K Number
K203792

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-01-20

(753 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sterile, single use needles intended for use with pen injector devices for the subcutaneous injection of drugs

Device Description

The Insupen Pen Needle/ Insupen ORIGINAL Pen needle/ Insupen ADVANCED Pen Needle are sterile, single use, single patient pen needles intended for use with pen injector devices for the subcutaneous injection of drugs. Pen needles are used by consumers, caregivers and healthcare professionals. The intended patients for this device is the population from newborn to elderly. The device can be used both in domestic and professional environment. It is considered MR unsafe. The list of drugs tested and intended to be used with Pikdare's pen needle are Insulin, peptic hormone, peptic glucagon-like-1, and drug for the treatment of osteoporosis. All compatibility is checked following the requirements of ISO 11608-2:2012.

The pen needle assembly consists of a double-ended cannula that is assembled into an injection molded hub by gluing. The hub has internal threads which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient end and the cartridge end of the cannula are lubricated using a silicone based lubricant for ease of injection and pen cartridge septum penetration. There is an inner needle shield assembled over the patient end of the cannula to protect the needle tip from damage and accidental needle sticks. Each pen needle is protected with a sealed medical paper which together with the primary container provide a sterility barrier. The pen needle is packaged in boxes and sterilized with ETO. It is intended for single patient and single use only. To use a pen needle, the user needs to remove the medical paper, screw the needle onto the pen injector device. Then the user removes both the outer cap and the inner protective cap to expose the needle and make an injection. After the injection, the needle is unscrewed from the pen injector device and disposed in an appropriate container respecting applicable regulations and laws.

AI/ML Overview

The document describes the acceptance criteria and study results for the Pikdare Insupen Pen Needle, Insupen ORIGINAL Pen Needle, and Insupen ADVANCED pen needle.

1. Table of acceptance criteria and the reported device performance:

Test ParameterAcceptance Criteria (Requirement – ISO 11608-2:2012)Reported Device Performance (Subject Device)
MaterialsThe needle shall be made of tubing materials specified in ISO 9626.Requirement met
DimensionsThe tubing features used in the needles shall meet the requirements of ISO 9626. For G34 needle, the limits of stiffness have been chosen to comply with the same limits of G33 TW needle. The needles shall fit the test apparatus specified in item 7.3 of ISO 11608-2. The dimension shall comply with Table 1 of clause 4.2.2 of ISO 11608-2.Requirement met
Determination of flow rate through the needleThe needle was tested in accordance with Annex A to ISO 11608-2 to determine flow rate through the needle.Requirement met
Bond between hub and needle tubeThe union of the hub and needle tube shall not break when tested in accordance with Clause 9 of ISO 11608-2.Requirement met
Needle pointsNeedle points shall appear sharp and free from featheredges, burrs and hocks when examined under magnification x2.5. The needle point at the cartridge end is designed to minimize coring and fragmentation when penetrating the cartridge set.Requirement met
Freedom from defectsThe needle tube shall fulfill the requirements of ISO 7864, clause 11.3.Requirement met
LubricationThe needle tube should be lubricated at both the patient end and the cartridge end. The lubricant shall not, under normal or corrected-to-normal vision, be visible as droplets of fluid on the outside surface of the needle tube.Requirement met
Dislocation of measuring point at patient endDislocation of the cannula point at the patient end shall be in accordance with Table 2 when tested in accordance with Clause 8 (of ISO 11608-2).Requirement met
Determination of functional compatibility with NISCompatibility with any NIS (Needle-based Injection System) shall be claimed only after testing in accordance with Clause 11.Requirement met
Ease of assembly and disassemblyAttachment of the needle shall be possible without removing the needle from its opened unit packaging. Compliance is checked according to the requirements of Clause 11.Requirement met
SterilityThe needles in its unit packaging is subjected to a validated sterilization process. (SAL = 10^-6)Requirement met (SAL = 10^-6 achieved)
Pre-conditioning of needlesThe needles satisfy all the requirements of ISO 11608-2 after preconditioning according to clauses 6.1, 6.2 and 6.3 of the same standard.Requirement met
Biocompatibility Tests
Cytotoxicity Study Using the ISO Elution MethodISO 10993-5 (Tests for in vitro cytotoxicity)No evidence of causing cell lysis or toxicity.
ISO Guinea Pig Maximization Sensitization TestISO 10993-10 (Tests for irritation and skin sensitization)No evidence of causing delayed dermal contact sensitization; not considered a sensitizer.
ISO Intracutaneous Study in RabbitsISO 10993-10 (Tests for irritation and skin sensitization)Met the requirements of the test.
ASTM Hemolysis StudyASTM F756 and ISO 10993-4 (Assessment of Hemolytic Properties of Materials)Non-hemolytic.
ISO Two Week Toxicity Study in the RatISO 10993-11 (Tests for systemic toxicity)No microscopic changes considered test article related; no systemic toxicity.
ISO Systemic Toxicity Study in MiceISO 10993-11 (Tests for systemic toxicity)No mortality or evidence of systemic toxicity.
USP Rabbit Pyrogen Study, Material-mediatedISO 10993-11 (Tests for systemic toxicity)Total rise of rabbit temperatures within acceptable USP limits; non-pyrogenic.
USP Pyrogen Study - Material MediatedUSP, General Chapter <151>, Pyrogen Test as recommended by ISO 10993-11Nonpyrogenic.
EO Residual ValuesMaximum EO residual values after degassing and before product release set at 1 mg per device (ISO 10993-7 limit: 4 mg/day, considering 4 uses/day).Met (values below 1 mg/device).
ECH Residual ValuesMaximum ECH residual values after degassing and before product release set at 2.25 mg per device (ISO 10993-7 limit: 9 mg/day, considering 4 uses/day).Met (values below 2.25 mg/device).

