K Number
K202211

Validate with FDA (Live)

Date Cleared
2020-12-22

(138 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Medical Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The Disposable Medical Surgical Face Masks are composed of mask body, nose clip and ear loop. The body of the mask is composed of three layers: the inner and outer layers are made of spun-bond nonwoven fabric, and the middle layer is made of melt blown non-woven fabric. The nose clip is made of PE and iron wire, ear loop is made of Nylon and Spandex. The size of the disposable surgical mask is 17.5*9.5cm with tolerance±1cm, the length of the ear loop is 18cm. The outer layer of disposable surgical mask will be provided in blue, the inner layer of the disposable surgical mask will be provided in white, and it will be provided with non-sterile and is intended to be single use, disposable devices.

AI/ML Overview

The provided document describes the acceptance criteria and study results for Disposable Medical Surgical Face Masks (K202211) manufactured by Guangdong Kaidi Garments Co., Ltd. The study focuses on demonstrating substantial equivalence to a predicate device (K153496) through non-clinical testing.

Here's the breakdown as requested:

1. Table of Acceptance Criteria and Reported Device Performance

ItemAcceptance Criteria (Level 2 Medical Mask)Reported Device Performance (K202211)Result
Fluid Resistance (ASTM F1862)29 out of 32 Pass at 120 mmHg32 out of 32 per lot pass at 120 mmHg, 3 non-consecutive lots testedPASS
Particulate Filtration Efficiency (ASTM F2299)≥ 98%Lot 1: 99.68%; Lot 2: 99.56%; Lot 3: 99.81% (3 non-consecutive lots tested)PASS
Bacterial Filtration Efficiency (ASTM F2101)≥ 98%Lot 1: 99.9%; Lot 2: 99.9%; Lot 3: 99.9% (3 non-consecutive lots tested)PASS
Differential Pressure (Delta P) (EN 14683 Annex C)< 6.0 mmH2O/cm²Lot 1: 3.6 mm H2O/cm²; Lot 2: 3.6 mm H2O/cm²; Lot 3: 3.7 mm H2O/cm² (3 non-consecutive lots tested)PASS
Flammability (16 CFR 1610)Class 1Class 1 (3 non-consecutive lots tested)PASS
Cytotoxicity (ISO 10993-5)Non-CytotoxicNon-cytotoxicPASS
Irritation (ISO 10993-10)Non-IrritatingNon-irritatingPASS
Sensitization (ISO 10993-10)Non-SensitizingNon-sensitizingPASS

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for performance tests: For each performance test (Fluid Resistance, Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability), 3 non-consecutive lots were tested, with a sample size of 32 units per lot. This means a total of 96 units were tested for each performance characteristic (3 lots * 32 samples/lot).
  • Sample Size for biocompatibility tests: Not explicitly stated but indicated as sufficient to determine cytotoxicity, irritation, and sensitization for the device materials, following ISO 10993 guidelines.
  • Data Provenance: The document does not explicitly state the country of origin of the data collection or if it was retrospective or prospective. However, given that Guangdong Kaidi Garments Co., Ltd is based in China, it is highly probable the testing was conducted in China. The data would be considered prospective as it was collected specifically for this 510(k) submission to demonstrate the device's performance against established standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This submission is for a medical mask and relies on non-clinical bench testing against established international and US standards (ASTM, EN, ISO, CFR). Therefore, there were no human experts establishing "ground truth" in the diagnostic sense (e.g., radiologists interpreting images). The "ground truth" is defined by the performance requirements outlined in the referenced standards for a Level 2 medical mask. The tests are objective, laboratory-based measurements.

4. Adjudication Method for the Test Set

Not applicable. As described above, the study involved objective bench testing using standardized protocols for physical and material properties, not human interpretation requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. A MRMC comparative effectiveness study was not performed. This type of study is typically relevant for diagnostic imaging AI algorithms to assess the impact of AI assistance on human reader performance. This submission is for a physical medical device (face mask) with performance validated via non-clinical testing.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not Applicable. This document pertains to a physical medical device (face mask), not an algorithm or AI software. Therefore, the concept of "standalone performance" of an algorithm is not relevant here.

