K Number
K192082

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-04-21

(263 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DROPLET® PEN NEEDLE 34G is intended for use with pen injector devices for the subcutaneous injection of drugs.

Device Description

DROPLET® PEN NEEDLES 34G are sterile, single use needles intended for use with pen injector devices for the subcutaneous injection of drugs. The pen needles are OTC devices. The pen needle assembly consists of a double-ended into an injection molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient end of the cannula are lubricated using a siliconebased lubricant for ease of injection and rubber septum penetration. There is an inner needle shield assembled over the cannula to protect the needle point from damage and accidental needle sticks. There is also an outer cover. Each pen needle assembly is protected with a peel away seal to provide a sterility barrier.

AI/ML Overview

The provided text is a 510(k) Summary for the DROPLET® PEN NEEDLE 34G. It details the device's characteristics and compares it to predicate devices to establish substantial equivalence, rather than describing a study proving the device meets specific acceptance criteria in the context of diagnostic performance or clinical outcomes.

Therefore, many of the requested elements for describing "acceptance criteria and the study that proves the device meets the acceptance criteria" are not applicable or cannot be extracted directly from this document. The document focuses on regulatory clearance based on substantial equivalence to existing devices, primarily through non-clinical performance and scientific rationale, rather than a clinical trial demonstrating diagnostic or treatment effectiveness against specific performance metrics.

However, I can extract the acceptance criteria related to non-clinical performance and the "study" (bench testing and validations) that demonstrates compliance with those criteria.


1. A table of acceptance criteria and the reported device performance

Test Parameter / Acceptance Criteria (from ISO 11608-2:2012)Reported Device Performance
Materials: Needle shall be made of tubing materials specified in ISO 9626.Meets requirements
Dimensions: Needles shall fit the test apparatus specified in item 7.3 of ISO 11608-2.Meets requirements
Determination of flow rate through the needle: Tested in accordance with Annex A to ISO 11608-2.Meets requirements
Bond between hub and needle tube: Union of the hub and needle tube shall not break when tested in accordance with Clause 9 of ISO 11608-2.Meets requirements
Needle points: When examined under a magnification of x2.5, needle points shall appear sharp and free from feather edges, burrs and hooks.Meets requirements
Freedom from defects: Needle tube shall fulfill the requirements of ISO 7864, 11.3.Meets requirements
Lubrication: Lubricant shall not, under normal or corrected-to-normal vision, be visible as droplets of fluid on the outside surface of the needle tube.Meets requirements
Dislocation of measuring point at patient end: Dislocation of the cannula point at the patient end shall be in accordance with Table 2 of ISO 11608-2, when tested as per Clause 8.Meets requirements
Determination of functional compatibility with needle-based injection systems: Compatibility with any NIS shall be claimed only after testing in accordance with Clause 11.Meets requirements
Ease of assembly and disassembly: Attachment of the needle shall be possible without removing the needle from its opened unit packaging. Compliance is checked according to the requirements of Clause 11.Meets requirements
Sterility: Needle in its unit packaging shall have been subjected to a validated sterilization process (SAL = 10^-6).Meets requirements (validated per ISO 11135-7:2014, SAL = 10^-6)
Pre-conditioning of needles: All requirements of the standard related to preconditioning of needles were met.Meets requirements
Biocompatibility (ISO 10993-1): Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Subacute/subchronic Toxicity, Material-Mediated Pyrogenicity, Hemocompatibility.Did not show any adverse biological / biocompatibility reactions.
Sterilization Residuals: Ethylene Oxide residuals0.0268 mg/device (acceptable level)
Package Integrity: Seal strength, Dye penetration, Bubble leak after environmental conditioning, simulated transportation, and accelerated aging.All packaging deemed acceptable for protection of product and sterility maintenance.
Shelf Life:5 years (validated using FDA recognized standard for Accelerated Aging)

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes (N numbers) for each individual non-clinical test (e.g., how many needles were tested for flow rate, or how many units for package integrity). It refers to compliance with ISO standards which typically involve specific sampling plans. The data provenance is internal testing performed by HTL-STREFA S.A. (manufacturer located in Poland). The studies are non-clinical (bench testing, lab validations).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a non-clinical device clearance based on engineering and material performance standards. There is no "ground truth" established by medical experts for diagnostic or clinical performance. Compliance with standards like ISO 11608-2:2012 is generally assessed by qualified technicians and engineers.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not a clinical study involving human reader interpretation. Test results are objective measurements against defined standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device (pen needle) and the submission is for non-clinical performance and substantial equivalence, not an AI-powered diagnostic or assistive technology.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm or AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For non-clinical performance, the "ground truth" is adherence to established international standards and specifications (e.g., ISO 11608-2:2012 for needle performance, ISO 10993-1 for biocompatibility, ISO 11135-7:2014 for sterilization). These standards define acceptable limits and methodologies.