2. Sample size used for the test set and the data provenance:

The document states that "Test were performed on both products, the total availability on sample of the fact that the Predicate Device is manufactured in Pikdare factory in Italy (Pikdare is the contract manufacturer of Droplet G34 Pen Needle)." This implies the testing was done on the actual manufactured products (both the subject device and the predicate device).

  • Sample size: Not explicitly stated with a specific number for each test, but it is implied that sufficient samples were used to meet the requirements of the listed ISO standards.
  • Data provenance: The predicate device is manufactured in Italy by Pikdare S.p.A. The testing described for both the subject device and the predicate device was presumably performed by or on behalf of Pikdare S.p.A. in Italy. All tests refer to recognized international standards (ISO).
  • Retrospective or prospective: The data appears to be prospective, collected specifically for the 510(k) submission to demonstrate compliance with standards and substantial equivalence.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. This is a medical device (pen needle) conformity assessment, not an AI or diagnostic imaging study that relies on expert interpretation for ground truth. The "ground truth" here is defined by meeting the objectively measurable criteria specified in ISO and ASTM standards.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

Not applicable, for the same reason as point 3. The assessment is based on objective measurements against predefined standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No. This is not an AI-assisted diagnostic device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This is not an AI-based device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" is defined by the technical specifications and performance requirements established by recognized international standards (e.g., ISO 11608-2, ISO 9626, ISO 10993-1, ISO 7864, ASTM F756, USP General Chapter <151>). The device performance is objectively measured against these established criteria.

8. The sample size for the training set:

Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.

January 20, 2023

Pikdare S.p.A Roberta Zanoni Official Correspondent Via Saldarini Catelli 10 Casnate con Bernate, COMO 22070 Italy

Re: K203792

Trade/Device Name: Insupen Pen needle, Insupen ORIGINAL Pen Needle, Insupen ADVANCED pen needle

Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: December 19, 2022 Received: December 22, 2022

Dear Roberta Zanoni:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Alan M.
Stevens -
S3

CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203792

Device Name Insupen Pen Needle Insupen ORIGINAL pen needle Insupen ADVANCED pen needle

Indications for Use (Describe)

Sterile, single use needles intended for use with pen injector devices for the subcutaneous injection of drugs

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "PIKDARE" in a bold, sans-serif font. The word is in blue, except for the first part of the "P", which is in a light brown color. The logo has a modern and professional look.

510(k) Summary

As required by the Medical Devices Act of 1990 and in accordance with 21 CFR §807.92(a).