7. Type of Ground Truth Used

The "ground truth" for the acceptance criteria and performance evaluation is based on established industry and regulatory standards for medical face masks. Specifically, for Level 2 masks, these include:

  • ASTM F2100: Standard Specification for Performance of Materials Used In Medical Face Masks
  • ASTM F1862: Resistance of Medical Face Masks To Penetration by Synthetic Blood
  • ASTM F2101: Bacterial Filtration Efficiency (BFE)
  • ASTM F2299: Particulate Penetration Using Latex Spheres
  • EN 14683: Medical Face Masks-Requirements and Test Methods (specifically for Differential Pressure)
  • 16 CFR 1610: Standard for the Flammability of Clothing Textiles
  • ISO 10993-5 and ISO 10993-10: Biological Evaluation of Medical Devices (Cytotoxicity, Irritation, Sensitization)

These standards define the objective performance metrics that the device must meet.

8. Sample Size for the Training Set

Not Applicable. This is a physical device, not an AI/ML algorithm that requires a training set. The term "training set" is typically used in the context of machine learning model development.

9. How the Ground Truth for the Training Set Was Established

Not Applicable. As there is no AI/ML algorithm or training set involved, the concept of establishing ground truth for a training set does not apply.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 22, 2020

Guangdong Kaidi Garments Co.,Ltd % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 13th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China

Re: K202211

Trade/Device Name: Disposable Medical Surgical Face Masks Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 30, 2020 Received: November 30, 2020

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202211

Device Name Disposable Medical Surgical Face Masks

Indications for Use (Describe)

The Disposable Medical Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

202211

ENXIYANG INDUSTRIAL ZONE, FENGSHAN STREET, HEPING TOWN, CHAOYANG, SHANTOU, CHINA 515000

510(K) Summary

Date of preparation: 2020-12-18

A. Applicant:

GUANGDONG KAIDI GARMENTS CO., LTD Address: SHENXIYANG INDUSTRIAL ZONE, FENGSHAN STREET, HEPING TOWN, CHAOYANG, SHANTOU, CHINA 515000 Contact Person: Vicky Chen Tel: +86- 754-83718111

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@@gmail.com

B. Device:

Trade Name: Disposable Medical Surgical Face Masks Common Name: Disposable Surgical Mask Model(s): NA

Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

C. Predicate device:

K153496 Disposable Surgical Face Mask

D. Indications for use of the device:

The Disposable Medical Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks

{4}------------------------------------------------

K202211

SHENXIYANG INDUSTRIAL ZONE, FENGSHAN STREET, HEPING TOWN, CHAOYANG, SHANTOU, CHINA 515000

are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

E. Device Description:

The Disposable Medical Surgical Face Masks are composed of mask body, nose clip and ear loop. The body of the mask is composed of three layers: the inner and outer layers are made of spun-bond nonwoven fabric, and the middle layer is made of melt blown non-woven fabric. The nose clip is made of PE and iron wire, ear loop is made of Nylon and Spandex.

The size of the disposable surgical mask is 17.5*9.5cm with tolerance±1cm, the length of the ear loop is 18cm. The outer layer of disposable surgical mask will be provided in blue, the inner layer of the disposable surgical mask will be provided in white, and it will be provided with non-sterile and is intended to be single use, disposable devices.

F. Comparison with predicate device

DeviceProposed DevicePredicate DeviceConclusion
ManufacturerGUANGDONG KAIDI GARMENTS CO., LTDXiantao Rayxin Medical Products Co., Ltd.NA
510K numberK202211K153496NA
ClassificationClass II Device, FXX (21 CFR878.4040)Class II Device, FXX (21 CFR878.4040)Same
Intended useThe Disposable Medical Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.Same
MaterialOuterSpun-bond nonwoven fabricSpun-bond polypropyleneSame
MiddleMelt blown non-woven fabricMelt blown polypropylene filterSame
Inner layerSpun-bond nonwoven fabricSpun-bond polypropyleneSame
Nose clipPE and iron wireMalleable aluminum wireDifferent
Ear loopsNylon and SpandexPolyesterDifferent
ColorBlueBlueSame
Dimension17.5cm+/-1cm17.5cm+/-1cmSame