8. The sample size for the training set

Not applicable. There is no "training set" as this is not a machine learning model.

9. How the ground truth for the training set was established

Not applicable. There is no "training set" as this is not a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

April 21, 2020

HTL-Strefa S.A Aleksandra Prazmowska-Wilanowska Regulatory Affairs Director Adamowek 7 Ozorkow, 95-035 Pl

Re: K192082

Trade/Device Name: Droplet Pen Needle 34G Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: March 20, 2020 Received: March 23, 2020

Dear Aleksandra Prazmowska-Wilanowska:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Alan Stevens Assistant Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192082

Device Name DROPLET PEN NEEDLE 34G

Indications for Use (Describe)

The DROPLET® PEN NEEDLE 34G is intended for use with pen injector devices for the subcutaneous injection of drugs.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

As required by the Safe Medical Devices Act of 1990 and in accordance with 21 CFR §807.92(a).

Summary [807.92 (a)(1,2)] Date Prepared: March 20, 2020 HTL-STREFA S.A. Submitted By: ul. Adamówek 7 95-035 Ozorków POLAND Primary Contact: Aleksandra Prażmowska-Wilanowska RA Director Phone: +48 42 270 00 16 ola.prazmowska-wilanowska@htl-strefa.pl Izabela Banaś Secondary Contact: Senior RA Specialist Phone: +48 42 270 02 93 izabela.banas@htl-strefa.pl Trade Name: DROPLET® PEN NEEDLE 34G Common Name: Pen Needle Regulation Number: 21 CFR § 880.5570 Regulation Name Needle, Hypodermic, Single Lumen

{4}------------------------------------------------

Product Code:FMI
Device Classification:II

Review Panel: General Hospital

Predicate Device

[807.92(a)(3)]

The legally marketed device to which substantial equivalence is claimed is:

Predicate device:

Manufacturer NameTrade Name510(k) Number
Terumo Medical CorporationTerumo® Pen Injector Needle 34K140516

Reference device:

Manufacturer NameTrade Name510(k) Number
HTL-Strefa S.A.DROPLET® PEN NEEDLEK171982

The reference device cleared under K171982 has been included in this submission to cover the indications for use of the subject device.

{5}------------------------------------------------

Description of Device:

[807.92(a)(4)]

DROPLET® PEN NEEDLES 34G are sterile, single use needles intended for use with pen injector devices for the subcutaneous injection of drugs. The pen needles are OTC devices.

The pen needle assembly consists of a double-ended into an injection molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient end of the cannula are lubricated using a siliconebased lubricant for ease of injection and rubber septum penetration.

There is an inner needle shield assembled over the cannula to protect the needle point from damage and accidental needle sticks. There is also an outer cover. Each pen needle assembly is protected with a peel away seal to provide a sterility barrier.

Indications for Use:

[807.92(a)(5)]

The DROPLET® PEN NEEDLES 34G are intended for use with pen injector devices for the subcutaneous injection of drugs.

Technological Characteristics:

[807.92(a)(6)]

A comparison of characteristics of DROPLET® PEN NEEDLE 34G, the predicate device and the reference device is shown in the table below:

