(30 days)
The Siemens Biograph Vision Quadra PET/CT systems are combined Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and /or PET diagnostic imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
This CT system can be used for low dose lung cancer screening in high risk populations *
- As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
The Biograph Vision Quadra PET/CT systems are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. These systems are designed for whole body oncology, neurology and cardiology examinations. The Biograph Vision and systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.
Biograph Vision Quadra software is a command-based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.
The Biograph Vision Quadra is an extended Field of View scanner based on the Biograph Vision 600 Edge scanner. The system incorporates the same CT on the Biograph Vision 600 Edge but incorporates a 106 cm axial PET FoV. The gantry mechanical frame has been updated to support four (4) times the detectors available on the Biograph Vision 600 systems.
The software for the Biograph Vision Quadra systems, which are the subject of this application, is substantially equivalent to the commercially available Biograph Vision software (K193248). Modifications have been made to the commercially available Biograph Vision software to provide for acquisition and reconstruction of the additional detectors associated with the extended axial FoV of the PET system.
Additionally, modifications have been made to the PET computers to ensure that the system computing can handle the amount of data received from the 1 m scan range system.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Criteria (NEMA NU2-2018) | Acceptance Criteria ("Acceptance") | Reported Device Performance ("Results") |
|---|---|---|
| For Maximum Ring Difference (MRD) of 85: | ||
| Transverse Resolution FWHM @ 1 cm | ≤ 4.0 mm | Pass |
| Transverse Resolution FWHM @ 10 cm | ≤ 4.8 mm | Pass |
| Transverse Resolution FWHM @ 20 cm | ≤ 5.2 mm | Pass |
| Axial Resolution FWHM @ 1 cm | ≤ 4.3 mm | Pass |
| Axial Resolution FWHM @ 10 cm | ≤ 5.4 mm | Pass |
| Axial Resolution FWHM @ 20 cm | ≤ 5.4 mm | Pass |
| Sensitivity @435 keV LLD | ≥ 70 cps/kBq | Pass |
| Count Rate peak NECR | ≥ 1060 kcps @ ≤ 20 kBq/cc | Pass |
| Count Rate peak trues | ≥ 2680 kcps @ ≤ 20 kBq/cc | Pass |
| Scatter Fraction at peak NECR | ≤ 43% | Pass |
| Mean bias (%) at NEC of 1060 kcps | [-6, 6] | Pass |
| Co-Registration Accuracy | ≤ 5 mm | Pass |
| Time of Flight Resolution at 5.3 kBq/cc | ≤ 249 ps | Pass |
| 10mm sphere (Contrast / Background Variability) | ≥ 55.0% / ≤ 10.0% | Pass |
| 13mm sphere (Contrast / Background Variability) | ≥ 60.0% / ≤ 9.0% | Pass |
| 17mm sphere (Contrast / Background Variability) | ≥ 65.0% / ≤ 8.0% | Pass |
| 22mm sphere (Contrast / Background Variability) | ≥ 70.0% / ≤ 7.0% | Pass |
| 28mm sphere (Contrast / Background Variability) | ≥ 75.0% / ≤ 6.0% | Pass |
| 37mm sphere (Contrast / Background Variability) | ≥ 80.0% / ≤ 5.0% | Pass |
| Lung Residual Error | ≤ 5.0% | Pass |
| For Maximum Ring Difference (MRD) of 322: | ||
| Sensitivity @435 keV LLD | ≥ 150 cps/kBq | Pass |
| Count Rate peak NECR | ≥ 1940 kcps @ ≤ 20 kBq/cc | Pass |
| Count Rate peak trues | ≥ 5130 kcps @ ≤ 20 kBq/cc | Pass |
| Scatter Fraction at peak NECR | ≤ 43% | Pass |
| Time of Flight Resolution at 5.3 kBq/cc | ≤ 249 ps | Pass |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). The performance testing was conducted according to NEMA NU2-2018 standards, which typically involve testing on phantoms rather than human or animal subjects.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not mention the use of experts to establish a ground truth for the test set. The performance testing appears to be based on objective measurements against NEMA NU2-2018 industrial standards, which are defined by physical phantoms and measurement protocols, not expert consensus on pathology or clinical outcomes.
4. Adjudication Method for the Test Set
No adjudication method is mentioned, as the testing involves objective physical measurements against established standards (NEMA NU2-2018) rather than subjective assessments by experts.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study is mentioned. The document describes performance testing of the device's physical and technical characteristics, not its performance in a clinical setting with human readers.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
The performance testing described is for the "algorithm only" (i.e., the Biograph Vision Quadra PET/CT System's technical performance) without a human reader in the loop. The "Results" column indicating "Pass" for all tested criteria confirms that the system, as a standalone device, met the predetermined acceptance values.
