Intended Use

It is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes. The safety shield is intended to reduce the risk of sharp injuries.

Device Description

Verifine Mechanical Safety Insulin Syringe is a sterile device consisting of a calibrated barrel with plunger intended to be used to administer an injection of insulin to a patient subcutaneously. It includes an attached needle with a safety mechanism, i.e, after the injection, the protective shield will permanently be locked in place by pushing forward till click, providing protection against needle sticks and rendering the device unusable.

AI/ML Overview

This document is a 510(k) summary for the Verifine Mechanical Safety Insulin Syringe. It outlines the device's characteristics, its intended use, and a comparison to a predicate device to demonstrate substantial equivalence.

Here's the information extracted and structured as requested:

1. Table of Acceptance Criteria and the Reported Device Performance:

The document refers to various ISO and ASTM standards, implying that compliance with these standards serves as the acceptance criteria for different aspects of the device's performance. The "reported device performance" is generally stated as "Complied" or "met the USP acceptance criteria," indicating successful adherence to these standards.

Performance CharacteristicAcceptance Criteria (Standard)Reported Device Performance
Needle Tubing (Stainless Steel)ISO 9626:2016Bench testing performed verifies performance
Sterile Hypodermic NeedlesISO 7864:2016Bench testing performed verifies performance
Sterile Single-Use Syringes for InsulinISO 8537:2016Bench testing performed verifies performance
Sharps Injury ProtectionISO 23908:2011Bench testing performed verifies performance
Sharps Injury Prevention FeaturesFDA guidance, Medical Devices with Sharps Injury Prevention FeaturesSimulated clinical use testing conducted
Biocompatibility (Cytotoxicity)ISO 10993-5:2009No cytotoxicity
Biocompatibility (Skin Irritation)ISO 10993-10:2010No evidence of skin irritation
Biocompatibility (Skin Sensitization)ISO 10993-10:2010No evidence of sensitization
Biocompatibility (Acute Systemic Toxicity)ISO 10993-11:2006No systemic toxicity
Biocompatibility (Subacute Systemic Toxicity)ISO 10993-11:2017No systemic toxicity
Biocompatibility (Hemocompatibility)ISO 10993-4:2017No evidence of hemolysis
Biocompatibility (Pyrogen)USP42-NF37<151>Non pyrogenic
Particulate MatterUSP <788>Met the USP acceptance criteria
Sterilization (Ethylene Oxide)ISO 11135:2014Complied with standard
Package IntegrityASTM D4169-16All packaging deemed acceptable for protection of product and sterility maintenance
Seal StrengthASTM F88/F88-15Performed
Dye PenetrationASTM F1929-15Performed
Sterility TestUSP38-NF33 C71Performed
Shelf LifeASTM F1980-165 years validated
Hub/Needle bond strength (≥0.33mm)Implicit: Minimum shearing strength 22NImplied compliance via "Same" to predicate
Hub/Needle bond strength (<0.33mm)Implicit: Minimum shearing strength 11NImplied compliance via "Same" to predicate
Dead spaceImplicit: Maximum dead space 0.01mlImplied compliance via "Same" to predicate

2. Sample Sizes Used for the Test Set and the Data Provenance:

The document does not specify exact sample sizes for each test set. It mentions "bench testing" and "simulated clinical use testing," but specific numbers of devices tested are not provided. The data provenance is not explicitly stated in terms of country of origin of the data or whether it was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

This information is not provided in the document. The tests performed are primarily bench testing against established engineering and biocompatibility standards, rather than tests requiring expert human interpretation for "ground truth."

4. Adjudication Method for the Test Set:

This information is not applicable/provided. The tests described are objective measurements against defined standards, not subjective assessments requiring adjudication by multiple experts.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done:

No, an MRMC comparative effectiveness study was not conducted or described in this 510(k) summary. The device is a physical medical instrument (insulin syringe) and the evaluation focuses on its physical and biological performance characteristics, not on diagnostic accuracy requiring human reader interpretation with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Not applicable. This device is a mechanical insulin syringe, not an AI algorithm.

7. The Type of Ground Truth Used:

The "ground truth" for the tests performed is based on established international and national standards (e.g., ISO, ASTM, USP) for medical device performance, biocompatibility, sterility, and packaging. For example, "No cytotoxicity" is the ground truth for the cytotoxicity test if the criteria of ISO 10993-5 are met.

8. The Sample Size for the Training Set:

Not applicable. This device is a mechanical insulin syringe, not an AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. As above, this device does not involve a training set.

