammography CAD SR protocol. This allows PACS workstations to implement the interface of Transpara™ in mammography reading applications.

Transpara™ automatically processes mammograms and the output of the device can be used by radiologists concurrently with the reading of mammograms. The user interface of Transpara™ has different functions:

  • a) Activation of computer aided detection (CAD) marks to highlight locations where the device detected suspicious calcifications or soft tissue lesions. Only the most suspicious soft tissue lesions are marked to achieve a very low false positive rate.
  • b) Regions can be queried using a pointer for interactive decision support. When the location of the queried region corresponds with a finding of Transpara™ a suspiciousness level of the region computed by the algorithms in the device is displayed. When Transpara™ has identified a corresponding region in another view of the same breast this corresponding region is also displayed to minimize interactions required from the user.
  • c) Display of the exam based Transpara™ Score which categorizes exams on a scale of 1-10 with increasing likelihood of cancer.

Transpara™ is configured as a DICOM node in a network and receives its input images from another DICOM node, such as a mammography device or a PACS archive. The image analysis unit includes machine learning components trained to detect calcifications and soft tissue lesions and a component to pre-process images in such a way that images from different vendors can be processed by the same algorithms.

AI/ML Overview

The provided text primarily focuses on the device's regulatory submission and comparisons to a predicate device, rather than a detailed clinical study demonstrating its performance against acceptance criteria. There is no explicit "acceptance criteria" table with performance metrics in the provided document, nor a detailed description of a multi-reader, multi-case (MRMC) study or standalone performance study with specific metrics and methodologies.

However, based on the available information, we can infer some aspects and highlight where information is missing for a complete response to your request.

Inferred Acceptance Criteria and Reported Device Performance (Table):

The document states, "Validation testing confirmed that algorithm performance has improved in comparison to Transpara 1.3.0 for the four manufacturers for which the device was already cleared and that for Fujifilm a similar performance is achieved." This implies that the acceptance criteria for the new version (1.5.0) were primarily:

  1. Improvement or maintenance of performance compared to the predicate device (Transpara 1.3.0) for existing compatible modalities.
  2. Achievement of similar performance for the newly supported Fujifilm modalities.

Without specific metrics (e.g., AUC, sensitivity, specificity at a defined operating point), it's impossible to create a quantitative table. The document only provides qualitative statements.

Acceptance Criteria (Inferred)Reported Device Performance (Qualitative)
Algorithm performance for existing compatible modalities (Hologic, GE, Philips, Siemens) to be improved compared to Transpara 1.3.0."algorithm performance has improved in comparison to Transpara 1.3.0 for the four manufacturers for which the device was already cleared"
Algorithm performance for newly supported Fujifilm modalities to be similar to Transpara 1.3.0."that for Fujifilm a similar performance is achieved."
Effectiveness in detection of soft lesions and calcifications at an appropriate safety level."Based on results of verification and validation tests it is concluded that Transpara™ is effective in the detection of soft lesions and calcifications at an appropriate safety level in mammograms acquired with mammography devices for which the software has been validated." "Standalone performance tests demonstrate that Transpara™ 1.5.0 achieves better detection performance compared to the predicate device." (This broadly supports improvement, but lacks specific metrics like sensitivity/specificity/AUC).

Study Details based on the provided text:

  1. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated. The document mentions "a multi-vendor test-set of mammograms acquired from multiple centers."
    • Data Provenance:
      • Country of Origin: Not specified.
      • Retrospective or Prospective: Not explicitly stated, but "a multi-vendor test-set of mammograms acquired from multiple centers" for "asymptomatic women" suggests a retrospective collection of existing screening mammograms.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided for the validation test set. The document refers to "biopsy proven examples of breast cancer, benign lesions and normal tissue" for training, but not explicitly how the ground truth for the test set was established or by how many experts.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not specified. The document only mentions the use of "biopsy proven examples" for training and does not detail the ground truth establishment process for the test set.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • An MRMC study was done, but not for the 1.5.0 version that is the subject of this 510(k) summary.
    • The document states: "A pivotal reader study was conducted with the predicate device Transpara 1.3.0. This study provided evidence for safety and effectiveness of Transpara™."
    • Therefore, no MRMC study details or effect sizes for human reader improvement with 1.5.0 assistance are provided in this document. The current submission relies on the standalone performance of 1.5.0 showing improvement over the predicate, and assumes the clinical benefit of the predicate carries over to the improved version.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone performance study was done. The document explicitly states: "Validation testing consisted of determining stand-alone performance of the algorithms in Transpara™ using a multi-vendor test-set..." and "Standalone performance tests demonstrate that Transpara™ 1.5.0 achieves better detection performance compared to the predicate device."
    • Specific performance metrics (e.g., AUC, sensitivity, specificity, FROC analysis) are not provided in this summary, only the qualitative statements of "improved" or "similar" performance compared to the predicate device.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the training set, it explicitly states "biopsy proven examples of breast cancer, benign lesions and normal tissue."
    • For the validation/test set, it states the test set included "mammograms of asymptomatic women," and while not explicitly stated, the context of cancer detection implies that cancer status (and thus ground truth) would be established through pathology or follow-up outcomes for positive cases, and freedom from cancer on follow-up for negative cases. The text doesn't explicitly detail the method of ground truth collection for the test set, but it implies a robust method to define "cancer" vs. "normal."
  7. The sample size for the training set:

