Blue Nitrile Exam Glove, Powder Free
K210058 · Great Harvest Trading Co., Ltd. · LZA · Mar 30, 2021 · General Hospital
Device Facts
| Record ID | K210058 |
| Device Name | Blue Nitrile Exam Glove, Powder Free |
| Applicant | Great Harvest Trading Co., Ltd. |
| Product Code | LZA · General Hospital |
| Decision Date | Mar 30, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.6250 |
| Device Class | Class 1 |
Intended Use
The Blue Nitrile Exam Glove, Powder Free is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
Device Story
Blue Nitrile Exam Glove, Powder Free; non-sterile, disposable, ambidextrous nitrile glove. Used by healthcare examiners in clinical settings to provide a protective barrier against contamination between patient and examiner. Device functions as a physical barrier; no electronic or software components. Performance verified through bench testing against ASTM and ISO standards for physical properties, biocompatibility, and freedom from holes.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified with ISO 10993-10 (irritation/sensitization), ISO 10993-5 (cytotoxicity), ASTM D6124 (residual powder), ASTM D5151 (freedom from holes), and ASTM D6319 (physical properties).
Technological Characteristics
Material: Nitrile. Form factor: Ambidextrous, powder-free, blue, non-sterile. Dimensions: Length 230mm, varying widths/thicknesses per size. Standards: ASTM D6319-19 (specifications), ASTM D5151-19 (holes), ASTM D6124-06 (powder), ISO 10993-5/10 (biocompatibility).
Indications for Use
Indicated for use as a non-sterile, disposable patient examination glove worn on the hands or fingers of healthcare examiners to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
- Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color (K171422)
Related Devices
- K143477 — Nitrile Patient Examination gloves, Powderfree, Blue color · Primus Gloves Pvt Limited · Apr 14, 2016
- K210057 — Blue Nitrile Exam Glove, Powder Free · Real Star Medical Technology Co., Ltd. · Mar 31, 2021
- K210276 — Disposable Nitrile Gloves · Fujian Ercon Medical Management Co., Ltd. · Apr 16, 2021
- K211131 — Disposable Nitrile Medical Examination Glove · Siyang Threeguard Medical Supplies Co., Ltd. · Jul 29, 2021
- K212716 — Dermoaroma Disposable Nitrile Examination Glove · Chuzhou Harmony Gloves Medical Technology Co.,Ltd · Dec 27, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 30, 2021
Great Harvest Trading Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608,No.738,Shangcheng Rd.,Pudong Shanghai, Shanghai 200120 China
Re: K210058
Trade/Device Name: Blue Nitrile Exam Glove, Powder Free Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: March 16, 2021 Received: March 22, 2021
Dear Boyle Wang:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jiangsong Jiang -S (Affiliate)
For Ryan Ortega, Ph D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
{3}------------------------------------------------
# 510(k) Summary
## K210058
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
#### 1.0 Submitter's Information
Great Harvest Trading Co., Ltd. Name: Xinghe Road, Dongying City 257100, Shandong, China Address: Phone Number: +86-13508943885 Contact: Chongfeng Zhao Date of Preparation: Mar.16,2021
#### Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn
#### 2.0 Device Information
Trade name: Blue Nitrile Exam Glove, Powder Free Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): S, M. L. XL
### 3.0 Classification
Production code: LZA Regulation number: 21CFR880.6250 Classification: Class | Panel: General Hospital
#### 4.0 Predicate Device Information
Manufacturer: Ever Global (Vietnam) Enterprise Corp Disposable Powder Free Nitrile Examination Glove, White/ Device: Blue/ Black/ Pink Color 510(k) number: K171422
### 5.0 Indication for Use
{4}------------------------------------------------
The Blue Nitrile Exam Glove, Powder Free is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
#### 6.0 Device Description
The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.
