K Number
K211591

Validate with FDA (Live)

Date Cleared
2021-09-30

(129 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

NAEOTOM Alpha:
This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning.
This CT system can be used for low dose lung cancer screening in high risk populations *

  • As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Scan&GO:
This in-room scan application is a planning and information system the necessary functions required for planning and controlling scans of supported SIEMENS CT scanners. It allows users to work in close proximity to the scanner.
The in-room scan application runs on standard information technology hardware and software, utilizing the standard information technology operating systems and user interface. Communication and data exchange are done using special protocols.

Device Description

Siemens intends to market a new CT scanner system NAEOTOM Alpha supporting software version, SOMARIS/10 syngo CT VA40 with mobile workflow options.
Dual Source CT Scanner System:

  • NAEOTOM Alpha
  • Scan&GO Mobile Medical Application (optional mobile workflow component) .
    The subject device NAEOTOM Alpha with SOMARIS/10 syngo CT VA40 is a dual-source Computed Tomography (CT) x-ray system featuring two detectors based on new photon counting technology. The CT scanner system algorithm is designed to allow image reconstruction by using photon counting data generated by the subject device. The reconstruction results are comparable with the predicate devices, but support with improved technological characteristics.
    The NAEOTOM Alpha with Software SOMARIS/10 synqo CT VA40 produces CT images in DICOM format, which can be used by trained staff for post-processing applications commercially distributed by Siemens and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.
    The Scan&GO mobile workflow is an optional planning and information software designed to perform the necessary functions required for planning and controlling of the NAEOTOM Alpha. Scan&GO can be operated on a Siemens provided various tablet hardware that meets certain minimum technical requirements.
    NOTE: Scan&GO does not support storage of images. Additionally, Scan&GO cannot trigger a scan or radiation release.
    The software version for the NAEOTOM Alpha, syngo CT VA40 (SOMARIS/10 syngo CT VA40), is a command-based program used for patient management, data manaqement, X-ray scan control, image reconstruction, and image archive/evaluation.
    The software platform provides a software plugin interface that allows for the use of specific commercially available post processing software algorithms in an unmodified form from the cleared stand-alone post processing version.
    New software version syngo CT VA40 (SOMARIS/10 syngo CT VA40) is a modified software version based on syngo CT VA30A (SOMARIS/10 syngo CT VA30) which was cleared for the secondary predicate device and supports the same plugin interfaces for the subject device Scan&GO mobile workflow and integration of post-processing tasks as the secondary predicate device Scan&GO cleared in (K200524).
AI/ML Overview

The provided text describes a 510(k) premarket notification for the NAEOTOM Alpha CT scanner and Scan&GO application. It includes performance data from non-clinical testing.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The document provides a general overview of performance testing and states that "all of the software specifications have met the acceptance criteria" and that verification and validation testing was "found acceptable to support the claim of substantial equivalence." However, it does not present a specific table with quantitative acceptance criteria and corresponding reported device performance metrics in a structured format for each feature. Instead, it describes the type of performance testing conducted for various features.

For example, for "Detector - QuantaMax," it mentions "in-depth evaluation of NAETOM Alpha Image Quality for general CT imaging, based on phantom evaluation of Typical Modes, compared to the predicate device SOMATOM Force. It also includes parameters for supporting the suitability of the subject device for low dose lung cancer screening." For "CARE keV," it states "The test procedure includes phantom measurements with clinically relevant phantom diameters and contrast materials to support the contrast, noise, and radiation dose related CARE keV information."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document primarily discusses non-clinical testing which involves phantom studies. There is no mention of patient data (test set) being used for performance evaluation in the context of proving substantial equivalence, nor any information about data provenance (country of origin, retrospective/prospective). The testing described is performed on the device itself and phantoms.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Since the testing described is primarily non-clinical (phantom studies) and doesn't explicitly involve human reader interpretation for a test set, there is no mention of experts establishing ground truth for a test set or their qualifications.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Given that no human reader-based test set evaluation is described for performance, there is no information on an adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document does not describe a multi-reader multi-case (MRMC) comparative effectiveness study. The focus is on the substantial equivalence of the NAEOTOM Alpha CT scanner and its components to existing predicate devices, primarily through non-clinical performance testing and technical comparisons, not on measuring human reader performance with or without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the testing described appears to be a form of standalone algorithm-only (device-only) performance evaluation through phantom studies. The document details testing of features like "Quantum Iterative Reconstruction," "Detector - QuantaMax," "CARE keV," and "Quantum Pure Lumen" using phantom measurements and technical analyses to assess their performance characteristics, independent of human interpretation in a clinical setting.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical phantom studies, the ground truth would be the known physical properties and configurations of the phantoms (e.g., known material compositions, shapes, sizes, and concentrations of inserts). For example, for "Always Dual Energy," it mentions "phantoms with iodine inserts," where the known concentration and distribution of iodine would serve as ground truth for assessing accuracy.

8. The sample size for the training set

The document refers to the NAEOTOM Alpha as a new CT scanner system with modified software. It mentions that "software version SOMARIS/10 syngo CT VA40 is a further development of the SOMARIS/10 syngo CT VA30 software version," implying an evolution rather than a de novo AI algorithm that requires a separate training set. While the algorithms (like Quantum Iterative Reconstruction, CARE keV) have been optimized, the document does not specify a sample size for a training set in the context of distinct machine learning model training as one might expect for a typical AI/ML device. The focus is on the device's inherent imaging capabilities and reconstruction algorithms.

