K Number
K210815

Validate with FDA (Live)

Date Cleared
2021-05-13

(56 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

Device Description

The Disposable Face Mask is single use, three-layer, flat-pleated mask with ear loops and nose clip. The inner and outer layers of the mask are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene. The ear loops are held in place over the users' mouth and nose by two elastic ear straps. The ear loops are made of spandex and nylon. The nose clip is a malleable iron wire covered by polyethylene. Users can adjust the nose clip according to the shape of the nose and fix the mask on the bridge of the nose to prevent the mask from falling off. The Disposable Face Mask is blue and provided non-sterile.

AI/ML Overview

The document describes the non-clinical tests performed on the Disposable Face Mask (K210815) to demonstrate its safety and effectiveness, and its substantial equivalence to the predicate device (K202211).

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

No.Performance TestAcceptance CriteriaReported Device Performance (Average)
1Bacterial Filtration (BFE)Level 2: >98%Lot 1: 99.40%, Lot 2: 99.45%, Lot 3: 99.46% (All >98%)
2Differential PressureLevel 2: <6.0 mm H2O/cm²Lot 1: 2.05 mm H2O/cm², Lot 2: 1.95 mm H2O/cm², Lot 3: 2.03 mm H2O/cm² (All <6.0 mm H2O/cm²)
3FlammabilityLevel 2: Class IClass I (All lots)
4Particulate FiltrationLevel 2: >98%Lot 1: 99.50%, Lot 2: 99.46%, Lot 3: 99.47% (All >98%)
5Resistance to PenetrationLevel 2: 120 mmHgPass at 120 mmHg (All lots)
Biocompatibility
In Vitro Cytotoxicity TestNo Cytotoxicity ObservedPassed: No Cytotoxicity Observed
Skin Sensitization TestNo Skin Sensitization ObservedPassed: No Skin Sensitization Observed
Skin Irritation TestNo Skin Irritation ObservedPassed: No Skin Irritation Observed

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size for performance tests: 32 samples per lot for 3 non-consecutive lots (total 96 samples for each performance test).
  • Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. It refers to "Non-Clinical Test Conclusion," implying laboratory testing rather than human subject studies. Given the manufacturer's location in China (Biobase Scientific (Shandong) Co., Ltd.), it's highly probable the testing was conducted in China.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • Not Applicable (N/A): This study involves non-clinical laboratory testing of a physical device (face mask) against established performance standards, not diagnostic or interpretive tasks requiring human expert ground truth.

4. Adjudication Method for the Test Set

  • Not Applicable (N/A): This study does not involve human adjudication for ground truth. It relies on objective laboratory measurements against defined performance criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable (N/A): This document describes the testing of a medical device (face mask), not an AI-powered diagnostic tool. Therefore, an MRMC comparative effectiveness study is irrelevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable (N/A): This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

  • Objective Laboratory Measurements: The "ground truth" for the device's performance is established through quantitative measurements obtained from standardized laboratory tests (e.g., Bacterial Filtration Efficiency, Differential Pressure, Particulate Filtration, Flammability, Resistance to Penetration, and Biocompatibility tests) according to recognized international and national standards (ASTM, EN, ISO, CFR).

8. The Sample Size for the Training Set

  • Not Applicable (N/A): This is a physical medical device, not a machine learning model, so there is no "training set."

9. How the Ground Truth for the Training Set was Established

  • Not Applicable (N/A): There is no training set for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

May 13, 2021

Biobase Scientific (Shandong) Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box. 120-119 Shanghai, 200120 China

Re: K210815

Trade/Device Name: Disposable Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: March 15, 2021 Received: March 18, 2021

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence w. Murray, III, Ph.D. Acting Assistant Director Sterility Device Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K210815 Device Name Disposable Face Mask

Indications for Use (Describe)

The Disposable Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMBnumber."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The

assigned 510(k) Number: K210815

    1. Date of Preparation: 05/06/2021
    1. Sponsor Identification

Biobase Scientific (Shandong) Co., Ltd.

NO.1777 Dazheng Road, High-Tech District, Jinan City, Shandong Province, China. Establishment

Registration Number: 3016747986.

Contact Person: Qingxiang Hou Position: RA Specialist Tel: +86-531-81219858 Fax: +86-531-81219804 Email: 17806167195@163.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)

Mid-Link Consulting Co., Ltd.

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Disposable Face Mask Common Name: Surgical Face Mask Model: BK-DSFM Size: 175mm×95mm

Regulatory Information Classification Name: Mask, Surgical Classification: II; Product Code: FXX; Regulation Number: 21CFR 878.4040 Review Panel: General Hospital

Indication for use:

The Disposable Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

Device Description:

The Disposable Face Mask is single use, three-layer, flat-pleated mask with ear loops and nose clip. The inner and outer layers of the mask are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene. The ear loops are held in place over the users' mouth and nose by two elastic ear straps. The ear loops are made of spandex and nylon. The nose clip is a malleable iron wire covered by polyethylene. Users can adjust the nose clip according to the shape of the nose and fix the mask on the bridge of the nose to prevent the mask from falling off. The Disposable Face Mask is blue and provided non-sterile.