#### 7.0 Technological Characteristic Comparison Table
| Table1-General Comparison | | | |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Item | Subject Device<br>(K210058) | Predicated Device<br>(K171422) | Remark |
| Product Code | LZA | LZA | Same |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same |
| Class | | | Same |
| Intended Use | The Blue Nitrile<br>Exam Glove,<br>Powder Free is a<br>non-sterile<br>disposable device<br>intended for<br>medical purposes<br>that is worn on the<br>examiner's hands<br>or finger to prevent<br>contamination<br>between patient<br>and examiner. | The Disposable<br>Powder Free Nitrile<br>Examination Glove,<br>White/ Blue/ Black/<br>Pink Color is a<br>disposable device<br>intended for<br>medical purposes<br>that is worn on the<br>examiner's hands<br>to prevent<br>contamination<br>between patient<br>and examiner. | Same |
| Powdered or<br>Powered free | Powdered free | Powdered free | Same |
| Design Feature | Ambidextrous | Ambidextrous | Same |
| Labeling Information | Single-use<br>indication, powder<br>free, device color,<br>device name, glove<br>size and quantity,<br>Nitrile Glove<br>Powder Free Blue, | Single-use<br>indication, powder<br>free, device color,<br>device name, glove<br>size and quantity,<br>Disposable Powder<br>Free Nitrile<br>Examination Glove | Same |
#### Table1-General Comparison
{5}------------------------------------------------
| | Non-Sterile | Non-Sterile | |
|--|-------------|-------------|--|
|--|-------------|-------------|--|
| Predicate<br>Device(K171422) | Designation | Size | | | | | Tolerance | |
|------------------------------|----------------|------|-----|-----|-----|-----|------------|--|
| | | XS | S | M | L | XL | | |
| | Length, mm | 230 | 230 | 230 | 230 | 230 | min | |
| | Width, mm | 75 | 85 | 95 | 105 | 115 | $\pm 5$ | |
| | Thickness, mm | | | | | | | |
| | Finger | 0.05 | | | | | min | |
| | Palm | 0.05 | | | | | min | |
| Subject Device<br>(K210058) | Designation | Size | | | | | Tolerance | |
| | | S | M | L | XL | | | |
| | Length, mm | 230 | 230 | 230 | 230 | | $\pm 10$ | |
| | Width, mm | 80 | 95 | 110 | 120 | | $\pm 10$ | |
| | Thickness, mm: | | | | | | | |
| | Finger | 0.08 | | | | | $\pm 0.03$ | |
| | Palm | 0.08 | | | | | $\pm 0.03$ | |
| Remark | Similar | | | | | | | |
#### Table2 Device Dimensions Comparison
Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319-19,so the differences do not raise any new safety or performance questions.
#### Table3 Performance Comparison
| Item | Subject device<br>(K210058) | Predicated<br>device<br>(K171422) | Remark | | |
|------------------------|---------------------------------------------------------------|---------------------------------------------------------------|------------|------------|------|
| Colorant | Blue | White/ Blue/<br>Black/ Pink | Same | | |
| Physical<br>Properties | Before<br>Aging | Tensile<br>Strength | 14MPa, min | 14MPa, min | Same |
| | | Ultimate<br>Elongation | 500% min | 500% min | Same |
| | After<br>Aging | Tensile<br>Strength | 14MPa, min | 14MPa, min | Same |
| | | Ultimate<br>Elongation | 400%min | 400%min | Same |
| | Comply with ASTM D6319 | Comply with ASTM D6319 | Same | | |
| Freedom from Holes | Be free from<br>holes when<br>tested in<br>accordance<br>with | Be free from<br>holes when<br>tested in<br>accordance<br>with | Same | | |
{6}------------------------------------------------
| | ASTMD5151 | ASTMD5151 | |
|----------------|----------------------|-------------------------------------------|------|
| | AQL=2.5 | AQL=2.5 | |
| Powder Content | 0.05 mg per<br>glove | Meet the<br>requirements of<br>ASTM D6124 | Same |
| Item | | Subject<br>device<br>(K210058) | Predicated<br>device<br>(K171422) | Remark |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|---------|
| Material | | Nitrile | Nitrile | Same |
| Biocompatibility | Irritation (ISO<br>10993-10:2010<br>Biological<br>Evaluation of<br>Medical Devices -<br>Part 10: Tests For<br>Irritation And Skin<br>Sensitization) | Under the<br>conditions of<br>the study, not<br>an irritant | Comply with<br>ISO10993-<br>10 | Same |
| | Sensitization<br>(ISO 10993-<br>10:2010<br>Biological<br>Evaluation of<br>Medical Devices -<br>Part 10: Tests For<br>Irritation And Skin<br>Sensitization)<br>Cytotoxicity (ISO<br>10993-5:2009<br>Biological<br>Evaluation of<br>Medical Devices -<br>Part 5: Tests For<br>In Vitro<br>Cytotoxicity) | Under<br>conditions of<br>the study, not<br>a sensitizer.<br>Under<br>conditions of<br>the study,<br>device<br>extract is not<br>cytotoxic | / | Similar |
#### Table4 Safety Comparison
#### 8.0 Discussion of Non-clinical and Performance Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
{7}------------------------------------------------
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
#### 9.0 Discussion of Clinical and Performance Testing
Clinical testing is not needed for this device.
#### 10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device Blue Nitrile Exam Glove, Powder Free is as safe, as effective, and performs as well as or better than the legally marketed predicated device Disposable Powder Free Nitrile Examination Glove cleared under K171422.