K Number
K210463

Validate with FDA (Live)

Date Cleared
2022-01-23

(341 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

Device Description

Thermoplastic Elastomer (TPE) Hybrid Examination Glove is a patient examination glove made from polyethylene resin and styrene-ethylene-styrene (SEBS) compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the Thermoplastic Elastomer (TPE) Hybrid Examination Glove:

1. Table of Acceptance Criteria and Reported Device Performance:

TestTest MethodPurposeAcceptance CriteriaReported Performance
DimensionASTM D3767Determine the geometrical dimension of glovesLength: 230 mm min.Thickness: Palm - 0.05 mm min., Finger - 0.05 mm min.Palm Width: Small 104 ± 5 mm, Medium 107 ± 5 mm, Large 115 ± 5 mm, X large 123 ± 5 mmPass
Freedom from holes (Water leak)21 CFR 800.20. & ASTM D5151-19Detect the holes on the gloves.G-I/AQL 2.5Pass
Tensile strength (Before aging/After aging)ASTM D412-16 & ASTM D573-04Evaluate the tensile properties of the gloves and the influence of elevated temperature.Before Aging: 14 MPa, min.After Aging: 14 MPa, min.Pass
Elongation (Before aging/After aging)ASTM D412-16 & ASTM D573-04Evaluate the elongation properties of the gloves and the influence of elevated temperature.Before Aging: 500% min.After Aging: 400% min.Pass
Powder ResidualASTM D6124-06Determine the average powder mass found on the gloves< 2mg per glovePass
Biocompatibility - Skin IrritationISO 10993-10:2010Determine the potential of glove to promote skin sensitization & irritation reactionsNegative ResponsePass
Biocompatibility - Skin SensitizationISO 10993-10:2010Determine the potential of glove to promote skin sensitization & irritation reactionsNo contact sensitizationPass
Biocompatibility - CytotoxicityISO 10993-5:2009Determine the cytotoxicity potential of gloveNo in vitro cytotoxicityPass

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample sizes used for each individual test. It states that "the following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device."

The provenance of the data is from Xuzhou Full Sun Medical Products Ltd. (China), as indicated by the submitter's address and the company branding throughout the document. The tests are non-clinical bench tests, meaning they were performed in a lab setting rather than on human subjects.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not applicable to this type of device and study. The "ground truth" for examination gloves is established by objective engineering and biocompatibility standards (e.g., ASTM, ISO standards) and not by expert consensus in the typical sense of medical image interpretation.

4. Adjudication Method for the Test Set:

This information is not applicable. The tests are objective measurements against defined acceptance criteria from established standards. There is no human interpretation or adjudication involved in the test results beyond the standard practices of laboratory testing and quality control.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. This device is a physical medical product (examination glove), not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and AI assistance are irrelevant to its evaluation.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable. As stated above, this is a physical medical product, not an algorithm.

7. The type of ground truth used:

The ground truth used for this device is based on established industry standards and regulatory requirements for medical examination gloves. Specifically:

  • Engineering Standards: ASTM D3767, ASTM D5151-19, ASTM D412-16, ASTM D573-04, ASTM D6124-06 for dimensions, freedom from holes, tensile strength, elongation, and powder residue.
  • Biocompatibility Standards: ISO 10993-10:2010 (Skin Irritation and Sensitization) and ISO 10993-5:2009 (Cytotoxicity).
  • Regulatory Standards: 21 CFR 800.20 (Freedom from holes).

These standards define the acceptable physical, mechanical, and biological properties a glove must possess to be deemed safe and effective for its intended use.

8. The Sample Size for the Training Set:

This information is not applicable. This is a physical product, not a machine learning model, so there is no concept of a "training set" in this context.