{0}------------------------------------------------

September 8, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Promisemed Hangzhou Meditech Co., Ltd. % Wei Hsu Regulatory Manager Vee Care (Asia) Limited 17th Chung Pont Commercial Building, 300 Hennessy Road Hong Kong, Hong Kong China

Re: K210712

Trade/Device Name: Verifine Mechanical Safety Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, FMF, FMI Dated: August 9, 2021 Received: August 9, 2021

Dear Wei Hsu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210712

Device Name

Verifine Mechanical Safety Insulin Syringe

Indications for Use (Describe)

It is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes. The safety shield is intended to reduce the risk of sharp injuries.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K210712 510(k) Summary

Date Prepared 1

Sep 7th, 2021

Submitter's Information 2

Name of Sponsor:

Promisemed Hangzhou Meditech Co., Ltd.

Address:

No. 1388 Cangxing Street, Canggian Community, Yuhang District, Hangzhou

City, 311121 Zhejiang, China

Contact Name:

Zearou Yang

Telephone No.:

+86 571 88772985

Fax No.:

+86 571 88772985

Email Address:

zearou.yang@promisemed.ca

Trade Name, Common Name, Classification ന

Trade/Product Name: Verifine Mechanical Safety Insulin Syringe

Classification name: Syringe, Piston

Regulation Number: 21 CFR 880.5860

Device Class: Class II

Product Code: FMF, FMI (hypodermic needle)

ldentification of Predicate Device ব

K193273: Insulin Syringe

Classification name: Syringe, Piston

Regulation Number: 21 CFR 880.5860

Device Class: Class II

Product Code: FMF

{4}------------------------------------------------

5 Description of the Device

Verifine Mechanical Safety Insulin Syringe is a sterile device consisting of a calibrated barrel with plunger intended to be used to administer an injection of insulin to a patient subcutaneously.

It includes an attached needle with a safety mechanism, i.e, after the injection, the protective shield will permanently be locked in place by pushing forward till click, providing protection against needle sticks and rendering the device unusable.

GaugeLengthWall Type
29G6 mm
29G8 mm
29G10 mm
29G13mm
30G6 mm
30G8 mm
30G10 mmRegular Wall
30G13mmThin Wall
31G4 mm
31G6 mm
31G8 mm
31G10 mm
31G13mm
32G4 mm

It can be used by health care personnel. This is a single-use device.

Indication for Use 6

CharacteristicsSubject DevicePredicate Device
Verifine Mechanical SafetyInsulin Syringe K210712Insulin SyringeK193273

{5}------------------------------------------------

Indication for UseIt is intended forInsulin Syringe is intended
subcutaneous injection offor subcutaneous injection
U-40 and U-100 insulin inof U-40 and U-100 insulin
the treatment of diabetes.in the treatment of diabetes.
The safety shield is
intended to reduce the
risk of sharp injuries.
Prescription Only orOver the counterPrescription OnlyPrescription Only

7 Similarities and Differences of the Proposed Devices to the Predicate Devices

The Verifine Mechanical Safety Insulin Syringe is substantially equivalent to the predicate device, the Insulin Syringe (K193273) in that these devices have same intended use and technological characteristics. The basic technological and operating principles are the same for both devices. Both the subject and predicate devices are disposable, sterile, single patient use devices. The differences above between the subject device and predicate device do not affect the basic design principle, usage of the subject device.

Subject DevicePredicate Device(K193273)
Trade NameVerifineMechanicalSafety Insulin SyringeInsulin SyringeComments
ManufacturerPromisemed HangzhouMeditech Co., LtdPromisemed HangzhouMeditech Co., Ltd
Device ClassClass IIClass IISame
Product CodeFMFFMFSame
Regulation number880.5860880.5860Same
Regulation NamePiston syringePiston syringeSame
Intended Use/Indications for UseIt is intended forsubcutaneous injectionof U-40 and U-100insulin in the treatmentof diabetes. The safetyshield is intended toreduce the risk of sharpinjuries.Insulin Syringe isintended forsubcutaneous injectionof U-40 and U-100 insulinin the treatment ofdiabetes.SimilarThe subjectdevice hasadditional safetyshield to preventsharp injury.
OperatingPrincipleThe insulin is injected tosubcutaneous tissue bypushing force generatedThe insulin is injected tosubcutaneous tissue bypushing force generatedDifferentThe subjectdevice has
through pushing plungerrod of the insulinsyringe. The protectiveshield will permanentlybe locked in place bypulling forward till click,providing protectionagainst needle sticks.through pushing plungerrod of the insulinsyringe.additional safetyshield to preventsharp injury.
Volume0.3ml,0.5ml, 1.0ml0.3ml,0.5ml, 1.0mlSame
Gauge32G, 31G, 30G, 29G32G, 31G, 30G, 29G, 28GDifferentThe gauge rangeof subject deviceis within that ofpredicate device.
Dead SpaceMaximum dead space0.01mlMaximum dead space0.01mlSame
Wall TypeRegular wall, Thin wallRegular wall, Thin wallSame
Needle Length4mm, 6mm, 8mm, 10mm,13mm6mm, 8mm, 12mmDifferentSubject devicehas wider needlelength thanpredicatedevice. Thespecification of4mm and 13mmmeets therequirement ofISO9626, ISO7864and ISO8537.
Needle tipconfiguration3 bevels3 bevelsSame
Nozzle typeNot applicableNot applicableSame
Numbering of scaleAt every five units forthe 0.3mL and 0.5mLsyringes, and at every 5or 10units for 1.0mLsyringeAt every five units forthe 0.3mL and 0.5mLsyringes, and at every10units for 1.0mLsyringeSame
GradationslegibilityLegibleLegibleSame
Needle cover colorRed (U-40) and orange(U-100)Red (U-40) and orange(U-100)Same
LubricantSilicone oilSilicone oilSame
Lubricantamount/cm²The lubricant is not formpools of fluid on theinterior surface of thesyringe or outsidesurfaces of the needletubeThe lubricant is not formpools of fluid on theinterior surface of thesyringe or outsidesurfaces of the needletubeSame
BarreltransparencyTransparentTransparentSame
Reuse durabilitySingle UseSingle UseSame
Hub/needlebondstrengthBond between hub andneedle tube should beBond between hub andneedle tube should beSame