9. How the ground truth for the training set was established

As there is no explicit mention of a training set or a distinct AI/ML model being trained with external data, there is no information on how ground truth for a training set was established. The development appears to be based on engineering principles, physics of CT imaging, and optimization of established reconstruction techniques for the new photon-counting detector technology.

{0}------------------------------------------------

September 30, 2021

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized depiction of human figures. To the right of it is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Siemens Medical Solutions USA, Inc. % Tabitha Estes Regulatory Affairs Specialist 810 Innovation Drive KNOXVILLE TN 37932

Re: K211591

Trade/Device Name: NAEOTOM Alpha, Scan&GO Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: August 19, 2021 Received: August 20, 2021

Dear Tabitha Estes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

2

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

, for

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K211591

Device Name NAEOTOM Alpha Scan&GO

Indications for Use (Describe) NAEOTOM Alpha

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning.

This CT system can be used for low dose lung cancer screening in high risk populations *

  • As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Scan&GO

This in-room scan application is a planning and information system the necessary functions required for planning and controlling scans of supported SIEMENS CT scanners. It allows users to work in close proximity to the scanner.

The in-room scan application runs on standard information technology hardware and software, utilizing the standard information technology operating systems and user interface. Communication and data exchange are done using special protocols.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K211591

510(K) SUMMARY FOR NAEOTOM ALPHA CT SCANNER AND SCAN&GO Submitted by:

Siemens Medical Solutions USA, Inc. 810 Innovation Drive Knoxville, TN 37932 Date Prepared: February 21, 2021

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

I. Submitter

Siemens Medical Solutions USA, Inc. 810 Innovation Drive Knoxville, TN 37932 Establishment Registration Number 1034973

Importer/Distributor

Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number 2240869

Location of Manufacturing Site (1)

Siemens Healthcare GmbH Siemensstr. 1 D-91301 Forchheim, Germany Establishment Registration Number 3004977335

Location of Manufacturing Site (2) SIEMENS SHANGHAI, MEDICAL EQUIPMENT LTD 278 Zhou Zhu Rd Shanghai, CHINA, 201318 Establishment Registration Number: 3003202425

Note: Description in this submission use the short company name Siemens. It covers both manufacturing locations and names as listed above. Brand name on all products is Siemens Healthineers.

Contact Person:

Tabitha Estes Requlatory Affairs Siemens Medical Solutions USA, Inc. (865) 804-4553 (work cell) (865) 218-3019 FAX

{4}------------------------------------------------

II. Device Name and Classification

Product Name:NAEOTOM Alpha
Trade Name:NAEOTOM Alpha
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
Regulation Number:21 CFR §892.1750
Device Class:Class II
Product Code:JAK
Product Name:Scan&GO
Propriety Trade Name:Scan&GO
Classification Name:Computed Tomography X-ray System
Secondary Classification Name:Picture Archiving and Communications System
Classification Panel:Radiology
CFR Section:21 CFR §892.1750
Secondary CFR Section:21 CFR §892.2050
Device Class:Class II
Product Code:JAK
Secondary Product Code:LLZ

III. Predicate Device

Primary Predicate Device: SOMATOM Force Trade Name:

510(k) Number:K190578
Clearance Date:June 27, 2019
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
Regulation Number:21 CFR §892.1750
Device Class:Class II
Product Code:JAK
Recall Information:All predicate device recalls have been considered in the subject devicedesign.

Secondary Predicate Device:

Trade Name:

SOMATOM X.cite, Scan&GO

510(k) Number:K200524
Clearance Date:April 01, 2020
Classification Name:Computed Tomography X-ray System
Classification Panel:Radiology
Regulation Number:21 CFR §892.1750
Device Class:Class II
Product Code:JAK
Recall Information:All predicate device recalls have been considered in the subject devicedesign.

Note: K200524 was a bundle submission with various Siemens CT Scanner Systems, including SOMATOM go.Now, SOMATOM go.Up, SOMATOM go.All, SOMATOM go.Top, SOMATOM go.Sim, SOMATOM go.Open Pro, SOMATOM X.cite and Scan&GO software.

In this submission, the primary predicate device SOMATOM Force and the secondary predicate device SOMATOM X.cite including Scan&GO are being used, to demonstrate substantial equivalence of technological characteristics.

{5}------------------------------------------------

IV. Device Description

Siemens intends to market a new CT scanner system NAEOTOM Alpha supporting software version, SOMARIS/10 syngo CT VA40 with mobile workflow options.

Dual Source CT Scanner System:

  • NAEOTOM Alpha
  • Scan&GO Mobile Medical Application (optional mobile workflow component) .

The subject device NAEOTOM Alpha with SOMARIS/10 syngo CT VA40 is a dual-source Computed Tomography (CT) x-ray system featuring two detectors based on new photon counting technology. The CT scanner system algorithm is designed to allow image reconstruction by using photon counting data generated by the subject device. The reconstruction results are comparable with the predicate devices, but support with improved technological characteristics.

The NAEOTOM Alpha with Software SOMARIS/10 synqo CT VA40 produces CT images in DICOM format, which can be used by trained staff for post-processing applications commercially distributed by Siemens and other vendors as an aid in diagnosis, treatment preparation and therapy planning support (including, but not limited to, Brachytherapy, Particle including Proton Therapy, External Beam Radiation Therapy, Surgery). The computer system delivered with the CT scanner is able to run optional post processing applications.

