K Number
K211586

Validate with FDA (Live)

Date Cleared
2021-11-01

(161 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nitrile Examination Gloves is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

The proposed device, Nitrile Examination Gloves (Blue, Violet) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size and color. The color of the proposed device is Blue and Violet. The proposed device is not provided as sterilized. The proposed device is made of Nitrile.

AI/ML Overview

This document describes the performance testing for Nitrile Examination Gloves (Blue, Violet). Here's a breakdown of the acceptance criteria and study details:

Acceptance Criteria and Reported Device Performance

Device: Nitrile Examination Gloves (Blue, Violet)
Purpose: Non-sterile disposable device worn to prevent contamination between patient and examiner.

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D5151Testing for Freedom from HolesFreedom from holes AQL 2.5Blue Color: No water leakage inspected from 200 samplesViolet Color: No water leakage inspected from 200 samples
ASTM D6124Determine the powder residue for powder free gloves< 2.0 mg per gloveBlue Color: Residual Powder: Average 0.05 – 0.10 mgViolet Color: Residual Powder: Average 0.04 – 0.10 mg
ASTM D412, ASTM D573Testing for Physical Property CharacteristicsBefore Aging: Tensile Strength: 14 MPa min., Ultimate Elongation: 500% min.After Aging: Tensile Strength: 14 MPa min., Ultimate Elongation: 400% min.Blue Color:- XS: Before Aging: Tensile Strength: ≥ 31.1 MPa, Ultimate Elongation: ≥ 505 %; After Aging: Tensile Strength: ≥ 30.1 MPa, Ultimate Elongation: ≥ 460%.- S: Before Aging: Tensile Strength: ≥ 31.3 MPa, Ultimate Elongation: ≥ 507 %; After Aging: Tensile Strength: ≥ 32.1 MPa, Ultimate Elongation: ≥ 457%.- M: Before Aging: Tensile Strength: ≥ 31.2 MPa, Ultimate Elongation: ≥ 506 %; After Aging: Tensile Strength: ≥ 28.2 MPa, Ultimate Elongation: ≥ 447%.- L: Before Aging: Tensile Strength: ≥ 30.9 MPa, Ultimate Elongation: ≥ 506 %; After Aging: Tensile Strength: ≥ 27.9 MPa, Ultimate Elongation: ≥ 426%.- XL: Before Aging: Tensile Strength: ≥ 31.0 MPa, Ultimate Elongation: ≥ 508 %; After Aging: Tensile Strength: ≥ 25.1 MPa, Ultimate Elongation: ≥ 408%. Violet Color:- XS: Before Aging: Tensile Strength: ≥ 24.1 MPa, Ultimate Elongation: ≥ 510 %; After Aging: Tensile Strength: ≥ 28.6 MPa, Ultimate Elongation: ≥ 408%. - S: Before Aging: Tensile Strength: ≥ 26.5 MPa, Ultimate Elongation: ≥ 515 %; After Aging: Tensile Strength: ≥ 29.8 MPa, Ultimate Elongation: ≥ 436%.- M: Before Aging: Tensile Strength: ≥ 26.3 MPa, Ultimate Elongation: ≥ 515 %; After Aging: Tensile Strength: ≥ 26.7 MPa, Ultimate Elongation: ≥ 462%.- L: Before Aging: Tensile Strength: ≥ 25.9 MPa, Ultimate Elongation: ≥ 510 %; After Aging: Tensile Strength: ≥ 25.6 MPa, Ultimate Elongation: ≥ 457%.- XL: Before Aging: Tensile Strength: ≥ 26.5 MPa, Ultimate Elongation: ≥ 517 %; After Aging: Tensile Strength: ≥ 25.0 MPa, Ultimate Elongation: ≥ 453%.
ASTM D412, ASTM D3767Testing for Physical Dimensions SpecificationLength: 220 mm min. for size (XS, S); 230 mm min. for size (M, L, XL); Width: 70±10 mm for XS; 80±10 mm for S; 95±10 mm for M; 110±10 mm for L; 120±10 mm for XL; Finger Thickness: ≥0.05 mm; Palm Thickness: ≥0.05 mm. All acceptance criteria meet ASTM D6319.Blue Color:- XS: Length: ≥ 225mm; Width: 70±8 (77-78) mm; Palm Thickness: ≥0.06 mm; Finger Thickness: ≥0.07 mm.- S: Length: ≥ 246mm; Width: 80±7 (86-87) mm; Palm Thickness: ≥0.06 mm; Finger Thickness: ≥0.07 mm.- M: Length: ≥ 240 mm; Width: 95±4 (98-99) mm; Palm Thickness: ≥0.06 mm; Finger Thickness: ≥0.08 mm.- L: Length: ≥ 243mm; Width: 110±3 (107-108) mm; Palm Thickness: ≥0.06 mm; Finger Thickness: ≥0.08 mm.- XL: Length: ≥ 243mm; Width: 120±6 (114-116) mm; Palm Thickness: ≥0.06 mm; Finger Thickness: ≥0.08 mm.Violet Color:- XS: Length: ≥ 225mm; Width: 70±8 (77-78) mm; Palm Thickness: ≥0.06 mm; Finger Thickness: ≥0.07 mm.- S: Length: ≥ 246mm; Width: 80±7 (86-87) mm; Palm Thickness: ≥0.06 mm; Finger Thickness: ≥0.07 mm.- M: Length: ≥ 240 mm; Width: 95±4 (98-99) mm; Palm Thickness: ≥0.06 mm; Finger Thickness: ≥0.08 mm.- L: Length: ≥ 242mm; Width: 110±3 (107-108) mm; Palm Thickness: ≥0.06 mm; Finger Thickness: ≥0.08 mm.- XL: Length: ≥ 243mm; Width: 120±6 (114-116) mm; Palm Thickness: ≥0.06 mm; Finger Thickness: ≥0.08 mm.
ISO 10993-11Evaluate systemic toxicity for biocompatibilityThe test article showed no evidence of systemic toxicity potential from the extract.The test article showed no evidence of systemic toxicity from the extract.
ISO 10993-10Evaluate irritation for biocompatibility; Evaluate sensitization for biocompatibilityThe response of the test article extract is negligible. The test article showed no evidence of causing delayed dermal contact sensitization.The test result showed that the response of the test article extract was categorized as negligible under the test condition. The test article showed no evidence of causing delayed dermal contact sensitization in the guinea pig.