{6}------------------------------------------------

FeatureSubject DeviceDROPLET® PENNEEDLE 34GPredicate DeviceTerumo® Pen InjectorNeedle 34Reference DeviceDROPLET® PENNEEDLEConclusion
510(k) NumberpendingK140516K171982n/a
Product CodeFMIFMIFMISame
DesignPrimary ContainerSame
Needle ShieldSame
Needle Tube and HubSame
Indications for useThe DROPLET® PENNEEDLE 34G isintended for use with apen injector device forthe subcutaneousinjection of drugs.The Terumo® Pen InjectorNeedle 34 is intended for usewith a pen injector devicefor the subcutaneousinjection of drugs, includinginsulin.The Droplet® PenNeedle is intendedfor use with a peninjector device for thesubcutaneousinjection of drugs.Similar topredicateSame asreference
Length3.5mm-0.4mm +0.5mm4mm ±1.25mm12 mm, 10 mm, 8mm, 6mm, 5 mm,4mmSeeDiscussion
FeatureSubject DeviceDROPLET® PENNEEDLE 34GPredicate DeviceTerumo® Pen InjectorNeedle 34Reference DeviceDROPLET® PENNEEDLEConclusion
Gage34G34G29G, 30G, 31G, 32GSame aspredicate
Method ofattachment to peninjectorScrew threadsScrew threadsScrew threadsSame
BiocompatibilityConforms to ISO 10993-1Conforms to ISO 10993-1Conforms toISO 10993-1Same
SterilitySAL = 10-6SAL = 10-6SAL = 10-6Same
Sterilization methodEtOE-beam radiationGamma irradiationSeeDiscussion
Unit PackagingPolypropylene containerwith seal made of medicalgrade paperPolypropylene container withseal made of medical gradepaperPolypropylene containerwith seal made ofmedical grade paperSeeDiscussion
User PackagingCardboard sales boxCardboard sales boxCardboard sales boxSeeDiscussion
FeatureSubject DeviceDROPLET® PENNEEDLE 34GPredicate DeviceTerumo® Pen InjectorNeedle 34Reference DeviceDROPLET® PENNEEDLEConclusion
Materials
Needle TubeStainless steelAISI 304LStainless steelStainless steelAISI 304SeeDiscussion
HubPrimary ContainerNeedle ShieldPlastic resinsPlastic resinsPlastic resinsSeeDiscussion
LubricantMedical grade siliconeSilicon oilMedical grade siliconeSeeDiscussion

Device Comparison between Subject Device, Predicate Device Including Indications for Use

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Discussion of differences:

HTL-STREFA S.A. has determined that the Droplet® Pen Needle 34G is substantially equivalent to a predicate device currently cleared for marketing in the United States.

The Droplet® Pen Needle 34G is substantially equivalent to the Terumo® Pen Injector Needle 34 cleared under K140516 in terms of indications for use, compositions, material, design and performance. Specifically, the following performance comparisons were made to determine equivalence to the predicate devices pen need: length, gauge, biocompatibility, materials, shelf life and sterility. Based on the comparisons above of the predicate devices Pen Needle 34G has raised no different questions of safety and effectiveness.

The intended use of the devices are identical (the injection of fluids subcutaneously), and the indications only differ slightly between the subject device and the predicate. The subject device only differs from Terumo® Pen Injector Needle 34 cleared under K140516 in that insulin is not specified. This difference is adequate because the indications for use is within the scope of the predicate, which specifies for the delivery of drugs.

The difference in needle length between the Droplet® Pen Needle 34G and the predicate device - 3.5 mm (Droplet 34G) vs. 4 mm (Terumo 34G) has also been analyzed. Based on all available design, technological information as well as the published scientific and clinical data, we have concluded that the Droplet 34G, 3.5 mm, with tightened tolerance limits (-0.4mm +0.5mm), is within the same needle length range as the predicate 4mm (±1.25 mm) and thereby substantially equivalent. The results of the bench testing indicate that the Droplet 34G, 3.5mm needle ensures the same level of confidence for needle safety with regard to ensuring subcutaneous injection as the Terumo 34G, 4mm needle.

Finally, available scientific data confirms that a 3.5 mm needles provide reliable subcutaneous drug delivery, as confirmed in clinical study reported by de Berardis et al. in 20181.

¹ De Berardis G, Scardapane M, Lucisano G, Abbruzzese S, Bossi AC, Cipponeri E, Lancione R, Marelli G, Sciangula L, Nicolucci A. Efficacy, safety and acceptability of the new per randomized non-inferiority trial: AGO 02 study. Curr Med Res Opin. 2018 Sep:34(9):1699-1704.

{10}------------------------------------------------

Based on the above analysis of the needle length, it has been concluded that difference in needle length between the Droplet 34G device and the Terumo 34G device does not impact safety or performance of the former device and raises no different questions of safety and performance.

The difference in sterilization was addressed through validation per ISO 11135-7:2014 and reaching assurance level(SAL) of 10-6. The ethylene oxide residual levels were found to be 0.0268 mg/device, an acceptable level for the device type and use duration.

The difference in materials of construction were addressed through biocompatibility testing leveraged from the reference device per ISO 10993-1.

HTL-STREFA S.A. has determined that the Droplet® Pen Needle 34G is substantially equivalent to a predicate device currently cleared for marketing in the United States.