The Scan&GO mobile workflow is an optional planning and information software designed to perform the necessary functions required for planning and controlling of the NAEOTOM Alpha. Scan&GO can be operated on a Siemens provided various tablet hardware that meets certain minimum technical requirements.

NOTE: Scan&GO does not support storage of images. Additionally, Scan&GO cannot trigger a scan or radiation release.

The software version for the NAEOTOM Alpha, syngo CT VA40 (SOMARIS/10 syngo CT VA40), is a command-based program used for patient management, data manaqement, X-ray scan control, image reconstruction, and image archive/evaluation.

The software platform provides a software plugin interface that allows for the use of specific commercially available post processing software algorithms in an unmodified form from the cleared stand-alone post processing version.

New software version syngo CT VA40 (SOMARIS/10 syngo CT VA40) is a modified software version based on syngo CT VA30A (SOMARIS/10 syngo CT VA30) which was cleared for the secondary predicate device and supports the same plugin interfaces for the subject device Scan&GO mobile workflow and integration of post-processing tasks as the secondary predicate device Scan&GO cleared in (K200524).

V. Indications for Use

NAEOTOM Alpha:

This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning.

This CT system can be used for low dose lung cancer screening in high risk populations *

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

{6}------------------------------------------------

Scan&GO:

This in-room scan application is a planning and information system designed to perform the necessary functions required for planning and controlling scans of supported SIEMENS CT scanners. It allows users to work in close proximity to the scanner.

The in-room scan application runs on standard information technology hardware and software, utilizing the standard information technology operating systems and user interface. Communication and data exchange are done using special protocols.

VI. Comparison of Technological Characteristics with the Predicate Device

The NAEOTOM Alpha scanner provides comparable technological characteristics in terms of materials, energy source, and control mechanisms when compared to the primary predicate device SQMATOM Force. The software and hardware components of this scanner have been modified or improved in comparison to the predicate devices to support enhanced device functionality and a to introduce a new detector technology.

The subject device NAEOTOM Alpha is building on the dual-source CT system configuration of the primary predicate device SOMATOM Force.

NAEOTOM Alpha features comparable technological characteristics of a Siemens dual-source CT system, such as fast scan speed and temporal resolution down to 66ms. It uses the same general geometric dimensions of the measurement field (such as detector z coverage and scan-field-of-view) in comparison to the primary predicate device SOMATOM Force CT scanner system.

As a dual source CT scanner system, the primary predicate device SOMATOM Force features two conventional energy-integrating detectors made of scintillator ceramic (product marketing name "Ultra Fast Ceramic", abbreviated UFC, based on scintillator ceramic). The subject device NAEOTOM Alpha features two photon counting detectors built from Cadmium-Telluride (CdTe).

Supported by the subject device, SOMARIS/10 syngo CT VA40 software version is a further development of the SOMARIS/10 syngo CT VA30 software version which is cleared in K200524. It reuses all unmodified software features of the legacy software syngo CT VA30 as described below in the section "Unmodified Features used in the Subject Devices".

As with the secondary predicate device SOMATOM X.cite, the subject device NAEOTOM Alpha is supported by an optional mobile workflow, Scan&GO has been modified to add new iPad hardware and operating software that supports the application software installation.

Software version SOMARIS/10 synqo CT VA40 is designed to reuse hardware independent extended functionalities and GO technologies provided by Siemens cleared software applications. The intended use and fundamental scientific technology for the NAEOTOM Alpha remains unchanged from the cleared predicate devices.

At a high level, the subject device NAEOTOM Alpha and the primary predicate device SOMATOM Force or the secondary predicate device SOMATOM X.cite are based on the same subset of technological elements:

  • Scanner Principle- Whole body X-Ray Computed Tomography Scanner .
  • System Acquisition Continuously rotating tube detector system ●
  • Iterative Reconstruction Support of various iterative reconstruction principles ●
  • Workplaces – Support of workplaces that include reconstruction and image evaluation software
  • Patient table ●
  • Patient table foot switch for movement
  • Tin filtration technology ●
  • Vectron X-ray Tube
  • . Power Generator
  • Mobile Medical application Software functionality (Scan&GO) .
  • Mobile workflow (Tablet) .
  • Support 3D Camera operation for fast patient positioning workflow

{7}------------------------------------------------

  • Scanner display and control functionality ●
  • Remote Scan Control ●
  • Long scan range

The following technological differences exist between the subject device NAEOTOM Alpha and the primary predicate device SOMATOM Force or the secondary predicate device SOMATOM X.cite:

  • Software version SOMARIS/10 syngo CT VA40
  • Support updated cybersecurity features
  • Additional options for Inline and GO technologies ●
  • Iterative Reconstruction Methods .
  • . QuantaMax Detector based on Quantum Technology

The following technological differences exist between the subject device Scan&GO mobile application software and the secondary predicate device Scan&GO mobile application software cleared in K200524:

  • Software version SOMARIS/10 syngo CT VA40
  • iPad hardware to support mobile workflow options .

The NAEOTOM Alpha will support the following modifications/further developments in comparison to the predicate devices as listed in the tables below.