Table 1 General Comparison

{5}------------------------------------------------

202211

GUANGDONG KAIDI GARMENTS CO., LTD

ENXIYANG INDUSTRIAL ZONE, FENGSHAN STREET, HEPING TOWN, CHAOYANG, SHANTOU, CHINA 515000

Dimension9.5cm+/-1cm9.5cm+/-1cmSame
OTC useYesYesSame
SterilityNon-SterileNon-SterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
ASTMF2100Level 2Level 2Same
BiocompatibilityISO10993ISO10993Same
Fluid Resistance32 out of 32 per lot pass at 120mmHg, 3 non-consecutive lots tested32 out of 32 Pass at 120 mmHgSimilar
ParticulateFiltrationEfficiency3 non-consecutive lots tested, using asample size of 32/lot.Lot 1: 99.68%Lot 2: 99.56%Lot 3: 99.81%98.46%Similar
BacterialFiltrationEfficiency3 non-consecutive lots tested, using asample size of 32/lot.Lot 1: 99.9%Lot 2: 99.9%Lot 3: 99.9%98.7%Similar
DifferentialPressure,3 non-consecutive lots tested, using asample size of 32/lot.Lot 1: 3.6 mm H2O/cm²Lot 2: 3.6 mm H2O/cm²Lot 3: 3.7 mm H2O/cm²4.2 mm H2O/cm2Similar
FlammabilityClass1, 3 non-consecutive lots tested,using a sample size of 32/lot.Class1Similar

From the comparison we found the material of the current nose clip and the ear loop were different from the predicate device. The biocompatibility tests were conducted to both components to ensure their compliance to the ISO10993-5 and ISO10993-10. There is no new risk generated from the difference of the material.

For the Performance testing, the test results are not identical to each other, but they are similar and they both meet the requirement of Level 2 medical mask according to the ASTM F 2100.

G. Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as to the predicate device. The test results demonstrated that the proposed device complies with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:

{6}------------------------------------------------

K202211

GUANGDONG KAIDI GARMENTS CO., LTD

SHENXIYANG INDUSTRIAL ZONE, FENGSHAN STREET, HEPING TOWN, CHAOYANG, SHANTOU, CHINA 515000

  • ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity A
  • ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization

  • A ASTM F2100, Standard Specification for Performance of Materials Used In Medical Face Masks
  • ASTM F1862, Standard Test Method for Resistance of Medical Face Masks To Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume At A Known Velocity);

  • EN 14683. Medical Face Masks-Requirements and Test Methods;

  • ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;

  • ASTM F2299, Standard test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;

  • 16 CFR 1610, Standard for the Flammability of clothing textiles;

ItemProposed deviceAcceptance Criteria (level 2)Result
Fluid ResistancePerformance ASTMF186232 out of 32 per lot pass at120 mmHg, 3 non-consecutive lots tested29 out of 32 Pass at 120 mmHgPASS
Particulate FiltrationEfficiency ASTM F22993 non-consecutive lots tested,using a sample size of 32/lot.Lot 1: 99.68%Lot 2: 99.56%Lot 3: 99.81%≥ 98%PASS
Bacterial FiltrationEfficiency ASTM F21013 non-consecutive lots tested,using a sample size of 32/lot.Lot 1: 99.9%Lot 2: 99.9%Lot 3: 99.9%≥ 98%PASS
Differential Pressure(Delta P) EN 14683Annex C3 non-consecutive lots tested,using a sample size of 32/lot.Lot 1: 3.6 mm H2O/cm²Lot 2: 3.6 mm H2O/cm²Lot 3: 3.7 mm H2O/cm²< 6.0mmH2O/cm²PASS
Flammability 16 CFR1610Class1, 3 non-consecutivelots tested, using a samplesize of 32/lot.Class 1PASS

Table 2 - Performance Testing

{7}------------------------------------------------

<202211

SHENXIYANG INDUSTRIAL ZONE, FENGSHAN STREET, HEPING TOWN, CHAOYANG, SHANTOU, CHINA 515000

ItemProposed deviceAcceptance CriteriaResult
CytotoxicityUnder the conditions of the study, thedevice is non-cytotoxic.Non-CytotoxicPASS
IrritationUnder the conditions of the study, thedevice is non-irritating.Non-IrritatingPASS
SensitizationUnder the conditions of the study, thedevice is non-sensitizingNon-SensitizingPASS

Table 3 Biocompatibility Comparison

H. Clinical Test Conclusion

No clinical study is included in this submission.

I. Conclusion

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, K153496 Xiantao Rayxin Medical Products Co., Ltd. Disposable Surgical Face Mask.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.