7. Type of Ground Truth Used
The ground truth used for the performance testing is based on the NEMA NU2-2018 standard. This standard specifies phantoms and measurement protocols to objectively evaluate the physical performance characteristics of PET systems. It is not based on expert consensus, pathology, or outcomes data, but rather on direct physical measurements against defined benchmarks.
8. Sample Size for the Training Set
The document does not mention a "training set" in the context of the described performance testing. The Biograph Vision Quadra PET/CT System is a medical imaging device (hardware and associated software), and the performance testing focuses on its physical and technical specifications rather than an AI model that undergoes a training phase. While there is software development involved ("Modifications have been made to the commercially available Biograph Vision software"), the document does not detail specific training data for that software in the context of this performance evaluation.
9. How the Ground Truth for the Training Set Was Established
As no training set is described in the context of the performance testing for this device's NEMA NU2-2018 criteria, information on how its ground truth was established is not provided.
{0}------------------------------------------------
March 3, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it.
Siemens Medical Solutions USA, Inc. % Alaine Medio Regulatory Affairs Professional 810 Innovation Drive KNOXVILLE TN 37932
Re: K210262
Trade/Device Name: Biograph Vision Quadra PET/CT System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: KPS, JAK Dated: January 29, 2021 Received: February 1, 2021
Dear Alaine Medio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K210262
Device Name Biograph Vision Quadra PET/CT System
Indications for Use (Describe)
The Siemens Biograph Vision Quadra PET/CT systems are combined Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and /or PET diagnostic imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
This CT system can be used for low dose lung cancer screening in high risk populations *
- As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
as required by 21 CFR Part 807.87(h)
Identification of the Submitter
| Submitter: | Alaine MedioRegulatory AffairsSiemens Medical Solutions USA, Inc.Molecular Imaging810 Innovation DriveKnoxville, TN 37932 |
|---|---|
| Alternative Contact: | Tabitha EstesRegulatory Affairs |
| Manufacturer: | Siemens Medical Solutions USA, Inc.Molecular Imaging2501 North Barrington RoadHoffman Estates, IL 60192 |
| Telephone Number: | (865)206-0337 |
| Fax Number: | (865)218-3019 |
| Date of Submission: | January 29, 2021 |
| Identification of the product | |
| Device ProprietaryName: | Biograph Vision Quadra PET/CT System |
| Common Name: | Positron Emission Tomography (PET) SystemComputed Tomography (CT) System |
| Classification Name: | Emission Computed Tomography System per 21 CFR892.1200Computed Tomography X-Ray System per 21 CFR 892.1750 |
| Product Code: | KPS and JAK |
| Classification Panel: | Radiology |
| Device Class: | Class II |
{4}------------------------------------------------
Marketed Devices to which Equivalence is claimed
| Primary PredicateDevice: | |
|---|---|
| Device ProprietaryName: | Biograph Vision PET/CT Systems |
| Manufacturer: | Siemens Medical Solutions USA, Inc |
| Product Code: | KPS and JAK |
| Device Class: | Class II |
| 510(k) Number: | K193248 |
Reference Devices:
| Device Name and 510(k) numbers: | SOMATOM Definition CT | K190578 |
|---|---|---|
| --------------------------------- | ----------------------- | --------- |
Device Description:
The Biograph Vision Quadra PET/CT systems are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. These systems are designed for whole body oncology, neurology and cardiology examinations. The Biograph Vision and systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.
Biograph Vision Quadra software is a command-based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.
The Biograph Vision Quadra is an extended Field of View scanner based on the Biograph Vision 600 Edge scanner. The system incorporates the same CT on the Biograph Vision 600 Edge but incorporates a 106 cm axial PET FoV. The gantry mechanical frame has been updated to support four (4) times the detectors available on the Biograph Vision 600 systems.
The software for the Biograph Vision Quadra systems, which are the subject of this application, is substantially equivalent to the commercially available Biograph Vision software (K193248). Modifications have been made to the commercially available Biograph
{5}------------------------------------------------
Vision software to provide for acquisition and reconstruction of the additional detectors associated with the extended axial FoV of the PET system.
Additionally, modifications have been made to the PET computers to ensure that the system computing can handle the amount of data received from the 1 m scan range system.
These changes have no impact on safety or effectiveness.
Intended Use:
The Siemens Biograph Vision Quadra systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.
The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
{6}------------------------------------------------
Performance Testing / Safety and Effectiveness:
Performance testing for the CT subsystem was included in the original premarket notification for the CT subsystems and there have been no changes affecting this testing.