Summary [807.92 (a)(1,2)]

Date Prepared: January 20, 2023 Submitted By: Pikdare S.p.A Via Saldarini Catelli 10 22070 Casnate con Bernate - COMO - Italy

Primary Contact: Roberta Zanoni - Market Access Phone: +39 031 7297757 roberta.zanoni@pikdare.com

Secondary Contact: Marco De Nardi Regulatory Affairs Phone: +39 031 7297832 marco.denardi@pikdare.com

Trade Name: Insupen Pen Needle Insupen ORIGINAL Pen needle Insupen ADVANCED Pen Needle

Device models: Insupen® Pen Needles 12 mm x 29G Insupen®Pen Needles 8 mm x 30G Insupen®Pen Needles 5 mm x 31G Insupen®Pen Needles 6 mm x 31G Insupen®Pen Needles 8 mm x 31G Insupen®Pen Needles 4 mm x 32G Insupen®Pen Needles 6 mm x 32G Insupen®Pen Needles 8 mm x 32G Insupen®Pen Needles 4 mm x 33G Insupen®Pen Needles 3,5 mm x 34G Insupen Original® Pen Needles 12 mm x 29G Insupen Original ®Pen Needles 8 mm x 30G Insupen Original ®Pen Needles 5 mm x 31G Insupen Original ®Pen Needles 6 mm x 31G Insupen Original ®Pen Needles 8 mm x 31G Insupen Original ®Pen Needles 4 mm x 32G Insupen Original ®Pen Needles 6 mm x 32G Insupen Original ®Pen Needles 8 mm x 32G Insupen Original®Pen Needles 4 mm x 33G Insupen Advanced®Pen Needles 5 mm x 31G Insupen Advanced®Pen Needles 4 mm x 32G Insupen Advanced®Pen Needles 6 mm x 32G Insupen Advanced®Pen Needles 4 mm x 33G Insupen Advanced®Pen Needles 3,5 mm x 34G

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "PIKDARE" in blue letters. To the left of the word is a gray symbol that looks like a backwards capital L with a horizontal line extending from the top. The letters are bold and sans-serif.

Common Name: Pen Needle for drugs subcoutaneous injection Regulation Number: 21 CFR § 880.5570 Product Code: FMI Device Classification: II Review Panel: 80 General Hospital Predicate Device [807.92(a)(3)]

The legally marketed device to which substantial equivalence is claimed are:

ManufacturerTrade name510K number
Reference deviceArtsana SpAInsupen pen needleK051783
Primary DeviceHTL-STREFA S.A.DROPLET® PENNEEDLE 34GK192082

In 2019 Artsana SpA has moved the proprietary of its 510K (K051783) to Pikdare S.p.A

Description of Device: [807.92(a)(4)]

The Insupen Pen Needle/ Insupen ORIGINAL Pen needle/ Insupen ADVANCED Pen Needle are sterile, single use, single patient pen needles intended for use with pen injector devices for the subcutaneous injection of drugs. Pen needles are used by consumers, caregivers and healthcare professionals. The intended patients for this device is the population from newborn to elderly. The device can be used both in domestic and professional environment. It is considered MR unsafe. The list of drugs tested and intended to be used with Pikdare's pen needle are Insulin, peptic hormone, peptic glucagon-like-1, and drug for the treatment of osteoporosis. All compatibility is checked following the requirements of ISO 11608-2:2012.

The pen needle assembly consists of a double-ended cannula that is assembled into an injection molded hub by gluing. The hub has internal threads which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient end and the cartridge end of the cannula are lubricated using a silicone based lubricant for ease of injection and pen cartridge septum penetration. There is an inner needle shield assembled over the patient end of the cannula to protect the needle tip from damage and accidental needle sticks. Each pen needle is protected with a sealed medical paper which together with the primary container provide a sterility barrier. The pen needle is packaged in boxes and sterilized with ETO. It is intended for single patient and single use only. To use a pen needle, the user needs to remove the medical paper, screw the needle onto the pen injector device. Then the user removes both the outer cap and the inner protective cap to expose the needle and make an injection. After the injection, the needle is unscrewed from the pen injector device and disposed in an appropriate container respecting applicable regulations and laws.

Indications for Use: [807.92(a)(5)]

Sterile, single use needles intended for use with pen injector devices for the subcutaneous injection of drugs. Technological Characteristics: [807.92(a)(6)]-

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for "PIKDARE". The logo is in a sans-serif font, with the letters in dark blue except for the first character which is a gray square with a line extending to the right. The logo is simple and modern.