    • Not specified. The document only mentions that the algorithms were "trained with large databases of biopsy proven examples of breast cancer, benign lesions and normal tissue."
  8. How the ground truth for the training set was established:

    • "biopsy proven examples of breast cancer, benign lesions and normal tissue." This indicates that the ground truth for training data was established through histopathological confirmation from biopsies.

{0}------------------------------------------------

December 10, 2019

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration".

Screenpoint Medical B.V. % Umar Waqas, Ph.D. Regulatory Officer Mercator II, 7th floor, Toernooiveld 300 6525 EC Nijmegen, Gelderland THE NETHERLANDS

Re: K192287

Trade/Device Name: Transpara™ Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological computer assisted detection and diagnosis software Regulatory Class: Class II Product Code: QDQ Dated: October 30, 2019 Received: November 1, 2019

Dear Dr. Waqas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192287

Device Name Transpara

Indications for Use (Describe)

The ScreenPoint Transpara™ system is intended for use as a concurrent reading aid for physicians interpreting screening mammograms from compatible FFDM systems, to identify regions suspicious for breast cancer and assess their likelihood of malignancy. Output of the device includes marks placed on suspicious and suspicious calcifications; region-based scores, displayed upon the physician's query, indicating the likelihood that cancer is present in specific regions, and an overall score indicating the likelihood that cancer is present on the mammogram. Patient management decisions should not be made solely on the basis of analysis by Transpara™.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to rouio inotructiono grob yigting dato gother and maintain the doto nooded and comploto

time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary Transpara™ K192287

This 510(k) summary of safety and effectiveness information is prepared in accordance with the requirements of 21 CFR § 807.92.

1. Submitter

Manufacturer:

ScreenPoint Medical B.V.

Mercator II, 7th floor

Toernooiveld 300

6525 EC Nijmegen

Netherlands

www.screenpoint-medical.com

Contact person:

Umar Waqas

Office: +31 24 3030045 | +31 24 2020020

Mobile: +31 6 44077104

Mercator II, 7th floor, Toernooiveld 300, 6525 EC Nijmegen, Netherlands

Date:

December 10, 2019

{4}------------------------------------------------

2. Device

Device trade nameTranspara™
DeviceRadiological Computer Assisted Detection andDiagnosis Software
Classification regulation21 CFR 892.2090
PanelRadiology
Device classII
Product codeQDQ
Submission typeTraditional 510(k)

3. Legally marketed predicate device

Device trade nameTranspara™ 1.3.0 (K181704)
Legal ManufacturerScreenPoint Medical B.V.
DeviceRadiological Computer Assisted Detection andDiagnosis Software
Classification regulation21 CFR 892.2090
PanelRadiology
Device classII
Product codeQDQ

4. Device description

Transpara™ is a software-only device for aiding radiologists with the detection and diagnosis of breast cancer in mammograms. The product consists of a processing server and an optional viewer. The software applies algorithms for recognition of suspicious calcifications and soft tissue lesions, which are trained with large databases of biopsy proven examples of breast cancer, benign lesions and normal tissue. Processing results of Transpara™ can be transmitted to external destinations, such as medical imaging workstations or archives, using the DICOM mammography CAD SR protocol. This allows PACS workstations to implement the interface of Transpara™ in mammography reading applications.

{5}------------------------------------------------

Transpara™ automatically processes mammograms and the output of the device can be used by radiologists concurrently with the reading of mammograms. The user interface of Transpara™ has different functions:

  • a) Activation of computer aided detection (CAD) marks to highlight locations where the device detected suspicious calcifications or soft tissue lesions. Only the most suspicious soft tissue lesions are marked to achieve a very low false positive rate.
  • b) Regions can be queried using a pointer for interactive decision support. When the location of the queried region corresponds with a finding of Transpara™ a suspiciousness level of the region computed by the algorithms in the device is displayed. When Transpara™ has identified a corresponding region in another view of the same breast this corresponding region is also displayed to minimize interactions required from the user.
  • c) Display of the exam based Transpara™ Score which categorizes exams on a scale of 1-10 with increasing likelihood of cancer.

