(133 days)
The MAGNETOLITH Muscle Stimulator is indicated for:
-Relaxation of muscle spasms,
-Prevention or retardation of disuse atrophy,
-Increase local blood circulation.
-Muscle re-education,
-Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis,
-Maintaining or increasing range of motion.
The MAGNETOLITH delivers electromagnetic pulses to stimulate muscle tissues by applying the Magnetic Induction Field Principle (high-energy magnetic pulses) directly over (but not in direct contact with) the patient in order to achieve the same powered muscle stimulation without having to place adhesive electrode pads directly onto the patient's body. Key components of the device are the Control Unit, Handpiece Applicator, Applicator holder, a Mains cable, and a Trolley. These device components are provided non-sterile and are intended to be reusable.
The principle of operation of the MAGNETOLITHis by the application of a treatment coil to generate electromagnetic induction. i.e., the production of voltage across an electrical conductor in a changing magnetic field. The duration of the individual pulses is short and therefore the system does not increase temperature within the tissues. In addition, the MAGNETOLITH Applicator is cooled by circulating water.
It seems there are no specific acceptance criteria or details of a study proving the device meets those criteria provided in the document. The document is a 510(k) summary for the Storz Medical MAGNETOLITH Muscle Stimulator, demonstrating substantial equivalence to predicate devices, rather than a report on specific performance acceptance criteria.
The "PERFORMANCE TESTING" section (Page 4-5) states that: "Verification and validation testing was performed and demonstrated that the MAGNETOLITH meets the design specifications and is safe and effective for its intended use. All tests required by the verification and validation plan were completed and passed." However, it does not explicitly list the acceptance criteria or refer to a specific clinical study with patient data and ground truth establishment.
Instead, the performance testing focuses on:
- Software Validation: The MAGNETOLITH software was validated and demonstrated to be of a moderate level of concern.
- Risk Management: Hazard analysis/risk management was performed, and all risks were mitigated to an acceptable level.
- Biocompatibility: Not performed as there are no body-contacting components.
- Conformity to Standards: The device was tested and demonstrated to conform to general safety requirements of IEC 60601-1:2012, IEC 60601-6:2013, electromagnetic compatibility requirements of IEC 60601-1-2:2014, ISO 14971:2000/A1:2007, IEC 62304:2015, and ISO 13485:2016.
The document then provides a comparison table of the MAGNETOLITH with its predicate devices (HPM-6000 and Bemer Classic & Pro) to establish substantial equivalence based on technological characteristics and safety standards. This comparison is not a study proving specific performance criteria with clinical outcomes.
Therefore, I cannot provide the requested information for acceptance criteria and a study that proves the device meets those criteria based on the provided text. The document focuses on demonstrating substantial equivalence through technical comparisons and adherence to regulatory standards, not on a clinical performance study with specific acceptance criteria, sample sizes, expert ground truth, or MRMC studies.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo, which features a stylized eagle. To the right of the eagle is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
May 3, 2021
Storz Medical AG % Michael Dayton President & CEO Biomed Research, Inc. 3959 Van Dyke Road Suite 245 Lutz, Florida 33558
Re: K203710
Trade/Device Name: Storz Medical MAGNETOLITH Muscle Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, NGX Dated: February 1, 2021 Received: February 3, 2021
Dear Michael Dayton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning vour device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 53 1 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-ind-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Patrick Antkowiak -S
Patrick Antkowiak, PhD Assistant Director (Acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K203710
Device Name
MAGNETOLITH Muscle Stimulator
Indications for Use (Describe)
The MAGNETOLITH Muscle Stimulator is indicated for:
-Relaxation of muscle spasms,
-Prevention or retardation of disuse atrophy,
-Increase local blood circulation.