The configuration table and comparison table use the following terms to describe various technological characteristics in comparison to the predicate device information:

Table 01: Overview term definition
TermDefinition
N/AThe feature is not supported for the subject device
NewThe feature is newly supported for Siemens CT Scanners and the subject device
ModifiedThis feature is modified from the predicate devices
UnmodifiedThis feature remains unchanged from the predicate device
enabledThis feature is currently supported by other cleared Siemens CT systems or clearedSiemens stand-alone software applications. This feature will be supported for the subjectdevice with software version SOMARIS/10 syngo CT VA40 and is substantially equivalentcompared to the cleared version.

New/modified hardware features:

Table 02: Overview of hardware modifications of NAEOTOM Alpha

Subject Device
Hardware properties supported by SOMARIS/10 syngo CT VA40NAEOTOM Alpha
01Detector - QuantaMaxnew
02Tablet hardware for Scan&GO - iPadmodified
03Patient Table Configurationmodified

New software version SOMARIS/10 syngo CT VA40 with enabled, modified and new software features:

Table 03: Overview of software modifications of NAEOTOM Alpha

Subject Device
Software properties supported by SOMARIS/10 syngo CT VA40NAEOTOM Alpha
01Precision Matrix (10242) (large image matrices)enabled
02Quantum Iterative Reconstructionnew
03Always Dual Energynew
04Calcium scoring with keV imagesmodified
05Imaging - Cardio BestPhaseenabled
06CARE keVnew
07Quantum Pure Lumennew
08Cardiac CT imaging - Motion artifact reduced ECG-gated imagingmodified
09Motion artifact reduced non-gated imagingmodified

{8}------------------------------------------------

Subject Device
Software properties supported by SOMARIS/10 syngo CT VA40NAEOTOM Alpha
10myExam Console and myExam Satellitemodified
11Recon&GO Inline Results - DE SPPmodified
12Recon&GO Spectral Reconmodified

A tabular summary of the comparable hardware and software properties between the subject device NAEOTOM Alpha with software version syngo CT VA40 and predicate devices are listed in Table 04 and Table 05 below (only the changes from the predicate devices are highlighted in gray in the sections below sections).

