K Number
K211435

Validate with FDA (Live)

Date Cleared
2021-08-06

(88 days)

Product Code
Regulation Number
888.3040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PAO cortical screws are intended to refixate the acetabular fragment to the ilium, after Periacetabular osteotomy.

Device Description

The PAO cortical screws are implantable devices for acetabular refixation after a Periacetabular osteotomy. They are provided sterile and single-packaged, for single use only. The PAO cortical screws are designed in two different sizes (Ø3.5mm and Ø4.5mm), both available in different working lengths (from 20mm to 140mm and from 50mm to 140mm respectively) to be suitable to treat all the desired population. The PAO Cortical Screws are made of Stainless Steel AISI 316 LVM according to ISO 5832-1.

AI/ML Overview

This document pertains to the Medacta International SA PAO Cortical Screw, a medical device intended for refixating the acetabular fragment to the ilium after a Periacetabular osteotomy. As such, the information provided does not describe an AI/ML powered device, but rather a physical implantable device. Therefore, many of the requested categories related to AI/ML device studies are not applicable.

Here's the relevant information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Since this is a physical medical device (cortical screw) and not an AI/ML powered device, the "acceptance criteria" and "device performance" are related to mechanical and material properties, not diagnostic or predictive performance metrics. The document references performance testing according to ASTM standards.

Acceptance Criteria (Test)Reported Device Performance (Summary from input)
Design ValidationSpecific design validation was conducted for PAO Cortical Screws.
Performance Testing (ASTM F543-17):
- Torsional propertiesEvaluated and tested according to ASTM F543-17.
- Driving torqueEvaluated and tested according to ASTM F543-17.
- Axial pull-out strengthEvaluated and tested according to ASTM F543-17.
- Self-tapping performanceEvaluated and tested according to ASTM F543-17.
Pyrogenicity (Bacterial endotoxin test)Performed according to European Pharmacopoeia §2.6.14 (equivalent to USP chapter <85>). Results not explicitly stated, but implies compliance.
Pyrogenicity (Pyrogen test)Performed according to USP chapter <151>. Results not explicitly stated, but implies compliance. Device is not labeled as non-pyrogenic or pyrogen-free.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: Not explicitly stated in the document. The performance tests (e.g., ASTM F543-17) would have involved a certain number of screws, but the specific quantity is not reported.
  • Data Provenance: The tests are non-clinical, meaning they were conducted in a laboratory setting. There is no patient data involved in these non-clinical studies.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • Not applicable as this is a physical device and not an AI/ML device requiring expert ground truth for classification or diagnosis. The "ground truth" for mechanical properties is established by the specifications of the ASTM standard.

4. Adjudication Method for the Test Set:

  • Not applicable for a physical device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • Not applicable. This is not an AI/ML system being compared to human readers.

6. Standalone Performance Study (Algorithm only without Human-in-the-Loop):

  • Not applicable. This is a physical medical device.

7. Type of Ground Truth Used:

  • For the non-clinical performance and material tests, the "ground truth" is defined by the specifications and methodologies outlined in the referenced international standards (e.g., ASTM F543-17, ISO 5832-1, European Pharmacopoeia §2.6.14, USP chapter <85>, USP chapter <151>).

8. Sample Size for the Training Set:

  • Not applicable. There is no AI/ML algorithm involved, therefore no "training set."

9. How the Ground Truth for the Training Set was Established:

  • Not applicable. There is no AI/ML algorithm involved, therefore no "training set" or corresponding ground truth establishment.

Summary of the Study that Proves the Device Meets Acceptance Criteria:

The device's substantial equivalence to predicate devices and its safety and effectiveness are supported by non-clinical studies focused on its mechanical performance and material properties. These studies include:

  • Design Validation: A specific design validation for the PAO Cortical Screws was conducted.
  • Performance Testing: This involved static tests on the PAO Cortical Screws according to ASTM F543-17 Standard Specification and Test Methods for Metallic Medical Bone Screws. The evaluation specifically included torsional properties, driving torque, axial pull-out strength, and self-tapping performance.
  • Pyrogenicity Testing: Bacterial endotoxin tests were conducted according to European Pharmacopoeia §2.6.14 (equivalent to USP chapter <85>), and pyrogen tests were performed according to USP chapter <151>.

The document also states that the material used (Stainless Steel AISI 316 LVM according to ISO 5832-1) is shared with reference devices, addressing potential new safety and effectiveness questions related to material differences.