-Muscle re-education,
-Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis,
-Maintaining or increasing range of motion.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Storz Medical AG
MAGNETOLITH Muscle Stimulator
1. SPONSOR
Storz Medical AG Lohstampfestr. 8 CH-8274 Tägerwilen Switzerland
Contact Person: Pavel Novak, Ph.D. Telephone: 41 (0) 71 677 45 13 FAX: 41 (0) 71 677 45 05
Date Prepared: May 3, 2021
DEVICE NAME 2.
Proprietary Name: MAGNETOLITH Regulation Name: Powered Muscle Stimulator
| Regulation Number: | 21 CFR 890.5850 |
|---|---|
| Product Code: | IPF, NGX |
| Regulatory Class: | II |
3. PREDICATE DEVICES
Equivalence is claimed to the following predicate devices: HPM-6000 (K160992, primary), and Bemer Classic & Pro (K151834, secondary).
4. DEVICE DESCRIPTION
The MAGNETOLITH delivers electromagnetic pulses to stimulate muscle tissues by applying the Magnetic Induction Field Principle (high-energy magnetic pulses) directly over (but not in direct contact with) the patient in order to achieve the same powered muscle stimulation without having to place adhesive electrode pads directly onto the patient's body. Key components of the device are the Control Unit, Handpiece Applicator, Applicator holder, a Mains cable, and a Trolley. These device components are provided non-sterile and are intended to be reusable.
The principle of operation of the MAGNETOLITHis by the application of a treatment coil to generate electromagnetic induction. i.e., the production of voltage across an
{4}------------------------------------------------
electrical conductor in a changing magnetic field. The duration of the individual pulses is short and therefore the system does not increase temperature within the tissues. In addition, the MAGNETOLITH Applicator is cooled by circulating water.
5. INTENDED USE
The MAGNETOLITH is indicated for
- · Relaxation of muscle spasms,
- Prevention or retardation of disuse atrophy, .
- · Increasing local blood circulation,
- · Muscle re-education,
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, .
- Maintaining or increasing range of motion. .
6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The technological characteristics of the MAGNETOLITH device and the predicate device is substantially equivalent in that each device is designed to non-invasively use electromagnetic induction to initiate action potential of nerves resulting in muscle contraction.
7. PERFORMANCE TESTING
Verification and validation testing was performed and demonstrated that the MAGNETOLITH meets the design specifications and is safe and effective for its intended use. All tests required by the verification and validation plan were completed and passed.
The MAGNETOLITH software was validated and demonstrated to be of a moderate level of concern; while hazard analysis / risk management was performed and demonstrated that all risks are mitigated to an acceptable level. While the MAGNETOLITH has cleaning and sanitation requirements, there are no body contacting components and therefore biocompatibility testing was not performed. The MAGNETOLITH was tested and demonstrated to conform to the general safety requirements of IEC 60601-1:2012; IEC 60601-6:2013; and electromagnetic compatibility requirements of IEC 60601-1-2:2014.
No performance standards applicable to this device have been adopted under Section 514 of the Act. The MAGNETOLITH complies with the applicable requirements of the following international consensus standards:
- ISO 14971:2000/A1:2007: Medical devices: Application of risk management to . medical devices.
{5}------------------------------------------------
- IEC 60601-1:2012, (Ed. 3.1): Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance.
- IEC 60601-1-2: 2014(4th Ed.): Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
- . IEC 60601-1-6: 2012 (3.1 Ed.): Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- IEC 62304:2015: Medical Device Software - Software Life CycleProcess.
- ISO 13485:2016: Medical devices Ouality management systems -. Requirements for regulatory purposes.
