K Number
K211374

Validate with FDA (Live)

Date Cleared
2021-10-13

(162 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free.

AI/ML Overview

This document is a 510(k) Premarket Notification for "Proclean - Nitrile Powder Free Examination Gloves", indicating it is a Class I medical device. The information provided heavily focuses on non-clinical performance and comparison to a predicate device, rather than a clinical study involving human patients or complex algorithms.

Here's a breakdown based on your request, with the understanding that this is for a physical medical device (gloves) and not an AI/software device:

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodStandardPurpose of TestingAcceptance CriteriaReported Device PerformanceStatus
Dimension (Length)ASTM D6319-19To determine the length of the glovesMin 230 mm for all sizesExtra Small: 230 mm, Small: 230 mm, Medium: 240 mm, Large: 236 mm, Extra Large: 236 mmPass
Dimension (Width)ASTM D6319-19To determine the width of the glovesExtra Small: 70 ± 10 mm, Small: 80 ± 10 mm, Medium: 95 ± 10 mm, Large: 110 ± 10 mm, Extra Large: 120 ± 10 mmExtra Small: 68mm, Small: 79 mm, Medium: 96 mm, Large: 110 mm, Extra Large: 120 mmPass
Dimension (Thickness)ASTM D6319-19To determine the thickness of the glovesPalm 0.05 mm min, Finger 0.05 mm min for all sizesExtra Small: Palm 0.11 mm, Finger: 0.11 mm; Small: Palm 0.09 mm, Finger: 0.10 mm; Medium: Palm: 0.08 mm, Finger: 0.08 mm; Large: Palm 0.10 mm, Finger: 0.11 mm; Extra Large: Palm 0.10 mm, Finger: 0.11 mmPass
Watertight testASTM D5151-19To determine the holes in the glovesSample size: 200 pcs, Inspection level: GI, AQL 1.5, Acceptance Number 7, Rejection Number 8The batch size for this sampling is 35,001-150,000. Sample code L equivalent to 200 pcs with accept 7 and reject 8. Results: Extra Small: 0 (Zero), Small: 0 (Zero), Medium: 0 (Zero), Large: 0 (Zero), Extra Large: 0 (Zero)Pass
Residual powderASTM D6124-06 (Reapproved 2017)To determine the residual powder2 mg per glove or lessSample size: 5 pcs. Results: Extra Small: 1.35 mg/glove, Small: 1.35 mg/glove, Medium: 0.50 mg/glove, Large: 1.34 mg/glove, Extra Large: 1.36 mg/glovePass
Physical Properties (Tensile Strength - Before Aging)ASTM D6319-19To determine physical properties14 Mpa Minimal for all sizesExtra Small: 18.08 MPa, Small: 18.37 MPa, Medium: 35.52 MPa, Large: 18.30 MPa, Extra Large: 18.20 MPaPass
Physical Properties (Tensile Strength - After Aging)ASTM D6319-19To determine physical properties14 Mpa Minimal for all sizesExtra Small: 14.01 MPa, Small: 14.08 MPa, Medium: 34.50 MPa, Large: 14.00 MPa, Extra Large: 14.00 MPaPass
Physical Properties (Ultimate Elongation - Before Aging)ASTM D6319-19To determine physical properties500% Min for all sizesExtra Small: 611 %, Small: 612%, Medium: 601 %, Large: 620%, Extra Large: 587%Pass
Physical Properties (Ultimate Elongation - After Aging)ASTM D6319-19To determine physical properties400% Min for all sizesExtra Small: 412 %, Small: 408%, Medium: 529%, Large: 403%, Extra Large: 416%Pass
Biocompatibility (Primary Skin Irritation)ISO 10993-10To assess skin irritationUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritantSame
Biocompatibility (Dermal Sensitization)ISO 10993-10To assess skin sensitizationUnder the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizerSame
Biocompatibility (In vitro cytotoxicity)ISO 10993-5To assess cytotoxicityUnder the conditions of the study, non-cytotoxicUnder the conditions of the study, non-cytotoxicSame
Biocompatibility (Acute Systemic Toxicity)ISO 10993-11To assess acute systemic toxicityUnder the conditions of study, the device extracts do not pose a systemic toxicity concernUnder the conditions of the study, did not induce any systemic toxicity.Same