Predicate Devices: [K051783; K192082]

A comparison of characteristics among Pikdare Pen Needles (New Device) and Predicate Device is shown in the table below, in which it is included the respective Indications for Use

New DevicePredicate DeviceConclusion
ManufacturerPikdare SpAHTL-STREFA S.A.n/a
510(k) NumberPendingK192082n/a
Product CodeFMIFMIsame
Proprietary nameInsupen Pen NeedleInsupen Original PenNeedleInsupen AdvancedPen needleDroplet 34Gsame
PrimaryContainerImage: Primary Container New DeviceImage: Primary Container Predicate Devicesame
Needle ShieldImage: Needle Shield New DeviceImage: Needle Shield Predicate Devicesame
Needle Tube andHubImage: Needle Tube and Hub New DeviceImage: Needle Tube and Hub Predicate Devicesame
Indication for usePen Needle for drugssubcutaneousinjectionSterile, single useneedles intended foruse with pen injectordevices for thesubcutaneous injectionof drugsSame
Intended usePen Needle for drugssubcutaneous injectionThe patients target ofthis device is thepopulation fromnewborn to elderly.The device can beused both in domesticand professionalenvironment. MRUnsafe.Pen Needle for drugssubcutaneous injectionSame
Principle ofoperationManualManualsame
Method ofattachment to peninjectorScrew threadsScrew threadssame
New DevicePredicate DeviceConclusion
Length12mm4mm5mm6mm,8mm(tolerance ±1.25mm)3.5mm(tolerance-0.4mm/+0.5mm)3.5mm(tolerance-0.4mm /+0.5mm)See discussion below
Gauge29G, 30G, 31G, 32G,33G, 34G34GSee discussion below
BiocompatibilityConforms toISO 10993-1:2019Conforms toISO 10993-1:2019same
SterilitySAL = $10^{-6}$SAL = $10^{-6}$same
SterilizationmethodETOETOsame
Shelf Life5 years5 yearssame
Unit PackagingBlister composed byPrimary containermade of Polyethylenenon-toxic for medicaluse (HDPE2) andseal made of medicalgrade paper (C/PAP22)Blister composed byPrimary containermade of Polyethylenenon-toxic for medicaluse (HDPE2) and sealmade of medical gradepaper (C/PAP22)same
User Packaging100 pcs Cardboardsales box (PAP 21)Cardboard sales box(PAP 21)same
Transportpackaging12 cardoboards salesboxes inside acorrugated cartoncontainer.(1200 penneedles in total) -(PAP 20)12 cardoboards salesboxes inside acorrugated cartoncontainer.(1200 penneedles in total) (PAP20same
Materials
Needle TubeMedical GradeStainless SteelAISI304L and AISI305Medical GradeStainless SteelAISI 304LSee Discussion
HubPolypropylenePolypropylenesame
Needle Shield andPrimary containerPolyethylene non-toxicfor medical usePolyethylene non-toxicfor medical usesame
LubricantMedical grade siliconeMedical grade siliconesame
Glue (for hub andneedle tubebonding)UV glue - non-toxicUV glue - non-toxicsame

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "PIKDARE" in a stylized font. The first letter "P" is made up of a gray square and a blue line. The rest of the letters are in blue. The font is bold and sans-serif.

The subject device and the Predicate devices are classified under 21 CFR 880.5570, which states: "A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin."

There are no differences in the main design, intended use (see details in the table above), principle of operation, method of attachment to pen injector devices, biocompatibility (see details in the table above), sterilization method, materials (see details in the table above), between the Pikdare pen needle device and the Predicate devices. Thus as detailed above the substantial equivalence is demonstrated

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for PIKDARE. The logo is in blue and gray. The word "PIKDARE" is written in blue, with the first letter "P" being partially gray. The gray part of the "P" is a square with a line extending from the top and left sides.

The Pikdare pen needle differs from the predicate device as shown in the table below:

GAUGESLENGTHAvailabilityStainless SteelINSUPENWall typeAvailabilityStainless SteelINSUPEN ORIGINALAvailabilityINSUPEN ADVANCEDWall typePREDICATE DEVICEK192082(HTL-S.A.)Wall type
G2912mmXAISI 304LRWXAISI 304LXAISI 304LRW-
G308mmXAISI 304LRWXAISI 304L---
5mmXAISI 304L&AISI 305ETWXAISI 304L&AISI 305--UTW
G316mmXAISI 304L&AISI 305ETWXAISI 304L&AISI 305XAISI304LETW(1)-
8mmXAISI 304L&AISI 305ETWXAISI 304L&AISI 305XAISI304LETW(1)-
4mmXAISI 304L&AISI 305ETWXAISI 304L&AISI 305--UTW
G326mmXAISI 304L&AISI 305TWXAISI 304L&AISI 305--ETW
8mmXAISI 304L&AISI 305TWXAISI 304L&AISI 305---
G334mmXAISI 304L&AISI 305TWXAISI 304L&AISI 305--ETW
G343,5mmXAISI 304L&AISI 305TW---ETWXAISI 304LTW

(1) Wall definition are declared based on the current revision of the standard.