Table 04: NAEOTOM Alpha comparable hardware properties (modified)
Hardwarepropertiessubject deviceNAEOTOM AlphaDual Sourceprimary predicate deviceSOMATOM ForceDual Sourcesecondary predicate deviceSOMATOM X.citeSingle Source
Scannerwhole body X-ray computedtomography scannerwhole body X-ray computedtomography scannerwhole body X-ray computedtomography scanner
Generatormax. power2x 120 kW2x 120 kW105 kW
DetectortechnologyDirect Conversion with"Quantum Technology"Energy-integratingEnergy-integrating
Detectorvolume coverage2 x 57.6 mm2 x 57.6 mm38.4 mm
Detectorphysical rows2 x 2882 x 9664
Detectorslice width0.2 mm0.6 mm0.6 mm
Detectorchannel No.2752 (A system)1984 (B system)920 (A system)640 (B system)840
X-ray TubeVectronVectronVectron
TubekV steps90 kV, 100 kV,120 kV, 140 kV(in 10kV steps)70 kV to 150kV(in 10kV steps)70 kV to 150kV
Tubemax. current2 x 1300 mA2 x 1300 mA1200 mA
Tubetube focus$0.4×0.5$ mm / $0.6x0.7$ mm /$0.8x1.1$ mm(for both tubes)$0.4×0.5$ mm / $0.6×0.7$ mm /$0.8x1.1$ mm(for both tubes)$0.6x0.7$ mm / $0.8x1.1$ mm
Tubeheat capacityhigher than 30 MHUhigher than 30 MHUhigher than 30 MHU
TubeCollimatoradaptive dose shiedsupportedadaptive dose shiedsupportedadaptive dose shiedsupported
Gantrybore size82 cm78 cm82 cm
GantryScan FoV50 cm50 cm50 cm
Gantryrotation time (sec)0.25s; 0.5s; 1.0s;0.25s, 0.285s, 0.33s, 0.5s,1.0s0.3s, 0.5s, 1.0s
GantryTilt (degree)N/AN/A+/- 25
Patient TabletypeVario 2.D:2000 mm,Vitus:2000 mmPHS5, MPT4:1600 mm or 2000 mmVario RT: 1600 mm,Vario 2: 2000 mm
Max. Scan lengthTopogramVario 2.D:2080 mm,Vitus:2080 mm1600 mm, 2000 mm1680 mm, 2080 mm
Max. Scan lengthImage acquisitionVario 2.D:2000 mm,Vitus:2000 mm1600 mm, 2000 mm1600 mm, 2000 mm
Spectral filtrationOption*Tin Filter for both tubesTin Filter for both tubesCombined Split Filter / TinFilter supported, plus extraTin Filter
3D Camerafor patient positioningoption for patient positioningwith 3D Cameraoption for patient positioningwith 3D Cameraoption for patient positioningwith 3D Camera
Propertiessoftwaresubject deviceNAEOTOM Alpha(syngo CT VA40 )primary predicate deviceSOMATOM Force(K190578) (syngo CT VB20)secondary predicate deviceSOMATOM X.cite(K200524)
OperatingSystemWindows basedSOMARIS/10syngo CT VA40Windows basedSOMARIS/7syngo CT VB20Windows basedSOMARIS/10syngo CT VA30 A
AcquisitionWorkplacesyngo Acquisition Workplace (AWP)syngo Viewing, syngo Filmingand syngo Archiving &Networking2nd Acquisition Workplacesupported with myExamSatellitesyngo Acquisition Workplace (AWP)syngo Viewing, syngo Filmingand syngo Archiving &Networking2nd Acquisition Workplace:RRWPsyngo Acquisition Workplace (AWP)syngo Viewing, syngo Filmingand syngo Archiving &NetworkingNo 2nd Acquisition Workplacesupported
IRSImage Reconstruction fromphoton counting dataImage Reconstruction fromclassic Siemens dual sourceCT scannerImage Reconstruction fromclassic Siemens single sourceCT scanner
DetectorQuantaMax detector firmwaresupportedStellar detector firmwaresupportedStellar detector firmwaresupported
TeamplaySupport teamplay ProtocolsSupport of:Support teamplay ProtocolsSupport of:Support teamplay ProtocolsSupport of:
ProtocolsProtocol supportingcontrast bolus-triggereddata acquisition Contrast media protocols Pediatric Protocols Flex Dose Profile Turbo Flash Spiral Dual Energy acquisitionProtocols for RadiationTherapy Planning Protocol supportingcontrast bolus-triggereddata acquisition Contrast protocols – CAREContrast III Pediatric Protocols Flex Dose Profile Turbo Flash Spiral Dual Energy acquisition Adaptive 4D Spiral Protocols that allowscanning with support ofan 3rd party respiratorygating system (ANZAI,Varian RGSC)Protocols for RadiationTherapy Planning supportpatient marking Protocol supportingcontrast bolus-triggereddata acquisition Contrast media protocols Pediatric Protocols Flex Dose Profile TwinBeam DE TwinSpiral DE Flex 4D Spiral Protocols that allowscanning with support ofan 3rd party respiratorygating system (ANZAI,Varian RGSC)
AdvancedReconstructionRecon&GO –- Spectral Recon (Dual EnergyReconstruction from photoncounting data) / includingVirtual Unenhanced,Monoenergetic plus- Inline Results DE SPP(Spectral Post-Processing withphoton counting image data)Advanced reconstruction toolssupported:The syngo acquisitionworkplace provides, imagereconstruction, and routinepostprocessing. Variousadvanced reconstructionfeatures supported by the CTScanner, e.g., FAST and CAREapplications.Advanced reconstruction toolssupported via Recon&GO –- Spectral Recon (Dual EnergyReconstruction)- Inline Results DE SPP(Spectral Post-Processing) /includingVirtual Unenhanced,Monoenergetic plus- Inline Results DE Ranges(Parallel/Radial) / Inline DE
Post-ProcessingSoftware Plugin functionsenabled via software interfaceRecon&GO - Inline Resultsvarious methods of clearedsoftware applicationssyngo.via - Wide Range ofindividual applications,syngo.via is a software solutionintended to be used forviewing, manipulation,communication, and storage ofmedical images. It can be usedas a standalone device orSoftware Plugin functionsenabled via software interfaceRecon&GO - Inline Resultsvarious methods of clearedsoftware applications
subject deviceprimary predicate devicesecondary predicate device
PropertiessoftwareNAEOTOM Alpha(syngo CT VA40)SOMATOM Force(K190578) (syngo CT VB20)SOMATOM X.cite(K200524)
cleared and unmodified syngo based software options.
CybersecurityIT HardeningIT HardeningIT Hardening
StandardTechnologiesFAST Features CARE Features GO technology CARE keVFAST Features CARE Features CARE kVFAST Features CARE Features GO technology CARE kV
IterativeReconstructionMethodsQuantum IterativeReconstructioniMAR, SAFIRE, ADMIREiMAR, ADMIRE
PrecisionMatrixPrecision Matrix resolutionsupport image matrix sizes of512x512, 768x768 pixels and1024x1024 pixel (auto modesupported)Precision Matrix resolutionsupport image matrix sizes of512x512, 768x768 pixels and1024x1024 pixel (auto modesupported)Matrix resolution 512x512(auto mode not supported)
Ca ScoringPhoton counting technologyoffers monoenergetic imageswhich can be used as a basefor calcium scoringindependent from tube voltagekV and beam filtration settings.Classic SOMATOM Forcetechnology offers imageswhich can be used for calciumscoring, independent fromtube voltage kV and beamfiltration settings based on adedicated single energy imagereconstruction method.Classic SOMATOM X.citetechnology offers imageswhich can be used for calciumscoring, independent fromtube voltage kV and beamfiltration settings based on adedicated single energy imagereconstruction method.

Table 04: NAEOTOM Alpha comparable hardware properties (modified)

{9}------------------------------------------------

Table 05: NAEOTOM Alpha comparable software properties

{10}------------------------------------------------

Any differences in technological characteristics do not raise different questions of safety and effectiveness. Siemens believes that the subject device is substantially equivalent to the predicate devices. Testing and validation are completed. Test results show that the subject device, the NAEOTOM Alpha, is comparable to the predicate devices in terms of technological characteristics and safety and effectiveness and therefore is substantially equivalent to the primary predicate SOMATOM Force and secondary predicate device SOMATOM X.cite.

VII. Performance Data

Non Clinical Testing

Non-clinical testing, (integration and functional) including phantom tests were conducted for the NAEOTOM Alpha during product development. The modifications described in this Premarket Notification were supported with verification and validation testing.

The general purpose of each tests is to verify and validate the functionality of the subject device modifications.