2. Sample Size Used for the Test Set and Data Provenance

The document provides sample sizes for specific tests:

  • Watertight test: 200 pieces (sample code L) for a batch size of 35,001-150,000.
  • Residual powder: 5 pieces.
  • Other physical properties (Dimension, Tensile Strength, Ultimate Elongation): The document doesn't explicitly state a "sample size" for these in the same way as the watertight or residual powder tests, but results are provided across all five glove sizes (Extra Small, Small, Medium, Large, Extra Large). It is implied that typical statistical sampling methods according to ASTM D6319-19 were followed.

Data Provenance: The document does not explicitly state the country of origin of the data for these tests. The manufacturer is Dr.Boo Company Limited, located in Chonburi, Thailand. These are non-clinical, in-vitro/bench testing results, not human data. The testing is retrospective in the sense that it was performed and then summarized for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device is a physical product (gloves), not an AI/software device that requires expert review for ground truth establishment. The "ground truth" for these tests is defined by the objective measurements according to the specified ASTM and ISO standards.

4. Adjudication Method for the Test Set

Not applicable. This is not a clinical study involving human judgment that would require adjudication. Test results are objective measurements against established standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

Not applicable. This is not an AI/software device and no MRMC study was performed or is relevant for this type of product.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Not applicable. This is not an AI/software device. The tests performed are standalone in the sense that they measure the physical properties of the gloves directly against objective standards.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The ground truth for this device is based on objective measurements and performance against recognized international standards (ASTM D6319-19, ASTM D5151-19, ASTM D6124-06, ISO 10993-10, ISO 10993-5, ISO 10993-11). These standards define acceptable ranges or limits for various physical and biological properties of the gloves.

8. The Sample Size for the Training Set

Not applicable. There is no training set for this type of physical medical device. The "data" refers to manufacturing quality control and performance testing, not machine learning model training.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 13, 2021

Dr.Boo Company Limited Darunee Varabuntoonvit Marketing Director 662/1-4 Moo 3 Phanomsarakarm-Sattahip Rd, Kaokansong Sriracha. Chonburi 20110 Thailand

Re: K211374

Trade/Device Name: Proclean - Nitrile Powder Free Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: July 30, 2021 Received: September 7, 2021

Dear Darunee Varabuntoonvit:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K211374

Device Name

PRO CLEAN - NITRILE POWDER FREE EXAMINATION GLOVES

Indications for Use (Describe)

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Dr. BOO, a company that has been around since 1988. The logo features a hand with the words "Dr. BOO" written in the palm. The logo is blue and white, and it has a clean, modern design. The text "Since 1988" is written below the logo.

662/1-4 Moo 3 Phanomsarakarm-Sattahip Rd., Kaokansong, Sriracha, Chonburi, 20110 Thailand Tel: (66) 86 341 9085-6, (66) 38 290 244 Fax: (66) 38 290 245 Email: marketing@drboo.co.th Website: www.drboo.co.th

510(k) SUMMARY Nitrile Powder Free Examination Gloves

1.0 Appliance Information

Applicant:Address:Dr.Boo Company Limited662/1-4 Moo 3 Phanomsarakarm-Sattahip Rd,Kaokansong, Sriracha. Chonburi, 20110 Thailand
Phone Number:+66 86 341 9085
Fax Number:+66 38 290 245
Email:marketing@drboo.co.th
Name of Contact Person:Ms. Darunee Varabuntoonvit
Designation:Marketing Director
Contact Number:+66 81 617 2961
Contact Email:darunee@drboo.co.th
Date 510(k) summary prepared:July 30, 2021