Study Information:

Based on the provided document, the device in question is Nitrile Examination Gloves, and the studies performed are non-clinical performance tests for medical gloves. The document focuses on demonstrating substantial equivalence to a predicate device through adherence to established ASTM and ISO standards for physical properties and biocompatibility.

Here's the breakdown of the requested information, where applicable:

  1. A table of acceptance criteria and the reported device performance: This is provided in the comprehensive table above.

  2. Sample size used for the test set and the data provenance:

    • Freedom from holes (ASTM D5151): 200 samples were inspected for both Blue and Violet colored gloves.
    • Other tests (Physical properties, Dimensions, Biocompatibility): The exact sample sizes for these tests are not explicitly stated in the provided text, beyond the "200 samples" for the pinhole test. However, it's inferred that sufficient samples were tested to meet the requirements of the respective ASTM and ISO standards.
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. Given it's a 510(k) submission, these are typically pre-market, prospective tests conducted by the manufacturer or accredited labs.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For performance testing of medical gloves, "ground truth" is established by adherence to specified scientific and engineering standards (ASTM and ISO). This does not typically involve human expert consensus in the way an AI diagnostic algorithm for medical imaging would. The expertise lies in the certified laboratories and personnel performing the standardized tests according to the laid-out protocols. The document does not specify the number or qualifications of experts involved in setting the ground truth beyond conforming to the cited standards.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable for this type of performance testing. Adjudication methods like 2+1 or 3+1 are used for expert consensus on clinical data (e.g., medical image interpretation) to establish ground truth for algorithm training or evaluation. Here, the ground truth is defined by objective physical and chemical testing standards.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a submission for a medical device (nitrile gloves), not an AI algorithm for diagnosis or interpretation. Therefore, no MRMC study or AI assistance evaluation would be performed.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is for a non-diagnostic physical medical device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth is based on established industry standards and test methodologies for physical properties, dimensions, and biocompatibility, as defined by ASTM International and ISO. This is a form of objective, standardized measurement, not expert consensus or clinical outcomes data in the traditional sense for diagnostic devices.
  8. The sample size for the training set:

    • Not applicable. This device is not an AI algorithm that requires a "training set." The listed tests are for product validation, not algorithm development.
  9. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this type of device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Nov 1, 2021

Lyncmed Medical Technology (Beijing) Co., Ltd. Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China

Re: K211586

Trade/Device Name: Nitrile Examination Gloves (Blue, Violet) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 15, 2021 Received: September 22, 2021

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K211586

Device Name

Nitrile Examination Gloves (Blue, Violet)

Indications for Use (Describe)

The Nitrile Examination Gloves is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number: K211586

    1. Date of Preparation: 10/30/2021
    1. Sponsor

Lyncmed Medical Technology (Beijing) Co., Ltd.