The performance testing demonstrated that the MAGNETOLITH Muscle Stimulator is substantially equivalent to the predicate devices and that it is as safe, as effective, and performs as well as or better than the predicate devices. The table below compares the characteristics of the MAGNETOLITH to the predicate devices.
| 510(k) numberDevice nameCompany name | K160992HPM-6000BTL Industries, Inc. | K151834Bemer Classic & ProBemer Int'l AG | K203710MAGNETOLITHStorz Medical AG | Comparison |
|---|---|---|---|---|
| Product Code andRegulation | Physical Medicine21 CFR 890.5850IPF - Stimulator,Muscle, Powered | Physical Medicine21 CFR 890.5850NGX - Stimulator, Muscle,Powered | Physical Medicine21 CFR 890.5850IPF - Stimulator,Muscle, PoweredNGX - Stimulator, Muscle,Powered | Same |
| Intended Use /Indications for Use | The HPM-6000 device is intendedto be used under medicalsupervision for adjunctive therapyfor the treatment of medicaldiseases and conditions.The HPM-6000 device is indicatedfor use in stimulatingneuromuscular tissue for bulkmuscle excitation in the legs orarms for rehabilitative purposes.Indications for Use for MuscleStimulators:-Relaxation of muscle spasms-Prevention or retardation of disuseatrophy-Increasing local blood circulation-Muscle re-education-Immediate post- surgicalstimulation of calf muscles toprevent venous thrombosis-Maintaining or increasing rangeof motion | Indications for Use:-To temporarily increase localblood circulation in healthy legmuscles-To stimulate healthy muscles inorder to improve and facilitatemuscle performance | Indications for Use:-Relaxation of muscle spasms-Prevention or retardation ofdisuse atrophy-Increasing local bloodcirculation-Muscle re-education-Immediate post- surgicalstimulation of calf muscles toprevent venous thrombosis-Maintaining or increasing rangeof motion-To stimulate healthy muscles inorder to improve and facilitatemuscle performance | Similar |
8. SUBSTANTIAL EQUIVALENCE COMPARISON
{6}------------------------------------------------
| Principle of Action | Non- invasive tissue stimulationvia magnetic field induction | Non- invasive tissue stimulationvia magnetic field induction | Non- invasive tissue stimulationvia magnetic field induction | Same |
|---|---|---|---|---|
| Clinical Use | Prescription use | Over the counter use | Prescription use | Same |
| ElectricalProtection | Class II, BF | unknown | Class I, B | Similar¹ |
| UserInterface | Touch screen | Touch screen | Touch Screen | Same |
| FirmwareControlled | Yes | Yes | Yes | Same |
| Type ofEnergy | Magnetic field | Magnetic field | Magnetic field | Same |
| Numberof outputs | 1 | 2 | 1 | Same |
| Number ofMagnetic Coils inthe Applicator | 1 | unknown | 1 | Same |
| Magnetic FieldIntensity | BTL 299-1 applicator: 0.5 - 1.8 TBTL 299-2 applicator: 0.7 - 2.5 T | 35-100 μT | 0.4T ±20% at surface0.08T ±20% at center of coil | Similar |
| Pulse RepetitionRate | 1 - 150 Hz | 10-30 Hz | 1-10Hz | Similar |
| Pulse Duration | BTL 299-1 applicator: 280μs ±20%BTL 299-2 applicator: 280μs ±20% | 10-33μs | 125μs ± 20% | Similar |
| Therapy Time | Up to 60 min | 20 minutes | 10-20 minutes | Similar |
| Energy Source | 100 - 240 VAC,50-60 Hz | 100 - 240 VAC,50/60 Hz | 100 - 240 VAC,50/60 Hz | Same |
| System Dimensions(W×H×D) | 500×970×580 mm (20×38×23 in) | 320 x 320 x 70 mm | 454 x 187 x 460 mm | Similar |
| AmbientTemperature | 10° - 30°C | 10° - 30°C | 10° - 30°C | Same |
| Relative Humidity | 30 - 75% | unknown | 5-55% | Similar |
| EnvironmentalSpecifications | For indoor use only | For indoor use only | For indoor use only | Same |
'Electrical Protection: The electrical security measures in devices are slightly different, but both comply with theIEC60601-1 standard for basic safety and essential performance.
N/A