2.0 Identification of the subject device

Trade/Proprietary Name(s):PROCLEAN - NITRILE POWDER FREE EXAMINATION GLOVES
Common Name:Nitrile Powder Free Examination Gloves
Classification Name:Patient Examination Gloves
Device Class:I
product codeLZA
Regulation Number:21 CFR 880.6250
Review Panel:General Hospital

3.0 Predicate Device

Device Name:Palm Care Blue Nitrile Examination Gloves Powder Free
510(k):K202384
Common Name:Patient Examination Gloves
Classification Name:Patient Examination Gloves
Device ClassificationI
Product Code:LZA
Regulation Number:21 CFR 880.6250

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for "Dr. BOO" and includes a hand graphic. The hand is in a light blue color and is positioned behind the text "Dr. BOO". The text is also in light blue and is inside of a circle. Below the logo, the text "Since 1988" is displayed.

DR.BOO COMPANY LIMITED 662/1-4 Moo 3 Phanomsarakarm-Sattahip Rd., Kaokansong, Sriracha, Chonburi, 20110 Thailand Tel: (66) 86 341 9085-6, (66) 38 290 244 Fax: (66) 38 290 245

Email: marketing@drboo.co.th Website: www.drboo.co.th

4.0 Description of the Device

Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free.

5.0 Indication for Use of the Device

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

6.0 Summary of Technological Characteristics of the Device Compared to the Predicate Device

CharacteristicsStandardsDevice performanceComparison
PredicateCurrent
510(k) Number-K202384New 510(k) submission----
Manufacturer(s)-Hi-Care Thai Gloves Co. LtdDr. Boo Company Limited----
Name of device-Blue NitrileExamination GlovesPowder freeBlue NitrileExamination GlovesPowder free----
Indication for UseMedical GlovesGuidance Manual -Issued on January22, 2008Blue NitrileExamination GlovesPowder-free is disposabledevices intended for medicalpurpose that are won on theexaminer's hand to preventcontaminationbetweenpatient and examinerA powder-free patientExamination Gloves is adisposable device intendedfor medical purposes that isworn on the examiner's handor finger to preventcontamination betweenpatient and examinerSimilar
MaterialASTM D6319-19Nitrile (NBR)Nitrile (NBR)Same
Color-BlueBlueSame
Texture-Textured FingersTextured FingersSame
SizeASTM D6319-19Extra SmallSmallMediumLargeExtra LargeExtra SmallSmallMediumLargeExtra LargeSame

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for "Dr. BOO" and includes a hand graphic. The hand is teal and is positioned above the text "Dr. BOO" which is inside of a circle. Below the circle is the text "Since 1988" also in teal.

662/1-4 Moo 3 Phanomsarakarm-Sattahip Rd., Kaokansong, Sriracha, Chonburi, 20110 Thailand Tel: (66) 86 341 9085-6, (66) 38 290 244 Fax: (66) 38 290 245 Email: marketing@drboo.co.th Website: www.drboo.co.th

CharacteristicsStandardsDevice performanceComparison
PredicateCurrent
Single UseMedical GlovesGuidance Manual -Issued on January22, 2008-LabelingSingle useSingle useSame
DimensionASTM D6319-19Meet ASTM D6319-19Length 230 mm min.Width 95 ± 10 mm min(for Medium size)Meet ASTM D6319-19Length 230 mm min.Width 95 ± 10 mm min(for Medium size)Same
ThicknessASTM D6319-19Meet ASTM D6319-19Finger: 0.05 mm minPalm: 0.05 mm minMeet ASTM D6319-19Finger: 0.05 mm minPalm: 0.05 mm minSame
PhysicalPropertiesASTM D6319-19Meet ASTM D6319-19Before agingTensile Strength:14 MPa minUltimate Elongation:500% minAfter agingTensile Strength14 MPa minUltimate Elongation:400% minMeet ASTM D6319-19Before agingTensile Strength:14 MPa minUltimate Elongation:500% minAfter agingTensile Strength14 MPa minUltimate Elongation:400% minSame
Watertight test(1000 ml)ASTM D5151-19Pass AQL 1.5Pass AQL 1.5Similar
Powder ResidueASTM D6124-06(Reapproved 2017)Meet≤ 2.0 mg/gloveMeet≤ 2.0 mg/gloveSimilar

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for "Dr. BOO" and it includes a hand graphic. The hand is in the background and the text "Dr. BOO" is in a circle in front of the hand. The text "Since 1988" is at the bottom of the image.