Room 1601, Building No. 2, Zhubang 2000 Business Building, Balizhuangxili 99, Chaoyang District, BeiJing, P.R. China 100022

Contact Person: Jiali Zhou Position: Regulation Affair Manager Tel: +86-15117964608 Email: naomizhou@lyncmed.com

    1. Submission Correspondent
      Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, China,102401

Mr. Ray Wang Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang(@believe-med.com

  • Proposed Device Identification 4.
    Trade Name: Nitrile Examination Gloves (Blue, Violet) Common Name: NITRILE Patient Examination Gloves (Powder Free)

Regulatory Information: Classification: I Product Code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

{4}------------------------------------------------

Indication For Use Statement:

The Nitrile Examination Gloves is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    1. Predicate Device Identification
      510(k) Number: K150340 Product Name: POWDER FREE Nitrile GLOVES (White, Cobalt Blue, Black, Ice Blue) Manufacturer: HEBEI HONGSEN PLASTICS TECHNOLOGY CO., LTD
  • Device Description 6.
    The proposed device, Nitrile Examination Gloves (Blue, Violet) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size and color. The color of the proposed device is Blue and Violet.

The proposed device is not provided as sterilized The proposed device is made of Nitrile.

Size
DesignationXSSMLXLTolerance
Length, mm220220230230230min
Width, mm708095110120±10
Thickness, mm:
Finger0.05min
Palm0.05min

Table 1 Device Size Specifications

Before AgingAfter AgingPinhole AQL
TensileStrengthUltimateElongationTensileStrengthUltimateElongation1.5
14 MPa, min500 % min14 MPa, min400 % min

The above data of size, performance, and physical specifications of proposed gloves meet all the current specifications listed in the ASTM standard D6319.

{5}------------------------------------------------

7. Comparison of Technological Characteristics

ITEMProposed Device (K211586)Nitrile Examination Gloves (Blue, Violet)Predicate Device (K150340)POWDER FREE Nitrile GLOVES (White,Cobalt Blue, Black, Ice Blue)Remark
Product CodeLZALZASAME
Regulation No.21 CFR 880.625021 CFR 880.6250SAME
ClassIISAME
Intended UseThe Nitrile Examination Gloves is anon-sterile disposable device intended formedical purpose that is worn on theexaminer's hand to prevent contaminationbetween patient and examiner.The POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, Ice Blue) is adisposable device intended for medicalpurposes that is worn on the examiner'shands to prevent contamination betweenpatient and examiner.SAME
Powdered orPowered freePowdered freePowdered freeSAME

Table 1 General Comparison

{6}------------------------------------------------

Proposed Device (K211586)Nitrile Examination Gloves(Blue, Violet)SizeTolerance
DesignationXSSMLXL
Length, mm220220230230230min
Width, mm708095110120±10
Thickness, mm:
Finger0.05min
Palm0.05min
Predicate Device (K150340)POWDER FREE NitrileGLOVES (White, Cobalt Blue,Black, Ice Blue)SizeTolerance
DesignationXSSMLXL
Length, mm230230230230230min
Width, mm708095110120±10
Thickness, mm:
Finger0.10-0.12±0.03
Palm0.08-0.10±0.03

Table 2 Device Dimensions Comparison

Different Analysis 1:

The proposed device has different size specification to the predicate device, but all proposed devices are meet the specifications of ASTM D 6319.