PET Testing in accordance with NEMA NU2-2018 was conducted on the Biograph Vision Quadra systems.
| Performance Criteria | Results | Acceptance |
|---|---|---|
| Resolution - Full Size | ||
| Transverse Resolution FWHM @ 1 cm | Pass | ≤ 4.0 mm |
| Transverse Resolution FWHM @ 10 cm | Pass | ≤ 4.8 mm |
| Transverse Resolution FWHM @ 20 cm | Pass | ≤ 5.2 mm |
| Axial Resolution FWHM @ 1 cm | Pass | ≤ 4.3 mm |
| Axial Resolution FWHM @ 10 cm | Pass | ≤ 5.4 mm |
| Axial Resolution FWHM @ 20 cm | Pass | ≤ 5.4 mm |
| Sensitivity @435 keV LLD | Pass | ≥ 70 cps/kBq |
| Count Rate peak NECR | Pass | ≥1060 kcps @ ≤ 20kBq/cc |
| Count Rate peak trues | Pass | ≥2680 kcps @ ≤ 20kBq/cc |
| Scatter Fraction at peak NECR | Pass | ≤43% |
| Mean bias (%) at NEC of 1060 kcps | Pass | [-6,6] |
| Co-Registration Accuracy | Pass | ≤ 5 mm |
| Time of Flight Resolution at 5.3kBq/cc | Pass | ≤249 ps |
| 10mm sphere (Contrast / Background Variability) | Pass | ≥ 55.0% / ≤ 10.0% |
| 13mm sphere (Contrast / Background Variability) | Pass | ≥ 60.0% / ≤ 9.0% |
| 17mm sphere (Contrast / Background Variability) | Pass | ≥ 65.0% / ≤ 8.0% |
| 22mm sphere (Contrast / Background Variability) | Pass | ≥ 70.0% / ≤ 7.0% |
| 28mm sphere (Contrast / Background Variability) | Pass | ≥ 75.0% / ≤ 6.0% |
| 37mm sphere (Contrast / Background Variability) | Pass | ≥ 80.0% / ≤ 5.0% |
| Lung Residual Error | Pass | ≤ 5.0% |
Table 1 PET NEMA 2018 Performance Summary Biograph Vision Quadra with a Maximum Ring Difference (MRD) of 85.
Table 2 PET NEMA 2018 Performance Summary Biograph Vision Quadra with a Maximum Ring Difference (MRD) of 322
| Performance Criteria | ResultsMRD 322 | Acceptance |
|---|---|---|
| Sensitivity @435 keV LLD | Pass | ≥ 150 cps/kBq |
| Count Rate peak NECR | Pass | ≥1940 kcps @ ≤ 20 kBq/cc |
| Count Rate peak trues | Pass | ≥5130 kcps @ ≤ 20 kBq/cc |
| Scatter Fraction at peak NECR | Pass | ≤43% |
| Time of Flight Resolution at5.3kBq/cc | Pass | ≤249 ps |
{7}------------------------------------------------
All Performance testing met the predetermined acceptance values.
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards such as IEC 60601-1 series and 21 CFR 1020.30 and 21 CFR 1020.33 to minimize electrical, mechanical and radiation hazards.
Siemens claims compliance with the following product standards for the Biograph Vision Quadra systems:
- . IEC 60601-1: 2005+ A1:2012
- IEC 60601-1-2: 2014
- IEC 60601-1-3: 2013
- IEC 60601-1-6:2010 +A1:2013
- IEC 60601-2-28:2017
- IEC 60601-2-44: 2016
- IEC 60825-1: 2014
- IEC 62366-1: 2015
- ISO 10993-1: 2018
- IEC 61223-2-6:2006
- IEC 61223-3-5:2004
- NEMA XR 25: 2019
- NEMA XR 28: 2013
- NEMA XR 29: 2013
- NEMA PS3.1-3.20
Additionally, the Biograph Vision Quadra systems have been developed in accordance with the requirements of the following standards:
- IEC 62304:2006 +A1:2015
- ISO 14971:2019
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Cybersecurity information in accordance with FDA Guidance documents issued October 2, 2014 has been provided. The Biograph Vision Quadra software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use.
{8}------------------------------------------------
Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the Biograph Vision Quadra and external devices.
Verification and validation of Siemens systems is performed in accordance with documented procedures, design and code reviews, test plans and specifications. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.
Statement regarding Substantial Equivalence:
There have been no changes implemented in the modifications to the Biograph Vision Quadra system that impacts either the fundamental technology or the indications for use. The Biograph Vision Quadra PET/CT system outlined in this Premarket Notification is substantially equivalent to the currently commercially available predicate device.
N/A