662/1-4 Moo 3 Phanomsarakarm-Sattahip Rd., Kaokansong, Sriracha, Chonburi, 20110 Thailand Tel: (66) 86 341 9085-6, (66) 38 290 244 Fax: (66) 38 290 245 Email: marketing@drboo.co.th Website: www.drboo.co.th

CharacteristicsStandardsDevice performance
PredicateCurrentComparison
BiocompatibilityPrimary SkinIrritation -ISO 10993-10 ThirdEdition 2010-08-01Under the conditionsof the study, not anirritantUnder the conditionsof the study, not anirritantSame
DermalSensitization -ISO 10993-10 ThirdEdition 2010-08-01Under the conditionsof the study, not asensitizerUnder the conditionsof the study, not asensitizerSame
In vitro cytotoxicityISO10993-5:2009(E)Under the conditionsof the study, noncytotoxicUnder the conditionsof the study, non-cytotoxicSame
Acute SystemicToxicity ISO10993-11:2017(E)Under the conditions ofstudy, the device extracts donot pose a systemic toxicityconcernUnder the conditions of thestudy, did not induce anysystemic toxicity.Same

There are no significant differences between the two products and are identical in terms of intended use, materials, design, manufacturing methods. Both devices meet the ASTM standards.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image is a logo for "Dr. BOO" and includes a hand with a circle around the brand name. The text "Since 1988" is located below the logo. The logo is a light blue color and has a clean, modern design.

662/1-4 Moo 3 Phanomsarakarm-Sattahip Rd., Kaokansong, Sriracha, Chonburi, 20110 Thailand Tel: (66) 86 341 9085-6, (66) 38 290 244 Fax: (66) 38 290 245 Email: marketing@drboo.co.th Website: www.drboo.co.th

