K Number
K210775

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-06-28

(105 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

White Nitrile Powder Free Patient Examination Glove, Non-Sterile meet all requirements of ASTM standard D6319 and FDA 21 CFR 880.6250. The powder free nitrile examination glove is manufactured from synthetic rubber latex. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves with ease without using any lubricant such as powder on the glove surface. The glove is ambidextrous, i.e., can be worn on right hand or left hand.

AI/ML Overview

The provided text describes the acceptance criteria and performance of the White Nitrile Powder Free Patient Examination Glove, Non-Sterile (K210775).

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicStandard/Test MethodAcceptance CriteriaReported Device Performance (Current White)Comparison Analysis
Manufacturer(s)--Onetexx Sdn BhdSame
MaterialASTM D6319NitrileNitrileSame
Color-WhiteWhiteSame
Physical Properties
Before AgingASTM D412Tensile Strength: Min 14.0 MPa Ultimate Elongation: Min 500%Tensile Strength: 32.35 MPa Ultimate Elongation: 568%Different but within the ASTM standard
After AgingASTM D412Tensile Strength: Min 14.0 MPa Ultimate Elongation: Min 400%Tensile Strength: 36.10 MPa Ultimate Elongation: 551%Different but within the ASTM standard
Thickness (Finger)ASTM D3767Min 0.05 mm0.10 mmDifferent but within the ASTM standard
Thickness (Palm)ASTM D3767Min 0.05 mm0.07 mmDifferent but within the ASTM standard
Powder FreeASTM D6124Less than 2 mg per glove0.24 mg/gloveDifferent but within the ASTM standard
Biocompatibility
Primary Skin IrritationISO 10993-10:2010 (E) & Consumer Product Safety Commission Title 16, Chapter II, Part 1500Non-irritant or non-sensitizerThe test material did not cause an irritant response. The Primary Irritant Response Category is deemed 'Negligible'.Similar
Dermal SensitizationISO 10993-10: 2010 (E) & Consumer Product Safety Commission, Title 16, Chapter II, Part 1500.3 (c) (4)Non-irritant or non-sensitizerThe test material did not produce a skin sensitization effect in the guinea pigs.Similar
Cytotoxicity - MEM ElutionISO 10993-5: 2009 (E)(Predicate: Not Tested)The test material demonstrated a cytotoxic effect under the condition of this study. Additional test i.e. Acute Systemic Toxicity was tested.Different
Acute Systemic ToxicityISO 10993-11: 2017 (E)(Predicate: Not Tested)The test item did not induce any systemic toxicity.Different
WatertightASTM D5151:2019Sample size: 500 pcs, Inspection level: G1, AQL: 1.5, Acceptance No. 10 (or equivalent based on batch size for sampling plan)Batch size 150,001 to 500,000, sample drawn: 315 pieces, accept: 10, reject: 11. 1 piece found with leaks. Falls within acceptance criteria. Gloves passed AQL 1.5.Similar
Dimension (Length)ASTM D3767Min 240 mmXS: 250mm, S: 250mm, M: 249mm, L: 249mm, XL: 250mmPass
Dimension (Width)ASTM D3767X-Small - 70 ± 10 mm, Small - 80 ± 10 mm, Medium - 95 ± 10 mm, Large - 110 ± 10 mm, X-Large – 120 ± 10 mmXS: 78mm, S: 87mm, M: 98mm, L: 107mm, XL: 118mmPass
Intended use-Disposable device for medical purposes, worn on hand/finger to prevent contamination.A patient examination glove is a disposable device made of synthetic rubber latex intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.Same
SizeMedical Glove Guidance Manual - LabelingSmall, Medium, Large, X Large (Predicate)Extra Small, Small, Medium, Large, Extra LargeSimilar
Single useMedical Glove Guidance Manual - LabelingSingle Patient Use (Predicate)Single UseSame

2. Sample size used for the test set and the data provenance

  • Physical Properties (Tensile Strength, Ultimate Elongation_): No specific sample size for a test set is provided beyond the general understanding that testing was conducted per ASTM D412. The results are reported as single values (e.g., 32.35 MPa), implying these are average or representative values from tests.
  • Dimension: Not explicitly stated, but measurements for XS, S, M, L, XL sizes are provided.
  • Watertight:
    • Sample Size: A batch size of 150,001 to 500,000 gloves was considered for sampling.
    • Actual Sample Drawn: 315 pieces as per single sampling plan GI for AQL 1.5.
  • Residual Powder:
    • Sample Size: 5 pieces.
  • Biocompatibility:
    • Specific sample sizes for Primary Skin Irritation, Dermal Sensitization, Cytotoxicity, and Acute Systemic Toxicity are not provided in the summary but typically involve animal testing (e.g., guinea pigs for sensitization).