In conclusion, analyzing the main differences between the subject device and the Predicate device, below these are the main consideration:

  1. Pikdare Pen Needles are manufactured using both AISI304L and AISI 305 while the Predicate devices use AISI 304L only. AISI 305 is mentioned in the ISO 9626:2016 where it is stated that "tubing shall be made in the stainless steels listed in the ISO 15510" and our material is included in the standard, therefore there is no new impact on safety and performances.

  2. Needle features and wall are confirmed to be compliant with ISO 9626 considering all the possible dimensions defining the wall. So there is no risk about the safety of the subject device With reference to needle's size G34, which is not covered by ISO 9626 requirements for stiffness, Pikdare SpA chooses to characterize the tube of this needle according to thestiffness test limits of G33 TW (as defined in ISO 9626:2016).

  3. The Pikdare pen needle will be under OTC as well as the Predicate device K192082,

{8}------------------------------------------------

  1. The Pikdare pen needle intended use is for injection of drugs in general as well as the Predicate device K192082. This feature does not introduce critical differences and new risks (migration and transfer tests has been carried out on needles with both intended uses).

The differences above indicated do not raise new questions of safety and/or functionality for the subject device. Considering that the Predicate device is covering G34 and 3,5mm only, while the Subject device is intended to cover the range of pen needles of G29 - G30 - G31 - G32 - G33 and G34 with lengths varying from 12mm to 3,5mm. we can declare that the products with extended gauges and lengths do not pose any new concerns in terms of safety and performance, since they have been submitted to tests according to the following recognized consensus standards:

-ISO 11608-2 Needle based injection system for medical use - Requirements and test methods - part 2: needles

-ISO 9626 Stainless steel needle tubing for the manufacture of Medical Devices – Requirements and Test Methods

-ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a Risk management Process

and the risk assessment carried out on them hasn't identified further risks not reduced to unacceptable level.

So, Pikdare pen needle product line has been demonstrated not having any impact on functional compatibility related to the combination of the pen needle with the pen injector devices including dose accuracy and limited differences in flow rates directly related to the differences of needle's internal diameter.

Based upon the above comparisons to the Predicate device. Pikdare Pen Needles are substantially equivalent to the Predicate device.

Non-Clinical Performance Data: [(807.92(b)(1)]

Pikdare pen needles successfully passed all the required non-clinical testing which included the following testing for compliance with the requirements of ISO 11608-2:2012 Needle-based injection systems for medical use -- Requirements and test methods -- Part 2: Needles

Test ParameterRequirement – ISO 11608-2:2012Result SubjectdeviceResult PredicatedeviceConclusion
MaterialsThe needle shall be made of tubingmaterials specified in ISO 9626.Requirement metRequirement metBoth products usetubes specified inISO 9626
DimensionsThe tubing features used in theneedles shall meet the requirementsof ISO 9626. For G34 needle thelimits of stiffness has been chosen tocomply with the same limits of G33TW needle. The needles shall fit thetest apparatus specified in item 7.3 ofISO 11608-2. The dimension shallcomply with Table 1 of clause 4.2.2of ISO 11608-2Requirement metRequirement metBoth productscomply with therequirements of ISO9626
Determinationof flow ratethrough theneedleThe needle was tested in accordancewith Annex A to ISO 11608-2 todetermine flow rate through theneedle.Requirement metRequirement metBoth products havethe same flow rate
Bond between huband needletubeThe union of the hub and needle tubeshall not break when tested inaccordance with Clause 9 of ISO11608-2.Requirement metRequirement metBoth products passthe breakage test

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "PIKDARE" in a bold, sans-serif font. The letters are a dark blue color, except for the "P", which is a lighter gray color. The "P" also has a small, horizontal line extending to the left, making it look like a plus sign.

