K Number
K210339

Validate with FDA (Live)

Date Cleared
2021-08-02

(178 days)

Product Code
Regulation Number
870.1210
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For Bullfrog devices capable of treating <6 mm diameter vessels: In selective areas of peripheral and coronary vessels, the Bullfrog Micro-Infusion Device is intended for the infusion of diagnostic and therapeutic agents into the vessel wall and perivascular area, or intraluminally.

For Bullfrog devices capable only of treating ≥6 mm diameter vessels: In selective areas of peripheral vessels, the Bullfrog Micro-Infusion Device is intended for infusion of diagnostic and therapeutic agents into the vessel wall and perivascular area, or intraluminally.

Device Description

The Mercator MedSystems Bullfrog Micro-Infusion Device is a wire-guided, singleoperator, endovascular catheter that consists of a perpendicular microneedle, which is sheathed by and contained within a semi-rigid polymer actuator balloon. The device is designed to be advanced to target vasculature and hydraulically actuated to move the microneedle through the external elastic lamina to deliver substances to adventitial and perivascular tissues. A compliant stabilizing balloon inflates with the actuator to provide a force opposite the needle tip for proper seating of the needle is retracted within the sheathing structure by vacuuming the hydraulic actuator.

AI/ML Overview
{
  "acceptance_criteria_and_performance_table": {
    "title": "Device Performance against Pre-determined Specifications and Acceptance Criteria",
    "headers": ["Test Category", "Test Performed", "Acceptance Criteria", "Reported Device Performance"],
    "rows": [
      ["Biocompatibility", "Biocompatibility Assessment", "Determined in accordance with ISO 10993", "Met"],
      ["Sterilization", "Sterilization testing", "Per ISO 11137-1:2006 and ISO 11137-2:2013", "Met"],
      ["Packaging", "Distribution testing", "Per ASTM D4169", "Met"],
      ["Packaging", "Pouch bubble testing", "Per ASTM F2096", "Met"],
      ["Packaging", "Pouch seal strength testing", "Per ASTM F88/F88M", "Met"],
      ["Labeling", "Label adhesion", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "System fluid delivery rate", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Catheter stiffness", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Actuator inflation and deflation time", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Actuator diameter minimum/maximum", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Actuator integrity after tortuous anatomy", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Actuator burst strength", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Drug line integrity", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Drug line burst strength", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Inflation line leakage under vacuum", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "System leakage under pressure", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Pressure relief valve cracking pressure and leak rate", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Radio-opaque markers intact", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Microneedle sheathing", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Microneedle insertion force at failure", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Microneedle durability", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Microneedle active length", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Actuator retraction force to failure and in vacuum", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Catheter bend kink resistance", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Tensile strength", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Torsional strength", "Not specified in document", "Met"],
      ["Mechanical and Fluid Delivery Performance", "Repetitive function and fatigue", "Not specified in document", "Met"]
    ]
  },
  "study_information": {
    "description": "The Bullfrog Micro-Infusion Device underwent non-clinical performance testing to demonstrate substantial equivalence and safety. This testing included mechanical and fluid delivery performance, biocompatibility, sterilization, and in-vivo safety studies. The testing adopted methodologies from prior versions of the device ([K161402](https://510k.innolitics.com/device/K161402)) for biocompatibility due to no material changes, and functional testing as defined in ISO 10555-1 with Amendments 1 and 2, along with customized testing for performance for changed components.",
    "sample_size_test_set": "Not specified in the provided document. The document refers to 'non-clinical testing of components' and 'Testing performed on the device' without detailing specific sample sizes for each test.",
    "data_provenance": "Not explicitly stated for all tests, but given the nature of the device (medical device) and the regulatory submission, the data is likely from laboratory testing conducted by Mercator MedSystems, Inc. The document refers to 'in-vivo safety studies' but does not provide details on their nature or origin. It is a retrospective summary of testing data for a 510(k) submission.",
    "number_of_experts_ground_truth_test_set": "Not applicable. The reported 'ground truth' or acceptance criteria for non-clinical performance tests are based on established standards (e.g., ISO, ASTM) and internal engineering specifications, not expert consensus on medical images or patient outcomes.",
    "qualifications_experts_ground_truth_test_set": "Not applicable. Ground truth for these non-clinical tests is based on published standards and engineering specifications.",
    "adjudication_method_test_set": "Not applicable. Adjudication methods are typically for clinical studies with subjective interpretations. Non-clinical tests usually have objective pass/fail criteria.",
    "mrmc_comparative_effectiveness_study": "No. The document details non-clinical performance testing of a medical device (catheter), not AI software. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant or reported.",
    "standalone_algorithm_performance_study": "No. This is a physical medical device, not an algorithm or AI product. References to 'performance data' and 'performance testing' relate to the device's physical and functional attributes.",
    "type_of_ground_truth_used": "Engineering specifications, established international standards (ISO 10993, ISO 10555-1, ISO 11137-1, ISO 11137-2), and American Society for Testing and Materials (ASTM) standards (ASTM D4169, ASTM F2096, ASTM F88/F88M).",
    "sample_size_training_set": "Not applicable. This is a physical device, not an AI model, so there is no 'training set' in the context of machine learning.",
    "how_ground_truth_training_set_established": "Not applicable. There is no training set for this type of device."
  }
}