9. How the Ground Truth for the Training Set was Established:

This information is not applicable. As explained in point 8, there is no training set for a physical examination glove.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

January 23, 2022

Xuzhou Full Sun Medical Products Ltd. Elizabeth Deng U.S. Representative 5748 Eaglewood Place Rancho Cucamonga, California 91739

Re: K210463

Trade/Device Name: Thermoplastic Elastomer (TPE) Hybrid Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: December 20, 2021 Received: December 23, 2021

Dear Elizabeth Deng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210463

Device Name

Thermoplastic Elastomer (TPE) Hybrid Examination Glove

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Xuzhou Full Sun Medical Products LTD. The logo consists of the letters 'FS' in blue and green, with a green plus sign below the 'F'. To the right of the logo is the company name in Chinese, followed by the English translation 'XUZHOU FULL SUN MEDICAL PRODUCTS LTD.'

ounty. Jiangsu Province. 221316. China

510(k) SUMMARY

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number: K210463

0.0 Summary Preparation Date: Jan. 20th, 2022

1.0 Submitter:

Submitter's name:Xuzhou Full Sun Medical Products Ltd.
Submitter's address:Bihe Industry Area, Yitang Town, Pizhou CountyJiangsu, China 221316
Phone number:86 - 516 - 67680555
Fax number:86 - 516 - 67680098
Name of contact person:Hsun-Hui Huang

Elizabeth Deng

909 4659188

5748 Eaglewood Place

Rancho Cucamonga, California Rancho Cucamonga, CA 91739

baxianunited48(@yahoo.com

2.0 US Agent:

US Representative Name: Company Address:

Telephone Number: Contact Email Address:

3.0 Name of the Device

Proprietary/Trade name:

Common Name: Classification Name: Device Classification: Regulation Number: Product Code:

Thermoplastic Elastomer (TPE) Hybrid Examination Glove Polymer Examination Gloves Non-powdered Patient Examination Glove Class I 21 CFR 880.6250 LZA

4.0 Predicate device

Device Name:

Company name: 510(K) Number: Disposable Powder Free Nitrile Examination Glove, Pink/Black Color Ever Growth (Vietnam) Co., Ltd. K190942

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Xuzhou Full Sun Medical Products LTD. The logo consists of a stylized "FS" in blue and green, with a plus sign incorporated into the "S". To the right of the logo is the company name in both Chinese and English. The Chinese characters are larger and bolder than the English text.

Binhe Industry Area. nty, Jiangsu Province, 221316, Chi +86-516-67680090 · FAX :

5.0 Device Description:

Thermoplastic Elastomer (TPE) Hybrid Examination Glove is a patient examination glove made from polyethylene resin and styrene-ethylene-styrene (SEBS) compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

6.0 Device Indications for use:

A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

DevicePredicate DeviceSubject DeviceComparison
Characteristic
Product NameDisposable Powder Free NitrileExamination Glove, Pink/BlackColorThermoplastic Elastomer (TPE)Hybrid Examination GloveN/A
510(K) No.K190942K210463N/A
Product OwnerEver Growth (Vietnam) Co., Ltd.Xuzhou Full Sun MedicalProducts Ltd.N/A
Product CodeLZALZAsame
Regulation21 CFR 880.625021 CFR 880.6250same
ClassIIsame
Indications forUseThe Nitrile Powder Free patientexamination glove is a non-sterile disposable deviceintended for medical purposesthat is worn on the examiner'shands or finger to preventcontamination between patientand examiner.Thermoplastic Elastomer (TPE)Hybrid Examination Glove is anon-sterile disposable deviceintended for medical purposethat is worn on the examiner'shands or fingers to preventcontamination between patientand examiner.same
Power freeYesYessame
SizeX Small /Small/Medium/Large/X LargeSmall/Medium/Large/X Largesimilar
Single UseYESYESsame
Non-SterileYESYESSame

7.0 Comparison of device technological characteristics:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains a logo for Xuzhou Full Sun Medical Products LTD. The logo consists of a blue and green graphic on the left, followed by the company name in Chinese characters and English. The English translation of the company name is "XUZHOU FULL SUN MEDICAL PRODUCTS LTD."