Non-Clinical Performance Data:

[(807.92(b)(1)]

DROPLET® PEN NEEDLE 34G successfully passed all the required non-clinical testing which included the following:

  • · Testing for compliance with the requirements of 11608-2:2012 Needle-based injection systems for medical use -Requirements and test methods -- Part 2: Needles
  • The table below presents the requirements of the 11608-2:2012 Needle-based injection systems for medical use --● Requirements and test methods -- Part 2: Needles standard and the result of the testing conducted.

{11}------------------------------------------------

Test ParameterClause no. & requirement of ISO 11608-2:2012Result
Materials4.1 The needle shall be made of tubing materials specified in ISO 9626.Meets requirements
Dimensions4.2 The needles shall fit the test apparatus specified in item 7.3 of ISO 11608-2.Meets requirements
Determination of flowrate through the needle4.3 The needle was tested in accordance with Annex A to ISO 11608-2 to determineflow rate through the needle.Meets requirements
Bond between hub andneedle tube4.4 The union of the hub and needle tube shall not break when tested in accordance withClause 9 of ISO 11608-2.Meets requirements
Needle points4.5 When examined under a magnification of x2,5, needle points shall appear sharp andfree from feather edges, burrs and hooks.Meets requirements
Freedom from defects4.6 The needle tube shall fulfill the requirements of ISO 7864, 11.3.Meets requirements
Lubrication4.7 The needle tube should be lubricated at both the patient end and the cartridge end.The lubricant shall not, under normal or corrected-to-normal vision, be visible asdroplets of fluid on the outside surface of the needle tube.Meets requirements
Dislocation of measuringpoint at patient end4.8 Dislocation of the cannula point at the patient end shall be in accordance with Table2 below when tested as per Clause 8 (of ISO 11608-2).Meets requirements
Determination offunctional compatibilitywith needle-basedinjection systems4.9 Compatibility with any NIS shall be claimed only after testing in accordance withClause 11.Meets requirements
Test ParameterClause no. & requirement of ISO 11608-2:2012Result
Ease of assembly anddisassembly4.10 Attachment of the needle shall be possible without removing the needle from itsopened unit packaging. Compliance is checked according to the requirements of Clause11.Meets requirements
Sterility4.11 The needle in its unit packaging shall have been subjected to a validatedsterilization process.Meets requirements
Pre-conditioning ofneedles6 All requirements of the standard related to preconditioning of needles were met.Meets requirements

{12}------------------------------------------------

Biocompatibility testing:

Selection of Biological Evaluation Tests

The Biological Tests selected to be performed on the has direct contact with the end-user according to the 2016 FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" contact type (externally communicating device, blood path indirect) and duration (prolonged contact (>24h to 30 days) were as follows:

    1. Cytotoxicity
    1. Sensitization
    1. Irritation or Intracutaneous Reactivity
    1. Acute Systemic Toxicity
    1. Subacute/subchronic Toxicity
    1. Material-Mediated Pyrogenicity
    1. Hemocompatibility

Biocompatibility tests selected as per the requirements of 2016 FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" considering contact type and duration for the DROPLET® PEN NEEDLE 34G did not show any adverse biological / biocompatibility reactions.

{13}------------------------------------------------

Sterility testing:

The DROPLET® PEN NEEDLE 34G is sterilized through Ethylene Oxide. The sterilization methodology has been validated per ISO 11 135-7:2014 overkill method (half cycle) with residuals of 0.0267 mg/device at a sterility assurance level (SAL) of 10°. The device is labeled non-pyrogenic and has been tested through LAL (USP <85>).

  • . Package integrity testing, after environmental conditioning and simulated transportation in accordance with ASTM D4169-16 and after accelerated aging was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection of product and sterility maintenance.
  • Sterile Barrier Packaging Testing performed on the proposed device: .
    • 0 Seal strength ASTM F88/F88-15
    • o Dye penetration ASTM F1929-15
    • o Bubble leak per ASTM 2096
  • Shelf life of 5 years is validated using the FDA recognized standard Guide for Accelerated Aging of ● Sterile Barrier Systems for Medical Devices

Clinical Performance Data:

[(807.92(b)(2)]

Clinical data is not required.

Conclusion:

[(807.92(b)(3)]

DROPLET® PEN NEEDLE 34G is concluded to be substantially equivalent in the intended use, technology/principle of operation. materials and performance to the predicate device.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).