Transpara™ is configured as a DICOM node in a network and receives its input images from another DICOM node, such as a mammography device or a PACS archive. The image analysis unit includes machine learning components trained to detect calcifications and soft tissue lesions and a component to pre-process images in such a way that images from different vendors can be processed by the same algorithms.

5. Indications for use

Transpara™ is a software medical device for use in a healthcare facility or hospital with the following indications for use:

The ScreenPoint Transpara™ system is intended for use as a concurrent reading aid for physicians interpreting screening mammograms from compatible FFDM systems, to identify regions suspicious for breast cancer and assess their likelihood of malignancy. Output of the device includes marks placed on suspicious soft tissue lesions and suspicious calcifications; region-based scores, displayed upon the physician's query, indicating the likelihood that cancer is present in specific regions; and an overall score indicating the likelihood that cancer is present on the mammogram. Patient management decisions should not be made solely on the basis of analysis by Transpara™.

Intended user population

Intended users of Transpara™ are physicians qualified to read screening mammograms.

{6}------------------------------------------------

Intended patient population

The device is intended to be used in the population of women undergoing screening mammography.

Warnings and precautions

Transpara™ is an adjunct tool and not intended to replace a physicians' own review of a mammogram. Decisions should not be made solely based on analysis by Transpara™.

6. Predicate device comparison

The indication for use of Transpara™1.5.0 is the same as that of the predicate device. Both devices are intended to be used by clinicians interpreting mammograms, to help them with localizing and characterizing abnormalities. The devices are both intended to be used concurrently with the reading of images and are not intended as a replacement for the review of a clinician or their clinical judgement.

The overall design of Transpara™1.5.0 is similar to that of the predicate device. The main differences in technological characteristics of Transpara™ 1.5.0 and the predicate device are algorithmic improvements leading to better detection performance. Additionally, the 1.5.0 version now supports Fujifilm modalities. These changes do not raise different questions of safety and effectiveness.

7. Summary of non-clinical performance data

In the design and development of Transpara™, ScreenPoint applied the following voluntary FDA recognized standards and guidelines:

Standard IDStandard TitleFDA Recognition #
ISO 14971:2007Medical Devices - Application Of RiskManagement To Medical Devices5-40
IEC 62304:2015Medical Device Software - Software Life -Cycle Processes13-79
DEN180005Decision summary with special controlsfor class II radiology device

The following guidance documents were used to support this submission:

{7}------------------------------------------------

  • . Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (Issued on May 11, 2005)
  • Computer-Assisted Detection Devices Applied to Radiology Images and . Radiology Device Data – Premarket Notification [510(k)] Submissions (Issued on July 3, 2012)
  • Guidance for Industry and FDA Staff Clinical Performance Assessment: ● Considerations for Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data - Premarket Approval (PMA) and Premarket Notification [510(k)] Submissions (Issued on July 3, 2012)

Software performance testing

Transpara™1.5.0 is a software-only device. The level of concern for the device is determined as Moderate Level of Concern.

Verification testing consisted of software unit testing, software integration testing and software system testing. The verification test showed that the software application satisfied the software requirements.

Validation testing consisted of determining stand-alone performance of the algorithms in Transpara™ using a multi-vendor test-set of mammograms acquired from multiple centers. This test dataset was not used for training of Transpara™ algorithms and included mammograms of asymptomatic women acquired with devices from five manufacturers: Hologic, GE, Philips, Siemens, and Fujifilm. Validation testing confirmed that algorithm performance has improved in comparison to Transpara 1.3.0 for the four manufacturers for which the device was already cleared and that for Fujifilm a similar performance is achieved.

Based on results of verification and validation tests it is concluded that Transpara™ is effective in the detection of soft lesions and calcifications at an appropriate safety level in mammograms acquired with mammography devices for which the software has been validated.

8. Summary of clinical tests

A pivotal reader study was conducted with the predicate device Transpara 1.3.0. This study provided evidence for safety and effectiveness of Transpara™

9. Conclusions

The data presented in this 510(k) includes all required information to support the review by FDA. Standalone performance tests demonstrate that Transpara™ 1.5.0 achieves better detection performance compared to the predicate device.

{8}------------------------------------------------

ScreenPoint has applied a risk management process in accordance with FDA recognized standards to identify, evaluate, and mitigate all known hazards related to Transpara™. These hazards may occur when accuracy of diagnosis is potentially affected, causing either false-positives or false-negatives. All identified risks are effectively mitigated and it can be concluded that the residual risk is outweighed by the benefits.

Considering all data in this submission, the data provided in this 510(k) supports the safe and effective use of Transpara™ for its indications for use and substantial equivalence to the predicate device.

N/A