Needle pointsNeedle points shall appear sharp andfree from featheredges, burrs andhocks when examinate undermagnification x2,5. The needlepoint at the cartridge end is designedto minimize coring andfragmentation when penetrating thecartridge set.Requirement metRequirement metBoth productsaccomplish theneedle requestedfeatures
Freedom fromdefectsThe needle tube shall fulfillthe requirements of ISO7864,clause11.3.Requirement metRequirement metBoth productsfulfill the ISO7864
LubricationThe needle tube should be lubricatedat both the patient end and thecartridge end. The lubricant shall not,under normal or corrected-to-normal vision, be visible asdroplets of fluid on the outsideRequirement metRequirement metBoth productsrespect theappropriate quantityof lubrication

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for PIKDARE. The logo consists of the word "PIKDARE" in a bold, sans-serif font. The letters are in a dark blue color. To the left of the word is a symbol that looks like a stylized "T" in a light brown color.

surface of the needle tube
Dislocation ofmeasuringpoint at patientendDislocation of the cannula point atthe patient end shall be inaccordance with Table 2 whentested in accordance with Clause 8(of ISO 11608-2).Requirement metRequirement metBoth products arecompliant with table2.
Determination offunctionalcompatibilitywith needle-based injectionsystemsCompatibility with any NIS(Needle-based Injection System)shall be claimed only after testing inaccordance with Clause 11.Requirement metRequirement metBoth products arecompatible to thesame list of peninjector device
Ease ofassemblyanddisassemblyAttachment of the needle shall bepossible without removing theneedle from its opened unitpackaging. Compliance is checkedaccording to the requirements ofClause 11.Requirement metRequirement metBoth products are incompliance withclause 11
SterilityThe needles in its unitpackaging is subjected to avalidatedsterilization processRequirement metRequirement metBoth products areETO sterilized inthe samesterilization place
Pre-conditioning ofneedlesThe needles satisfy all therequirements of ISO 11608-2 afterpreconditioning according toclauses 6.1, 6.2 and 6.3 of the samestandardRequirement metRequirement metBoth productsrespect ISO 11608-2

Based on the comparison table on Non-Clinical Performance Data, we can declare that there is no difference between Subject device and the Predicate device.

Test were performed on both products, the total availability on sample of the fact that the Predicate Device is manufactured in Pikdare factory in Italy (Pikdare is the contract manufacturer of Droplet G34 Pen Needle).

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "PIKDARE" in a bold, sans-serif font. The letters are primarily blue, except for the left part of the "P", which is a light brown color. The logo appears to be for a company or organization named "PIKDARE".

Biocompatibility testing:

Based on the test conduction according to ISO 10993-1 the medical device is categorized as follow:

  • Nature of the body contact: -
    • 0 Category: externally communicating medical device
    • 0 Contact: Blood path indirect
  • Contact Duration: B - prolonged (>24 h to 30 d)

List of constituents (extract from the Biological Evaluation Plan)

The list of the device constituents provided by the sponsor is reported in Table 2.

ComponentMaterial Type TypeMaterial Trade nameSupplierType ofcontact*
PrimaryContainerHigh DensityPolyethyleneCopolymerERACLENE ® MR 80 UVersalisNone/Instant
ERACLENE ® MS 80 UVersalis
RIGIDEX HD5218EAIneos
RIGIDEX HD5226EAIneos
HubHomopolymerPolypropyleneEltex® MED 100-MG25IneosNone/Instant
HEALTHCARE HPP25GRepsol
PPM H250 or ACESO® PPMH250Total
Hub MasterbatchPigment for polymers Remafin Blue PE52080094Pigment for polymers Remafin Yellow PE11078232Pigment for polymers Remafin Grey PE71077043Pigment for polymers Remafin Red PE31079060Pigment for polymers Remafin Red PE31079091Pigment for polymers Remafin Purple PE41076243Pigment for polymers Remafin PP-All coloursClariantNone/Instant
Pigment for polymers Sicolen Green 90-0755 SNPigment for polymers Sicolen Orange 28-5504BasfNone/Instant
CannulaeAISI 305 stainless steelBwsDirect
AISI 305 stainless steelOutokumpu
AISI 304 L stainless steelKobayashi
UV GlueUV3028LoxealDirect
AA3921Loctite
LubricantMedical silicon 360 12.500cpsMedical silicone MDX4-4159Q7 9180Dow CorningDirect
ShieldHigh DensityPolyethyleneCopolymerERACLENE ® MR 80 UVersalisNone/Instant
ERACLENE ® MS 80 UVersalis
RIGIDEX HD5218EAIneos
RIGIDEX HD5226EAIneos
Medical paperGrid Coated Paper0K08xxxx05xAmcorNone/Instant
BoxRecycled cardboardAutajonNone/Instant
MasterCarton B waveSaicaNone/Instant

Table 2: Constituents list.(") instant contact is referred to the contact with the intact skin in order to use the device, therefore negligible

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "PIKDARE" in a stylized font. The "P" is formed by a gray square with a blue vertical line and a horizontal line extending to the right. The rest of the letters are in blue and in a bold, sans-serif font.