Testing will cover all related subsystems that contribute to the device modifications. Test levels are defined. For each test level several test activities are performed. The test specification and acceptance criteria are related to the corresponding requirements. Various test activities are performed to specific modifications on different test levels to ensure safe and effective integration in the system. Three test levels are defined:

System Validation test:

  • Acceptance test (workflow and user manual test) .
  • Legal and Regulatory test ●

System Verification test:

  • System Integration Test (functional) ●
  • Functionality verification ●
  • Image Quality (IQ) Evaluation ●

{11}------------------------------------------------

Tests are conducted for all software components developed in product development and for the complete product itself. Several activities are considered for this process, including creation of test specifications that relate to software/hardware requirements including tests to address risk mitigations that are identified, documented and traced by hazard keys.

Additional evaluation tests are performed as bench tests to support the new device or device modification on Non-Clinical Performance Testing as listed in table 07 below.

#FeatureNon-Clinical Performance Testing
01Quantum Iterative ReconstructionQuantum Iterative Reconstruction - QIR is routinely used in all reconstructions on the subject device NAEOTOM Alpha.As with the primary predicate device SOMATOM Force supporting ADMIRE, the subject device supports iterative reconstruction principle.NAEOTOM Alpha supports Quantum Iterative Reconstruction with image data generated with Quantum Technology (QuantaMax Detector) and optimized for photon counting data.Iterative Beam Hardening Correction - IBHC is an iterative Beam Hardening Correction algorithm optimized for the subject device NAEOTOM Alpha.Performance testing was done to support all technological characteristics, that are associated with this new technology.
02Detector - QuantaMaxThe subject device NAEOTOM Alpha supports two "QuantaMax" detectors for CT data acquisition. The QuantaMax detectors are built on the new "Quantum Technology" which is Siemens Healthineers' implementation of photon counting technology for CT detectors.The performed tests include the measurement and analysis of raw data for single photon counting, the nonexistence of classical noise and the ability to measure multiple energies simultaneously.Furthermore, the performance evaluation provides an in-depth evaluation of NAETOM Alpha Image Quality for general CT imaging, based on phantom evaluation of Typical Modes, compared to the predicate device SOMATOM Force. It also includes parameters for supporting the suitability of the subject device for low dose lung cancer screening.
05Cardiac CT imaging - Motion artifact reduced ECG-gated imagingAs with the primary predicate device the subject device supports ECG-gated imaging based on ECG-controlled dose modulation and triggering derived from continuous monitoring of the ECG. With same technological principle as cleared with the primary predicate device, the subject device supports with an unmodified algorithm, the desired prediction of the target start- and stop phase in retrospective spiral and in sequence mode. The prediction of the start phase in ECG triggered high pitch scanning is modified to improve the prediction over a wider range of heart rates.The associated test evidence contains a basic test scenario and the results of measurements based on an andromorphic motion heart phantom to support the clinical approach based on technical phantom measurements, which are to some extend clinically realistic.
06CARE keVCARE kV is supported by the primary predicate device SOMATOM Force. The subject device NAEOTOM Alpha based on Quantum Technology supports CARE keV. CARE keV takes the dose efficiency of Monoenergetic reconstructions into account, by automatically adjusting the appropriate kV and effective mAs settings to optimize the applied dose while the image quality is maintained. The test procedure includes phantom measurements with clinically relevant phantom diameters and contrast materials to support the contrast, noise, and radiation dose related CARE keV information.
07Quantum Pure LumenThe subject device NAEOTOM Alpha supports with its new feature "VCR - Pure Lumen" a modified algorithm, which allows to reconstruct with a virtual removal of calcium in spectral generated images with a possible application in heavily calcified vessels.Algorithm, clinical references and example images derived from phantom based measurements. Performance testing was done to support all technological characteristics, that are associated with this new technology.
#FeatureNon-Clinical Performance Testing
08Always Dual EnergyThe subject device NAEOTOM Alpha supports virtual monochromaticimages, virtual non-contrast images and iodine images for allQuantumPlus scans, i.e., all scans with 120kV or 140kV without tinfiltration and not in UHR mode.Performance testing was done to assess the accuracy of these threeimage types on phantoms with iodine inserts.
09Motion artifact reduced non-gated imagingAs with the primary predicate device SOMATOM Force the subjectdevice supports spiral scanning at different pitch value of up to 3.2 in adual source acquisition and up to 1.5 in a single source acquisition.These acquisitions at high pitch values and therefore table speedshave the potential to minimize motion artifacts. This feature name anddetector hardware is modified. In comparison to the primary predicatedevice SOMATOM Force, other technological characteristics tosupport the feature remains unmodified.Performance testing was done to confirm the characteristics to allowimprovement of motion artifacts. It also contains a basic test scenarioand the results of measurements based on an andromorphic motionheart and respiratory phantom to support the clinical approach basedon technical phantom measurements, which are clinically realistic.
10Calcium Scoring with keV ImagesAs with the primary predicate device SOMATOM Force, the subjectdevice support CaScoring. With its photon counting detectortechnology, the subject device software is modified to use itsgenerated monoenergetic images as a base for calcium scoring.The performance evaluation used monoenergetic imagesreconstructed at 70 keV with different QIR strength settings derivedfrom a counting detector technology acquisition of the subject deviceNAEOTOM Alpha, which are suitable in the application of AgatstonScore base Coronary Calcium scoring.