Data Provenance: Not explicitly stated, but the testing would have been conducted by Onetexx Sdn Bhd (the submitter) or a contracted lab to meet the specified ASTM and ISO standards. Given the manufacturer's address in Malaysia, it is likely the testing was conducted in Malaysia or a recognized testing facility accessible to them. The studies are non-clinical (device performance tests).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This submission pertains to physical and biocompatibility testing of a medical glove, which relies on standardized objective test methods (ASTM, ISO) rather than expert consensus on interpretation. The "ground truth" is defined by the acceptance criteria set forth in these standards.

4. Adjudication method for the test set

Not applicable. The tests are objective measurements against established technical standards, not subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a submission for a patient examination glove, not an AI-powered diagnostic or assistive device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is for a physical medical device (glove), not a software or algorithm.

7. The type of ground truth used

The "ground truth" for the device's performance is based on objective measurements and adherence to established industry standards (ASTM, ISO) and regulatory requirements (21 CFR 880.6250). For example:

  • Physical properties (tensile strength, elongation, thickness) are measured against numerical thresholds defined by ASTM D412 and ASTM D3767.
  • Watertight integrity is assessed against the Acceptance Quality Limit (AQL) defined by ASTM D5151.
  • Residual powder is measured against a quantitative limit in ASTM D6124.
  • Biocompatibility (skin irritation, sensitization, cytotoxicity, systemic toxicity) is evaluated based on the criteria specified in ISO 10993 series, which typically involve observing biological responses in test models and determining if they meet "non-irritant," "non-sensitizer," or "no systemic toxicity" classifications.

8. The sample size for the training set

Not applicable. There is no "training set" as this is not a machine learning or AI device.

9. How the ground truth for the training set was established

Not applicable. There is no "training set" for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

June 28, 2021

Onetexx SDN BHD % Michael Van Der Woude U.S. Agent Emergo Global Representation 2500 Bee Cave Road, Building 1 Suite 300 Austin, Texas 78746

Re: K210775

Trade/Device Name: White Nitrile Powder Free Patient Examination Glove, Non Sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: April 1, 2021 Received: April 5, 2021

Dear Michael Van Der Woude:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ryan Ortega, Ph D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210775

Device Name WHITE NITRILE POWDER FREE PATIENT EXAMINATION GLOVE, NON STERILE

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1.0 Submitter:

Name:Freddy Low
Address:Onetexx Sdn BhdNo: 73-86, Jalan Logam 5, Perindustrian Kamunting 3,Kamunting Raya Industrial Estate, 34600 Kamunting, PerakDarul Ridzuan, Malaysia
Phone No.:+60 5 8070 666
Fax No.:+60 5 8070 666

Date of Summary Prepared: 24th June 2021

2.0 Identification of the subject device:

Trade Name: White Nitrile Powder Free Patient Examination Glove, Non-Sterile
Common Name: Patient Examination Gloves
Classification Name: Patient Examination Gloves
Device Classification: 1
Regulation Number: 21 CFR 880.6250
Product Code: LZA.

3.0 Predicate Device:

K171615

Xingyu Nitrile Powder Free Patient Examination Gloves, White Color Company: Shandong Xingyu Gloves Co., Ltd

4.0 Description of The Device:

White Nitrile Powder Free Patient Examination Glove, Non-Sterile meet all requirements of ASTM standard D6319 and FDA 21 CFR 880.6250.

The powder free nitrile examination glove is manufactured from synthetic rubber latex. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves with ease without using any lubricant such as powder on the glove surface. The glove is ambidextrous, i.e., can be worn on right hand or left hand.

5.0 Indication for use:

A patient examination qlove is a disposable device made of synthetic rubber latex intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

{4}------------------------------------------------

6.0 Comparison of the Technological Characteristics of the Device:

The White Nitrile Powder Free Patient Examination Glove, Non-Sterile are summarized with the following technological characteristics compared to ASTM D6319 or equivalent standards as shown in Table 1.

{5}------------------------------------------------

Table 1

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCECOMPARISONANALYSIS
PREDICATEWHITECURRENTWHITE
510(k) Number-K171615
Manufacturer(s)-Shandong Xingyu Gloves Co., LtdOnetexx Sdn BhdSame
MaterialASTM D6319NitrileNitrileSame
Color-WhiteWhiteSame
Physical PropertiesASTM D6319
Before AgingTensile Strength:Ultimate Elongation:18-25Mpa560-600%32.35Mpa568%Different butwithin the ASTMstandard
After AgingTensile Strength:Ultimate Elongation:17-21Mpa460-560%36.10Mpa551%Different butwithin the ASTMstandard
Thickness:- Finger- PalmASTM D63190.08mm0.08mm0.10mm0.07mmDifferent butwithin the ASTMstandard
Powder FreeASTM D6124Below 2mg of residual powder0.24 mg/gloveDifferent butwithin the ASTMstandard