ITEMProposed Device (K211586)Nitrile Examination Gloves (Blue,Violet)Predicate Device (K150340)POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, IceBlue)Remark
ColorantBlue, VioletWhite, Cobalt Blue, Black, Ice BlueDifferentAnalysis 2
PhysicalPropertiesBeforeAgingTensileStrength14MPa, min15 MPa, minDifferentAnalysis 1
UltimateElongation500 % min500 % minSAME
AfterAgingTensileStrength14 MPa, min14 MPa, minSAME
Ultimate400 % min400 % minSAME

Table 3 Performance Comparison

{7}------------------------------------------------

Elongation
Comply with ASTM D6319Comply with ASTM D6319SAME
Freedom from HolesBe free from holes when tested inaccordance with ASTM D5151Be free from holes when tested inaccordance with ASTM D5151SAME
Powder ContentLess than 2 mg per glove when testedin accordance with ASTM D6124Meet the requirements ofASTM D6124SAME

Different Analysis 2:

The proposed device has different color to the predicate device, this different may causes potential biocompatibility risk, for this risk we conducted the biocompatibility test according to the ISO 10993-10, the test results showed that the proposed devices with blue colorant did not induce skin irritation and showed no significant evidence of causing skin sensitization.

ITEMProposed Device (K211586)Disposable Nitrile ExaminationGloves (Blue, Violet)Predicate Device (K150340)POWDER FREE Nitrile GLOVES(White, Cobalt Blue, Black, Ice Blue)Remark
MaterialNitrileNitrileSAME
BiocompatibilityIrritationUnder the conditions of the study,not an irritantUnder the conditions of the study,not an irritantSAME
SensitizationUnder conditions of the study, not asensitizer.Under conditions of the study, not asensitizer.
Acute System toxicityUnder the conditions of study, nomortality or evidence of systemictoxicityN.ADifferentAnalysis 3
Label and LabelingMeet FDA's RequirementsMeet FDA's RequirementsSAME

Table 4 Safety Comparison

Different Analysis 3:

The proposed device has conducted the systemic toxicity testing than predicated device for toxicity evaluation, the test results shown that there were no mortality or evidence of systemic toxicity.

  1. Summary of Non-Clinical performance testing

The test results demonstrated that the proposed device complies with the following standards:

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. ASTM D6124-17, Standard Test Method for Residual Powder on Medical Gloves. ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.

{8}------------------------------------------------

Please add a table here. The table should include four columns: Test method, purpose, acceptance criteria and results.

Table 5 Performance Test Results Summary
Test MethodPurposeAcceptance CriteriaResults
ASTM D5151Testing for Freedom from holesFreedom from holesAQL 2.5Blue Color:No water leakage is inspected form 200 samplesViolet Color:No water leakage is inspected form 200 samples
ASTM D6124Determine the powder residue for powder free gloves<2.0 mg per gloveBlue Color:Residual Powder: Average 0.05 – 0.10 mg;Violet Color:Residual Powder: Average 0.04 – 0.10 mg;
ASTM D412ASTM D573Testing for Physical property characteristicsBefore AgingTensile Strength: 14 MPa min.Ultimate Elongation: 500% min.Blue Color:
Before AgingTensile Strength: 14 MPa min.Ultimate Elongation: 400% min.XS Before AgingTensile Strength: ≥ 31.1 MPa;Ultimate Elongation: ≥ 505 %.After AgingTensile Strength: ≥ 30.1 MPa;Ultimate Elongation: ≥ 460%.S Before AgingTensile Strength: ≥ 31.3 MPa;Ultimate Elongation: ≥ 507 %.After AgingTensile Strength: ≥ 32.1 MPa;Ultimate Elongation: ≥ 457%.M Before AgingTensile Strength: ≥ 31.2 MPa;Ultimate Elongation: ≥ 506 %.After AgingTensile Strength: ≥ 28.2 MPa;Ultimate Elongation: ≥ 447%.L Before AgingTensile Strength: ≥ 30.9 MPa;Ultimate Elongation: ≥ 506 %.After AgingTensile Strength: ≥ 27.9 MPa;Ultimate Elongation: ≥ 426%.XL Before AgingTensile Strength: ≥ 31.0 MPa;Ultimate Elongation: ≥ 508 %.After AgingTensile Strength: ≥ 25.1 MPa;Ultimate Elongation: ≥ 408%.Violet Color:XS Before AgingTensile Strength: ≥ 24.1 MPa;Ultimate Elongation: ≥ 510 %.After AgingTensile Strength: ≥ 28.6 MPa;Ultimate Elongation: ≥ 408%.S Before AgingTensile Strength: ≥ 26.5 MPa;Ultimate Elongation: ≥ 515 %.After AgingTensile Strength: ≥ 29.8 MPa;Ultimate Elongation: ≥ 436%.M Before AgingTensile Strength: ≥ 26.3 MPa;Ultimate Elongation: ≥ 515 %.After AgingTensile Strength: ≥ 26.7 MPa;Ultimate Elongation: ≥ 462%.L Before AgingTensile Strength: ≥ 25.9 MPa;
Table 5 Performance Test Results Summary