Table 06: Non-Clinical Performance Testing

{12}------------------------------------------------

Electrical Safety and Electromagnetic Compatibility (EMC) testing were conducted on the subject device NAEOTOM Alpha in accordance with the following standards: 60601-2-44, and 60601-1-2. A list of recognized and general consensus standards considered for the subject devices is provided as Table 07 and Table 08 below.

Date ofRecognitionRecognitionNumberStandardDevelopingOrganizationStandardDesignationNumber and DateTitle of Standard
06/27/201612-300NEMAPS 3.1 - 3.20 (2016)Digital Imaging and Communicationsin Medicine (DICOM) Set
07/06/202012-325NEMAXR 25-2019Computed Tomography Dose Check
01/27/201512-287NEMAXR 28-2013Supplemental Requirements for UserInformation and System FunctionRelated to Dose in CT
06/27/20165-40ANSI AAMIISO14971:2007/(R)2010(Corrected 4October 2007)Medical Devices - Applications OfRisk Management To MedicalDevices
ISO14971 Secondedition 2007-03-01Medical Devices - Applications OfRisk Management To MedicalDevices
01/14/201913-79IEC62304 Edition 1.12015-06CONSOLIDATEDVERSIONMedical Device Software - SoftwareLife Cycle Processes
07/09/201419-4ANSI AAMIES60601-1:2005/(R)2012 AndA1:2012,C1:2009/(R)2012 andA2:2010/(R)2012 (ConsolidatedText) Medical electrical equipment -Part 1: General requirements forbasic safety and essentialperformance (IEC 60601-1:2005,MOD)
09/17/201819-8ANSI AAMIIEC60601-1-2:2014Medical electrical equipment - Part 1-2: General requirements for basic
Date ofRecognitionRecognitionNumberStandardDevelopingOrganizationStandardDesignationNumber and DateTitle of Standard
12/23/20165-114ANSI AAMIIEC62366-1:2015safety and essential performance -Collateral Standard: Electromagneticdisturbances - Requirements andtests
07/09/201412-273IEC60825-1 Edition 2.02007-03Medical devices - Part 1: Applicationof usability engineering to medicaldevices
06/27/201612-302IEC60601-2-44 Edition3.2: 2016Safety Of Laser Products - Part 1:Equipment Classification, AndRequirements [Including: TechnicalCorrigendum 1 (2008), InterpretationSheet 1 (2007), Interpretation Sheet2 (2007)]Note: Requirements according toIEC 60825-1:2014 (Ed.3.0) areimplemented.
01/14/201412-269IEC60601-1-3 Edition2.1 2013-04Medical electrical equipment - Part 2-44: Particular requirements for thebasic safety and essentialperformance of x-ray equipment forcomputed tomography
06/27/20165-89IEC60601-1-6 Edition3.1 2013-10Medical electrical equipment - Part 1-3: General requirements for basicsafety and essential performance -Collateral Standard: Radiationprotection in diagnostic X-rayequipment
03/14/201112-226IEC61223-2-6 SecondEdition 2006-11Medical Electrical Equipment - Part1-6: General Requirements ForBasic Safety And EssentialPerformance - Collateral Standard:Usability
01/30/201412-270IEC61223-3-5 FirstEdition 2004-08Evaluation and routine testing inmedical imaging departments - Part2-6: Constancy tests - Imagingperformance of computedtomography X-ray equipment
12/23/201912-328IEC61223-3-5 Edition2.0 2019-09Evaluation and routine testing inmedical imaging departments - Part3-5: Acceptance tests - Imagingperformance of computedtomography X-ray equipment[Including: Technical Corrigendum 1(2006)]
06/07/201812-309IEC60601-2-28 Edition3.0 2017-06Evaluation and routine testing inmedical imaging departments - Part3-5: Acceptance tests - Imagingperformance of computedtomography X-ray equipment[Including: Technical Corrigendum 1(2006)]
06/27/201612-299IEC62563-1 Edition 1.1Medical Electrical Equipment - Part2-28: Particular Requirements ForThe Basic Safety And EssentialPerformance Of X-Ray TubeAssemblies For Medical Diagnosis
Medical electrical equipment -Medical image display systems - Part1: Evaluation methods

Table 07: Recognized Consensus Standards

{13}------------------------------------------------

{14}------------------------------------------------

Table 08: General Use Consensus Standards

StandardDevelopingOrganizationStandardDesignationNumber andDateTitle of StandardHow was Standard Used
IEC60601-1:2005+A1:2012Medical electricalequipment - part 1: generalrequirements for basicsafety and essentialperformanceCovered by ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012 as part ofEMC testing.
IEC/ISO17050-1Conformity Assessment -Supplier's declaration ofconformity - Part 1:General requirementsDeclaration of conformance to FDArecognized consensus standards.
IEC/ISO17050-2Conformity assessment -Supplier's declaration ofconformity - Part 2:Supporting documentation.General consensus standards notcurrently recognized by FDA.

A list of applicable guidance documents considered for this submission is provided as Table 09
below.