Biocompatibility Test Summary – no adverse biocompatibility effects were observed:

Test methodCompliance withResult
Biological RiskAssessmentEvaluationISO 10993-1- Biologicalevaluation of medical devices - Part 1:Evaluation and testing within a riskmanagement processThe following end-points have beenidentified and verified by testing.
Cytotoxicity StudyUsing the ISO ElutionMethodISO 10993-5 - Biologicalevaluation of medical devices - Part 5:Tests for in vitro cytotoxicityThe test article showed no evidenceof causing cell lysis or toxicity.
ISO Guinea PigMaximization SensitizationTestISO 10993-10 - Biologicalevaluation of medical devices -- Part 10:Tests for irritation and skinsensitizationThe test article extracts showed noevidence of causing delayed dermalcontact sensitization in the guinea pig. Thetest article was not considered a sensitizerin the guinea pig maximization test.
ISO IntracutaneousStudy in RabbitsISO 10993-10 - Biologicalevaluation of medical devices -- Part 10:Tests for irritation andskin sensitizationThe test article met the requirementsof the test.
ASTM HemolysisStudyASTM F756, Standard Practicefor Assessment of Hemolytic Propertiesof Materials and ISO10993-4 - Biological evaluation ofmedical devices -- Part 4:Selection of tests for interactions withbloodBoth the test article in direct contactwith blood and the test article extract werenon-hemolytic
ISO Two WeekToxicity Study in the Rat,RepeatedParenteralAdministration of TwoExtractsISO 10993-11 - Biologicalevaluation of medical devices -- Part 11:Tests for systemic toxicityThere were no microscopic changesconsidered to be a test article relatedresponse. Parenteraladministration of the test articleextract did not produce systemic toxicity inrats.
ISO Systemic Toxicity Studyin MiceISO 10993-11 - Biologicalevaluation of medical devices -- Part 11:Tests for systemic toxicityThere was no mortality or evidenceof systemic toxicity from the extractsinjected into mice. Each test article met therequirements of the study.
USP Rabbit PyrogenStudy, Material-mediatedISO 10993-11 Biologicalevaluation of medical devices --Part 11: Tests for systemic toxicityThe total rise of rabbit temperaturesduring the 3 hours observation periodwas within acceptable USP limits.The test article was judged as non-pyrogenic
USP Pyrogen Study -Material MediatedUSP, General Chapter <151>,Pyrogen Test as recommended by ISO10993-11 Biologicalevaluation of medical devices --Part 11: Tests for systemic toxicityThe test article was judged asnonpyrogenic.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "PIKDARE" in a bold, sans-serif font. The letters are primarily in a dark blue color, except for the left part of the "P", which is in a light brown color. The logo has a modern and corporate look.

Sterilization:

Product is EO sterilized in house by manufacturer. Sterilization process is validated using current standard ISO 11135-1:2014 + ISO 11135:2014/Amd 1:2018.

The sterilization cycle assures a SAL of 10-6.

The maximum EO residual values after the degassing period, and before the product release, have been set at 1 mg for each device (ISO 10993-7 limit: 4 mg/day): these limits are set considering a maximum use of the devices equal to 4 times a day.

The maximum ECH residual values after the degassing period, and before the product release, have been set at 2,25 mg for each device (ISO 10993-7 limit: 9 mg/day): these limits are set considering a maximum use of the devices equal to 4 times a day.

The shelf life of the product, considering the integrity of the sterility and physical properties, is 5 years from the production date. The expiring date and sterilization methods are clearly indicated on the pack. The product shelf life is ensured if the product is stored and transported in compliance with the environmental condition stated on the pack and on the delivery boxes.

Clinical Performance Data: [(807.92(b)(2)]

Clinical data are not required.

Conclusion: ](807.92(b)(3)]

Pikdare Pen Needles (Insupen ORIGINAL/ Insupen ADVANCED) are concluded to be substantially equivalent in the intended use, technology/principle of operation, materials and performance to the predicate device.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).