No clinical studies were conducted as part of this submission for substantial equivalence. The conclusion is that "the PAO cortical screws implants are substantially equivalent to the predicate devices" based on the provided non-clinical data, comparison of technological characteristics, and risk analysis.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 6, 2021

Medacta International SA % Chris Lussier Senior Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis. Tennessee 83118

Re: K211435

Trade/Device Name: PAO Cortical Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: May 7, 2021 Received: May 10, 2021

Dear Chris Lussier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, MPH Assistant Director DHT6A: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K211435

Device Name PAO Cortical Screw

Indications for Use (Describe)

The PAO cortical screws are intended to refixate the acetabular fragment to the ilium, after Periacetabular osteotomy.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. Submitter

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director of Quality and Regulatory, Medacta USA Date Prepared: May 7, 2021 Date Revised: August 5, 2021

II. Device

Device Proprietary Name:PAO Cortical Screw
Common or Usual Name:Screw, Fixation, Bone
Classification Name:Smooth or threaded metallic bone fixation fastener
Primary Product Code:HWC
Regulation Number:21 CFR 888.3040
Device Classification:II

III. Predicate Device

Substantial equivalence is claimed to the following devices:

Primary predicate device

  • DePuy Synthes 4.5 mm Cortex Screws, K161616, Synthes USA Products, LLC
    Other predicate devices:

  • DePuy Synthes 3.5 mm Cortex Screws, K112583, Synthes USA Products, LLC A

  • AAP CANNULATED SCREWS, K111316, AAP IMPLANTATE AG

In addition the following reference devices are cited within the submission:

  • Pelvic Trauma Screws, K171595, Medacta International SA

  • Ligament Staple, K210456, Medacta International SA A

IV. Device Description

The PAO cortical screws are implantable devices for acetabular refixation after a Periacetabular osteotomy. They are provided sterile and single-packaged, for single use only.

{4}------------------------------------------------

The PAO cortical screws are designed in two different sizes (Ø3.5mm and Ø4.5mm), both available in different working lengths (from 20mm to 140mm and from 50mm to 140mm respectively) to be suitable to treat all the desired population.

The PAO Cortical Screws are made of Stainless Steel AISI 316 LVM according to ISO 5832-1.

Indications for Use V.

The PAO cortical screws are intended to refixate the acetabular fragment to the ilium, after Periacetabular osteotomy.

VI. Comparison of Technological Characteristics

The PAO Cortical Screw implants and the predicate devices share the following characteristics:

  • shape and design; ●
  • . biocompatibility:
  • . device usage;
  • . sterility; and
  • packaging.

The PAO Cortical Screw implants differ from the predicate devices as follow:

  • sizes: and ●
  • . material.

Discussion

Medacta International SA has not made any change to the shape and design, device usage, biocompatibility, sterility, shelf life, and packaging of the subject devices respect to the predicate devices.

The indications for use of the subject and predicate devices can be considered substantially equivalent since the indications for use of the subject devices is totally included in the predicate device indications for use: both the subject and the predicate devices share the same indications with regards to the fixation of osteotomies.

The expanded range of sizes of the predicate devices with respect to the subject devices does not raise new questions of safety and effectiveness since the subject device's sizes are included in the predicate devices' range of products.

The different material of the subject and predicate devices does not introduce new questions of safety and effectiveness as well, since the AISI 316LVM according to ISO 5832-1 of the PAO Cortical Screw implants is shared with the reference devices, Pelvic Trauma Screws (K171595).

The comparison of technological characteristics and performance data provided within this submission, shows that there are no new risks associated with the subject devices design, and supports the substantial equivalence of the PAO cortical screws implants to the identified predicate devices.

{5}------------------------------------------------

VII. Performance Data

Based on the risk analysis, design validation and performance testing were conducted to written protocols. The following validation and tests are being provided in support of the substantial equivalence determination:

Non-Clinical Studies

  • . DESIGN VALIDATION
    • o PAO cortical Screws Design validation
  • PERFORMANCE TESTING ●
    • 0 PAO Cortical Screws Static test according to ASTM F543-17 Standard Specification and Test Methods for Metallic Medical Bone Screws including the evaluation and testing of:
      • Torsional properties,
      • Driving torque, ●
      • Axial pull-out strength, and ●
      • Self-tapping performance. ●
  • . PYROGENICITY:
    • o Bacterial endotoxin test (LAL test) according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>)
    • 0 Pyrogen test according to USP chapter <151> for pyrogenicity determination
    • o The subject devices are not labeled as non-pyrogenic or pyrogen free.

Clinical Studies:

  • No clinical studies were conducted.

VIII. Conclusion

The information provided above supports that the PAO cortical screws implants are substantially equivalent to the predicate devices.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.