A detailed comparison to the predicate is provided in Table 1.

{6}------------------------------------------------

{7}------------------------------------------------

Nominaloutsidediameterof needle(mm)MinimumshearingstrengthNominaloutsidediameterof needle(mm)Minimumshearingstrength
≥0.3322N≥0.3322N
<0.3311N<0.3311N
BiocompatibilityComplied with ISO10993series standards, and thefollowing tests areperformed- Cytotoxicity: Nocytotoxicity- Skin Irritation: Noevidence of skinirritation- Skin Sensitization: Noevidence of sensitization-Acute and SubacuteSystemic Toxicity: Nosystemic toxicity-Hemolysis: No evidenceof hemolysis-Pyrogen: Non pyrogenicSame
Configuration andMaterialsNeedle: Stainless Steel(SUS304)Barrel: PolypropylenePlunger: PolypropylenePush-bottom:PolypropyleneNeedle cap:PolyethylenePiston: PolyisoprenerubberSliding Sleeve, Cover:PolypropylenePlug, Supporting bar:Acrylonitrile ButadieneStyreneSame
LabelDevice name, indication,instruction for use,precaution, warning,shelf life, manufacturerSame
SterilizationEO SterilizationSame

8 Discussion of differences in technological characteristics

  • . Considering the skin thickness of different populations, additional length is added for patients to choose from. The additional length does not affect the effectiveness and safety of the device.

{8}------------------------------------------------

  • . The gauge range of subject device is within that of predicate device. The different gauge range does not impact the effectiveness and safety of the device.
  • . The operating procedure for insulin injection process is the same as predicate device. The subject device has additional safety shield to protect against sharps injury.

Performance Testing Summary の

The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:

  • ISO 9626:2016, Stainless steel needle tubing for the manufacture of . medical devices - Requirements and test methods
  • . ISO 7864:2016, Sterile hypodermic needles for single use -Requirements and test methods
  • ISO 8537:2016, Sterile single-use syringes, with or without needle, for ● insulin
  • ISO 23908:2011, Sharps injury protection Requirements and test . methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
  • Simulated clinical use testing was conducted as recommended in FDA . guidance, Medical Devices with Sharps Injury Prevention Features.

Biocompatibility

In accordance with ISO 10993-1 the device is classified as External communicating device, Blood path, indirect, with prolonged contact duration (> 24h to 30d). It is leverage for the predicate the following testing;

  • ISO 10993-5:2009 Biological Evaluation of Medical Devices -- Part 5: . Tests for in Vitro Cytotoxicity
    • ISO 10993-10:2010 Biological Evaluation of Medical Devices -- Part 10: Tests for Irritation and Skin Sensitization

{9}------------------------------------------------

  • ISO 10993-11:2006, Biological evaluation of medical devices Part 11: Tests for systemic toxicity-Acute systemic toxicity and pyrogen test
  • ISO 10993-11:2017, Biological evaluation of medical devices Part 11: Tests for systemic toxicity-Subacute systemic toxicity
  • ISO 10993-4:2017, Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood, Hemocompatibility
  • USP42-NF37<151> Pyrogen Test
  • Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.

Sterility, Shipping, and Shelf-Life

  • ISO 11135:2014, Sterilization of health-care products Ethylene . oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
  • Package integrity testing, after environmental conditioning and . simulated transportation in accordance with ASTM D4169-16, was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection of product and sterility maintenance.
  • Sterile Barrier Packaging Testing performed on the proposed device: .
    • Seal strength ASTM F88/F88-15
    • Dye penetration ASTM F1929-15
    • Sterility test USP38-NF33 C71
  • Shelf life of 5 years is validated using the FDA recognized standard . ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices

10 Conclusion

The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Verifine Mechanical Safety Insulin Syringe is substantially equivalent to the Insulin Syringe with respect to the indications for use, target populations, treatment method, and technological characteristic.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).