5. Identification of Predicate Device

510(k) Number: K202211 Product Name: Disposable Medical Surgical Face Masks

{5}------------------------------------------------

  • Summary of Technological Characteristics 6.
    Provided below is the technological characteristic comparing the subject device and the predicate device.
ITEMSubject Device K210815Predicate Device K202211Comparison
Product CodeFXXFXXSame
Regulation No.21 CFR 878.404021 CFR 878.4040Same
ClassIIIISame
Indication for UseThe Disposable Face Mask isintended to be worn to protect boththe patient and healthcarepersonnel from transfer ofmicroorganisms, body fluids andparticulate material. The facemask is intended for use ininfection control practices toreduce potential exposure to bloodand body fluids. The face mask isa single use, disposabledevice, provided non-sterile.The Disposable Medical SurgicalFace Masks are intended to beworn to protect both the patientand healthcare personnel fromtransfer of microorganisms, bodyfluids and particulate material.These face masks are intended foruse in infection control practicesto reduce the potential exposure toblood and body fluids. This is asingle use, disposable device(s),provided non-sterile.Same
Design featureEar loopEar loopSame
ColorBlueBlueSame
Dimension175mm×95mm175mm×95mmSame
LevelLevel 2Level 2Same
SterilityNon-SterileNon-SterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
Fluid resistance32 out of 32 per lot pass at 120mmHg, 3 non-consecutive lotstested32 out of 32 per lot pass at 120mmHg, 3 non-consecutive lotstestedSame
ParticulateFiltrationEfficiency3 non-consecutive lots tested,using a sample size of 32/lot.Lot 1: 99.30%Lot 2: 99.46%Lot 3: 99.47%3 non-consecutive lots tested,using a sample size of 32/lot.Lot 1:99.68%Lot 2:99.56%Lot 3:99.81%Different

Table 1 Comparison of Disposable Face Mask

Page 3 of 6

{6}------------------------------------------------

510(k) Summary
3 non-consecutive lots tested, using a sample size of 32/lot.3 non-consecutive lots tested, using a sample size of 32/lot.
BacterialFiltrationEfficiencyLot 1: 99.40%Lot 1:99.9%Different
Lot 2: 99.45%Lot 2:99.9%
Lot 3: 99.46%Lot 3:99.9%
3 non-consecutive lots tested, using a sample size of 32/lot.Lot 1: $2.05 mm H_2O/cm^2$3 non-consecutive lots tested, using a sample size of 32/lot.Lot 1: $3.6 mm H_2O/cm^2$Different
Differential PressureLot 2: $1.95 mm H_2O/cm^2$Lot 2: $3.6 mm H_2O/cm^2$
Lot 3: $2.03 mm H_2O/cm^2$Lot 3: $3.7 mm H_2O/cm^2$
Class 1, 3 non-consecutive lotsClass 1, 3 non-consecutive lotsSame
Flammabilitytested, using a sample size of32/lot.tested, using a sample size of32/lot.
Label/LabelingComplied with 21 CFR part 801Complied with 21 CFR part 801Same
Patient Contacting Material
Outer facing layerSpun-bond polypropyleneSpun-bond nonwoven fabric
Middle layerMelt blown polypropylene filterMelt blown non-woven fabric
Inner facing layerSpun-bond polypropyleneSpun-bond nonwoven fabricDifferent
Nose clipMalleable polyethylene wire andironPE and iron wire
Ear loopSpandex and nylonSpandex and nylon
Biocompatibility
CytotoxicityISO 10993-5 and ISO 10993-10;ISO 10993-5 and ISO 10993-10;
Under the conditions of the study,Under the conditions of the study,
Sensitizationthe proposed device extract wasthe proposed device extract was
determined to be non-cytotoxic,determined to be non-cytotoxic,Same
non-sensitizing, andnon-sensitizing, and
Irritationnon-irritating.non-irritating.

7. Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

  • 16 CFR Part 1610 Standard for the Flammability of Clothing Textiles;

  • ♪ ASTM F1862/F1862M-17 Standard Test Method for Resistance of Medical Face Masks to

Page 4 of 6

{7}------------------------------------------------

510(k) Summary

Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)

  • ASTM F2299/F2299M-03 (2017) Standard Test Method for Determining the Initial Efficiency of Material Used in medical Face Masks to Penetration by Particulates using Latex Spheres

  • ASTM F2101-2019 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus

  • ハ ASTM F2100-2019 Standard Specification for Performance of Materials Used in Medical Face Masks
  • EN 14683: 2019 Medical face masks- Requirements and test methods

  • ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity

  • ISO 10993-10:2010 Biological evaluation of medical device- Part 10: Tests for irritation and skin sensitization

No.PerformanceTestSampleSizeTestMethAcceptanceCriteriaTest ResultEvidence
Lot #1:20070721Lot #2:20071021Lot #3:20072221Reports:
1BacterialFiltrationn32ASTMF2100-19Level 2: >98%Min:99.03%Avg:99.40%Min:99.22%Avg:99.45%Min:99.27%Avg:99.46%041 to 043;Passed
2DifferentialPressure32Annex C,EN14683:2019Level 2: <6.0mm H2O/cm²Max: 2.55Avg: 2.05Max: 2.75Avg: 1.95Max: 2.86Avg: 2.03044 to 046;Passed

Table 1: Summary of Performance Testing

{8}------------------------------------------------

510(k) Summary
3Flammability16 CFRPart 1610Level 2: Class IClass IClass IClass I047 to 049;Passed
4ParticulateFiltrationF2299/F2299M-03Level 2: >98%Min:99.24%Avg:99.50%Min:99.25%Avg:99.46%Min:99.33%Avg:99.47%050 to 052;Passed
5Resistance toPenetrationASTMF1862/F1826M-17Level 2: 120 mmHgPass at120mmHgPass at120mmHgPass at120mmHg053 to 055;Passed

Table 2: Biocompatibility Test Results

TestStandardReportResult
In Vitro Cytotoxicity TestISO 10993-5-20092020-0451-001Passed: No Cytotoxicity Observed
Skin SensitizationTestISO 10993-10:20102020-0451-002Passed: No SkinSensitization Observed
Skin irritation TestISO 10993-10:20102020-0451-003Passed: No Skinirritation Observed

8. Clinical Test Conclusion

No clinical study is included in this submission.

9. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K202211.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.