Table 09: FDA Guidance Document and Effective Date

FDA Guidance Document and Effective Date
1.Guidance for Industry and FDA Staff – User Fees and Refunds for Premarket NotificationSubmissions 510(k)Document issued on October 2, 2017
2.Guidance for Industry and Food and Drug Administration Staff: Refuse to Accept Policy for 510(k)sDocument issued on February 21, 2019
3.Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated 510(k)s - Guidance forIndustry and FDA StaffDocument issued on August 12, 2005
4.Guidance for Industry and FDA Staff: Deciding when to submit a 510(k) for a change to an existingdevice.Document issued on October 25, 2017
5.Guidance for Industry and Food and Drug Administration Staff: The 510(k) Program: EvaluatingSubstantial Equivalence in Premarket Notifications [510(k)]Document Issued on July 28, 2014
6.Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submission forSoftware in Medical DevicesDocument issued on May 11, 2005
7.Guidance for Industry and FDA Staff: Guidance for Off-The-Shelf Software Use in Medical DevicesDocument issued on September 9, 1999
8.Guidance for Industry and FDA Staff: Applying Human Factors and UsabilityEngineering to Medical Devices.Document issued February 3, 2016
9.Guidance for Industry and FDA Staff: Pediatric Information for X-ray ImagingDevice Premarket Notifications.Document issued on November 28, 2017
10.Guidance for Industry and FDA Staff: Content of Premarket Submissions forManagement of Cybersecurity in Medical devices.Document issued on October 2, 2014
11.Guidance for Industry and FDA Staff: Information to Support a Claim ofElectromagnetic Compatibility (EMC) of Electrically-Powered Medical DevicesDocument issued on July 11, 2016
12.Guidance for Industry and Food Drug Administration Staff: Design considerations and Pre-MarketSubmission recommendations for Interoperable Medical devicesDocument Issued on September 6, 2017
13.Guidance for Industry and Food Drug Administration Staff:Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for MedicalDevicesDocument issued on September 14, 2018

{15}------------------------------------------------

Verification and Validation

Software documentation for a moderate level of concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The Risk Analysis was completed, and risk control implemented to mitigate identified hazards. The testing supports that all software specifications have met the acceptance criteria. Testing for verification and validation support the claims of substantial equivalence.

Siemens conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with quidance document "Content of Premarket Submissions for Management of Cybersecurity Medical Devices issues on October 2, 2014" is included within this submission.

Additionally, Siemens conforms to the requirements for Radio Frequency Wireless Technology as defined in FDA quidance document "Radio Frequency Wireless Technology in Medical Devices, Guidance for Industry and Food and Drug Administration Staff, issued on August 14, 2013" by adhering to the EMC and risk-based verification and validation requirements in design, testing, and labeling of the wireless remote control components of the subject devices.

The Radio Frequency Wireless Technology of the optional Remote Scan Control and supporting Control Device iPad for Scan&GO complies to 47 CFR part 15 subpart c - Intentional Radiators. All Radio device labels will show an FCC ID code to show compliance. Shielding requirement applicable to the NAEOTOM Alpha and respective Scatter Radiation diagrams for typical room installations are provided in the User Documentation and Planning Guide of the intended Scanners in accordance with IEC60601-2-44.

Wireless Coexistence Testing

Siemens has considered several measures to address wireless coexistence by design to ensure the safe operation of the wireless components in combination with the applicable system supported functionality. Wireless technology in the system setup to perform a task in a given shared environment where other systems have an ability to perform their tasks and may or may not be using the same set of rules has been considered. According to FDA quidance 'Radio Frequency Wireless Technology in Medical Devices" Siemens has addressed the safety, effectiveness, and high likelihood of coexistence with other devices of this technology in our product design by our Risk Management Process, Failure Mode and Effects Analysis (FMEA) Process, and Requirement Engineering Process. As part of the risk management process, hazardous situations associated with the Scan&GO and its connection to the host system via Wi-Fi were addressed as part of the Risk Management process.

Testing for co-existence considered for following scenarios:

  • Co-Channel Testing ●
  • Adjacent Channel Testing
  • RF Interference Testing .
  • . Separation Distance/Location Testing

Scan&GO is designed to allow dynamic frequency selection and transmission power control by default in accordance with IEEE 802.11h. Adjacent channel testing is addressed by the fact that Scan&GO does not support shared medium access to Siemens Wi-Fi network. RF interference was tested by successfully ensuring that wireless communications were actively transmitting in situations where possible interference may exist. Recommended distance and router locations requirements are documented in the user documentation.

Summary

The features described in this premarket notification are supported with verification and validation testing, dosimetry and imaging performance, and analysis of phantom images to assess device and feature performance during product development. The risk analysis was completed, and risk control implemented to mitigate identified hazards. The test results show that all of the software specifications have met the acceptance criteria. Verification and validation testing of the device was found acceptable to support the claim of substantial equivalence.

{16}------------------------------------------------

General Safety and Effectiveness Concerns

The device labeling contains instructions for use as well as necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a system related risk analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing according to the risk management process. In order to minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.

VIII. Conclusions

The predicate devices were cleared based on the results of non-clinical testing including verification and validation, phantom tests, and supportive literature. The subject device is also tested using the same test methods and workflows as used for the predicate devices. The non-clinical data supports the safety of the device and the hardware and software verification demonstrates that the NAEOTOM Alpha should perform as intended in the specified use conditions. The data included in this submission demonstrates that the NAEOTOM Alpha with described modifications performs comparably to the predicate devices currently marketed for the same intended use. Since the subject and predicate devices were tested using the same methods, Siemens believes that the data generated from the NAEOTOM Alpha testing supports a finding of substantial equivalence.

N/A