{9}------------------------------------------------

ASTM D412ASTM D3767Testing For physicaldimensions specificationLength: 220 mm min. for size (XS, S);Length: 230 mm min. for size (M, L, XL);70±10 mm for XS; 80±10 mm for S;95±10 mm for M; 110±10 mm for L;120±10 mm for XL.Finger Thickness: ≥0.05 mm;Palm Thickness: ≥0.05 mm;All acceptance criteria above meet therequirements in Table 1 Dimensions andTolerances of ASTM D6319XL Ultimate Elongation: ≥ 510 %.After AgingTensile Strength: ≥ 25.6 MPa;Ultimate Elongation: ≥ 457%.Before AgingTensile Strength: ≥ 26.5 MPa;Ultimate Elongation: ≥ 517 %.After AgingTensile Strength: ≥ 25.0 MPa;Ultimate Elongation: ≥ 453%.Blue Color:Length of Size XS: ≥ 225mm;Width of Size XS: 70±8 (77-78) mm;Palm Thickness of Size S: ≥0.06 mm;Finger Thickness of Size S: ≥0.07 mm.Length of Size S: ≥ 246mm;Width of Size S: 80±7 (86-87) mm;Palm Thickness of Size S: ≥0.06 mm;Finger Thickness of Size S: ≥0.07 mm.Length of Size M: ≥ 240 mm;Width of Size M: 95±4 (98-99) mm;Palm Thickness of Size M: ≥0.06 mm;Finger Thickness of Size M: ≥0.08 mm.Length of Size L: ≥ 243mm;Width of Size L: 110±3 (107-108) mm;Palm Thickness of Size L: ≥0.06 mm;Finger Thickness of Size L: ≥0.08 mm.Length of Size XL: ≥ 243mm;Width of Size XL: 120±6 (114-116) mm;Palm Thickness of Size XL: ≥0.06 mm;Finger Thickness of Size XL: ≥0.08 mm.Violet Color:Length of Size XS: ≥ 225mm;Width of Size XS: 70±8 (77-78) mm;Palm Thickness of Size S: ≥0.06 mm;Finger Thickness of Size S: ≥0.07 mm.Length of Size S: ≥ 246mm;Width of Size S: 80±7 (86-87) mm;Palm Thickness of Size S: ≥0.06 mm;Finger Thickness of Size S: ≥0.07 mm.Length of Size M: ≥ 240 mm;Width of Size M: 95±4 (98-99) mm;Palm Thickness of Size M: ≥0.06 mm;Finger Thickness of Size M: ≥0.08 mm.Length of Size L: ≥ 242mm;Width of Size L: 110±3 (107-108) mm;Palm Thickness of Size L: ≥0.06 mm;Finger Thickness of Size L: ≥0.08 mm.Length of Size XL: ≥ 243mm;Width of Size XL: 120±6 (114-116) mm;Palm Thickness of Size XL: ≥0.06 mm;Finger Thickness of Size XL: ≥0.08 mm.
ISO 10993-11Evaluate the endpoint ofSystemic toxicity forbiocompatibilityThe test article showed no evidence ofsystemic toxicity potential from theextract.The test article showed no evidence ofsystemic toxicity from the extract.
ISO 10993-10Evaluate the endpoint ofirritant forbiocompatibilityEvaluate the endpoint ofsensitization forbiocompatibilityThe response of the test article extract isnegligible.The test article showed no evidence ofcausing delayed dermal contact sensitization.The test result showed that the response of thetest article extract was categorized asnegligible under the test condition.The test article showed no evidence of causing delayed dermal contact sensitization in the guinea pig.

{10}------------------------------------------------

9. Conclusion:

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, POWDER FREE Nitrile GLOVES (White, Cobalt Blue, Black, Ice Blue) cleared under K150340.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.