Binhe Industry Area, Yitang Town, Pizhou County, Jiangsu Province, 221316, China TEL:+86-516-67680090・FAX:+86-516-67680098

DevicePredicate DeviceSubject DeviceComparison
Characteristic
Dimensions-LengthComplies with ASTM D6319-10230 mm min.Complies with ASTM D6319-19230 mm min.same
Dimensions -Palm WidthComplies with ASTM D6319X Small70 ± 10 mmSmall80 ± 10 mmMedium95 ± 10 mmLarge110±10 mmX large120 ±10 mmInternal StandardSmall104 ± 5 mmMedium107 ± 5 mmLarge115 ± 5 mmX large123 ± 5 mmSimilar. Dueto thecharacteristic of PEmaterial, weuse longerdimensionfor palmwidth.
Dimensions -ThicknessComplies with ASTM D6319Palm - 0.05 mm min.Finger - 0.05 mm min.Complies with ASTM D6319Palm - 0.05 mm min.Finger - 0.05 mm min.same
PhysicalPropertiesTensile StrengthBefore Aging: 14 MPa, min.After Aging: 14 MPa, min.Tensile StrengthBefore Aging: 14 MPa, min.After Aging: 14 MPa, min.same
Elongation:Before Aging: 500% min.After Aging: 400% min.Elongation:Before Aging: 500% min.After Aging: 400% min.same
ResidualpowderComplies with ASTM D6319< 2mg per gloveComplies with ASTM D6319< 2mg per glovesame
Freedom fromHolesIn accordance with ASTMD6319 and ASTM D5151 (G-1,with AQL 2.5)In accordance with ASTMD6319 and ASTM D5151 (G-1with AQL 2.5)same
BiocompatibilityISO 10993-10Skin Sensitization and SkinIrritation test: PassesISO 10993-5In vitro cytotoxicity test: PassesISO 10993-10Skin Sensitization and SkinIrritation test: PassesISO 10993-5In vitro cytotoxicity test: Passessame

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Xuzhou Full Sun Medical Products LTD. The logo consists of the letters 'FS' in blue and green, with a plus sign at the bottom left. To the right of the logo is the company name in Chinese characters, followed by the English translation 'XUZHOU FULL SUN MEDICAL PRODUCTS LTD.'

Binhe Industry Area, Yitang Town, Pizhou County, Jiangsu Province, 221316, China TEL:+86-516-67680090・FAX:+86-516-67680098

8.0 Assessment of Non-Clinical Performance Data:

The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device. Thermoplastic Elastomer (TPE) Hybrid Examination Glove made by Xuzhou Full Sun Medical Products Ltd. met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing:

TestTestMethodPurposeAcceptance CriteriaResults
DimensionASTMD3767Determine thegeometrical dimensionof glovesLength: 230 mm min.Thickness:Palm - 0.05 mm min.Finger - 0.05 mm min.Palm Width:Small 104 ± 5 mmMedium 107± 5 mmLarge 115 ± 5 mmX large 123 ± 5 mmPass
Freedom fromholes (Water leak)21 CFR800.20. &ASTMD5151-19Detect the holes on thegloves.G-I/AQL 2.5Pass
Tensile strength(Beforeaging/After aging)ASTMD412-16 &ASTMD573-04Evaluate the tensile(tenson) properties ofthe gloves. In addition,it also determines theinfluence of elevatedtemperature on theBefore Aging: 14 MPa, min.After Aging: 14 MPa, min.Pass
Elongation(Beforeaging/After aging)ASTMD412-16 &ASTMD573-04temperature on thephysical properties ofgloves.Before Aging: 500% min.After Aging: 400% min.Pass
Powder ResidualASTMD6124-06Determine the averagepowder mass found onthe gloves< 2mg per glovePass
Biocompatibility-Skin IrritationISO10993-10:2010determine the potentialof glove to promoteskin sensitization &Negative ResponsePass
Biocompatibility-Skin SensitizationISO10993-10:2010irritation reactions afterrepeated applicationsNo contact sensitizationPass
Biocompatibility-cytotoxicityISO10993-5:2009determine thecytotoxicity potentialof gloveNo in vitro cytotoxicityPass

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image displays the logo for Xuzhou Full Sun Medical Products LTD. The logo features a stylized "FS" in blue and green on the left. To the right of the emblem is the company name in Chinese characters, followed by the English translation "XUZHOU FULL SUN MEDICAL PRODUCTS LTD."

Binhe Industry Area, Yitang Town, Pizhou County, Jiangsu Province, 221316, China TEL:+86-516-67680090・FAX:+86-516-67680098

9.0 Assessment of Clinical Performance Data:

Clinical data is not needed for this type of device.

10.0 Conclusion:

The conclusion drawn from the nonclinical tests demonstrate that the subject device Thermoplastic Elastomer (TPE) Hybrid Examination Glove is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.