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 2, 2021

Mercator MedSystems, Inc. Kirk Seward President and Chief Science & Technology Officer 1900 Powell Street Suite 800 Emeryville, California 94608

Re: K210339

Trade/Device Name: Bullfrog® Micro-Infusion Device Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: June 29, 2021 Received: July 1, 2021

Dear Kirk Seward:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210339

Device Name Bullfrog Micro-Infusion Device

Indications for Use (Describe)

For Bullfrog devices capable of treating <6 mm diameter vessels:

In selective areas of peripheral and coronary vessels, the Bullfrog Micro-Infusion Device is intended for the infusion of diagnostic and therapeutic agents into the vessel wall and perivascular area, or intraluminally.

For Bullfrog devices capable only of treating ≥6 mm diameter vessels:

In selective areas of peripheral vessels, the Bullfrog Micro-Infusion Device is intended for infusion of diagnostic and therapeutic agents into the vessel wall and perivascular area, or intraluminally.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K210339 510(K) SUMMARY

510(k) Applicant Mercator MedSystems, Inc. 1900 Powell Street, Suite 800 Emeryville, CA 94608 Telephone: (510) 614-4550 Facsimile: (510) 614-4560

Contact Person:Phone:Email:Monica Barrett(916) 849-5399monica@mercatormed.com
Date of Summary:20 April 2021
Device OverviewTrade Name:Bullfrog® Micro-Infusion Device
Common Name:Continuous Flush Infusion Catheter (per 21 CFR 870.1210)
Classification:Continuous Flush Infusion Catheter21 CFR 870.1210Product Code KRA
Panel:Cardiovascular

Predicate Device

510(k)NumberTrade or Proprietary or Model NameManufacturer
K161402Bullfrog® Micro-Infusion DeviceMercator MedSystems

Reference Devices

510(k)NumberTrade or Proprietary or Model NameManufacturer
K153501Bullfrog® Micro-Infusion DeviceMercator MedSystems
K062752MicroSyringe II/Bullfrog Micro-InfusionDeviceMercator MedSystems

Device Description

The Mercator MedSystems Bullfrog Micro-Infusion Device is a wire-guided, singleoperator, endovascular catheter that consists of a perpendicular microneedle, which is sheathed by and contained within a semi-rigid polymer actuator balloon. The device is designed to be advanced to target vasculature and hydraulically actuated to move the microneedle through the external elastic lamina to deliver substances to adventitial and perivascular tissues. A compliant stabilizing balloon inflates with the actuator to provide

{4}------------------------------------------------

a force opposite the needle tip for proper seating of the needle is retracted within the sheathing structure by vacuuming the hydraulic actuator.

Indications for Use

For Bullfrog devices capable of treating <6 mm diameter vessels: In selective areas of peripheral and coronary vessels, the Bullfrog Micro-Infusion Device is intended for the infusion of diagnostic and therapeutic agents into the vessel wall and perivascular area, or intraluminally.