7.0 Non-clinical testing summary

Performance Data

Test MethodStandardPurpose of testingAcceptance CriteriaResultStatus
ASTM D6319-19Standard Specification for NitrileExamination Gloves for MedicalApplicationTo determine the lengthof the glovesMin 230 mm for all sizesExtra Small: 230 mmSmall: 230 mmMedium: 240 mmLarge: 236 mmExtra Large: 236 mmPass
DimensionASTM D6319-19Standard Specification for NitrileExamination Gloves for MedicalApplicationTo determine the widthof the glovesExtra Small: 70 $\pm$ 10 mmSmall: 80 $\pm$ 10 mmMedium: 95 $\pm$ 10 mmLarge: 110 $\pm$ 10 mmExtra Large: 120 $\pm$ 10 mmExtra Small: 68mmSmall: 79 mmMedium: 96 mmLarge: 110 mmExtra Large: 120 mmPass
ASTM D6319-19Standard Specification for NitrileExamination Gloves for MedicalApplicationTo determine thethickness of the glovesPalm 0.05 mm minFinger 0.05 mm min for all sizesExtra Small: Palm 0.11 mm, Finger: 0.11 mmSmall: Palm 0.09 mm, Finger: 0.10 mmMedium: Palm: 0.08 mm, Finger: 0.08 mmLarge: Palm 0.10 mm, Finger: 0.11 mmExtra Large: Palm 0.10 mm, Finger: 0.11 mmPass
Test MethodStandardPurpose of testingAcceptance CriteriaResultStatus
Watertight testASTM D5151-19Standard Test Method for Detectionof Holes in MedicalGlovesTo determine the holes inthe glovesSample size: 200 pcsInspection level : GlAQL 1.5Acceptance Number 7Rejection Number 8The batch size for this sampling is35,001-150,000. Hence, according to thesingle sampling plan GI, the sample to bedrawn is under code L equivalent to 200pcs with accept 7 and reject 8 to beaccept under AQL 1.5Extra Small: : 0 (Zero)Small: 0 (Zero)Medium: 0 (Zero)Large:0 (Zero)Extra Large:0 (Zero).Pass
ResidualpowderASTM D6124-06 (Reapproved 2017)Standard Test Method for ResidualPowder on Medical GlovesTo determine the residualpowder in the gloves2 mg per glove or lessSample size : 5 pcsRequirement: 2 mg per glove or lessResultExtra Small: 1.35 mg/gloveSmall:1.35 mg/gloveMedium: 0.50 mg/gloveLarge: 1.34 mg/gloveExtra Large:1.36 mg/glovePass
PhysicalPropertiesASTM D6319-19Standard Specification for NitrileExamination Gloves for MedicalApplicationTo Determine thephysical properties-Tensile strengthBefore AgeingTensile Strength 14MpaMinimal for all sizesAfter Ageing TensileStrength 14MpaMinimal for all sizesBefore AgeingExtra Small: 18.08 MPaSmall: 18.37 MPaMedium: 35.52 MPaLarge: 18.30 MPaExtra Large: 18.20 MPaAfter Ageing:Extra Small: 14.01 MPaSmall: 14.08 MPaMedium: 34.50 MPaLarge: 14.00 MPaExtra Large: 14.00 MPaPass
Test MethodStandardPurpose of testingAcceptance CriteriaResultStatus
PhysicalPropertiesASTM D6319-19Standard Specification for NitrileExamination Gloves for MedicalApplicationTo Determine thephysical properties-Ultimate ElongationBefore AgeingUltimate Elongation500% Min for all sizesAfter Ageing UltimateElongation 400% Minfor all sizesBefore Ageing:Extra Small: 611 %Small: 612%Medium: 601 %Large:620%Extra Large:587%After Ageing:Extra Small: 412 %Small:408%Medium: 529%Large:403%Extra Large:416%Pass

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a logo for Dr. BOO, a company that has been around since 1998. The logo is a square with rounded corners, and it features a hand with the fingers spread apart. The text "Dr. BOO" is written in a bold, sans-serif font inside of a circle. The text "Since 1998" is written below the square.

662/1-4 Moo 3 Phanomsarakarm-Sattahip Rd., Kaokansong, Sriracha, Chonburi, 20110 Thailand Tel: (66) 86 341 9085-6, (66) 38 290 244 Fax: (66) 38 290 245 Email: marketing@drboo.co.th Website: www.drboo.co.th

Since 1988

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image is a logo for "Dr. BOO". The logo is a square with rounded corners, and it is colored in a light blue. Inside the square is a hand with the words "Dr. BOO" written on top of it. The hand is also colored in light blue.

662/1-4 Moo 3 Phanomsarakarm-Sattahip Rd., Kaokansong, Sriracha, Chonburi, 20110 Thailand Tel: (66) 86 341 9085-6, (66) 38 290 244 Fax: (66) 38 290 245 Email: marketing@drboo.co.th Website: www.drboo.co.th

Since 1988

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image is a logo for "Dr. BOO" and includes a hand graphic. The hand is white with a blue outline and is positioned inside of a blue circle. The text "Dr. BOO" is in blue and is located in the center of the circle. Below the circle, the text "Since 1988" is written in blue.

662/1-4 Moo 3 Phanomsarakarm-Sattahip Rd., Kaokansong, Sriracha, Chonburi, 20110 Thailand Tel: (66) 86 341 9085-6, (66) 38 290 244 Fax: (66) 38 290 245 Email: marketing@drboo.co.th Website: www.drboo.co.th

8.0 Clinical Testing Summary

Not applicable - Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

9.0 Conclusion

The conclusions drawn from the non-clinical test demonstrate that the subject device Nitrile Powder Free Examination Glove is as safe, as effective, and performs as well as or better than the legally marketed predicate device K202384.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.