{6}------------------------------------------------

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCECOMPARISONANALYSIS
PREDICATEWHITECURRENTWHITE
BiocompatibilityPrimary Skin Irritation- ISO 10993-10:2010(E) & ConsumerProduct SafetyCommission Title 16.Chapter II, Part 1500Under the conditions of thisstudy, the test article was anon-irritant or non-sensitizer.SKIN IRRITATION DERMAL andSENSITIZATION STUDIES MeetsISO 10993-10: Third Edition2010-08-01The test material did not causean irritant response. ThePrimary Irritant ResponseCategory is deemed 'Negligible'Similar
Dermal Sensitization-ISO 10993-10: 2010(E) & ConsumerProduct SafetyCommission, Title 16,Chapter II, Part1500.3 (c) (4)Under the conditions of thisstudy, the test article was anon-irritant or non-sensitizer.SKIN IRRITATION DERMAL andSENSITIZATION STUDIESMeets ISO 10993-10: ThirdEdition 2010-08-01The test material did notproduce a skin sensitizationeffect in the guinea pigs.Similar
BiocompatibilityCytotoxicity - MEMElution, ISO 10993-5:2009 (E)Not TestedThe test materialdemonstrated a cytotoxiceffect under the condition ofthis study. Additional test i.e.Acute Systemic Toxicity wastestedDifferent
DEVICE PERFORMANCE
CHARACTERISTICSSTANDARDSPREDICATECURRENTCOMPARISONANALYSIS
WHITEWHITE
BiocompatibilityAcute SystemicToxicity, ISO 10993-11:2017 (E)Not TestedThe test item did not induceany systemic toxicity.Different
Watertight (1000ml)ASTM D5151:2019Meets• 21 CFR 800.20• ASTM D6319-10(Reapproved 2015)Tested in accordance with ASTMD5151 (Reapproved 2015) withacceptable resultsGloves passed AQL 1.5Similar
Intended use-Powder Free Nitrile PatientExamination Gloves, WhiteColor is a disposable deviceintended for medical purposesthat is worn on the examiner'shand or finger to preventcontamination between patientand examiner.A patient examination glove is adisposable device made ofsynthetic rubber latex intendedfor medical purposes that isworn on the examiner's hand orfinger to prevent contaminationbetween patient and examiner.Same
SizeMedical GloveGuidance Manual -LabelingSmallMediumLargeX LargeExtra SmallSmallMediumLargeExtra LargeSimilar
Single useMedical GloveGuidance Manual -LabelingSingle Patient UseSingle UseSame

{7}------------------------------------------------

{8}------------------------------------------------

There are no significant differences between the two products and they are the same or similar in terms of intended use, materials design, physical properties, thickness and biocompatibility test.

7.0 Summary of Non-Clinical Testing

The performance test data of the non-clinical test for this powder free nitrile examination glove is summarized as per below.

{9}------------------------------------------------

TestMethodStandardPurpose of TestingAcceptance CriteriaResultsStatus
PhysicalPropertiesASTM D412(Standard TestMethod forVulcanized Rubberand ThermoplasticElastomers-Tension)To evaluate thetensile (tension)properties of glove.TensilestrengthMin 14.0MPaMin 14.0MPa32.35Mpa36.10MpaPass
UltimateelongationMin500%Min400%568%551%Pass
TestMethodStandardPurposeof TestingAcceptance CriteriaResults (in mm)Status
XSSMLXL
DimensionASTMD3767StandardPractice forRubber-Measurement ofDimensionsTo measurethe length,width andthickness ofgloveLengthMin 240 mm250250249249250Pass
WidthX-Small - 70 ± 10 mmSmall - 80 ± 10 mmMedium - 95 ± 10 mmLarge - 110 ± 10 mmX-Large – 120 ± 10 mm788798107118Pass
ThicknessFinger - min 0.05mm0.100.100.100.100.10Pass
Palm - min 0.05mm0.060.060.070.070.07

{10}------------------------------------------------

TestMethodStandardPurpose ofTestingAcceptance CriteriaResultsStatus
WatertightASTM D5151(Standard TestMethod forDetection of Holesin Medical Gloves)To detect holesthat leak waterand therebycompromise theusefulness of theglove.Sample size: 500 pcsInspection level: G1AQL: 1.5, Acceptance No. 10The batch size for thissampling is 150,001 to500,000. Hence,according to the singlesampling plan GI, thesample to be drawn isunder code Mequivalent to 315pieces with accept 10and reject 11 to beaccepted under AQL1.5. During the test, 1piece was found withleaks. Hence it fallswithin the acceptancecriteria.Pass
TestMethodStandardPurpose ofTestingAcceptance CriteriaResultsStatus
ResidualPowderASTM D6124(Standard TestMethod for ResidualPowder on MedicalGloves)To determine theamount ofresidual powderand non-powdersolids found ongloves.Less than 2 mg per gloveSample size : 5 pcsRequirement : <2mg/gloveResult:0.24mg/glovePass

{11}------------------------------------------------

8.0 Summary of Clinical Testing:

No clinical study is included in this submission.

9.0 Conclusion

The conclusion drawn from the non-clinical tests demonstrate that the subject White Nitrile Powder Free Patient Examination Glove, Non-Sterile is as safe, as effective, and performs as well as or better than the legally marketed predicate devices Xingyu Nitrile Powder Free Patient Examination Gloves, White Color (K171615).

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.