For Bullfrog devices capable only of treating ≥6 mm diameter vessels: In selective areas of peripheral vessels, the Bullfrog Micro-Infusion Device is intended for infusion of diagnostic and therapeutic agents into the vessel wall and perivascular area, or intraluminally.

Technological Characteristics

All materials used in the manufacture of the Bullfrog Micro-Infusion Device are suitable for this use and have been used in several previously cleared products.

Performance Data

Performance testing of the Bullfrog Micro-Infusion Device included mechanical and fluid delivery performance, biocompatibility, sterilization, and in-vivo safety studies. All tests met the pre-determined specifications and acceptance criteria.

The Bullfrog Micro-Infusion Device labeling contains instructions for use and any necessary cautions and warnings to assure that there are no new questions of safety and effectiveness with use of the device compared to the predicate. The biocompatibility assessment was conducted in accordance with ISO 10993. Biological Evaluation of Medical Devices.

Comparison to Predicate and Reference Devices

The Bullfrog Micro-Infusion Device is available in 5 models, each having a specified operating diameter range and in total spanning from 2 mm to 16 mm operating diameter. The predicate and reference devices span the range from 2 mm to 50 mm operating diameter. Notably, the Rex Medical Quadra-Fuse™ Multi-pronged Injection Needle, which is 510(k) exempt, is a multi-needle device with larger diameter needles than the Bullfrog device and is indicated for infusion and aspiration of fluids with ability to target up to 50 mm diameter tissue zone. Reference devices utilizing various grades of Pebax® including 40D and 72D durometer were provided as evidence of Pebax® material biocompatibility.

The following changes to the Bullfrog device were implemented since K161402 clearance.

    1. External laser-cut steel hypotube replaced the internal bonded wire to provide torque control.

{5}------------------------------------------------

    1. An additional device (BF102R) was added to the product family with the same operating diameter as prior model BF102S (2-4mm), but smaller crossing profile for smaller introducer sheath and guide catheter compatibility.
    1. Manufacturing method changes were verified.
    1. Re-ordering of manufacturing processes.
    1. Addition of BF102P to the Bullfrog product family.
    1. Updates to Instructions for Use.
    1. Intended Use and Indications for Use identical to K161402, but without allowance for coronary use for devices intended to operate only in vessels with 6mm or larger diameter.

Non-Clinical Performance Data

Biocompatibility testing was adopted from prior versions of the device (K161402) since there were no material changes and the changes in percent composition of the complete device were determined to be insignificant. Non-clinical testing of components of the device that have changed since K161402 included functional testing as defined in ISO 10555-1 with Amendments 1 and 2, and customized testing for performance. Testing performed on the device included:

  • Sterilization testing per ISO 11137-1:2006 and ISO 11137-2:2013 ●
  • Distribution testing per ASTM D4169 ●
  • Pouch bubble testing per ASTM F2096 ●
  • Pouch seal strength testing per ASTM F88/F88M ●
  • Label adhesion ●
  • System fluid delivery rate
  • Catheter stiffness ●
  • Actuator inflation and deflation time
  • Actuator diameter minimum/maximum
  • Actuator integrity after tortuous anatomy
  • Actuator burst strength
  • Drug line integrity
  • Drug line burst strength ●
  • Inflation line leakage under vacuum ●
  • System leakage under pressure
  • Pressure relief valve cracking pressure and leak rate ●
  • Radio-opaque markers intact
  • Microneedle sheathing
  • Microneedle insertion force at failure
  • Microneedle durability
  • Microneedle active length
  • Actuator retraction force to failure and in vacuum ●
  • Catheter bend kink resistance .
  • . Tensile strength
  • Torsional strength
  • Repetitive function and fatigue ●

{6}------------------------------------------------

Conclusions

The Bullfrog Micro-Infusion Device is substantially equivalent to the predicate and reference devices. Risk analysis has shown that the Bullfrog Micro-Infusion Device has the same risk profile as the predicate.

§ 870.1210 Continuous flush catheter.

(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).